K153059 · Minxray, Inc. · IZL · Feb 2, 2016 · Radiology
Device Facts
Record ID
K153059
Device Name
HF1202H PowerPlus Portable X-ray Equipment
Applicant
Minxray, Inc.
Product Code
IZL · Radiology
Decision Date
Feb 2, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1720
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This radiographic system is intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
Device Story
Portable high-frequency X-ray generator; includes tubehead/control, light beam collimator, mounting trunnion, exposure cord, and AC power cord. Operated by qualified physician or technician. Inputs: manual technique factor selection (kVp, mAs) via pushbuttons. Outputs: X-ray radiation for diagnostic imaging. Features serial/Bluetooth ports for transmitting technique factors to external PC; PC cannot initiate exposure. Designed for use with third-party FDA-cleared digital receptor panels. Enhances clinical workflow by providing portable diagnostic imaging capabilities; supports patient care in various clinical settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, software validation, and compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-28, 60601-2-54, IEC 62304, and IEC 62366 standards. System operation verified via test images.
Technological Characteristics
High-frequency (85 kHz) X-ray generator; 3.0 kW peak power. Materials: standard medical-grade components. Dimensions: 460 x 305 x 225 mm; weight 19.5 kg. Energy: 120V 50-60Hz AC. Connectivity: RS-232 and Bluetooth for technique factor transmission. Collimator: Collimare. X-ray tube: Superior X-ray Tube Company SXR-130 (1.2 mm focal spot, 65 kHU).
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic x-rays. Not for mammographic use.
Regulatory Classification
Identification
A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 2, 2016
MinXray, Inc. % Daniel Kamm Principal Engineer Kamm & Associates 8870 Ravello Ct Naples, Florida 34114
Re: K153059
Trade/Device Name: HF1202H PowerPlus™ Portable X-ray Equipment Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile X-Ray System Regulatory Class: Class II Product Code: IZL Dated: October 19, 2015 Received: October 21, 2015
Dear Daniel Kamm:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153059
Device Name
HF1202H PowerPlus(tm) Portable X-ray Equipment
Indications for Use (Describe)
This radiographic system is intended for use by a qualified/trained physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Summary: 510(k) Number K153059 MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062, USA Toll Free 1-800-221-2245 (USA & Canada) Tel. 1-847-564-0323 Fax 1-847-564-9040 Date Prepared: January 25, 2016 Keith Kretchmer, President Contact:
- 1. Identification of the Device: Proprietary-Trade Name: HF1202H PowerPlus™ Portable X-Ray Equipment
- Classification Name: Mobile x-ray system, Product Code IZL Regulation Number 892.1720 2. Common/Usual Name: Mobile Diagnostic X-Ray System
- 3. Equivalent legally marketed device: K040046, MinXray HF120/60 PowerPlus,™ made by Mikasa X-Ray Co. Ltd. (for MinXray). Mobile x-ray system, Product Code IZL Regulation Number 892.1720
- Indications for Use (intended use): This radiographic system is intended for use by a qualified/trained 4. physician or technician on both adult and pediatric subjects for taking diagnostic x-rays. Not for mammographic use.
- 5. Description of the Device: The MinXray HF1202 consists of an X-ray generator (tubehead/control), continuously adjustable light beam collimator, mounting trunnion, exposure cord with 2-stage exposure switch, and AC power cord. If a stand is purchased with the HF1202, such as the MinXray XGS series of gas spring portable mobile stands, instructions for assembly of the stand and the attachment of the HF1202 are included with the stand. This is a high-frequency generator of updated design. The unit has a serial port and a Bluetooth port for communication of technique factors to a PC. The PC cannot initiate an exposure. When used, the PC must employ FDA cleared software and digital receptor panels. Known compatible with Toshiba FDX4343RP and FDX4343RPW digital x-ray panels. The FDX4343RP was cleared in K130883 (and K131211) and the FDX4343RPW was cleared in K143257.
- 6. Safety and Effectiveness, comparison to predicate device. The results of bench testing indicates that the new device is as safe and effective as the predicate device. Proper system operation is fully verified upon installation.
