← Product Code [OMP](/productcode/OMP) · K153199

# Prevena Plus Incision Management System Kit with Peel & Place Dressing, Prevena Plus Peel & Place Dressing 5-Pack, Prevena Plus Incision Management System Kit with Customizable Dressing, Prevena Plus Customizable Dressing 5-Pack (K153199)

_Kci USA, Inc. · OMP · Feb 29, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153199

## Device Facts

- **Applicant:** Kci USA, Inc.
- **Product Code:** [OMP](/productcode/OMP.md)
- **Decision Date:** Feb 29, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment, and removing exudate via the application of negative pressure wound therapy.

## Device Story

System manages surgical incision environment post-closure; utilizes negative pressure wound therapy (NPWT) to remove exudate and maintain closed environment. Components include Prevena Plus Peel & Place or Customizable dressings and a KCI negative pressure therapy unit (Prevena Plus 125, ActiV.A.C., V.A.C. Freedom, InfoV.A.C., V.A.C. ATS, or V.A.C. Ulta). Dressings applied in OR; connected to therapy unit via SensaT.R.A.C. tubing and disposable canister. System provides continuous -125 mmHg pressure for 2-7 days. Therapy units feature visual/audible alarms for leaks, blockages, or full canisters. Clinicians use system to potentially reduce surgical site complications by drawing incision edges together and managing fluid. Prevena Plus 125 unit supports AC/battery power and larger canister volumes compared to previous iterations.

## Clinical Evidence

No clinical tests were necessary. Substantial equivalence supported by non-clinical bench testing, including pressure distribution analysis on simulated wound beds and alarm verification (leak, blockage, canister full) over 7-day continuous operation.

## Technological Characteristics

NPWT system; -125 mmHg continuous pressure. Components: occlusive dressings (Peel & Place/Customizable), SensaT.R.A.C. tubing, disposable canisters (150/250cc), and therapy units (AC/battery powered). Connectivity: pressure sensing via tubing. Sterilization: not specified. Software: embedded firmware for alarm management and pressure regulation.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- Prevena Incision Management System with Peel & Place and Customizable Dressings ([K141017](/device/K141017.md))

## Reference Devices

- V.A.C. Via Negative Pressure Wound Therapy System ([K132741](/device/K132741.md))
- V.A.C. Negative Pressure Wound Therapy Systems ([K120033](/device/K120033.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 29, 2016

Kinetic Concepts Incorporated USA Ms. Margaret Marsh Technical Director, Regulatory Affairs 6203 Farinon Drive San Antonio, Texas 78249

Re: K153199

Trade/Device Name: Prevena Plus Incision Management System Kit With Peel & Place Dressing, Prevena Plus Peel & Place Dressing 5-pack, Prevena Plus Incision Management System Kit With Customizable Dressing. Prevena Plus Customizable Dressing 5-pack

Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: OMP Dated: February 1. 2016 Received: February 3, 2016

Dear Ms. Marsh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

#### K153199

Device Name

Prevena Plus Incision Management System with Peel & Place and Customizable Dressings

Indications for Use (Describe)

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment, and removing exudate via the application of negative pressure wound therapy.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) SUMMARY Prevena Plus Incision Management System

