K153111 · Adin Dental Implants Systems , Ltd. · DZE · May 6, 2016 · Dental
Device Facts
Record ID
K153111
Device Name
TOUAREG CloseFit™ UNP 2.75mmD
Applicant
Adin Dental Implants Systems , Ltd.
Product Code
DZE · Dental
Decision Date
May 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
TOUAREG CloseFit™ UNP 2.75mmD implants are indicated to replace missing masticatory functional units (teeth) in single or multiple unit applications within the mandible or maxilla. For single-stage or two-stage procedures. For immediate implantation in extraction sites or implantation in partially healed or completely healed alveolar ridge situations. When an one-stage surgical is applied, the implant maybe immediately loaded when good primary stability is achieved and thefunctional load is appropriate. The TOUAREG CloseFit™ UNP 2.75mmDDental Implant shall only be used to replace maxillary lateral incisors and mandibular lateral and central incisors.
Device Story
TOUAREG CloseFit™ UNP 2.75mmD is a two-piece endosseous dental implant system; features tapered core, spiral tap, and triple lead thread design. Used by dental professionals in clinical settings for tooth replacement. Implant is surgically placed into mandible or maxilla; used with cover screws, healing abutments, and abutments. Provides a conical-hex connection. Benefits include restoration of masticatory function. Clinicians use the device to support crowns, bridges, or overdentures. Output is the physical implant structure serving as an artificial root.
Clinical Evidence
No clinical performance data is provided in this submission. Bench testing only (fatigue testing per ISO 14801:2007).
Technological Characteristics
Material: Ti6Al4V ELI (ASTM F 136-08). Design: Tapered core, spiral tap, triple lead thread (3x1mm). Connection: Conical-hex. Surface: OsseoFix™ and anodized. Dimensions: 2.75mm diameter; 10, 11, 13, 16, 18mm lengths. Sterilization: Sterile.
Indications for Use
Indicated for patients requiring replacement of missing teeth in mandible or maxilla. Specifically for maxillary lateral incisors and mandibular lateral and central incisors. Suitable for single or multiple unit applications, single or two-stage procedures, and immediate or delayed loading if primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Adin Dental Implant Systems Ltd. Touareg CloseFit™ NP (K140293)
Adin Dental Implant Systems Ltd. Touareg CloseFit™ (K112585)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three overlapping profiles facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 6, 2016
Adin Dental Implants Systems Ltd. c/o Iman Khorshid CEO, Founder ORS Industrial Park Tefen Tefen 2495900 ISRAEL
Re: K153111
Trade/Device Name: TOUAREG CloseFit™ UNP 2.75mmD Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: March 31, 2016 Received: April 6, 2016
Dear Iman Khorshid:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153111
Device Name TOUAREG CloseFit™ UNP 2.75mmD
#### Indications for Use (Describe)
TOUAREG CloseFit™ UNP 2.75mmD implants are indicated to replace missing masticatory functional units (teeth) in single or multiple unit applications within the mandible or maxilla. For single-stage or two-stage procedures. For immediate implantation in extraction sites or implantation in partially healed or completely healed alveolar ridge situations. When an one-stage surgical is applied, the implant maybe immediately loaded when good primary stability is achieved and thefunctional load is appropriate.
The TOUAREG CloseFit™ UNP 2.75mmDDental Implant shall only be used to replace maxillary lateral incisors and mandibular lateral and central incisors.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for ADIN Dental Implant Systems LTD. The logo features a stylized letter "A" inside of a circle, followed by the text "ADIN" in a bold, sans-serif font. Below the text "ADIN" is the text "DENTAL IMPLANT SYSTEMS LTD" in a smaller, sans-serif font.