| | K040046, MinXray HF120/60<br>PowerPlus™, made by Mikasa X-Ray<br>Co. Ltd. (made for MinXray) | HF1202H PowerPlus™ Portable X-Ray<br>Equipment |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | This radiographic system is intended<br>for use by a qualified/trained physician<br>or technician on both adult and<br>pediatric subjects for taking diagnostic<br>x-rays. (Not for mammographic use.<br>Language later required by FDA) | (SAME) This radiographic system is intended for<br>use by a qualified/trained physician or<br>technician on both adult and pediatric subjects<br>for taking diagnostic x-rays. Not for<br>mammographic use. |
| Size / weight: | 453 x 292 x 224 mm 17.94kgs | 460 x 305 x 225 mm 19.5kgs |
| Energy Source: | 120V 50 - 60AC | SAME |
| | K040046, MinXray HF120/60<br>PowerPlus,™ made by Mikasa X-Ray<br>Co. Ltd. (made for MinXray) | HF1202H PowerPlus™ Portable X-Ray<br>Equipment |
| Use Interface: | Up-Down pushbuttons for kVp<br>selections and exposure time<br>selections and exposure time<br>selections with LED indicators mAs<br>indicator. | SAME |
| Exposure times: | (0.01–0.2sec) in 0.01sec. Step<br>(0.2-0.4sec) in 0.02sec. Step<br>(0.4-1.0sec) in 0.05sec Step<br>(1.0-5.0sec) in 0.1sec Step | (0.02–0.2sec) in 0.01sec. Step<br>(0.2-0.4sec) in 0.02sec. Step<br>(0.4-1.0sec) in 0.05sec Step<br>(1.0-5.0sec) in 0.1sec Step |
| mA: | 60 mA(0.01-0.1sec),<br>42 mA (0.11–5.0sec) @ 40 - 50 kVDC<br>50 mA(0.01-0.1sec),<br>35 mA (0.11–5.0sec) @ 52 - 60 kVDC<br>45 mA(0.01-0.1sec),<br>31.5 mA (0.11–5.0sec) @ 62 - 70 kVDC<br>38 mA(0.01-0.1sec),<br>26.6 mA (0.11–5.0sec) @ 72 - 80 kVDC<br>33 mA(0.01-0.1sec),<br>23.1 mA (0.11–5.0sec) @ 82 - 90 kVDC<br>30 mA(0.01-0.1sec),<br>21 mA (0.11–5.0sec) @ 92 - 100 kVDC<br>20 mA(0.01-0.1sec),<br>14 mA (0.11–5.0sec) @ 102 - 120 kVDC | 60 mA(0.02-0.05sec),<br>45mA (0.06 -0.2sec),<br>33 mA(0.22-5.0sec), @ 40 - 60 kVDC<br>55 mA(0.02-0.05sec),<br>41.3mA (0.06 -0.2sec),<br>30.3 mA(0.22-5.0sec), @ 62 - 70 kVDC<br>50 mA(0.02-0.05sec),<br>37.5mA (0.06 -0.2sec),<br>27.5 mA(0.22-5.0sec), @ 72 - 80 kVDC<br>45 mA(0.02-0.05sec),<br>33.8mA (0.06 - 0.2sec),<br>24.8 mA(0.22-5.0sec), @ 82 - 90 kVDC<br>40 mA(0.02-0.05sec),<br>30.0mA (0.06 - 0.2sec),<br>22.0 mA(0.22-5.0sec), @ 92 - 100 kVDC<br>30 mA(0.02-0.05sec),<br>22.5mA (0.06-0.2sec),<br>16.5 mA(0.22-5.0sec), @ 102 - 110 kVDC<br>25 mA(0.02-0.05sec),<br>18.8mA (0.06-0.2sec),<br>13.8 mA(0.22-5.0sec), @ 112 - 120 kVDC |
| Memory settings<br>(technique) | 5 memories via pushbutton | 10 memories via pushbutton |
| HF Generator | 85 kHz | SAME |
| kW | 2.4 kW peak | 3.0 kW peak |
| kVp: | 40 – 120kVp | SAME |
| X-ray Tube | Superior X-ray Tube Company<br>SXR-130 1.2 mm, 65 kHU | SAME |
| Collimator | Advantech | Collimare® |
| Serial<br>Communication | Not available | RS-232 or Bluetooth for communication of<br>technique factors to a PC. |
| | K040046, MinXray HF120/60<br>PowerPlus™, made by Mikasa X-Ray<br>Co. Ltd. (made for MinXray) | HF1202H PowerPlus™ Portable X-Ray<br>Equipment |
| Photo | Image: K040046, MinXray HF120/60 PowerPlus | Image: HF1202H PowerPlus Portable X-Ray Equipment |
## 7. Substantial Equivalence Chart
{4}------------------------------------------------
{5}------------------------------------------------
- 8. Summary of non-clinical testing: Software validation and risk analysis was performed. Laboratory testing was performed according to the following standards:
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) Medical Electrical Equipment, Part 1: General Requirements for Safety
IEC 60601-1-2:2007 Medical Electrical Equipment-Part 1-2: General Requirements for Safety - 2. Collateral Standard-Electromagnetic compatibility - Requirements and tests
IEC 60601-1-3:2008 (Second Edition) for use with IEC 60601-1: 2005 (Third Edition) Medical electrical equipment Part 1-3: General Requirements for Radiation Protection in Diagnostic X-Ray Equipment IEC 60601-1-6:2010 (Third Edition) for use in conjunction with IEC 60601-1: 2005 (Third Edition)Medical electrical equipment Part 1-6: General requirements for safety - Collateral Standard: Usability including IEC 62366: Application of usability engineering to medical devices
IEC 60601-2-28:2010 (Second Edition) for use in conjunction with IEC 60601-1: 2005 (Third edition) Medical electrical equipment Part 2: Particular requirements for the safety of X-ray source assemblies and X-ray tube assemblies for medical diagnosis
IEC 60601-2-54 (First Edition): 2009 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy IEC 62304:2006 (First Edition) Medical device software: Software life-cycle processes IEC 62366: 2007 (First Edition) + A1: 2014 Medical devices – Application of usability engineering to medical devices. The test results showed compliance with these standards. We also confirmed overall operation by taking and reviewing test images; however, sample clinical images were not required to determine substantial equivalence. The Bluetooth® module is FCC approved and was validated for use with the HF1202H.
- 9. Summary of clinical testing: Not applicable. Clinical testing was not deemed to show substantial equivalence. We relied on non-clinical testing and compliance with standards.
- 10. Conclusion: After analyzing bench tests, it is the conclusion of MinXray Inc. that the HF1202H PowerPlus™ Portable X-Ray Equipment is as safe and effective as the predicate device, has the same indications for use, has few technological differences, which are addressed through performance testing and compliance with the standards listed above, thus rendering it substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.