K153054 · Nuga Medical Co., Ltd. · JFB · Feb 5, 2016 · Physical Medicine
Device Facts
Record ID
K153054
Device Name
N5-1
Applicant
Nuga Medical Co., Ltd.
Product Code
JFB · Physical Medicine
Decision Date
Feb 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5880
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the N5-1 is to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, it provides topical heating for; -Temporary relief of minor muscle and joint pain, and stiffness. - Temporary relief of minor joint pain associated with arthritis. - The temporary increase in local circulation where applied. - Relaxation of muscle.
Device Story
N5-1 is an electric multi-function physical therapy device for muscle relaxation and topical heating. System components include a main assembly (mat) with 7 internal heating roller-type ceramics, a 5-ball projector, and a remote control. Device operates by delivering heat and mechanical massage to the patient's body (cervical to lumbar region). Used in home or clinic settings; operated by the patient or a caregiver. The remote control allows users to select manual or automatic massage modes and adjust temperature (40-60°C). Heat is generated via Tourmanium ceramic elements. Output provides thermal and mechanical stimulation to muscles and joints, intended to alleviate pain and stiffness and improve local circulation. Benefits include non-invasive muscle relaxation and symptomatic pain relief.
Clinical Evidence
No clinical data. Bench testing only, including electrical safety (AAMI ES60601-1) and electromagnetic compatibility (IEC 60601-1-2) testing. Software verification and validation performed.
Technological Characteristics
Multi-function physical therapy table. Materials: Tourmanium ceramic heating elements. Energy source: Electric. Temperature range: 40-60°C. Max load: 170 kg. Components: Main assembly, 5-ball projector, remote control. Software: Moderate level of concern. Standards: AAMI ES60601-1, IEC 60601-1-2.
Indications for Use
Indicated for patients requiring muscle relaxation therapy, temporary relief of minor muscle/joint pain, stiffness, arthritis-related joint pain, and increased local circulation.
Regulatory Classification
Identification
A multi-function physical therapy table is a device intended for medical purposes that consists of a motorized table equipped to provide patients with heat, traction, and muscle relaxation therapy.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 5, 2016
Nuga Medical Co., Ltd Jong Soo Lee Vice President 185. Jiraeul-ro. Jijeong-myeon Wonju, Gangwon-do 220-821 Republic of Korea
Re: K153054 Trade/Device Name: N5-1 Regulation Number: 21 CFR 890.5880 Regulation Name: Multi-Function Physical Therapy Table Regulatory Class: Class II Product Code: JFB, ILY, ISA Dated: January 4, 2016 Received: January 5, 2016
Dear Jong Soo Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153054
Device Name NS-I
Indications for Use (Describe)
The intended use of the N5-1 is to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, it provides topical heating for;
-Temporary relief of minor muscle and joint pain, and stiffness.
- Temporary relief of minor joint pain associated with arthritis.
- The temporary increase in local circulation where applied.
- Relaxation of muscle.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
![]()
# 510(k) Summary
[As required by 21 CFR 807.92]
#### Date Prepared [21 CFR 807.92(a)(a)] 1.
February 4, 2016
#### Submitter's Information [21 CFR 807.92(a)(1)] 2.
| Name of Sponsor: | NUGA MEDICAL Co., Ltd. |
|-----------------------|----------------------------------------------------------------|
| Address: | 185, Jiraeul-ro, Jijeong-myeon, Wonju-si,<br>Gangwon-do, Korea |
| Contact Name: | Jong Soo Lee/Vice President |
| Telephone No.: | +82 33 730 0001 |
| Fax No.: | +82 33 730 0008 |
| Email Address: | oky7800@nuga.kr |
| Registration Number: | 300200626518 |
| Name of Manufacturer: | Same as Sponsor |
| Address: | Same as Sponsor |
#### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)] 3.
| Trade Name: | N5-1 |
|-------------|------|
|-------------|------|
- Common Name: Personal Heating Therapeutic Device ●
- Classification:
| Classification Name | 1. Multi-Function Physical Therapy Table<br>2. Massager, Therapeutic, Electric<br>3. Lamp, Infrared, Therapeutic Heating |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Classification Panel | Physical Medicine |
| Classification Regulation | 21 CFR 890.5880, 21 CFR 890.5660 and<br>21 CFR 890.5500 |
| Product Code | 1. JFB<br>2. ISA<br>3. ILY |
| Device Class | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NUGA BEST. The logo consists of a stylized graphic of three figures in pink on the left, followed by the text "NUGA BEST" in blue. The "BEST" portion of the text has a trademark symbol next to it.
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)[
The identified predicate device within this submission is shown as follow;
- 510(k) Number: K111329
- Applicant: NUGA MEDICAL Co., Ltd. ●
- Common Name: Personal Heating Therapeutic Device ●
- Device Name: NM-7000
There are no significant differences between the N5-1 and the predicate device that would adversely affect the use of the product. It is substantially equivalent to both device in design, operational principles, functions and technological characteristics.
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
The N5-1 Personal Heating Therapeutic Device is the electric multi-function physical energy device. Its use is to provide muscle relaxing therapy to patients via a thermal function and massage and this device consist of the following components:
- (1) Main assembly
- (2) 5-Ball Projector
- (3) Remote Control
- (4) Accessories
Inside the main assembly, 7 internal heating roller type ceramics, which can massage from the cervical vertebra to the lumbar, have been installed.
