K153021 · Amemo, Inc. · DQA · Nov 22, 2016 · Cardiovascular
Device Facts
Record ID
K153021
Device Name
Fingertip Pulse Oximeter A310
Applicant
Amemo, Inc.
Product Code
DQA · Cardiovascular
Decision Date
Nov 22, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Indications for Use
Fingertip Pulse Oximeter A310 is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
Device Story
Fingertip Pulse Oximeter A310 is a battery-powered, non-invasive, portable device for spot-checking SpO2 and pulse rate in adult hospital patients. It uses a finger-clip sensor containing a dual light source (660 nm red, 905 nm infrared) and a photodetector. The device measures light absorption in the pulsating arteriolar vascular bed; the photodetector converts light intensity into electronic signals. The ratio of light absorption during systole and diastole is calculated to determine SpO2. The device consists of a PCB, OLED screen, and plastic shell; it lacks alarms and is not for continuous monitoring or life support. Healthcare providers use the displayed numerical values to assess patient oxygenation and pulse rate, aiding in clinical spot-check assessments.
Clinical Evidence
Clinical validation performed on 12 healthy adult volunteers (8 male, 4 female) with varying skin pigmentations. 555 data points were collected across an SpO2 range of 70%-100% and compared against a laboratory CO-Oximeter. Results showed an overall root-mean-square (Arms) accuracy of 2.21% (range 1.58%-2.80% depending on saturation level). Bland-Altman analysis confirmed agreement within 95% limits. No adverse effects reported.
Technological Characteristics
Battery-powered (2 AAA) fingertip pulse oximeter. Dual light source (660nm red, 905nm infrared) and photodetector. OLED display. Biocompatibility per ISO 10993-5/10. Electrical safety per IEC 60601-1, IEC 60601-1-11. EMC per IEC 60601-1-2. Performance per ISO 80601-2-61. Reusable, non-sterile.
Indications for Use
Indicated for spot checking functional oxygen saturation (SpO2) and pulse rate (PR) in adult patients in hospital settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus-like symbol with three figures in profile, representing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 22, 2016
AMEMO INC. Hua Xie CEO 1154 Cadillac Court Milpitas. California 95035
Re: K153021
Trade/Device Name: Fingertip Pulse Oximeter A310 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: November 14, 2016 Received: November 14, 2016
Dear Hua Xie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153021
Device Name Fingertip Pulse Oximeter
Indications for Use (Describe)
Fingertip Pulse Oximeter A310 is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
AMEMO INC 1154 Cadillac Court Milpitas CA 95035 USA
#### Section III 510(K) Summary
### Date: 2016-11-17
## Sponsor:
AMEMO INC
1154 Cadillac Court Milpitas CA 95035 USA Contact person: Mr. Hua Xie Email: Raymond691116@gmail.com
#### Proposed Device:
| Model | Name |
|-------|--------------------------|
| A310 | Fingertip Pulse Oximeter |
Trade Name: Fingertip Pulse Oximeter
Common Name: Fingertip Pulse Oximeter
Product Classification: DQA; 870.2700; Class II
Classification Panel: Anesthesiology
Predicate Device: MD300C1 Fingertip Pulse Oximeter (K093757)
### Indications for use:
Fingertip Pulse Oximeter A310 is a non-invasive device intended for spot checking of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR). This portable device is indicated for use in adult patients in hospitals.
### Device description:
Fingertip Pulse Oximeter A310 is a battery powered device intended for use in
{4}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA measuring and displaying functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (PR).
The Finger Pulse Oximeter works by applying a sensor to a pulsating arteriolar vascular bed. The sensor contains a dual light source and photo detector. The one wavelength of light source is 660 nm, which is red light; the other is 940 nm, which is Infrared light. Skin, bone, tissue, and venous vessels normally absorb a constant amount of light over time. The photodetector in finger sensor collects and converts the light into electronic signal which is proportional to the light intensity. The arteriolar bed normally pulsates and absorbs variable amounts of light during systole and diastole, as blood volume increases and decreases. The ratio of light absorbed at systole and diastole is translated into an oxygen saturation measurement. This measurement is referred to as SpO2.
