M-Fix Acromioclavicular Device

K153211 · Coracoid Solutions, Inc. · HTN · Feb 2, 2016 · Orthopedic

Device Facts

Record IDK153211
Device NameM-Fix Acromioclavicular Device
ApplicantCoracoid Solutions, Inc.
Product CodeHTN · Orthopedic
Decision DateFeb 2, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The M-Fix Acromioclavicular Device is intended to provide fixation during the healing process following syndesmotic trauma such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions.

Device Story

M-Fix Acromioclavicular Device is a suture band system for reduction and stabilization of acromioclavicular joint following syndesmotic disruptions; consists of polyethylene terephthalate (PET) band tensioned and secured with integrated low-profile titanium or stainless steel connector/button; used by orthopedic surgeons in clinical settings; provided sterile for one-time use; includes single-use accessory instruments (Counter Traction Tool, Punch Tool) for securing connector; provides mechanical fixation to support healing; benefits patient by maintaining joint reduction during recovery.

Clinical Evidence

Bench testing only. Performance evaluated via ultimate tensile and cyclic testing to verify design and compare against predicate devices.

Technological Characteristics

Materials: Polyethylene terephthalate (PET) band, titanium or stainless steel connector/button. Form factor: Suture band system with low-profile connector. Sterilization: Provided sterile. Single-use. Mechanical fixation principle.

Indications for Use

Indicated for patients requiring fixation of acromioclavicular separations resulting from coracoclavicular ligament disruptions during the healing process.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them that resemble fabric or water. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 2, 2016 Coracoid Solutions, Inc. % Michael Kolber Consultant, Regulatory Affairs Michael Kolber 337 Ester Ave Campbell, California 94025 Re: K153211 Trade/Device Name: M-Fix Acromioclavicular Device Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN Dated: November 4, 2015 Received: November 5, 2015 Dear Michael Kolber: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/1 description: The image contains the logo for Coracoid Solutions. The logo consists of a stylized letter "C" and "S" intertwined, with the "C" in blue and the "S" in green. To the right of the logo are the words "CORACOID" and "SOLUTIONS" stacked on top of each other, both in a light blue color. ## Indications for Use 510(k) Number (if known): _K153211 Device Name: Coracoid Solutions, M-Fix™ Acromioclavicular Device Indications for Use: The M-Fix™ Acromioclavicular Device is intended to provide fixation during the healing process following syndesmotic trauma such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Coracoid Solutions. The logo consists of two intertwined letter C's, one in blue and one in green. To the right of the logo is the company name, "CORACOID" in blue, with "SOLUTIONS" underneath it, also in blue. #### 510(k) Summary (per 21CFR807.92) | General Company Information | | | |-----------------------------|-----------------------------------------------------------------------------------------------|---------------| | Name: | Coracoid Solutions, LLC | | | Contact: | Michael Kolber<br>Regulatory Affairs | | | Address: | 15 Normans Way<br>Park City, UT 84060 | | | Telephone: | 408-505-6626 | | | Fax: | 408-608-0338 | | | Date Prepared: | November 4, 2015 | | | General Device Information | | | | Product Name: | M-FixTM Acromioclavicular Device | | | Common Name: | Button/Suture | | | Classification Name: | Washer, Bolt Nut | | | Regulation Number: | 21CFR888.3030, Single/multiple component metallic bone<br>fixation appliances and accessories | | | Device Class: | Class II | | | Product Code: | HTN | | | Predicate Devices | | | | Manufacturer | Device Name | 510(k) Number | | Arthrex, Inc. | TightRopeTM Acromioclavicular (AC) Device | K052776 | | Surgicraft, Ltd. | LockdownTM Acromioclavicular (AC) Device | K091207 | | Dallen Medical, Inc. | TensynTM Band | K131850 | | Description | | | The M-Fix™ Acromioclavicular Device (M-Fix™) is a suture band system designed to provide reduction and stabilization of the acromioclavicular joint following syndesmotic disruptions. The M-Fix is a polyethylene terephthalate (PET) band that may be tensioned and secured with the integrated, low-profile Connector/Button which is made of titanium or stainless steel. The one-time use band with connector is provided sterile. Additional accessory instruments, including a Counter Traction Tool and a Punch Tool for securing the Connector/Button are single use devices provided sterile for use during the orthopedic procedure. #### Intended Use (Indications) The M-Fix Acromioclavicular Device is intended to provide fixation during the healing process following syndesmotic trauma such as fixation of acromioclavicular separations due to coracoclavicular ligament disruptions. #### Substantial Equivalence This submission supports the position that the Coracoid Solutions M-Fix is substantially equivalent to the Arthrex TightRope™ Acromioclavicular (AC) Device (K052776), the {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for Coracoid Solutions. The logo consists of two intertwined letter C's, one in blue and one in green, with the green C appearing to be in front of the blue C. To the right of the intertwined C's is the text "CORACOID" in blue, with the text "SOLUTIONS" in blue below it. The logo is simple and modern, and the use of blue and green gives it a clean and professional look. Surgicraft Lockdown™ Acromioclavicular (AC) Device (K091207), and the Dallen Medical Tensyn™ Band (K131850). The M-Fix™ has the following similarities to the previously cleared predicate devices: the same or similar intended use, same operating principle, similar technologies, and similar manufacturing process. Design verification activities were performed as a result of the risk analysis. The 510(k) notice contains summaries of bench studies, which were conducted to evaluate the performance characteristics of the M-Fix™ . Ultimate Tensile and Cyclic Tests were performed. The information presented demonstrates that the M-Fix™ Acromioclavicular Device is substantially equivalent to the predicate devices. #### Conclusions Coracoid Solutions, LLC believes that the information provided demonstrates that the proposed device is substantially equivalent to the predicate devices and does not raise any new issues of safety or efficacy. Based on the indications for use, technological characteristics, and comparison to predicate devices, the M-Fix has been shown to be substantially equivalent to predicate devices as described under the Federal Food, Drug and Cosmetic Act.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...