K153143 · Terumo Cardiovascular Systems Corporation · DTN · Dec 2, 2015 · Cardiovascular
Device Facts
Record ID
K153143
Device Name
Capiox Advance Hardshell Reservoir
Applicant
Terumo Cardiovascular Systems Corporation
Product Code
DTN · Cardiovascular
Decision Date
Dec 2, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Capiox® Advance Hardshell Reservoir is a hardshell reservoir used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal. The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return blood to the patient for blood volume replacement. The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass. The Hardshell Reservoir contains X-Coating, which is intended to reduce platelet adhesion on the surfaces of the device. The device may be used for procedures lasting up to 6 hours.
Device Story
Hardshell reservoir for extracorporeal circulation; stores venous and cardiotomy blood during/after cardiopulmonary bypass. Inputs: venous blood via gravity or vacuum; cardiotomy blood via suction. Processing: blood passes through internal defoamers to remove air bubbles and filters to remove particulates. Output: filtered blood returned to bypass circuit or patient. Used in OR/clinical settings by perfusionists/clinicians. Features X-Coating to reduce platelet adhesion. Benefits: facilitates air removal, particulate filtration, and blood volume management during bypass and autotransfusion.
Clinical Evidence
Bench testing only. No clinical data required. Performance evaluations included air handling, hemolysis, venous filter dynamic hold-up, pressure drop, vacuum integrity, burst/leak/circulation/pull testing, pre-connect evaluation, and package testing.
Indicated for patients undergoing cardiopulmonary bypass procedures requiring extracorporeal circulation, post-operative chest drainage, and autotransfusion for blood volume replacement. Suitable for use with vacuum-assisted venous return. May be used for procedures up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Terumo's FX15 Advance Oxygenator with Integrated Filter and Hardshell Reservoir (K151389)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, layered on top of each other. The profiles are black and have a flowing, abstract design. The logo is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in black, uppercase letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 02, 2015
Terumo Cardiovascular Systems Corporation Tierra Brown Regulatory Affairs Specialist 125 Blue Ball Road Elkton, Maryland 21921
Re: K153143
Trade/Device Name: Capiox Advance Hardshell Reservoir Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN, DTP Dated: October 22, 2015 Received: October 30, 2015
Dear Tierra Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
piox Advance Hardshell Reser U CLION 4 - Indications for U TERUM
10(li) Number (f known); Unknown at time of Submission
# Device Name: Capiones Advance Hardshall Reservo
# indications for Use:
സ്വന്തമാക്കും വിശ്വാസ്ത്രാക്കു വിവേശം നിന്നും നിന്നുവെന്നും അവന്നും കാവന്യത്.
പിരുന്നും വിവ്യ നിരക്കു വിമ്പേ ബന്ധം നിന്നും നിന്നുവെന്നും അവന്മാനും നിന്നും വല
ുന്നു. പിന്നു ന
olsunan bir əsasən qalınmışdır. Əhli olan qalınmışı ilk olan çıxır. Bu mərkəzi ilə bir mərkəzi ilə bir mərkəzi ilə bir mə
isquinonary bypasss Hardshell Reservoir is also used with the vacuum-assisted venous return technique during
surfaces of the device no noised bel a service of bebruari si doine si instrument of Services Nellebresses in
he is nay be used for proceedures lasting up to 6 hours
Prescription Use 21 CFR 801 Subpart D XX AO
as and results in the program in the first of O
SA NEEDED) F
oncurrence of CDRRH, Office of Device Evaluation (ODE
21
{3}------------------------------------------------
### TERUMO® SECTION 5 – 510(k) Summary Capiox® Advance Hardshell Reservoir
| Submitter Information | |
|-----------------------------------|-------------------------------------------|
| Date of Preparation | October 22, 2015 |
| Name | Terumo Cardiovascular Systems Corporation |
| Address | 125 Blue Ball Road<br>Elkton, MD 21921 |
| Establishment Registration Number | 1124841 |
| Name of Contact Person | Tierra A. Brown, MSRA, RAC |
| Contact Person's Title | Regulatory Affairs Specialist |
| Phone Number | (800) 262-3304 (x7021) |
| Fax Number | 410-392-7171 |
| Email | tierra.brown@terumomedical.com |
| Device Information | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification Type | Cardiopulmonary Bypass Blood Reservoir (Code DTN)<br>Cardiopulmonary Bypass Defoamer (Code DTP) |