# 510(k) Summary:
## TOUAREG CloseFit™ UNP 2.75mmD Implants
Company Name: Adin Dental Implants Systems Ltd. Alon Tavor Industrial Zone P.O.Box 1128, Afula Hazafon, 1811101 ISRAEL Telephone: + 972-4-6426732 Fax: +972-4- -6426733
#### Establishment Registration Number: 3007518363
| Contact Name: | Iman Khorshid – Regulatory Consultant<br>QRS |
|---------------|----------------------------------------------|
| | Mobile: +972-54-3023043 |
| | Telephone: + 972-4-6860006 |
| | Fax: +972-4-6860006 |
| | E-mail: iman@qrs-global.com |
US Agent: Markus Weitz Adin Implants USA 657 Central Ave Cedarhurst , NY 11516 Phone: 212 7642900 212 7642349 Fax: E-mail: info@adinimplants.com
Date prepared: May 5, 2016
Trade Name: TOUAREG CloseFit™ UNP 2.75mmD
Classification name: Endosseous Dental Implant
Common/usual name: Dental Implant
Product Code: DZE; NHA
Regulation No.: 872.3640
Class: II
Panel identification: Dental Devices Panel
510 (k) TOUAREG CloseFit™ UNP 2.75mmD Implants Section E - Page 1 of 5
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for ADIN Dental Implant Systems LTD. The logo features a stylized letter "A" inside of a blue circle, followed by the text "ADIN" in a bold, sans-serif font. Below the text "ADIN" is the text "DENTAL IMPLANT SYSTEMS LTD" in a smaller, sans-serif font.
# Predicate Devices:
Primary Predicate: Adin Dental Implant Systems Ltd. dental implants Touareg CloseFit™ NP, cleared under 510(k) K140293.
Reference Predicates: Anthogyr SAS' Axiom 2.8 implant cleared under 510(k) K141450 and Adin Dental Implant Systems Ltd. dental implants Touareg CloseFit™, cleared under 510(k) K112585.
# Description of the device:
The TOUAREG CloseFit™ UNP 2.75mmD spiral implant system is comprised of a tapered core implant with a spiral tap, with unique tip and triple lead thread design (3x1mm).
The TOUAREG CloseFit™ UNP 2.75mmD spiral implant system offers a unique, strong and solid conical-hex connection.
All Touareg CloseFit™ implants feature the OsseoFix™ surface treatment The TOUAREG CloseFit™ UNP 2.75mmD implant is provided with a diameter of 2.75mm and lengths of 10, 11, 13, 16 and 18 mm.
The TOUAREG CloseFit™ UNP 2.75mmD implants are two piece devices whereas the implant is to be used in combination with cover screws, healing abutments and abutments.
The TOUAREG CloseFit™ UNP 2.75mmD implants, abutments, and abutment fixation screw are made of Ti6AL4V ELI complying with standard ASTM F 136-08- Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant.
510 (k) TOUAREG CloseFit™ UNP 2.75mmD Implants Section E - Page 2 of 5
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for "ADIN DENTAL IMPLANT SYSTEMS LTD". The logo features a stylized letter "A" inside a blue circle on the left, followed by the word "ADIN" in large, bold, dark blue letters. Below the word "ADIN" is the text "DENTAL IMPLANT SYSTEMS LTD" in a smaller font size. There is also a curved line above the word "ADIN".
## Indications for Use:
TOUAREG CloseFit™ UNP 2.75mmD implants are indicated to replace missing masticatory functional units (teeth) in single or multiple unit applications within the mandible or maxilla. For single-stage or two-stage procedures. For immediate implantation in extraction sites or implantation in partially healed or completely healed alveolar ridge situations. When a one-stage surgical approach is applied, the implant maybe immediately loaded when good primary stability is achieved and thefunctional load is appropriate.
The TOUAREG CloseFit™ UNP 2.75mmDDental Implant shall only be used to replace maxillary lateral incisors and mandibular lateral and central incisors.
