# Galaxy UNYCO System (K153233)

_Orthofix Srl · KTT · Jan 6, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153233

## Device Facts

- **Applicant:** Orthofix Srl
- **Product Code:** [KTT](/productcode/KTT.md)
- **Decision Date:** Jan 6, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Galaxy UNYCO System is intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment. Temporary stabilization of the tibia and foot in conditions and procedures, such as: - comminuted open or closed tibial fractures - polytrauma patient - damage control orthopedics for fractures with severe soft tissue injuries - peri-prosthetic or peri-implant fractures - joint dislocations, intra- and extra-articular injuries where spanning fixation is needed - intra-operative fracture reduction - intermediate stabilization in staged surgery - infected non-union pending second stage treatment bone-loss or other reconstructive procedures The Galaxy UNYCO System is compatible with Galaxy Fixation System and bicortical screws. Galaxy Fixation System and bicortical screws must be used when Galaxy UNYCO is not indicated or available. The product is indicated for non-weight-bearing use.

## Device Story

Modular external fixation system for temporary stabilization of lower limb fractures; consists of sterile kits containing UNYCO cancellous screws, large multiscrew clamps, 12mm diameter rods, radiolucent foot unit, and application tools. Used in clinical settings (trauma/orthopedic procedures) by surgeons to provide spanning fixation for tibia and foot injuries prior to definitive treatment. System is non-weight-bearing; provides mechanical stability to bone segments during healing or staged reconstruction. Surgeon assembles frame configuration based on fracture needs. Benefits include stabilization of severe soft tissue injuries and complex fractures, facilitating subsequent definitive orthopedic care.

## Clinical Evidence

Bench testing only. Mechanical testing performed per ASTM F 1541 (Standard Specification and Test Methods for External Skeletal Fixation Devices) and MRI compatibility testing per ASTM F 2182. No clinical data was required to support safety and effectiveness.

## Technological Characteristics

Modular external skeletal fixation system. Components include UNYCO cancellous screws, large multiscrew clamps, 12mm diameter rods, and radiolucent foot unit. Materials and design are comparable to predicate devices. Mechanical performance validated per ASTM F 1541-02. MRI compatibility evaluated per ASTM F 2182.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes Large External Fixation, MR Conditional ([K082650](/device/K082650.md))
- Orthofix Galaxy Unyco Diaphyseal Tibia Kit ([K142052](/device/K142052.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 6, 2016

Orthofix SRL % Ms. Cheryl Wagoner Consultant Wagoner Consulting LLC P.O. Box 15729 Wilmington, North Carolina 28408

Re: K153233

Trade/Device Name: Galaxy UNYCO System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: KTT Dated: November 6, 2015 Received: November 9, 2015

Dear Ms. Wagoner:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Lori A. Wiggins -S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K153233

Device Name Galaxy UNYCO System

Indications for Use (Describe)

The Galaxy UNYCO System is intended to be used for temporary bone stabilization in trauma and orthopedic procedures of the lower limb prior to definitive treatment.

Temporary stabilization of the tibia and foot in conditions and procedures, such as:

- · comminuted open or closed tibial fractures
- · polytrauma patient
- · damage control orthopedics for fractures with severe soft tissue injuries
- peri-prosthetic or peri-implant fractures
- · joint dislocations, intra- and extra-articular injuries where spanning fixation is needed
- · intra-operative fracture reduction
- · intermediate stabilization in staged surgery
- · infected non-union pending second stage treatment bone-loss or other reconstructive procedures

The Galaxy UNYCO System is compatible with Galaxy Fixation System and bicortical screws. Galaxy Fixation System and bicortical screws must be used when Galaxy UNYCO is not indicated or available. The product is indicated for non-weight-bearing use.

Type of Use (Select one or both, as applicable)

| <span style="font-size:10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/2 description: The image shows the logo for Orthofix, followed by the text "510(k) Summary". The Orthofix logo is a blue, abstract shape. The text is in a sans-serif font and is black.