| Device Common Name | Blood Reservoir<br>Defoamer |
| Device Trade Name | Capiox® Advance Hardshell Reservoir |
| Predicate Device | Capiox® RX Hardshell Reservoir; Cleared by FDA with<br>K130359 |
| Classification Regulation | Cardiopulmonary Bypass Reservoir 21 CFR § 870.4400<br>Cardiopulmonary Bypass Defoamer 21 CFR § 870.4230 |
| Device Risk Classification | Cardiopulmonary Bypass Reservoir - Class II<br>Cardiopulmonary Bypass Defoamer – Class II |
| Reason for 510(k) Submission | Modification to Hardshell Reservoir |
| Substantial Equivalence Statement | The device submitted in this 510(k) maintains<br>characteristics that are substantially equivalent in<br>intended use, design, technology/principles of operation,<br>materials and specifications to the predicate device. |
{4}------------------------------------------------
# Capiox® Advance Hardshell Reservoir
### Intended Use:
The intended use remains the same as the intended use for the predicate device cleared with submission K130359. There have been no changes to the indications or intended use of the subject device as a result of the modifications.
The Capiox® Advance Hardshell Reservoir is a hardshell reservoir used to store blood during extracorporeal circulation from both the venous line and the cardiotomy line. The reservoir contains filters to remove particulate matter and defoamers to facilitate air bubble removal.
The Hardshell Reservoir is also used for post-operative chest drainage and autotransfusion procedures to aseptically return blood to the patient for blood volume replacement.
The Hardshell Reservoir is also used with the vacuum-assisted venous return technique during cardiopulmonary bypass.
The Hardshell Reservoir contains X-Coating, which is intended to reduce platelet adhesion on the surfaces of the device.
The device may be used for procedures lasting up to 6 hours.
### Principles of Operation and Technology:
The principles of operation and technological characteristics remain the same as that of the predicate device cleared with submission K130359. There have been no changes to the fundamental scientific technology of the subject device.
The Capiox® Advance Hardshell Reservoir device and the predicate Capiox® RX Hardshell Reservoir device are both used as a blood storage devices during and after cardiopulmonary bypass procedures. Venous blood enters the reservoir via gravity, or by way of external vacuum that may be applied to the reservoir.
Venous blood that is drawn from the patient enters the device via the venous blood inlet port. The blood passes through a defoamer to remove air bubbles from the blood and through a filter to remove particulates from the blood.
Blood may also be suctioned into the reservoir from the cardiotomy field. This blood enters the device through the cardiotomy blood suction ports. The blood passes through a defoamer to remove air bubbles from the blood and through a filter to remove particulates from the blood.
Blood exits the device via gravity through the blood outlet port and is subsequently pumped through the remainder of the cardiopulmonary bypass circuit.
### Design and Materials:
The design of the Capiox® Advance Hardshell Reservoir is comprised of a hardshell casing that serves as a blood containment system within the bypass circuit. The upper portion of the
{5}------------------------------------------------
### SECTION 5-510(k) Summary Capiox® Advance Hardshell Reservoir
reservoir consists of a hardshell lid assembly that contains the necessary inlet ports and vent ports. It contains a rotatable venous blood inlet port. The total capacity of the reservoir is 4000mL.
The venous section of the reservoir contains a filter for particulate matter removal and a defoamer to facilitate air bubble removal. The cardiotomy section of the reservoir contains a filter for particulate matter removal and a defoamer to facilitate air bubble removal.
### Differences in the design of the two Reservoirs are as follows:
- The design of the Capiox® Advance Hardshell Reservoir device will include a curved ● venous inlet port and flared venous inlet drop tubes. The design of the predicate Capiox® RX Hardshell Reservoir device has an angled venous inlet port and straight venous inlet drop tubes. The intent of the curved venous inlet port is to improve air handling performance of the port. A curved design is more conducive to smooth flow than an angled design. The intent of the flared venous inlet drop tubes modification is to help minimize air emboli generation in the venous section of the reservoir.
