K153014 · Mobius Medical Systems, LP · IYE · Apr 29, 2016 · Radiology
Device Facts
Record ID
K153014
Device Name
Mobius3D
Applicant
Mobius Medical Systems, LP
Product Code
IYE · Radiology
Decision Date
Apr 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K153014 · Apr 29, 2016
Mobius3D
Mobius Medical Systems, LP
Anonymized patient treatment data (163 fractions from 22 patients)
Retrospective clinical data was used to validate the CBCT module's ability to detect potential patient positioning or anatomical differences by comparing daily treatment images to planning images.
Ability to notify users of potential patient positioning or anatomical differences
Indications for Use
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography). Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
Device Story
Software-only QA tool for radiation therapy clinics; operated by trained radiation oncology personnel. Inputs: treatment plans from external treatment planning systems (TPS), linear accelerator delivery log files, planning CT images, and daily cone beam CT (CBCT) images. Core function: independent 3D radiation dose calculation using proprietary collapsed cone convolution superposition (CCCS) algorithm; dose delivery verification via linac log files (MobiusFX module); and patient alignment/anatomy analysis via image registration (CBCT Checks module). Outputs: dose verification reports and image analysis notifications. Healthcare providers use output to verify treatment plan accuracy and monitor patient positioning/anatomical changes before radiation delivery. Benefits: ensures treatment plan integrity and patient alignment, reducing risk of delivery errors.
Clinical Evidence
Bench testing only. Evaluated CBCT module using anonymized data from 22 patients across 163 treatment fractions. Testing validated the module's ability to detect and notify users of potential patient positioning or anatomical differences compared to planning images. Software verification and validation performed per FDA guidelines.
Technological Characteristics
Software-only QA tool. Performs independent 3D dose calculation using collapsed cone convolution superposition (CCCS) algorithm. Includes modules for dose delivery verification (MobiusFX) and image-based alignment analysis (CBCT Checks). Operates on read-in data from external TPS and imaging systems. No patient contact. No hardware components.
Indications for Use
Indicated for radiation oncology personnel to perform quality assurance, treatment plan verification, and patient alignment/anatomy analysis in radiation therapy. Not for use as a treatment planning system.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 29, 2016
Mobius Medical Systems, LP % Mr. Stan Eshelman COO 5012 Tamarisk Street BELLAIRE TX 77401
Re: K153014 Trade/Device Name: Mobius3D Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: April 7, 2016 Received: April 8, 2016
Dear Mr. Eshelman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known)
K153014
Device Name Mobius3D
#### Indications for Use (Describe)
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).
Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5 510(K) Summary
## 5.1 510(K) Owner
Mobius Medical Systems, LP 5012 Tamarisk St Bellaire, TX 77401 Tel: (888) 263-8541 Fax: (888) 263-8541
## 5.2 Contact Person
Stan Eshelman
## 5.3 Preparation Date
April 7, 2016 (revised from original October 14, 2015 submission)
# 5.4 Trade Name
Mobius3D
#### 5.5 Common Name
Secondary Check QA Software
# 5.6 CLASSIFICATION NAME
Accelerator, Linear, Medical 21 CFR 892.5050 Product Code - IYE
# 5.7 PREDICATE DEVICES
Primary Predicate:
- o K140660 Mobius3D (Mobius Medical Systems, LP)
Secondary Predicates:
- K090531 RadCalc Software (Lifeline Software, Inc.) o
- K081076 VelocityAIS (Velocity Medical Solutions, LLC) o
# 5.8 Device Description
Mobius3D is a software product used within a radiation therapy clinic for quality assurance and treatment plan verification. It is important to note that while Mobius3D operates in the field of radiation therapy, it is neither a radiation delivery device (e.g. a linear accelerator), nor is it a treatment planning system (TPS). Mobius3D cannot design or transmit instructions to a delivery device, nor does it control any other medical device. Mobius3D is an analysis tool meant solely for quality assurance (QA) purposes
{4}------------------------------------------------
when used by trained medical professionals. Being a software-only QA tool, Mobius3D never comes into contact with patients.
Mobius3D performs dose calculation verifications for radiation treatment plans by doing an independent calculation of radiation dose. Radiation dose is initially calculated by a treatment planning system (TPS), which is a software tool that develops a detailed set of instructions (i.e. a plan) for another system (e.g. a linear accelerator) to deliver radiation to a patient. The dose calculation performed by Mobius3D uses a proprietary collapsed cone convolution superposition (CCCS) algorithm.
Mobius3D also performs dose delivery quality assurance for radiation treatment plans by using the measured data recorded in a linear accelerator's delivery log files to calculate a delivered dose. This is presented to the end user in a software component of Mobius3D called MobiusFX. The MobiusFX component is available to users through licensing as an add-on to the core Mobius3D software features.
Compared to the previously cleared Mobius3D v 1.3.2 (K140660), Mobius3D v 2.0.0 contains the additional intended use of performing quality assurance of a patient's alignment and anatomy analysis. This analysis is based on comparison of Cone Beam Computed Tomography (CBCT) images taken immediately before treatment to the images used for treatment planning, which are typically acquired using standard Computed Tomography (CT). This analysis is presented to the end user in an add-on software module within Mobius3D called CBCT Checks.
# 5.9 Intended Use
Mobius3D software is used for quality assurance, treatment plan verification, and patient alignment and anatomy analysis in radiation therapy. It calculates radiation dose three-dimensionally in a representation of a patient or a phantom. The calculation is based on read-in treatment plans that are initially calculated by a treatment planning system, and may additionally be based on external measurements of radiation fields from other sources such as linac delivery log data. Patient alignment and anatomy analysis is based on read-in treatment planning images (such as computed tomography) and read-in daily treatment images (such as registered cone beam computed tomography).
Mobius3D is not a treatment planning system. It is only to be used by trained radiation oncology personnel as a quality assurance tool.
#### 5.10 TECHNOLOGICAL CHARACTERISTICS SUMMARY
The principle technological characteristic of Mobius3D is to perform quality assurance evaluations on the radiation treatment planning system (TPS) and radiation delivery systems (e.g. linear accelerators). Mobius3D (v 2.0.0) adds the technical characteristic of performing comparative analysis between patient images acquired at different times using different image modalities.
{5}------------------------------------------------
Detailed technological characteristics and indications for use presented within the full set of submitted documentation for this 510(k) application support the claim that Mobius3D is substantially equivalent to the predicate devices.
#### 5.11 Testing Summary
Software development, verification, and validation have been carried out in accordance with FDA quidelines. The software was tested against the established Software Design Specifications and passed all required tests. A Risk Management Report was completed which identified and verified the mitigation of all required hazards.
In addition, bench testing was performed and is presented in a report titled: Mobius3D CBCT Module Evaluation - Patient Positioning / Anatomical Changes Bench Testing. The report was generated by performing a full suite of tests utilizing anonymized data from 22 different patients over 163 different fractions with varying treatment sites to validate each portion of the analysis. The report demonstrates the Mobius3D CBCT module's ability to notify users to a potential patient positioning or anatomical difference.
## 5.12Conclusions
Based on the evidence presented in this 510k submission, Mobius Medical Systems believes that Mobius3D is substantially equivalent to its predicate devices both in terms of its indications for use and its technical characteristics, and that Mobius3D is a safe and effective medical device for its intended users.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.