The Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets is intended to deliver nutritional formula to the gastrointestinal system of a patient age infant and older who is physically unable to eat and swallow. Not for use with neonates. The feeding pump and feeding sets are intended to be used in clinical or home care settings by users ranging from laypersons to physicians.
Device Story
Enteral feeding pump delivers nutritional formula via rotary peristaltic tension loop pumping; uses disposable feeding sets. New wireless-enabled version includes communication hub and Kangaroo™ Connect Portal. Hub collects pump data via Zigbee; transmits via WiFi/cellular to cloud for remote monitoring. Used in clinical or home settings by laypersons or physicians. Clinicians access portal to monitor feeding rates and review delivery history; facilitates improved patient outcomes through remote data collection. Pump features menu-controlled OS, ultrasonic sensors for air/blockage detection, and anti-free flow valve.
Clinical Evidence
No clinical data submitted. Evidence consists of a safety assurance case, system verification and validation, wireless performance testing, human factors usability testing, electrical safety (IEC 60601-1), and electromagnetic compatibility (IEC 60601-1-2).
Indicated for delivery of nutritional formula to the gastrointestinal system of patients (infant and older) physically unable to eat/swallow. Contraindicated for neonates.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Plum 360™ Infusion system with Hospira MedNet™ / Smart Card Plug 'n' Play CE 3.0 Module (K141789)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, which is a staff with two snakes entwined around it, often associated with medicine and healthcare. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 6, 2016
Covidien Ms. Karin Desjardins Regulatory Affairs Manager 15 Hampshire Street Mansfiled, Massachusetts 02048
Re: K153074
Trade/Device Name: Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZH Dated: May 16, 2016 Received: May 17, 2016
Dear Ms. Desjardins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153074
Device Name
Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal
Indications for Use (Describe)
The Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets is intended to deliver nutritional formula to the gastrointestinal system of a patient and older who is physically unable to eat and swallow. Not for use with neonates. The feeding pump and feeding sets are intended to be used in clinical or home care settings by users ranging from laypersons to physicians.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 10pt;"> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
| | <span> <span style="font-size: 10pt;"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K153074 510(K) SUMMARY
| Manufacturer's Name: | Covidien<br>15 Hampshire Street<br>Mansfield, MA 02048 |
|------------------------------------|------------------------------------------------------------------------------------------------------------|
| Corresponding Official: | Karin Desjardins<br>Manager, Regulatory Affairs<br>Patient Monitoring & Recovery<br>Medtronic |
| Telephone Number:<br>E-mail: | (508) 452-1856<br>karin.d.desjardins@medtronic.com |
| Preparation Date: | June 28, 2016 |
| Trade Name: | Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™<br>Connect Feeding Sets and Kangaroo™ Connect Portal |
| Common or Usual<br>Name: | Enteral Feeding pump, Infusion pump |
| Classification Name and<br>Number: | Infusion Pump<br>21 CFR 880.5725<br>Class II<br>Product Code: LZH |
| Primary Predicate<br>Device: | K143263 - Kangaroo™ Connect Enteral Feeding Pump with<br>Kangaroo™ Connect Feeding Sets |
| Secondary Predicate<br>Device: | K141789 - Plum 360™ Infusion system with Hospira MedNet™<br>Smart Card Plug 'n' Play CE 3.0 Module |
#### Device Description
The Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets and Kangaroo™ Connect Portal consists of a wireless enabled enteral feeding pump and disposable enteral feeding sets that deliver formula via rotary peristaltic tension loop pumping to provide nutrition for those who do not have the ability to orally ingest food. The new Kangaroo™ Connect Enteral Feeding Pump wireless accessories include a communication hub and the Kangaroo™ Connect Portal. The communication hub collects short-range data from the pump through a Zigbee signal and sends the data through either WiFi or a cellular signal to the device cloud where the Kangaroo™ Connect Portal can access the data for monitoring. The wireless enabled pump, when used in conjunction with the wireless accessories, will provide a platform for remote data collection allowing clinicians to more efficiently monitor feeding rates and review actual history of nutritional delivery.
