Cerdak Basic

K153138 · Bu Polysciences Equipment & Trading, LLC · FRO · Jun 23, 2017 · SU

Device Facts

Record IDK153138
Device NameCerdak Basic
ApplicantBu Polysciences Equipment & Trading, LLC
Product CodeFRO · SU
Decision DateJun 23, 2017
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

For the management of the following: Minor Cuts Minor Abrasions Minor Lacerations Minor Burns (1st degree burns)

Device Story

Cerdak is a single-use, sterile, biocompatible wound dressing consisting of aluminum oxide ceramic granules contained within a non-woven fabric sachet. It is available in various configurations (Basic, Aerofilm, Aerocloth, Drydoc) and sizes. The device is applied topically to cover wounds and maintain a moist wound environment. It is intended for over-the-counter use by patients or caregivers for minor skin injuries. The dressing functions by providing a physical barrier and moisture management. It is not for use on dry wounds.

Clinical Evidence

Clinical study compared Cerdak to normal saline control. Results showed no statistically significant difference in infection rates, infection types, or complications between groups. Cerdak was found safe for covering wounds and maintaining a moist environment. Additional bench testing included biocompatibility (cytotoxicity, skin irritation, sensitization) per ISO 10993-5 and GPMT, confirming non-cytotoxic and non-sensitizing properties.

Technological Characteristics

Non-woven fabric sachet containing micro-porous aluminum oxide ceramic granules. Dimensions range from 15mm x 25mm to 100mm x 250mm. Biocompatibility testing performed per ISO 10993-1, ISO 10993-5, and ANSI/AAMI/ISO 10993. Sterile, single-use device.

Indications for Use

Indicated for management of minor cuts, abrasions, lacerations, and 1st degree burns. Contraindicated for dry wounds.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 23, 2017 BU Polysciences Equipment & Trading, LLC N. Umamaheswaran President 10855 Symphony Park Dr. North Bethesda, Maryland 20852 Re: K153138 Trade/Device Name: Cerdak Basic Regulatory Class: Unclassified Product Code: FRO Dated: May 16, 2017 Received: May 19, 2017 Dear N. Umamaheswaran: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # David Krause -S for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ See PRA Statement Below #### Indications for use 510(K) Number (if known) K153138 Device Name | CERDAK™ | |--------------------------------------| | Indications for Use (Describe) | | For the management of the following: | | Minor Cuts | | Minor Abrasions | | Minor Lacerations | | Minor Burns (1st degree burns) | Type of Use (Select one or both, as applicable) ______________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. X This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Page no Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to, a collection of information unles it displays a currently valid OMB number." {3}------------------------------------------------ 510(k) summary as described in § 807.92 | Submitter: | BU Polysciences Equipments & Trading, LLC<br>10855 Symphony Park Dr.<br>N. Bethesda, MD 20852<br>Phone: 301 309 1210 (Office)<br>301 325 5574 (Mobile) | |-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | N. T. Umamaheswaran (Uma) President,<br>BU Polysciences Equipments & Trading, LLC<br>Phone: 301 325 5574 (Mobile)<br>bupet2015@gmail.com | | Manufacturer: | CARBORUNDUM UNIVERSAL LIMITED<br>655 T.H.ROAD<br>THIRUVOTTIYUR | CHENNAI 600019 INDIA | Preparation Date: | June 19, 2017 | |--------------------------|-------------------| | Proprietary Device name: | CERDAK™ (K153138) | | Common Name | : DRESSING | #### Device Classification: | Product Code | : FRO (Dressing) | |------------------|------------------| | Regulatory Class | : Unclassified | - Predicate Device : WoundStat Wound Dressing (K071936) #### Description: Cerdak™ is a single-use, sterile, bio-compatible device made of Aluminum Oxide ceramic granules in a non-woven fabric sachet. Cerdak™ dressing is provided in four configurations, they are: Basic, Aerofilm, Aerocloth and Drydoc and in several sizes ranging from 15mm X 25mm to 100mm X 250 mm. CERDAK™ covers and provides moist wound environment. It is contra-indicated for dry wounds. {4}------------------------------------------------ ## Indications for use: For the management of the following: Minor Cuts Minor Abrasions Minor Lacerations Minor Burns (1st degree burns) # TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE: The purpose of this 510(k) is to obtain clearance to market CERDAK™ as a wound dressing. CERDAK™ is substantially equivalent to the predicate device with respect to design and intended use. We have chosen WoundStat Wound Dressing (K071936) as our primary predicate device. # COMPARISON OF TECHNOLOGICAL CHARACTERISTICS: CERDAK™ is a non-woven fabric sachet filled with micro-porous ceramic granules that covers and provides moist wound environment CERDAK™ is similar to WoundStat Wound Dressing (K071936) which is a clay based, granular haemostatic agent. WoundStat is offered in a moisture-impermeable, sterile, 5.5-ounze foil pouch. ## STATEMENT OF SUBSTANTIAL EQUIVALENCE: CERDAK™ is substantially equivalent in the intended use to WoundStat (K071936) CERDAK™ has been evaluated for its biocompatibility which meets requirements as stipulated in ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. CERDAK™ has the same intended use as the primary predicate device as they are both intended to cover a wound. Data presented in this submission substantiates equivalence of CERDAK™ to the predicate device. ## PERFORMANCE TESTING: CERDAK™ dressing has been subjected to testing to assess the Biocompatibility, as well as the physical performance of the device. CERDAKTM was evaluated through in-vitro and animal safety studies. {5}------------------------------------------------ Safety tests and the conclusions drawn for the test articles: CERDAK TM was evaluated through in-vitro and animal safety studies. Safety tests and the conclusions drawn for the test articles: 1. Cytotoxicity: (Balb/c/3T3/ cells in-vitro condition using elution method) CERDAK TM is non-cytotoxic and meets the requirements of the Test defined in ISO 10993-5 guidelines. 2. Primary Skin Irritation Test: Irritation (New Zealand white rabbits) CERDAK™ is considered a negligible irritant after a single topical application to the skin of New Zealand White rabbits. 3. Skin Sensitization :( Guinea Pig Maximization Test - GPMT) No mortality and morbidity were observed in any of the animals used in the study. No skin sensitization reactions were observed in both control and test sites of all the animals. CERDAK™ was found to be non-sensitizer. The biocompatibility of CERDAK™ has been demonstrated through appropriate in vivo and in vitro tests. The product has been assessed in accordance with ANSI/AAM VISO 10993 and does not introduce any additional safety risk. The bench testing includes visual inspection, weight and thickness of final product packaging testing. ## CLINICAL STUDY: A clinical study was conducted on CERDAK™ using normal saline as control. Results of culture and sensitivity test showed there was no statistically significant difference in infection rate and infection type between both the groups. There was no statistical difference in complications among both the groups. CERDAK™ was found to be safe in covering the wounds and creating a moist wound environment.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...