This equipment mainly composed of PCB board, On/Off button, mode button, OLED screen, battery compartment, and plastic shell.
The device is a spot-check pulse oximeter and does not include alarms.
The device does not support the measurement in the condition of low perfusion.
The device is not intended for life-supporting or life-sustaining.
The device is reusable and does not need sterilization.
{5}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
# Comparison list of the technological characteristics
| Comparison Items | Applicant Device | Predicate Device (K093757) | Remark | |
|---------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|------------|
| Indications for use | Fingertip Pulse Oximeter A310 is a<br>non-invasive device intended for spot<br>checking of functional oxygen saturation<br>of arterial hemoglobin (SpO2) and pulse<br>rate (PR). This portable device is<br>indicated for use in adult patients in<br>hospitals | Fingertip Pulse Oximeter MD300C1 is a<br>portable non-invasive, spot-check,<br>oxygen saturation of arterial hemoglobin<br>(SpO2) and pulse rate of adult and<br>pediatric patient at home, and hospital<br>(including clinical use in<br>internist/surgery, Anesthesia, intensive<br>care and etc). Not for continuously<br>monitoring | Different① | |
| Principle | The device displays numerical values for<br>functional oxygen saturation of arterial<br>hemoglobin (SpO2) and pulse rate by<br>measuring the absorption of red and<br>infrared (IR) light passing through<br>perfused tissue.<br>Changes in the absorption caused by<br>the pulsation of blood in the vascular<br>bed are used to determine oxygen<br>saturation and pulse rate. | The device displays numerical values<br>for functional oxygen saturation of<br>arterial hemoglobin (SpO2) and pulse<br>rate by measuring the absorption of red<br>and infrared (IR) light passing through<br>perfused tissue.<br>Changes in the absorption caused by<br>the pulsation of blood in the vascular<br>bed are used to determine oxygen<br>saturation and pulse rate. | Same | |
| Specifications | LED wavelength | Red= 660 nm; Infrared=905nm | Red= 660 nm; Infrared=940nm | Different② |
| | Power source | 2 AAA alkaline batteries. | 2 AAA alkaline batteries. | Same |
| Testing | Display data | SPO2%; PR | SPO2%; PR | Same |
| | SpO2 Measuring<br>Range | 0%-100% | 0%-100% | Same |
| | SpO2 Resolution | 1% | 1% | Same |
| | SpO2 Accuracy | 70~100%, ±3%;<br>0-69%, unspecified; | 70~100%, ±3%;<br>0-69%, unspecified; | Same |
| | PR Measuring<br>Range | 30-235BPM. | 30-235BPM | Same |
| | PR Resolution | 1 bpm | 1 bpm | Same |
| | PR Accuracy | ± 2 bpm (30-99bmp)<br>2% (100-235bmp) | ± 2 bpm (30-99bmp)<br>2% (100-235bmp) | Same |
| | Biocompatibility | ISO 10993-5 and ISO 10993-10 | ISO 10993-5 and ISO 10993-10 | Same |
| | Electrical Safety | IEC 60601-1<br>IEC 60601-1-11 | IEC 60601-1<br>IEC 60601-1-11 | Same |
| | Electromagnetic<br>Compatibility | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| | Performance | ISO 80601-2-61 | ISO 9919 | Different③ |
{6}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
{7}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
# [Discussion]
The proposed device has the same application site, performance testing, and accuracy as the differences are
- 1. The predicate device is intended for both adult and pediatric patients in hospital and home environments, however Oximeter A310 is intended only for adult in hospital environments. Both applicable population and environment of A310 are smaller than the predicate, which will not raise different questions of safety or effectiveness compared to the predicate device. All A310's indications for use is Substantially Equivalent (SE) to the predicate device in its declared scope.
