# LithoVue System (K153049)

_Boston Scientific Corporation · FGB · Nov 13, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153049

## Device Facts

- **Applicant:** Boston Scientific Corporation
- **Product Code:** [FGB](/productcode/FGB.md)
- **Decision Date:** Nov 13, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, 3rd-Party Reviewed

## Indications for Use

The LithoVue System is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

## Device Story

LithoVue System is a software-controlled digital flexible ureteroscope; consists of a workstation (touch PC, interface box, cart) and a sterile, single-use digital flexible ureteroscope. Device captures real-time video images of the urinary tract; transmits data to the workstation for display to the physician. Used by urologists in clinical settings to perform diagnostic and therapeutic procedures (e.g., stone management using baskets, laser fibers, forceps). Single-use design eliminates reprocessing requirements associated with reusable endoscopes; provides consistent image quality and mechanical performance for each procedure. Physician views output on a monitor to guide navigation and instrument manipulation; facilitates minimally invasive access to the upper urinary tract.

## Clinical Evidence

No clinical data. Substantial equivalence supported by extensive bench testing, including optical performance (resolution, field of view, latency), mechanical durability (deflection, tensile strength, fatigue), electrical safety (IEC 60601-1, 60601-1-2, 60601-2-18), and biocompatibility (ISO 10993-1).

## Technological Characteristics

Digital flexible ureteroscope; single-use disposable scope; workstation-based processing. Features include automatic light control (ALC), deflection control, and instrument channel. Electrical safety per IEC 60601-1/1-2/2-18; photobiologic safety per IEC 62471. Biocompatibility per ISO 10993-1 for mucosal membrane contact <24 hours.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- DUR-Digital Ureteroscope and Choledochoscope (DUR-D) ([K060269](/device/K060269.md))
- Olympus URF-V ([K033651](/device/K033651.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 13, 2015

Boston Scientific Corporation % Dave Yungvirt CEO Third Party Review Group, LLC 45 Rockefeller Plaza, Suite 2000 New York, NY 10111

Re: K153049

> Trade/Device Name: LithoVue System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FGB Dated: November 5, 2015 Received: November 9, 2015

Dear Dave Yungvirt,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K153049

Device Name LithoVue System

#### Indications for Use (Describe)

The LithoVue System is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CER 801 Subnart D)  | Over-The-Counter Use (21 C.FR 801 Subnart C |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## SECTION 6

## 510K SUMMARY

## 510(k) Summary for LithoVue™ System

### A. Date Prepared

August 12, 2015

#### B. Sponsor

Boston Scientific Corporation Urology and Women's Health Division 100 Boston Scientific Way Marlborough, MA 01756

## C. Contact

Jeanne O'Toole Senior Specialist, Regulatory Affairs 508-683-4271 jeanne.otoole@bsci.com or Nichole Riek

Manager, Regulatory Affairs 508-683-4175 nichole.riek@bsci.com

## D. Device Name

| Trade name:        | LithoVue System                                    |
|--------------------|----------------------------------------------------|
| Common usual/name: | Digital Flexible Ureteroscope                      |
| Classification:    | FGB – Ureteroscope and accessories, flexible/rigid |
|                    | 21 CFR 876.1500, Class II                          |

## E. Predicate Device(s)

| Trade name:             | DUR-Digital Ureteroscope and Choledochoscope<br>(DUR-D)              |
|-------------------------|----------------------------------------------------------------------|
| Common usual/name:      | Ureteroscope and Accessories, Flexible/Rigid                         |
| Classification:         | FGB – Endoscope, associated accessories<br>21 CFR 876.1500, Class II |
| Premarket Notification: | ACMI, K060269                                                        |

and

| Trade name:        | Olympus URF-V (cleared as the Video Ureteroscope<br>NTSC)            |
|--------------------|----------------------------------------------------------------------|
| Common usual/name: | Ureteroscope and Accessories, Flexible/Rigid                         |
| Classification:    | FGB – Endoscope, associated accessories<br>21 CFR 876.1500. Class II |

{4}------------------------------------------------

Premarket Notification: Olympus, K033651

## F. Device Description

The LithoVue™ System is a software-controlled digital flexible ureteroscope system that consists of the LithoVue System Workstation (Touch PC with installed Interface Box and Cart) and the Litho Vue Single-Use Digital Flexible Ureteroscope (sterile, single-use disposable).