### Substantial Equivalence:
### Technological Characteristics
The TOUAREG CloseFit™ UNP 2.75mmD implants have the same intended use as Adin Dental Implant Systems Ltd. TOUAREG NP CloseFit™ Dental Implants, cleared under 510(k) K140293 with the exception of the elimination of the specific indications for the TMA Abutment cleared in the primary predicate K140293 but is not included in the current submission device. Further the indications for the submission device are identical to the reference predicate K141450. The material used for the TOUAREG CloseFit™ UNP 2.75mmD dental implants and abutment (Ti6AL4V ELI), as well as the surface treatment and other manufacturing methods, are identical to Adin's TOUAREG NP CloseFit™ Dental Implants as well as previously cleared Adin's TOUAREG CloseFit™ Dental Implants, cleared under 510(k) K112585. The sterility status of the implant bodies and abutments are identical to Adin's previously cleared TOUAREG NP CloseFit Dental Implants and the previous sterilization validation supports the sterility of the current submission. The available abutment designs are similar to the applicant's own previously cleared dental abutments and the maximum angulation of the abutment is the same as the applicant's own predicate devices as well as supported for substantial equivalence by fatigue test results.
The main differences between the submission device and the primary predicate 510 (k) TOUAREG CloseFit™ UNP 2.75mmD Implants Section E - Page 3 of 5
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for ADIN Dental Implant Systems LTD. The logo features a stylized letter "A" inside of a circle, followed by the text "ADIN" in a bold, sans-serif font. Below the text "ADIN" is the text "DENTAL IMPLANT SYSTEMS LTD" in a smaller, sans-serif font.
are the reduction in implant body diameter for the submission device from 3.0mm to 2.75mm and the inclusion of an 18mm length implant body. The reference predicate Axiom 2.8 Implant K141450 supports the reduction in implant body diameter. The reference predicate K112585 supports the inclusion of a longer implant body.
| | Touareg CloseFitTM<br>UNP Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd. | Touareg NP<br>CloseFitTM Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd. | Axiom® 2.8<br>Implants from<br>Anthogyr SAS | Touareg CloseFitTM<br>Dental Implants<br>from Adin Dental<br>Implant Systems<br>Ltd. |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Predicate | | Primary Predicate | Reference Predicate | Reference Predicate |
| 510(k)<br>number | | K140293 | K141450 | K112585 |
| Product<br>Code | DZE | DZE | DZE | DZE |
| Indications<br>For Use | TOUAREG<br>CloseFitTM UNP<br>2.75mmD implants | Intended use for<br>Touareg NP<br>CloseFitTM Dental | ANTHOGYR<br>Axiom® implants<br>are intended for use | Touareg CloseFitTM<br>Dental Implants are<br>intended for |
| Touareg CloseFit™<br>UNP Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd. | Touareg NP<br>CloseFit™ Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd. | Axiom® 2.8<br>Implants from<br>Anthogyr SAS | Touareg CloseFit™<br>Dental Implants<br>from Adin Dental<br>Implant Systems<br>Ltd. | |
| are indicated to<br>replace missing<br>masticatory<br>functional units<br>(teeth) in single or<br>multiple unit<br>applications within<br>the mandible or<br>maxilla. For single-<br>stage or two-stage<br>procedures. For<br>immediate<br>implantation in<br>extraction sites or<br>implantation in<br>partially healed or<br>completely healed<br>alveolar ridge<br>situations. When an<br>one-stage surgical<br>approach is applied,<br>the implant maybe<br>immediately loaded<br>when good primary<br>stability is achieved<br>and the functional<br>load is appropriate.<br>The TOUAREG<br>CloseFit™ UNP<br>2.75mmDDental<br>Implant shall only<br>be used to replace<br>maxillary lateral<br>incisors and<br>mandibular lateral<br>and central incisors. | Implant System:<br>_ To replace<br>missing masticatory<br>functional units<br>(teeth) in single or<br>multiple unit<br>applications within<br>the mandible or<br>maxilla.<br>_ For single-stage<br>or two-stage<br>procedures.