K153146 · Sejoy Electronics & Instruments Co., Ltd. · FLL · Mar 31, 2016 · General Hospital
Device Facts
Record ID
K153146
Device Name
Infrared Ear Thermometer
Applicant
Sejoy Electronics & Instruments Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Mar 31, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Infrared Ear Thermometers, models ET-101D and ET-101H are indicated for the intermittent measurement and monitoring of human body temperature by consumers in a home use environment. It is intended for use on people of all ages.
Device Story
Infrared ear thermometers (models ET-101D, ET-101H) measure body temperature via infrared radiation emitted from the tympanic membrane and surrounding ear canal tissues. Device uses a thermopile infrared detector to capture thermal radiation; internal software converts analog signals into digital temperature values. Intended for home use by consumers of all ages. ET-101H includes optional disposable probe covers and additional modes (object/room temperature); ET-101D is a simpler model without probe covers or object/room modes. Results are displayed on an integrated LCD. Provides intermittent monitoring to assist users in tracking body temperature.
Clinical Evidence
Clinical evaluation conducted per ISO 80601-2-56 and ASTM E1965. ET-101D clinical bias: 0.07°C, repeatability: 0.11°C, limits of agreement: 0.76°C. ET-101H clinical bias: 0.08°C, repeatability: 0.13°C, limits of agreement: 0.76°C. No animal testing performed.
Technological Characteristics
Infrared thermopile sensor; ABS plastic and stainless steel construction; powered by CR2032 (ET-101D) or AAA batteries (ET-101H). Measures 32.0°C–43.0°C. Complies with AAMI/ANSI ES60601-1, ISO 80601-2-56, ASTM E1965-98, IEC 60601-1-11, and IEC 60601-1-2. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for intermittent measurement and monitoring of human body temperature for all ages in home use environments.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2016
Sejoy Electronics & Instruments Co., Ltd. Ren Yunhua General Manager Building 2, No. 202, Zhenzhong Road, West Lake Economy & Technology Zone Hangzhou, 310030 CHINA
Re: K153146
Trade/Device Name: Infrared Ear Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: February 25, 2016 Received: February 29, 2016
Dear Ren Yunhua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153146
Device Name Infrared Ear Thermometers ET-101D,ET-101H
#### Indications for Use (Describe)
Infrared Ear Thermometers, models ET-101D and ET-101H are indicated for the intermittent measurement and monitoring of human body temperature by consumers in a home use environment.It's intended for use on people of all ages.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Research Use Only (21 CFR 201.128) and/or |
|----------------------------------------------------------------------------------------|
| <input checked="checked" type="checkbox"/> Over-The-Counter Use (21 CFR 201.66) and/or |
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
The assigned 510(k) number is: K153146 Date Prepared: 2015.11.22
## I. Submitter:
Name: Sejoy Electronics & Instruments Co., Ltd. Add.: Building 2, No.202, Zhenzhong Road, West Lake Economy & Technology Zone, Hangzhou, China 310030 Contact Person: Yunhua Ren Phone: +86-571-81957767 Fax: +86-571-81957750 Email: renyh@sejoy.com
#### II. Device:
Trade Name: Infrared Ear Thermometer, Including models: ET-101D,ET-101H Common Name: Infrared Ear Thermometer Classification name: Clinical Electronic Thermometer
Classification Information: Product Code: FLL- Clinical Electronic Thermometer Device Class: II Panel: 80 Regulation number: 880.2910
## III.Predicate Device Information:
Infrared Ear Thermometer ET-101A, K082192 The predicate device is manufactured by Sejoy Electronics & Instruments Co., Ltd.
#### IV. Device Description:
The Infrared Ear Thermometers, models ET-101D and ET-101H are electronic thermometers using an infrared detector (thermopile detector) to detect body temperature from the auditory canal. The thermometer's operation is based on measuring the natural thermal radiation emanating from the tympanic membrane and the adjacent surfaces of the patient.
Both thermometer models include probes, which are used to measure ear canal temperature, plastic enclosures enclosing the display window as well as buttons and a battery cover. The disposable probe cover of ET-101H is optional when measuring temperature while ET-101D does not utilize a disposable probe cover.
The Infrared Ear Thermometers measure temperature by reading infrared radiation emitting from the eardrum tissue. The small cone-shape end of the thermometer is inserted into the ear canal, where the eardrum (tympanic membrane) and surrounding tissues give off heat. The thermometer converts the heat into a temperature value using software.