The LithoVue System is designed to allow physicians to access, visualize, and perform procedures in the urinary tract, using appropriate accessory devices (e.g., baskets, laser fibers, forceps).

## G. Intended Use

The LithoVue System is intended to be used to visualize organs, cavities and canals in the urinary tract (urethra, bladder, ureter, calyces and renal papillae) via transurethral or percutaneous access routes. It can also be used in conjunction with endoscopic accessories to perform various diagnostic and therapeutic procedures in the urinary tract.

This Intended Use is equivalent to that for the predicates, which are also used to perform various diagnostic and therapeutic procedures in the urinary tract using additional accessories.

## H. Technological Characteristics

The LithoVue System has the same technological characteristics and fundamental design as the predicate devices. The LithoVue System and the predicate devices are all designed to provide real-time images to the physician in order to facilitate diagnostic and therapeutic procedures in the urinary tract.

The main difference between the LithoVue System and the predicate devices is reusability. The LithoVue Single-Use Digital Flexible Ureteroscope is single-use, while the predicate devices are reusable.

## I. Substantial Equivalence

A direct comparison of key characteristics demonstrates that the LithoVue System is substantially equivalent to the predicate devices in terms of intended use, technological characteristics, and performance characteristics. The differences between the LithoVue System and predicate devices do not alter suitability of the proposed device for its intended use.

## J. Performance Testing (Bench Evaluation)

Boston Scientific has conducted performance testing with LithoVue Flexscope samples aged at T=0 and T=3 months accelerated aged.

{5}------------------------------------------------

The following testing was completed to demonstrate the LithoVue System functions as intended:

- · Working Distance Resolution in Air -Near, Typical and Far
- Field of View in Air
- · Direction of View
- · Image Orientation
- · Contrast Sensitivity
- · Image Signal to Noise
- · Uniformity of Image Brightness
- Total Flux Available at Tip
- · Image Latency
- Automatic Light Control (ALC)
- · ALC Response Time
- Width of Insertion Portion Distal Tip
- Maximum Width of Insertion Portion (Overall Shaft Diameter and Size Designation)
- Working Length (Shaft Length)
- Minimum Instrument Channel Width
- Tortuous Path Ability
- · Surface and Edges
- · SUD Working Channel Freedom from Leakage
- · SUD Shaft Leakage
- · Torsional Resistance of Tip
- · Tip Column Strength
- Shaft Flexural Resistance
- · Maximum Angle of Deflection
- Bend Radius
- · Secondary Deflection
- · Deflection Life Cycling
- · Shaft to Handle Tensile Strength
- Tip to Shaft Tensile Strength
- · Handle to Active Deflection Torque Angle at Break
- Umbilicus to Handle Tensile Strength
- Umbilicus to Connector Tensile Strength
- Critical Shaft Bend Radius
- · Shaft Bend Fatigue
- · SUD Durability (Procedure Duration)
- SUD Mechanical Durability
- · Neutral Position Marking
- · Deflection Direction Marking
- · Lever Force
- · Connector Retention Strength
- · Laser Aiming Beam Compatibility
- · Laser Lithotripsy Compatibility
- · Radiopacity
- · Working Channel Connection Compatibility
- Maximum Endotherapy Tool Blind Insertion Distance
- · Working Channel Length
- · Monitor and Switcher Compatibility
- SUD Packaging Integrity
- SUD Package Seal Strength
- · Photobiologic Safety per IEC 62471
- Software verification and validation

Biocompatibility of the LithoVue Single-Use Digital Flexible Ureteroscope was evaluated in accordance with ISO 10993-1:2009 for the body contact category of "Surface - Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed, as recommended: Cytotoxicity, Irritation, Sensitization, USP Physiochemical <661>, and Latex. All evaluation acceptance criteria were met.

Electrical safety testing of the System was evaluated in accordance with IEC 60601-1 (2005) Edition 3, IEC 60601-1-2 (2007) Edition 3, IEC 60601-1-4 (2000), Edition 1.1 Consolidated Edition, and IEC 60601-2-18 (2009), Edition 3. All evaluation acceptance criteria were met.

The results of the performance testing demonstrate that the LithoVue System is considered safe and effective for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K153049](https://fda-staging.innolitics.com/device/K153049)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).