(as required by 21 CFR 807.92)

| Submitter | Orthofix Srl                                         |
|-----------|------------------------------------------------------|
|           | Via delle Nazioni, 9   37012 Bussolengo (VR) - Italy |
| Telephone | + 39 045 6719.000                                    |
| Fax       | + 39 045 6719.380                                    |

| Contact Person | Gianluca Ricadona<br>Quality & Regulatory Affairs Manager |
|----------------|-----------------------------------------------------------|
| Address        | Via delle Nazioni, 9<br>37012 Bussolengo (VR) - Italy     |
| Telephone      | + 39 045 6719.000                                         |
| Fax            | + 39 045 6719.380                                         |
| email          | GianlucaRicadona@orthofix.it                              |
| Date Prepared  | November 6, 2015                                          |

| Trade Name             | Galaxy UNYCO System                                                             |
|------------------------|---------------------------------------------------------------------------------|
| Common Name            | External Fixation Device and Accessories                                        |
| Panel Code             | Orthopedic                                                                      |
| Classification<br>Name | Single/multiple component metallic bone fixation appliances and<br>accessories. |
| Class                  | Class II                                                                        |
| Regulation<br>Number   | 21 CFR 888.3030                                                                 |
| Product Code           | KTT                                                                             |

| Predicate Device Name                           | 510(k) Number | Manufacturer |
|-------------------------------------------------|---------------|--------------|
| Synthes Large External Fixation, MR Conditional | K082650       | Synthes      |
| Orthofix Galaxy Unyco Diaphyseal Tibia Kit      | K142052       | Orthofix Srl |

| Description | The Orthofix Galaxy UNYCO System consists of a series of sterile kits that<br>include UNYCO Cancellous Screws, Large Multiscrew Clamps for<br>UNYCO Screws, Rods Ø 12mm, a radiolucent Foot Unit and specific<br>application tools. External fixation systems are modular, therefore different<br>frame configurations are possible. The Orthofix components in the Galaxy<br>UNYCO System are not intended to replace normal healthy bone or to<br>withstand the stresses of weight bearing. |
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Indications and<br>Intended Use | The Galaxy UNYCO System is intended to be used for temporary bone<br>stabilization in trauma and orthopedic procedures of the lower limb prior<br>to definitive treatment.<br>Temporary stabilization of the tibia and foot in conditions and procedures,<br>such as:<br>• comminuted open or closed tibial fractures<br>• polytrauma patient<br>• damage control orthopedics for fractures with severe soft tissue injuries<br>• peri-prosthetic or peri-implant fractures<br>• joint dislocations, intra- and extra-articular injuries where spanning<br>fixation is needed |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

{4}------------------------------------------------

## Traditional 510(k) Premarket Notification
Galaxy Unyco System

| • intra-operative fracture reduction                                     |
|--------------------------------------------------------------------------|
| • intermediate stabilization in staged surgery                           |
| • infected non-union pending second stage treatment bone-loss or other   |
| reconstructive procedures                                                |
| The Galaxy UNYCO System is compatible with Galaxy Fixation System        |
| and bicortical screws. Galaxy Fixation System and bicortical screws must |
| be used when Galaxy UNYCO is not indicated or available.                 |
| The product is indicated for non-weight-bearing use.                     |

| Technological<br>Characteristics<br>and Substantial<br>Equivalence | Documentation was provided to demonstrate that the Galaxy Unyco<br>System is substantially equivalent to the legally marketed predicates. The<br>devices and accessories included in the Galaxy Unyco System and the<br>predicate devices are all external fracture fixation systems as defined in<br>21 CFR 888.3030. The Galaxy Unyco System is substantially equivalent<br>to the predicate devices in intended use, site of application, patient<br>population, conditions-of-use, mechanical performances, basic design,<br>operating principles, and materials. The Galaxy Unyco System is<br>comparable to its predicate in size and materials. Testing in accordance<br>with ASTM F 1541-02 shows the mechanical strength of the subject device<br>to be equivalent or better than the predicate devices. |
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|--------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

| Performance Data | The potential hazards have been evaluated and controlled through a Risk<br>Management Plan.<br><br>All testing met or exceeded the requirements as established by the test<br>protocols and applicable standards. A review of the mechanical data<br>indicates that the components of the Subject device are capable of<br>withstanding expected loads without failure. The Subject device was<br>therefore found to be substantially equivalent to the predicate devices.<br>Clinical data was not needed to support the safety and effectiveness of<br>the Subject Device.<br>The following mechanical testing was performed:<br>ASTM F 1541 "Standard Specification and Test Methods for External<br>Skeletal Fixation Devices"          MRI compatibility testing was also conducted per:  ASTM F2182 "Standard test method for measurement of radio<br>frequency induced heating near passive implants during magnetic<br>resonance imaging" |
|------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion       | Based on design, materials, intended use, technological characteristics,<br>and comparison to predicate devices, the Galaxy UNYCO System has<br>been shown to be substantially equivalent to legally marketed predicate<br>devices..                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

---

**Source:** [https://fda-staging.innolitics.com/device/K153233](https://fda-staging.innolitics.com/device/K153233)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).