| Submitter Information [21 CFR 807.92(a)(1)]                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                                                                                      | KCI USA, Inc. (Kinetic Concepts, Inc.)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Address                                                                                   | 6203 Farinon Drive<br>San Antonio, TX 78249                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Phone number                                                                              | 210-255-6481                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Fax number                                                                                | 210-255-6727                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Establishment<br>Registration Number                                                      | 3005178245                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Name of contact person                                                                    | Margaret Marsh, Technical Director, Regulatory Affairs                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Date prepared                                                                             | February 22, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Name of the device [21 CFR 807.92(a)(2)]                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Trade or proprietary<br>name                                                              | Prevena Plus Incision Management System with Peel & Place and<br>Customizable Dressings                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Common or usual name                                                                      | Negative Pressure Wound Therapy System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification name                                                                       | Negative Pressure Wound Therapy Powered Suction Pump (and<br>components)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Classification panel                                                                      | General and Plastic Surgery                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Regulation                                                                                | 878.4780                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulatory Class                                                                          | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code(s)                                                                           | OMP                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Legally marketed<br>device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | Prevena Incision Management System with Peel & Place and Customizable<br>Dressings (as cleared under K141017)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device description<br>[21 CFR 807.92(a)(4)]                                               | The Prevena Plus Incision Management System provides surgical incision<br>management via the application of negative pressure wound therapy over an<br>incision site that has been surgically closed with sutures or staples. The<br>system is applied to the incision site immediately after surgery for a minimum<br>of 2 days up to a maximum of 7 days depending on the surgeon's preference.<br>The Prevena Plus Incision Management System consists of:<br>A Prevena Dressing (Prevena Plus Peel & Place Dressing or Prevena Plus Customizable Dressing) and One of the following KCI Negative Pressure Wound Therapy Units:  Prevena Plus 125 Therapy Unit* ActiV.A.C. Therapy Unit V.A.C. Freedom Therapy Unit InfoV.A.C. Therapy Unit V.A.C. ATS Therapy Unit V.A.C. Ulta Therapy Unit            * The Prevena Plus 125 Therapy Unit is a slightly modified V.A.C.Via Therapy<br>Unit, previously cleared under K132741, that provides continuous negative<br>pressure at -125mmHg only. The V.A.C.Via Therapy Unit serves as a<br>reference device for 510(k) K153199. |

{4}------------------------------------------------

| <br><br><br>:                                                                                                                                                                                                                                                                                                    |                                                               | take on the street to MIN<br>.                                                                                                                                                      | <br>:<br>                             | 2008/02/2                                    | <br>.<br>.<br>**********<br><br><br>:                                             |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|----------------------------------------------|-----------------------------------------------------------------------------------|
| 11. Personal                                                                                                                                                                                                                                                                                                     | Comments of the consisted to the first to the first           | <br>:                                                                                                                                                                               | Contraction of the Research           | The submit the submit the subject to the<br> | <br>.<br>.<br><br>                                                                |
| .                                                                                                                                                                                                                                                                                                                | Control Concession Comparis<br><br>School proportunity<br>. . | <br>                                                                                                                                                                                | <br><br>                              | <br>                                         | Coupon Children<br>                                                               |
| 1994/04<br><br>. .<br>BARRAND MARKET                                                                                                                                                                                                                                                                             |                                                               | the successful the production of the first of the country of the county of<br>.<br>Concession Comprehensive Children<br>Comments of Children                                        | .<br>Production<br>.                  | PARTAL PROPERTY.<br>:<br>.                   | .<br><br>Actually as the first to the<br>.                                        |
| 000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000<br>.<br>were and the state the state of the specifically and the first of<br>for any the promise for the first from the first<br> | · September 1988 - 1988 - 1998<br>                            | .<br>****************************************************************************************************************************************************************************** | <br>:<br><br>Company of Children Star | :<br>1933 - 1<br>                            | Schoolstrapper<br>Children Children Children Children<br><br>Callery College<br>. |

The Prevena Plus Incision Management System is intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudate via the application of negative pressure wound therapy.

Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]

Negative Pressure Wound Therapy is the technological principal for both the subject and predicate Negative Produce Wound Though is the tosting on incision site that is closed via staples or sutures helps draw the incision edges together and removes fluid from the incision site. The occlusive drape of the dressing provides a negative pressure environment and protects the incision from external contamination.

The subject device system (Prevena Plus Incision Management System) is comprised of previously cleared components as shown below:

| Component                         | Predicate Device:<br>Prevena Incision<br>Management System<br>(K141017)                                                       | Reference Device:<br>V.A.C. Via Negative<br>Pressure Wound Therapy<br>System<br>(K132741)                                                                                                                 | Reference Device:<br>V.A.C. Negative Pressure<br>Wound Therapy Systems<br>(K120033) |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Indications<br>for Use            | Identical to Predicate                                                                                                        | Not applicable                                                                                                                                                                                            | Not applicable                                                                      |
| Dressings                         | Identical to the Predicate<br>Prevena Peel & Place and<br>Customizable Dressings                                              | Not applicable                                                                                                                                                                                            | Not applicable                                                                      |
| Therapy<br>Units and<br>Canisters | Identical to the Predicate<br>Acti V.A.C., Info V.A.C.,<br>V.A.C.Ulta and V.A.C.<br>Freedom Therapy Units                     | The Prevena Plus 125<br>Therapy Unit is identical to<br>the reference V.A.C. Via<br>Therapy Unit cleared<br>under (K132741), except for<br>removal of options for<br>intermittent and -75 mmHg<br>therapy | Not applicable                                                                      |
| Tubing Set                        | Not applicable                                                                                                                | Not applicable                                                                                                                                                                                            | SensaT.R.A.C. tubing set                                                            |
| Labeling                          | Includes Predicate<br>Prevena Incision<br>Management System<br>safety information and<br>dressing application<br>instructions | Includes Reference<br>V.A.C. Via Therapy Unit<br>operating instructions and<br>alarm resolution<br>information.                                                                                           | Not applicable                                                                      |

{5}------------------------------------------------

Summary of the technological characteristics of the device compared to the predicate device, continued

At a high level, the subject device and predicate device are based on the following same technological elements:

- The dressings that are applied over the incision site in the operating room are identical. One of the . following dressings may be selected by the surgeon, based on incision length and geometry:
	- The Prevena Plus Peel & Place Dressing which can be used for linear incisions up to 8 inches, or o
	- The Prevena Plus Customizable Dressing which can be configured for non-linear incisions or linear o incisions longer than 8 inches
- A negative pressure pump (therapy unit) is required that can provide -125 mmHg of negative pressure . continuously to the dressing for a maximum of 7 days.
- The dressings are connected to the therapy unit via a disposable canister and tubing set. .
- Incision fluid is collected into the disposable canister .
- The therapy unit provides alarms that indicate when negative pressure wound therapy may be . compromised (e.g., visual and audible alarms indicating an air leak in the system or when the canister is full).

The following technological differences exist between the subject and predicate devices:

- Connection between the dressings and therapy unit canisters is via a SensaT.R.A.C. tubing set rather . than by a Prevena tubing set. This allows for pressure sensing at the incision site for all of the indicated therapy units.
- Use of the Prevena Plus 125 Therapy Unit requires a connector (Prevena Plus Connector) to connect the . SensaT.R.A.C. tubing to the canister. No tubing connector is required for use of all of the other indicated therapy units.
- Comparing the Prevena Plus 125 Therapy Unit to the Prevena 125 Therapy Unit, there are the following . technological differences:
	- The Prevena Plus 125 Therapy Unit is both AC and battery powered, whereas the Prevena 125 o Therapy Unit is powered by batteries only.
	- The Prevena Plus 125 Therapy Unit is provided with either a 150 or 250 cc canister, whereas the O Prevena 125 Therapy Unit has a 45 cc canister.
	- The Prevena Plus 125 Therapy Unit provides a shared blockage/canister full alert whereas the O Prevena 125 Therapy Unit has a canister full alert.

#### Performance Data [21 CFR 807.92(b)]

Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]

The following performance data were provided in support of the substantial equivalence determination.

- The average distribution of negative pressure across the full length of the Prevena Plus Customizable . Dressing on a simulated wound bed under wet conditions and at maximum air leak demonstrated that the Prevena Plus 125 Therapy Unit and tubing set were able to provide negative pressure within specification at -125 mmHg, continuous mode for 7 days, equivalent to both the predicate and reference therapy units.
- The Prevena Plus 125 Therapy Unit was shown to provide alarms for leak, blockage and full canister . within specification when connected to the Prevena Plus Customizable Dressing and tubing set.

Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]

No clinical tests were necessary

Conclusions drawn [21 CFR 807.92(b)(3)]

The Prevena Pius Incision Management System and its predicate, the Prevena Incision Management System (K141017) are substantially equivalent in terms of safety, function and indications for use.

---

**Source:** [https://fda-staging.innolitics.com/device/K153199](https://fda-staging.innolitics.com/device/K153199)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com