{4}------------------------------------------------
## V. Intended use / Indication for Use:
Infrared Ear Thermometers, models ET-101D and ET-101H are indicated for the intermittent measurement and monitoring of human body temperature by consumers in a home use environment. It is intended for use on people of all ages.
| SE Comparisons | SUBJECT DEVICE : | PREDICATE DEVICE: | Comparison Result | Remark |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|-------------------|---------------------------------------|
| | Infrared Ear Thermometer<br>Model:ET-101D、ET-101H<br>Manufacturer:<br>SEJOY ELECTRONICS &<br>INSTRUMENTS CO., LTD | Infrared Ear Thermometer<br>ET-101A (K082192)<br>Manufacturer:<br>SEJOY ELECTRONICS &<br>INSTRUMENTS CO., LTD | Identical | / |
| Thermometer type | Infrared Ear Thermometer<br>Product code: FLL | | Identical | / |
| Measuring Site: | Ear canal | | Identical | / |
| Reference Body<br>Site: | Oral (This thermometer converts the ear temperature to display<br>its "oral equivalent") | | Identical | / |
| Intended use(s) | Indicated for the intermittent measurement and monitoring of human body<br>temperature by consumers in a home use environment.It's intended for use on<br>people of all ages. | | Identical | / |
| Operation mode | Ear mode(Adjusted mode) | | Identical | / |
| Components<br>& Materials | ET-101H<br>Case: ABS<br>Probe: Stainless steel & ABS<br>Probe Cover: PE<br><br>ET-101D<br>Case: ABS<br>Probe: Stainless steel & ABS | Case: ABS<br>Probe: Stainless steel & ABS<br>Probe Cover: PE | Identical | ET-101D do<br>not have<br>probe cover |
| Sensor | Infrared Sensor | | Identical | / |
| Signal processing<br>and display | Transform the analog signals to digital signals, then displays on the LCD | | Identical | / |
## VI. Comparison of Technological Characteristics with the Predicate Device
{5}------------------------------------------------
| Measurement range | ET-101D | $32.0°C~43.0°C (89.6°F~109.4°F)$ | $32.0°C~43.0°C (89.6°F~109.4°F)$ | Similar | Meet the<br>requirement<br>of ASTM<br>E1965-98 |
|----------------------------------------------|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------|
| | | Ear mode:<br>$28.0°C~43.0°C (82.4°F~109.4°F)$ | | | |
| | ET-101H | Object mode:<br>$-20°C~100°C(-4°F~212°F)$<br>Room temperature:<br>$0°C~50°C (32°F~122°F)$ | | | |
| | Operating<br>temperature range | ET-101D | $10°C~40°C (50°F~104°F)$ , 15%~85<br>%RH, non-condensing<br>Atmospheric Pressure:<br>800hPa ~ 1060hPa | $10°C~40°C (50°F~104°F)$ , 15%~85%RH, non-condensing<br>Atmospheric Pressure:<br>800hPa ~ 1060hPa | Identical |
| ET-101H | | Ear/Object mode:<br>$10°C~40°C(50°F~104°F)$ , 15%~85<br>%RH, non-condensing<br>Atmospheric Pressure:<br>800hPa ~ 1060hPa | | | |
| Storage and<br>transportation<br>environment | | | $-25°C~55°C(-13°F~131°F)$ , 15%~95%RH,non-condensing<br>Atmospheric Pressure:800hPa ~ 1060hPa | | Identical |
| Precision<br>(Laboratory accuracy) | ET-101D | $±0.2°C (0.4°F)$ during $35.0°C~42.0°C (95.0°F~107.6°F)$ at<br>$15°C~35°C (59.0°F~95.0°F)$<br>Operating temperature range<br>$±0.3°C (0.5°F)$ for other measuring<br>and operating<br>temperature range | $±0.2°C (0.4°F)$ during $35.0°C~42.0°C (95.0°F~107.6°F)$ at | | |
| | | | | | |
| | | | | | |
| | | | | | |
| | ET-101H | Ear mode:<br>$±0.2°C$ ( $0.4°F$ ) during operating temperature range<br>$35.5°C~42.0°C$ ( $95.9°F~107.6°F$ ) at $15°C~35°C$ ( $59.0°F~95.0°F$ ) operating temperature range<br>$±0.3°C$ ( $0.5°F$ ) for other measuring and operating temperature range<br>Object mode:<br>$±4%$ or $±2°C$ ( $4°F$ ) whichever is greater<br>Room temperature: $±2°C$ ( $±4°F$ ) | $15°C~35°C$ ( $59.0°F~95.0°F$ ) $±0.3°C$ ( $0.5°F$ ) for other measuring and operating temperature range | Similar | / |
| Biocompatibility | | Complied with the biocompatible requirements of FDA. | | Identical | / |
| | ET-101D | No accessory | Storage Case,Probe Cover | Different | / |
| Accessories | ET-101H | Storage Case, Probe Cover | Storage Case, Probe Cover | Identical | / |
| Response time | | Approximately 1 Second | | Identical | / |
| | ET-101D | Approx. 1 year/3000 readings | Approx. 1 year/3000 readings | Identical | / |
| Battery Life | ET-101H | Approx. 1 year/6000 readings | Approx. 1 year/3000 readings | Similar | / |
| Clinical Accuracy | ET-101D | Clinical bias: $0.07°C$ ( $0.1°F$ )<br>Clinical repeatability: 0.11( $0.2°F$ )<br>Limits of agreement: $0.76°C$ ( $1.4°F$ ) | Clinical bias: $0.12°C$ ( $0.2°F$ )<br>Clinical repeatability:<br>0.12( $0.2°F$ )<br>Limits of agreement<br>: $0.8°C$ ( $1.4°F$ ) | Similar | / |
| | ET-101H | Clinical bias: $0.08°C$ ( $0.14°F$ )<br>Clinical repeatability: 0.13( $0.23°F$ )<br>Limits of agreement: $0.76°C$ ( $1.31°F$ ) | | | |
| | ET-101D | One CR2032 battery | One CR2032 battery | Different | / |
| Power Sources | ET-101H | Two AAA batteries | One CR2032 battery | Different | / |
{6}------------------------------------------------
| Probe Cover | ET-101D | NO | YES | Different | / |
|------------------------------------------------|---------|----------------------------------------------------------|---------|-----------|---|