- 2. Infrared LED Wavelength of Oximeter A310 is 905nm, and predicate's wavelength is 940nm. These two wavelengths are both adopted widely in measurement of SPO2. This difference will affect the performance of the device, especially the accuracy. We conducted the ISO 80601-2-61 test to verify the performance and electrical safety of Oximeter A310, and the clinical evaluation to validate the accuracy of measurement. The clinical evaluation result shows that A310's accuracy complies with the requirement defined in the standard ISO 80601-2-61. All validation demonstrates that Oximeter A310 is substantially equivalent to the predicate device.
- 3. Application of the performance standard is different. The predicate device uses the ISO 9919 to demonstrate the performance. However, ISO 9919 has been withdrawn, and ISO 80601-2-61 is the newest international performance standard of oximeter, and state-of-the oximeter. And, ISO 80601-2-61 is also the Recognized Consensus Standards of FDA.
{8}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
#### Non-clinical testing summary:
The Fingertip Pulse Oximeter A310 was subjected to bench testing. The following non-clinical testing was performed to demonstrate substantial equivalence of A310 with its predicate. The test results demonstrated that the proposed device complies with the following standards and requirements:
IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
ISO 80601-2-61: Medical electrical equipment - Particular requirements for the basic safety and essential performance of pulse oximeter equipment for medical use.
IEC 60601-1-11: Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
FDA Guidance for Pulse Oximeters - Premarket Notification Submissions [510(k)s]
ISO 10993-5: Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
ISO 10993-10: Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization
Pulse Rate Accuracy testing has been performed by referring to the SpO2 simulator. Pulse rate accuracy was stated as the root-mean-square difference between paired pulse date recorded with A310 and simulator. The testing procedure and results are included in the report A310 Fingertip Pulse Rate Accuracy Test Report, which
{9}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
includes the results of PR measurement range, resolution, and measurement accuracy. The report demonstrates that the PR accuracy complies with the declaration of specification at the normal condition.
The testing to verify the performance after simulated lifetime was conducted. The testing conducted included 1424 cycles of cleaning of the device according to the description in the operation manual. After the cycles of cleaning, the accuracy testing of SpO2 and PR was performed by referring to the SpO2 simulator. The testing result demonstrates that the device continues to perform within specifications after a simulated lifetime of use.
### Clinical testing summary
The functional oxygen saturation (SpO2) measurement has been validated in accordance with ISO 80601-2-61. The clinical testing was completed on a total of 12 healthy adult volunteers (8 man and 4 women) with light to dark skin pigmentations in the range of 70% to 100% against a laboratory CO-Oximeter. The subjects include 6 people with medium skin, 1 with light skin, and 5 with dark skin pigmentation. Total 555 data points were sampled for analysis.
The measured arterial hemoglobin saturation value (SpO2) of the proposed device was compared with arterial hemoglobin oxygen (SaO2) value, determined from blood samples with a CO-oximeter. The accuracy of the device is in comparison with the CO-oximeter samples measured over the SpO2 range 70%-100%.
Accuracy data was calculated using the root-mean-squared (Arms value) for all subjects, per ISO 80601-2-61, the result showed that the error is far less than the scope specified in the standard; and the Agreement between Methods of Measurement with Multiple Observations per each subject was analyzed with the Bland and Altman statistics, the analysis demonstrated that the vast majority of data is within ±95% limit of agreement.
The SpO2 accuracy result showed that the root-mean-square (Arms) value of the Fingertip Pulse Oximeter is ±3% with the saturations from 70% to 100%. The following is the summary of the testing results:
| Items | 70-100 | 70-<80 | 80-<90 | 90-100 |
|-------|--------|--------|--------|--------|
| #pts | 555 | 192 | 237 | 126 |
| Bias | 0.40 | 0.98 | 0.41 | 1.06 |
| RMS | 2.21 | 2.42 | 1.58 | 2.80 |
{10}------------------------------------------------
1154 Cadillac Court Milpitas CA 95035 USA
In addition, there were no reported adverse effects during these investigations.
### Conclusion:
The Fingertip Pulse Oximeter A310 has the same intended use, the same technology as the predicate device. Thus we conclude the subject device to be Substantially Equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.