1
{4}------------------------------------------------
### Intended Use
The Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets is intended to deliver nutritional formula to the gastrointestinal system of a patient age infant and older who is physically unable to eat and swallow. Not for use with neonates. The feeding pump and feeding sets are intended to be used in clinical or home care settings by users ranging from laypersons to physicians.
| | Primary Predicate<br>Device | Secondary Predicate<br>Device | Subject Device |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K143263 | K141789 | K153074 |
| Device Name | Kangaroo™ Connect<br>Enteral Feeding Pump,<br>Kangaroo™ Connect<br>Enteral Feeding Sets | Plum 360™ Infusion system<br>with Hospira MedNet™ /<br>Smart Card Plug 'n' Play<br>CE 3.0 Module | Kangaroo™ Connect<br>Enteral Feeding Pump,<br>Kangaroo™ Connect<br>Enteral Feeding Sets, and<br>Kangaroo™ Connect Portal |
| Device Description | Enteral feeding pump and<br>disposable enteral feeding<br>sets | Infusion Pumps, Infusion<br>safety management<br>software & Intravascular<br>Administration Sets | Enteral feeding pump and<br>disposable enteral feeding<br>sets |
| Intended Use | The Kangaroo™ Connect<br>Enteral Feeding Pump with<br>Kangaroo Connect™ Feeding<br>Sets is intended to deliver<br>nutritional formula to the<br>gastrointestinal system of a<br>patient age infant and older<br>who is physically unable to<br>eat and swallow. Not for use<br>with neonates. The feeding<br>pump and feeding sets are<br>intended to be used in<br>clinical or home care<br>settings by users ranging<br>from laypersons to clinicians | Indicated for use in<br>parenteral, enteral and<br>epidural therapies and the<br>administration of whole<br>blood and blood products. | The Kangaroo™ Connect<br>Enteral Feeding Pump with<br>Kangaroo™ Connect Feeding<br>Sets is intended to deliver<br>nutritional formula to the<br>gastrointestinal system of a<br>patient age infant and older<br>who is physically unable to<br>eat and swallow. Not for use<br>with neonates. The feeding<br>pump and feeding sets are<br>intended to be used in<br>clinical or home care settings<br>by users ranging from<br>laypersons to physicians |
| Intent of Wireless<br>Connectivity | Not applicable - no wireless<br>features | Allows communication<br>between the Infusion<br>Pump, Hospira MedNet™<br>server and the facility's<br>communication systems. | Provide a platform for<br>remote data collection. |
| Sterility | Non Sterile feeding sets | Not known | Non Sterile feeding sets |
| | Primary Predicate<br>Device | Secondary Predicate<br>Device | Subject Device |
| Technological<br>Characteristics | The feeding sets are<br>based on peristaltic<br>pumping using a rotating<br>wheel which presses<br>against the tubing and<br>moves the fluid at a<br>controlled rate.<br>The connection to the<br>patient enteral access<br>device is an ENFit<br>Connector compliant to<br>ISO 80369-3. | Not known | The feeding sets are<br>based on peristaltic<br>pumping using a rotating<br>wheel which presses<br>against the tubing and<br>moves the fluid at a<br>controlled rate.<br>The connection to the<br>patient enteral access<br>device is an ENFit<br>Connector compliant to<br>ISO 80369-3. |
| Design (pump) | The pump incorporates a<br>menu controlled<br>operating system which<br>contains on board custom<br>software designed to<br>allow the user to set feed<br>rates and volumes as well<br>as other feeding options.<br>The device incorporates<br>ultrasonic sensors to<br>detect the air and<br>blockages in the feeding<br>set. | Not known | The pump incorporates a<br>menu controlled operating<br>system which contains on<br>board custom software<br>designed to allow the user<br>to set feed rates and<br>volumes as well as other<br>feeding options. The<br>device incorporates<br>ultrasonic sensors to<br>detect the air and<br>blockages in the feeding<br>set. |
| Rechargeable<br>Battery | lithium ion | Not known | lithium ion |
| Graphic Display | multi-color | Not known | multi-color |
| Water Ingress Rating | IP26 | Not known | IP26 |
| Design (feeding set) | The pump set incorporates 5<br>basic segments:<br>• Fluid reservoir(s), which<br>may be an attached bag<br>(500ml or 1000ml) or a<br>spike for connection to<br>a formula container<br>• Tubing from fluid<br>reservoir to pump (14<br>inch)<br>• Cassette containing<br>pump interface module<br>(peristaltic tubing)<br>• Tubing from pump to<br>patient connector (66<br>inches)<br>• Patient connector (ENFit<br>connector compliant to<br>ISO 80369-3) | Not known | The pump set incorporates 5<br>basic segments:<br>• Fluid reservoir(s), which<br>may be an attached bag<br>(500ml or 1000ml) or a<br>spike for connection to<br>a formula container<br>• Tubing from fluid<br>reservoir to pump (14<br>inch)<br>• Cassette containing<br>pump interface module<br>(peristaltic tubing)<br>• Tubing from pump to<br>patient connector (66<br>inches)<br>• Patient connector (ENFit<br>connector compliant to<br>ISO 80369-3) |