| | ET-101H | Optional | YES | Similar | / |
| Low Battery Detection | | YES | | Identical | / |
| Automatic Power-Off | | YES | | Identical | / |
| Beep Alarm | | YES | | Identical | / |
| °C/°F Switchable | | YES | | Identical | / |
| Object temperature<br>mode | ET-101D | NO | NO | Identical | / |
| | ET-101H | YES | NO | Different | / |
| Ambient temperature<br>mode | ET-101D | NO | NO | Identical | / |
| | ET-101H | YES | NO | Different | / |
| Real time clock | ET-101D | NO | NO | Identical | / |
| | ET-101H | YES | NO | Different | / |
| Room Temperature | ET-101D | NO | NO | Identical | / |
| | ET-101H | YES | NO | Different | / |
| Backlight function | ET-101D | One color | NO | Different | / |
| | ET-101H | Three colors in ear mode and one<br>color in object mode | NO | Different | / |
| Memory | ET-101D | 10 sets | 10 sets | Identical | / |
| | ET-101H | Each 10 sets memories for ear<br>and object measurements | 10 sets | Similar | / |
| Probe cover<br>automatic detection<br>function | ET-101D | NO | NO | Identical | / |
| | ET-101H | YES | NO | Different | / |
#### Summary of technological characteristics / performance
The Infrared Ear Thermometer, Model ET-101D, ET-101H have substantially equivalent indications for use and technological characteristics as the predicate device.
- 1) ET-101D,ET-101H have the same intended use/indication for use as the predicate device.
- 2) The same technologies and test principles are used in the ET-101D and ET-101H thermometers as the predicate device.
- 3) The identified differences in technological characteristics do not raise new or different questions of safety and effectiveness than the predicate devices.
- 4) Although some specifications are slightly different than the predicate device, ET-101A, changes have been verified and validated as part of performance testing and safety /EMC testing and are included as part of this submission. Performance information and evidence of compliance to recognized standards demonstrate the device is substantially equivalent to the predicate devices.
Based upon the intended use, and upon the similarity of materials, product configuration and administration, it can be concluded the Infrared Ear Thermometer, Model ET-101D and ET-110H devices are substantially equivalent to the identified predicate device.
## VII. Performance Data
The following performance data were provided in support of the substantial equivalence determination:
Performance testing was conducted to validate and verify that Infrared Electronic Thermometers, models ET-101D and ET-101H met all requirements of related international standards, including electrical safety, EMC, biocompatibility, software validation and product specifications. Results of these tests demonstrate compliance to the requirements of the below consensus standards.
{7}------------------------------------------------
Electrical Safety and performance requirements:
- AAMI/ANSI ES60601-1:2055/(R)2012 and C1:2009/(R)2012 and A2:2010/(R)2012 Medical Electrical Equipment
- . ISO 80601-2-56:2099 Medical electrical equipment Part 2-56 Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement
- ASTM E1965-98(Reapproved 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature
- ASTM E1104-98(Reapproved 2009) Standard Specification for clinical thermometer probe . covers and sheaths
Home-used medical equipment requirements and environmental test:
- IEC 60601-1-11:2010 General requirements for basic safety and essential performance --● Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
Electromagnetic compatibility requirements:
- IEC 60601-1-2:2007 Medical electrical equipment Part 1-2: General requirements for basic ● safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests
Biocompatibility Evaluation for patient contacting components:
- ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro ● cytotoxicity
- ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests for irritation and . skin sensitization
Guidance Document:
- Guidance on the content of Premarket Notifications [510(k)] Submissions for clinical electronic ● thermometers
The software/firmware verification and validation was provided in accordance with the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
Discussion of Animal Tests Performed: No animal testing was performed.
Discussion of Clinical Tests Performed:
Clinical evaluation of Infrared Ear Thermometers was conducted by Sejoy Electronics & Instruments Co., Ltd in compliance with ISO 80601-2-56 and ASTME1965.
## VIII. Conclusions:
The non-clinical data support the safety of the device and the hardware and software verifications and validation demonstrate that the Infrared Ear Thermometers, models: ET-101D and ET-101H should perform as intended in the specified use conditions. Based on the information provided in this submission, the Infrared Ear Thermometers, models: ET-101H are substantially equivalent to the predicate thermometer ET-101A.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.