#### 6. Intended Use [21 CFR 807.92(a)(5)]
The intended use of the N5-1 is to provide patients with muscle relaxation therapy by delivering heat and soothing massage. Additionally, it provides topical heating for;
- -Temporary relief of minor muscle and joint pain, and stiffness.
- -Temporary relief of minor joint pain associated with arthritis.
- -The temporary increase in local circulation where applied.
- -Relaxation of muscle.
# General Information
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Nuga Best. The logo consists of a pink circular graphic with white figures inside, followed by the text "NUGA BEST" in blue. The word "BEST" has a trademark symbol next to it.
# 7. Technological Characteristics [21 CFR 807.92(a)(6)]
The N5-1 is based on a technical feature comparison, the subject device was found to be similar to predicate device with regard to design, function, and technical characteristics.
| | Proposed Device | Predicate Device |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| K Number | Not known | K111329 |
| Common name | Personal Heating<br>Therapeutic Device | Personal Heating<br>Therapeutic Device |
| Model | N5-1 | NM-7000 |
| Manufacturer | NUGA MEDICAL Co., Ltd. | NUGA MEDICAL Co., Ltd. |
| Intended Use | The intended use of the N5-<br>1 is to provide patients<br>with muscle relaxation<br>therapy by delivering heat<br>and soothing massage.<br>Additionally, it provides<br>topical heating for;<br>-Temporary relief of minor<br>muscle and joint pain, and<br>stiffness.<br>- Temporary relief of minor<br>joint pain associated with<br>arthritis.<br>- The temporary increase in<br>local circulation where<br>applied.<br>- Relaxation of muscle. | The intended use of the<br>NM-7000 is to provide<br>patients with muscle<br>relaxation therapy by<br>delivering heat and<br>soothing massage.<br>Additionally, it provides<br>topical heating for;<br>- Temporary relief of minor<br>muscle and joint pain, and<br>stiffness.<br>- Temporary relief of minor<br>joint pain associated with<br>arthritis.<br>- The temporary increase in<br>local circulation where<br>applied.<br>- Relaxation of muscle. |
| Emission Source | Tourmanium ceramic | Tourmanium ceramic |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for NUGA BEST. The logo consists of a circular graphic on the left and the text "NUGA BEST" on the right. The circular graphic is pink and appears to depict three stylized figures with their arms raised. The text "NUGA BEST" is in a bold, sans-serif font and is colored blue. A small trademark symbol is located to the upper right of the word BEST.
| Design | | | |
|-------------------------|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------|------------------------------------------|
| Component | Main assembly (Mat),<br>Auxiliary Heating Part,<br>5-ball Projector,<br>Remote controller | Mat,<br>Auxiliary mat,<br>5-ball Projector,<br>9-ball Projector,<br>Remote controller,<br>Leg heater | |
| Accessory | Outside cover,<br>Projector cover,<br>User manual | Outside cover,<br>Projector cover,<br>User manual | |
| Material | Internal<br>Projector | Ceramic | Ceramic |
| | External<br>Projector | Ceramic | Ceramic |
| Technical Specification | | | |
| Temperature | | 40 - 60°C (104 - 140°F) | 40 - 60°C (104 - 140°F) |
| Mode set-up | | Manual Mode and Auto<br>Mode 1/2 | Manual Mode 1/2 and Auto<br>Mode 1/2/3/4 |
| Safe<br>Working Lode | 170 kg | | 170 kg |
The key differences between both devices are (1) the music player function on the remote controller, (2) the size of the main assembly, (3) the list of components (with or without the 9-ball projector and the leg heater), and (4) Mode set-up, and which do not raise any new safety and effectiveness issues.
# Non-Clinical Test Summary:
- 1) Electrical Safety, Electromagnetic Compatibility and Performance:
The N5-1 complies with the electrical safety and electromagnetic compatibility requirements established by the standards AAMI ES60601-1 and IEC 60601-1-2.
- · Testing to confirm compliance with the safety requirements of standard AAMI ES60601-1
- · Testing to confirm compliance with EMC requirements of standard IEC 60601-1-2
# General Information
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for NUGA BEST. The logo consists of a stylized image of three people holding hands in a circle on the left, and the words "NUGA BEST" in blue on the right. The logo is simple and modern, and it is likely used to represent a company or organization that is focused on health and wellness.
- 2) Software Validation:
The N5-1 contains MODERATE level of concern software. Software was designed and developed according to a software development process and was verified and validated.
Software information is provided in accordance with FDA guidance: The content of premarket submissions for software contained in medical devices, on May 11, 2005;
- 3) Clinical Test Summary:
No clinical studies were considered necessary and performed.
- 4) Biocompatibility
The N5-1 does not contain any parts which be contacted with the body of patient directly. Therefore the biocompatibility test has not been performed.
#### Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92] 8.
When compared to the predicate device (K111329), the N5-1 and the predicate are similar, with respect to indications for use and design, function, and technical characteristics.
#### Conclusion [21 CFR 807.92(b)(3)] 9.
In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification NUGA MEDICAL Co., Ltd., concludes that the N5-1 is substantially equivalent to predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.