ConforMIS iTotal Posterior Stabilized Knee Replacement System
K153217 · Conformis, Inc. · JWH · Jan 7, 2016 · Orthopedic
Device Facts
Record ID
K153217
Device Name
ConforMIS iTotal Posterior Stabilized Knee Replacement System
Applicant
Conformis, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jan 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Indications for Use
The iTotal® Posterior Stabilized (PS) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis or osteonecrosis of the knee. Post traumatic loss of joint function. Moderate varus, valgus or flexion deformity. Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans
Device Story
Patient-specific, tricompartmental, cruciate-sacrificing, semi-constrained, fixed-bearing, cemented knee replacement system. Input: patient CT scan data. Process: proprietary software (iTotal PS iView v1.0) designs patient-specific implants matching geometric/anatomic requirements; manual or automated generation of iView. Output: patient-specific femoral (CoCrMo alloy), tibial (CoCrMo tray, UHMWPE insert), and patellar (UHMWPE or Vitamin E infused iPoly XE™) components; patient-specific single-use manual surgical instruments for bone cutting and component positioning. Used in clinical/OR settings by orthopedic surgeons. Output assists surgeons in intraoperative alignment and bone preparation. Benefits: customized fit to patient anatomy.
Clinical Evidence
No clinical data. Substantial equivalence established via device description and non-clinical software verification and validation testing of the iTotal PS iView v1.0 module.
Technological Characteristics
Patient-specific cemented knee prosthesis. Materials: Cobalt-chromium-molybdenum (CoCrMo) alloy, ultra-high molecular weight polyethylene (UHMWPE), and Vitamin E infused polyethylene (iPoly XE™). System includes patient-specific single-use manual surgical instruments. Software: iTotal PS iView v1.0 for automated design/view generation. Connectivity: Proprietary software module. Sterilization: Provided sterile for single-use instruments.
Indications for Use
Indicated for patients with knee joint pain and disability due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis, osteonecrosis, post-traumatic loss of function, or failed prior implants (osteotomies, hemiarthroplasties, unicondylar/patellofemoral/bicompartmental implants) with moderate varus, valgus, or flexion deformity. Requires identifiable anatomic landmarks on imaging for revision procedures.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
ConforMIS iTotal PS Knee Replacement System (K152704)
ConforMIS iTotal PS Knee Replacement System (K142404)
ConforMIS iTotal PS Knee Replacement System (K142161)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure in profile, with three overlapping faces suggesting community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure, emphasizing the department's role and national scope. The logo is simple, using a single color, and is designed to be easily recognizable.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 7, 2016
ConforMIS, Incorporated Ms. Amita Shah Senior Vice President, Regulatory and Quality Affairs 28 Crosby Drive Bedford, Massachusetts 01730
Re: K153217
Trade/Device Name: ConforMIS iTotal® Posterior Stabilized Knee Replacement System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibal polymer/metal/polymer semi-contrained cemented prosthesis Regulatory Class: Class II Product Code: JWH, OOG, OIY Dated: November 3, 2015 Received: November 5, 2015
Dear Ms. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| гээгүйээ иамин она НТЈАЭН ТО ТИЭМТЯАЧЭО<br>സ്വലംബം<br>ອຢາ 107 ຂະໜາດ | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>les PRA Statement below. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 500(k) Number (if known<br>7733317 | |
| Device Name<br>ຕອນຂຽງຂັນທາງວັນຕຸລິກ ອອກ ກີ ອ້າງການປະກວດນາ ໂລກປີ 21M100m03 | |
| ອນ ໃນ ຂອງປະຈາບອາຣາ 10 ຊຶ່ງເປັນຮູບິດ ລ້າງການປະກວດນາງການປະກວດນາງ ເປັນປີ 10 ປີ 2000 ປີ ອານີ ອາຂອງປີ ນາມປູ່ ເມີນເຮົາ ຈ<br>ເຂົ້າປະດາປຸ ເຄກາອາການປະເທດຈີນປະເທດ ປະເທດຈາກບໍລິສັດ ແລະ ຂາກອນປະເທດລາວ ລອຍກາ ອາການ ຈຳນາ ງາມ 10 ທີ່ປີ 10 ທີ່ປີ<br>The Indications for Use include:<br>o ano yino ziren tatt saivel sitedirs on su ad yd besserbbs ylelos ad tournes and interessible in a uisd<br>triol sens this and in themes provins in the sur sour of bebreini is (287) mars and in the University of The<br>Indications for Use (Describe) | |