| | Primary Predicate<br>Device | Secondary Predicate<br>Device | Subject Device |
| Includes Anti-Free<br>Flow Valve | Yes | Not known | Yes |
| Materials/Chemical<br>Composition | Polyvinyl chloride (PVC)<br>• Feeding bags and caps<br>• Tubing | Not known | Polyvinyl chloride (PVC)<br>• Feeding bags and caps<br>• Tubing |
| | Silicone<br>• Peristaltic tubing<br>• Diaphragm valve | | Silicone<br>• Peristaltic tubing<br>• Diaphragm valve |
| | Polycarbonate<br>• Cassette body | | Polycarbonate<br>• Cassette body |
| | CoPolyester<br>• Patient connector | | CoPolyester<br>• Patient connector |
| | ABS<br>• Spike<br>• Tube holder and Cap | | ABS<br>• Spike<br>• Tube holder and Cap |
| | Strontium Ferrite / nylon<br>• Set ID magnets | | Strontium Ferrite / nylon<br>• Set ID magnets |
#### Substantial Equivalence Discussion
{5}------------------------------------------------
{6}------------------------------------------------
## Discussion of Differences
There are no differences between the indications for use of the predicate, K143263 and the subject device. The primary difference between the predicate Kangaroo™ Connect Enteral Feeding Pump cleared under K143263 and the proposed Kangaroo™ Connect Pump is an enabled wireless processor. To facilitate wireless capability, Covidien intends to introduce new accessories to be used in conjunction with the wireless-enabled Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets. the new accessories include a communication hub and Connect Portal. The wireless enabled pump, in conjunction with new wireless accessories, will provide a platform for remote data collection allowing clinicians to more efficiently monitor feeding rates and review actual history to help improve patient outcomes.
## Discussion of Performance Testing
A safety assurance case as recommended by the FDA quidance document, Infusion Pumps Total Product Life Cycle was provided for the Kangaroo™ Connect Enteral Feeding Pump with Kangaroo™ Connect Feeding Sets. The stated top-level claim of the assurance case is:
The Kangaroo" Connect system is acceptably safe for its intended use, within its environment of use, when being used by intended users, over the lifecvcle of the product.
The assurance case defined the device system, including the indications for use, system definition, operational description, patient populations, and use environments.
{7}------------------------------------------------
The following assurance claims were used to support the design safety goals:
- System design is acceptably safe .
- o All relevant hazards have been identified and controlled
- l System wise hazard analyses are adequate
- All identified hazards have at least one corresponding control .
- . All requirements have been implemented and verified
- o Device design ensures safe behavior
- Reliability requirements are met
- . System in use is acceptably safe
.
- Development and supporting processes are adequate ●
- Quality assurance is adequate o
- Software development is adequate o
- Risk Management process is adequate o
- . Hazard identification and mitigation are complete
- . All controls have been correctly implemented in the final device design
The assurance case also included additional goals, which were not relied on to support the design safety review, but were included to support the top level structure.
- Manufactured System is acceptably safe .
- System disposal is acceptably safe .
The following Performance Testing evidence was included in the assurance case:
- System Verification and Validation activities .
- Wireless enabled-disabled performance testing of essential performance attributes .
- o Flow Rate Accuracy
- o Downstream Occlusion Detection
- Pump Alarms (Feed Bag Empty, Supply Tube Blocked, Cassette Dislodged, Rotor o Stuck, Cassette Error)
- . Cleaning and disinfection validation and instructions
- . Human factors testing to evaluate the usability of the device
- . Electrical Safety testing in accordance with applicable IEC standards (IEC 60601-1:2005 AAMI ES 60601-1:2005, Medical electrical equipment-Part1: General requirements for basic safety and essential performance)
- Electromagnetic compatibility evaluation in accordance with applicable standards (IEC . 60601-1-2:2014, Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests.)
- Software verification and validation testing, as recommended by the FDA quidance . document Infusion Pump Total Product Life Cycle
## Clinical Testing
Clinical evaluations were not submitted and not needed.
### Conclusions
{8}------------------------------------------------
This information provided within this pre-market notification demonstrates that the Kangaroo™ Connect Enteral Feeding Pump, Kangaroo™ Connect Enteral Feeding Sets, and Kangaroo™ Connect Portal are substantially equivalent to the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.