<br>_ For immediate<br>implantation in<br>extraction sites or<br>implantation in<br>partially healed or<br>completely healed<br>alveolar ridge<br>situations. When a<br>one-stage surgical<br>approach is applied,<br>the implant may be<br>immediately loaded<br>when good primary<br>stability is achieved<br>and the functional<br>load is appropriate.<br>The Touareg NP<br>CloseFit™ Dental<br>Implant shall only<br>be used to replace<br>maxillary lateral<br>incisors and<br>mandibular lateral<br>and central incisors.<br>_ The Trans<br>Mucosal Abutment<br>(TMA) is indicated<br>for multiple-unit,<br>screw-retained | as artificial<br>root structures for<br>replacement of<br>missing teeth. They<br>can be<br>used for fixation of<br>single tooth<br>restorations.<br>ANTHOGYR<br>dental systems are<br>indicated for one-<br>stage or two stage<br>surgery. It is up to<br>the practitioner to<br>decide whether<br>immediate or<br>delayed loading is<br>most appropriate,<br>based on<br>clinical factors like<br>good primary<br>stability and<br>appropriate<br>occlusal loading.<br>Axiom® 2.8<br>implants are<br>indicated for single<br>replacement of<br>mandibular incisors<br>and lateral<br>maxillary incisors<br>in cases<br>presenting a<br>restricted<br>mesiodistal space.<br>The prosthetic<br>components of the<br>Axiom 2.8 product<br>line are<br>intended to ensure<br>support for single | surgical placement<br>in the maxillary<br>and/or mandibular<br>arch to support<br>crowns, bridges, or<br>overdentures in<br>edentulous or<br>partially edentulous<br>patients.<br>Touareg CloseFit™<br>Dental Implants<br>may be immediately<br>loaded when good<br>primary stability is<br>achieved and with<br>appropriate occlusal<br>loading. | |
| | Touareg CloseFitTM<br>UNP Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd. | Touareg NP<br>CloseFitTM Dental<br>Implants from Adin<br>Dental Implant<br>Systems Ltd.<br><br>restorations, and<br>may be used in<br>combination with<br>an implant level<br>framework design. | Axiom® 2.8<br>Implants from<br>Anthogyr SAS<br><br>crowns only. | Touareg CloseFitTM<br>Dental Implants<br>from Adin Dental<br>Implant Systems<br>Ltd. |
| Supplied<br>Sterile | Yes | Yes | Yes | Yes |
| Re-use | No | No | No | No |
| Material of<br>Construction | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI |
| Shape | Screw type | Screw type | Screw type | Screw type |
| Surface<br>Treatment | OsseoFixTM and<br>anodized | OsseoFixTM and<br>anodized | BCP® | OsseoFixTM and<br>anodized |
| Length | 10, 11.5, 13, 16 and<br>18 mm | 10, 11.5, 13 and 16<br>mm | 10, 12, and 14 mm | 8, 10, 11.5, 13, 15<br>and 18 mm |
| Thread<br>Diameter | 2.75 mm | 3.0 mm | 2.8 mm | 3.5, 4.3 and 5.0 mm |
| Abutment | Straight and up to 15° | Straight and up to 15° | Straight and up to 23° | Straight and up to 15° |
| Material of<br>Construction<br>of<br>Abutments | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI | Titanium Alloy –<br>Ti6Al4V ELI |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image is a logo for ADIN Dental Implant Systems LTD. The logo features a stylized letter "A" inside of a circle on the left side of the logo. To the right of the circle is the company name, "ADIN", in large, bold, blue letters, with the words "DENTAL IMPLANT SYSTEMS LTD" in smaller letters underneath.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is the logo for ADIN Dental Implant Systems LTD. The logo features a stylized letter "A" inside of a circle on the left, followed by the text "ADIN" in bold, dark blue letters. Below the text "ADIN" is the text "DENTAL IMPLANT SYSTEMS LTD" in a smaller font size.
# Non-Clinical Performance Data
Fatigue test was performed according to ISO 14801:2007 on the TOUAREG
CloseFit™ UNP 2.75mmD dental implants and showed equivalence to the predicate devices.
Clinical Performance Data
No clinical performance data is provided in this submission.
## Conclusion:
The evaluation of the TOUAREG CloseFit™ UNP 2.75mmD dental implants may therefore be considered substantially equivalent to its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.