| ຂດຂວາ ຊູຕາ້ງສາກເ ກາອນີສຸດ ແດ ອໄດ້ຮັກປີຕອນເ<br>ອາສາຫະກຸນ ອປີ 10 ຊຶກເກດກໍ່ຂອງ ປາຣ ກາຣະອວອກ ຂໍ້າເຣຍາວາ ຂໍ້າເຣນນາຄາຕາ ເຄຕີ ປອນ້ຳບາງ ຂອງມຸ່ງອາຫຼຸ ແລະ ເຂົ້າເຈົ້າ ຈ<br>•<br>knene<br>ຣາຕຣໄທນ໌ ໄສເຕີອິນເຕີຣ໌ເນຊອຟຕ໌ວໄວຣ໌ເນຊັນແນລໄທນ໌ ຄ.ສ. ຂອງເຮຣໄໝວາປຸກຄົນພອບ ຂອງພັດທີ່ຈັງ<br>Moders varus, suras, or flexion deformin | |
| ຽງໄທວ ອະນຸ ອາກອາກອວ າດ? ນ້ອຍການປະກວດນ້ຳ ຂໍ້າ ການປະກວດຄວາມງາມ ຄົນໄປ | |
| | |
| Type of Use (Select one or them, as applicable<br>(O ກຣຊ ປິນຣ 108 ກິ ຈ ວ 12 ກິ ຄ ປ 12 ກິ ຄ ປ 12 ກິ ຄ ປ 10 ການ ໄດ້ ຖາ ວ ຂອງ ( 2 | (วันธุ์ (วันธุ์ (ต้อง สิวิ) อยู่ปี เพื่อวันออกอาการออก ( |
| นักมีความ จิง 3 จิง 9 ริโลสิคคลิส ค.ศ. 3 มหาวิท 1 ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1994 ค.ศ. 1994 1994 1994 1994 1999 1999 1999 1999 1999 1999 1999 1999 1999 1999 | |
| .WOJBE 223RGOA JIAM3 37ATS AA9 3HT OT MA07 G3T3J9MO3 AU0Y GM32 TOM OG<br>sing to to to to the nowledges and to stremerings of your seiland in the single | |
| the<br>ələldir. Bu baş və əlab ədə mərkəzi və və yaşıyır. Şəhrəm şəhər çalışmış və və və və və və və və və və və və və və və və və və və və və<br>:of ,nebrud sint pubsubar 101 suoitzeggus guibulani ,noitsello noitemoni sidi to<br>suburni senodser red super et sognews of betermine and the notestions sint tot emit neburg edit<br>Paperwork Reduction Act (PRA) Stati<br>sessives namul bus dillself to thembred<br><br>Office of Chief Information Officer<br>ADS SUY EPI Official Style | |
| o no<br>" เอตเนนต BMO bilev บุนิกอนทว ย รงเธอร์เวท ii รรอเกม ทั้งเรียนบอกเ | |
| (AN8) 1888 AG3 M803<br>ារប្រជាជនជនជាតិ | 14.41 120 |
{3}------------------------------------------------
## 6.0 510(K) SUMMARY (PAGE 1 OF 5)
| Submitter's Name and<br>Address: | ConforMIS, Inc.<br>28 Crosby Drive<br>Bedford, MA 01730 |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration<br>Number: | 3009844603 and 3004153240 |
| Date of Summary: | November 3, 2015 |
| Contact Person:<br>Telephone Number:<br>Fax Number: | Amita S. Shah, Sr. Vice President, Regulatory and Quality Affairs<br>(781) 345-9164<br>(781) 345-0147 |
| Name of the Device: | ConforMIS iTotal® Posterior Stabilized Knee Replacement System<br>(iTotal PS KRS) |
| Common Name: | Total Posterior Stabilized Knee Replacement System |
| Regulatory Status and<br>Regulation Number: | Class II<br>21 CFR 888.3560 |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-constrained<br>cemented prosthesis |
| Device Classification: | Product Code:<br>JWH: Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis.<br><br>OOG: Knee Arthroplasty Implantation System.<br>Intended to be used to assist in the implantation of a specific knee<br>arthroplasty device or a set of specific knee arthroplasty devices.<br>Indicated to include guiding alignment, making or establishing cuts,<br>selecting, sizing, attaching, positioning or orienting implant<br>components.<br><br>OIY: prosthesis, knee, patellofemorotibial, semi-constrained, cemented,<br>polymer + additive/metal/polymer + additive.<br>This generic type of device includes prosthesis that have a femoral<br>component made of alloys, such as cobalt-chromium-molybdenum, and<br>a tibial component(s) and/or a retropatellar resurfacing component<br>made of ultra-high molecular weight polyethylene plus an additive, such<br>as a-tocopherol. |
| Indications for Use: | The iTotal® Posterior Stabilized (PS) Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.<br><br>The Indications for Use include:<br>Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis or osteonecrosis of the knee. Post traumatic loss of joint function. Moderate varus, valgus or flexion deformity. Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans |
| This implant is intended for cemented use only. | |
| Identification of the Legally<br>Marketed Device (Predicate<br>Device): | ConforMIS iTotal PS Knee Replacement System (ITOTAL PS KRS)<br>Device Class: II<br>Product Code: JWH, OOG, OIY<br>Regulation Number: 21 CFR 888.3560<br>510(k) Number: K152704, K142404, K142161 |
{4}------------------------------------------------
# 510(K) SUMMARY (PAGE 2 OF 5)
{5}------------------------------------------------
#### 510(K) SUMMARY (PAGE 3 OF 5)
The iTotal Posterior Stabilized Retaining Knee Replacement System Device Description: (hereafter referred to as the "iTotal PS KRS") is a patient specific tricompartmental faceted, cruciate sacrificing knee replacement system. The iTotal PS KRS is a semi-constrained, fixed bearing, cemented knee implant which consists of a femoral, tibial and patellar component.