- o The Capiox® Advance Hardshell Reservoir device has a full-length graduated side scale label (decal). The predicate Capiox® RX Hardshell Reservoir device has a half-length graduated side scale label (decal). The full-length graduated side scale label (decal) enhancement offers greater visibility of blood volume during cardiopulmonary bypass.
- . The Capiox® Advance Hardshell Reservoir device has a minimum operating volume of The predicate Capiox® RX Hardshell Reservoir device has a minimum 150mL. operating volume of 200mL. The minimum operating volume is lowered to 150mL to facilitate lower blood volumes during cardiopulmonary bypass.
### Difference in the materials of the two Reservoirs are as follows:
The materials used in the construction of the Capiox® Advance Hardshell Reservoir include, but are not limited to, nylon, polycarbonate, silicone, polyvinyl chloride, polyurethane, polyester, polypropylene, polymethyl methacrylate, polyhydroxyethyl methacrylate, polyethylene, and X-Coating™. PVC tubing with TOTM plasticizer has been cleared for use with Terumo's FX15 Advance Oxygenator with Integrated Filter and Hardshell Reservoir, K151389.
- The material used in the construction of the Capiox® Advance Hardshell Reservoir o device's venous inlet drop tubes and auxiliary tubing (blood-contacting) will be PVC tubing with Tris-(2-Ethylhexyl) Trimellitate, identified as TOTM plasticizer'. The material used in the construction of the predicate Capiox® RX Hardshell Reservoir device's venous inlet drop tubes and auxiliary tubing (blood-contacting) is PVC tubing with Di-(2-ethylhexyhl) phthalate, identified as DEHP plasticizer.
- . The intent of the PVC tubing plasticizer change to TOTM plasticizer is to address growing potential safety concerns associated with DEHP and to address increasing concerns related to the use of DEHP.
Terumo Cardiovascular Systems concludes that the differences between the subject device and the predicate device do not affect the intended use of the device nor do they affect safety and effectiveness of the device when used as labeled.
<sup>1</sup> PVC tubing with Tris-(2-Ethylhexyl) Trimellitate, identified as TOTM plasticizer, is used with Terumo's Capiox® FX15 Advance Oxygenator with Integrated Filter and Hardshell Reservoir, K151389.
{6}------------------------------------------------
## Capiox® Advance Hardshell Reservoir
### Performance Evaluations:
Clinical studies involving patients are not necessary to demonstrate substantial equivalence of the Capiox® Advance Hardshell Reservoir to the predicate device. Substantial equivalence is demonstrated with the following in-vitro performance evaluations:
- Air Handling Performance ●
- Hemolysis .
- Venous Filter Dynamic Hold Up and/or Pressure Drop .
- Vacuum Testing (Product Integrity, Large Venous Port-O-Ring Seal) .
- Burst, Leak, Circulation, and/or Pull Testing to Evaluate Venous Port and Tubing ● Connection Integrity
- Pre-Connect Evaluation ●
- . Package Testing
### Substantial Equivalence Comparison:
In demonstrating substantial equivalence of the Capiox® Advance Hardshell Reservoir to the predicate Capiox® RX Hardshell Reservoir, Terumo has considered the following aspects to ensure equivalence between the predicate and the subject devices:
- Intended use .
- . Target Population
- Duration of use ●
- Product labeling .
- Product design ●
- Materials .
- Principles of Operation and Technology ●
- Device Performance .
### Substantial Equivalence Statement:
The Capiox® Advance Hardshell Reservoir is substantially equivalent to the predicate Capiox® RX Hardshell Reservoir in intended use, duration of use, design, materials, principles of operation and technology, and performance and specifications. Any noted differences between the subject device and the predicate device do not raise new issues of safety and/or effectiveness.
### Conclusion:
Based upon the comparative studies and analyses, Terumo Cardiovascular Systems concludes that the Capiox® Advance Hardshell Reservoir is substantially equivalent in intended use, duration of use, design, materials, principles of operation and technology, and performance and specifications to the predicate Capiox® RX Hardshell Reservoir. It is further concluded that any recognized differences noted during the assessments do not raise new issues of patient/user safety or product effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.