> Using patient imaging (CT scan) and a combination of proprietary and off the shelf software, a patient specific implant is designed, that best meets the geometric and anatomic requirements of the specific patient. The femoral component is manufactured from cobalt chromium molybdenum ("CoCrMo") alloy. The tibial component includes a metal tray manufactured from CoCrMo alloy and a polyethylene insert that is manufactured from ultra-high molecular weight polyethylene (UHMWPE). The patellar component is manufactured from either UHMWPE or from a highly cross-linked Vitamin E infused polyethylene (iPoly XE™).
> For user convenience, and similar to the predicate iTotal PS KRS, patient specific accessory orthopedic manual surgical instruments designed for use with the modified iTotal PS KRS are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding the cutting of bone. These instruments are single use only and may be provided sterile. In addition, reusable orthopedic manual surgical instruments are provided separately.
The function and general design features of the patient specific implants and ancillary instruments remain similar to those described in the predicate 510(k)s K152704, K142404 and K142161.
{6}------------------------------------------------
#### 510(K) SUMMARY (PAGE 4 OF 5)
Comparison Summary of Technological Characteristics and Modifications Proposed:
The modified iTotal PS KRS and predicate iTotal PS KRS have the same technological characteristics. There have been no changes to the modified iTotal PS KRS device with respect to design, materials and methods of manufacture, packaging, or sterilization. The designs of the device remains similar to that cleared in K152704, K142404 and K142161.
Modifications which represent improving manufacturing efficiencies are primarily focused on adding the option of automating the generation of the iTotal PS iView using the proprietary software module iTotal PS iView v1.0.
#### Table 6.0-1: Comparison between the Modified and Predicate Device
| Characteristic | Predicate<br>iTotal PS KRS<br>(K152704, K142404<br>and K142161) | Modified iTotal PS<br>KRS<br>(This submission) |
|------------------------|-----------------------------------------------------------------|------------------------------------------------|
| Generation of<br>iView | Manual process | iTotal PS iView v1.0<br>or Manual process |
Nonclinical testing: The determination of substantial equivalence for this device was based on a detailed device description. The following non-clinical laboratory testing was performed demonstrating that the device can be considered substantially equivalent to the predicate device for the intended use:
- Detailed software description and software verification and . validation testing of proprietary software iTotal PS iView v1.0.
Substantial Equivalence: The product subject of this premarket notification is substantially equivalent to the iTotal PS KRS (K152704, cleared October 21, 2015, K142404, cleared December 11, 2014 and K142161, cleared November 14, 2014). The following testing was performed to establish substantial equivalence:
- . Software verification and validation testing of proprietary software iTotal PS iView v1.0.
{7}------------------------------------------------
### 510(K) SUMMARY (PAGE 5 OF 5)
Conclusion: The determination of substantial equivalence for this device was based on a detailed device description. Non-clinical laboratory testing was performed demonstrating that the device can be considered substantially equivalent to the predicate device for the intended use. Clinical data is not necessary to demonstrate substantial equivalence.
> Based on the testing conducted, it is concluded that the modified iTotal PS KRS is substantially equivalent to the predicate iTotal PS KRS (K152704, cleared October 21, 2015, K142404, cleared December 11, 2014 and K142161, cleared November 14, 2014).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.