Reprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology (EP) Catheter

K153006 · Sterilmed, Inc. · NLH · Jun 23, 2016 · Cardiovascular

Device Facts

Record IDK153006
Device NameReprocessed Lasso NAV eco and Lasso 2515 NAV eco Variable Electrophysiology (EP) Catheter
ApplicantSterilmed, Inc.
Product CodeNLH · Cardiovascular
Decision DateJun 23, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. stimulation and recording only. The catheters are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters with navigation (Nav) capabilities provide location information when used with compatible Carto® 3 EP Navigation Systems. (These catheters are not compatible with Carto® 3 EP Navigation Systems prior to version 2.3.)

Device Story

Deflectable EP catheters with circular distal spine and platinum/iridium electrodes; used for stimulation and recording of cardiac electrograms in atrial regions. NAV models provide location data via integration with Carto® 3 EP Navigation Systems (v2.3+). Operated by physicians in clinical settings; inserted into right or left atrium via 8F guiding sheath. Reprocessed by Sterilmed; limited to one reprocessing cycle. Output consists of electrical signals for mapping and location data for navigation systems; assists clinicians in diagnosing cardiac arrhythmias. Benefits include diagnostic mapping capabilities equivalent to original single-use devices.

Clinical Evidence

Bench testing only. Validation included cleaning and sterilization validation (ISO 11135, USP <71>), biocompatibility (ISO 10993-1), ethylene oxide residuals (ISO 10993-7), packaging (ASTM D 4169, ASTM F 88, ASTM F 2096), and shelf life (ASTM F 1980). Functional performance verified via simulated use, visual inspection, fatigue testing, and 100% functional testing of all reprocessed units.

Technological Characteristics

Deflectable catheter with circular distal spine; platinum/iridium electrodes. Compatible with Carto® 3 EP Navigation System. Materials identical to predicate. Sterilization via ethylene oxide. Reprocessed maximum of one time; 100% functional testing performed post-reprocessing. Standards: ISO 11135, ISO 10993-1/7, ASTM D 4169, ASTM F 88, ASTM F 2096, ASTM F 1980.

Indications for Use

Indicated for patients requiring multiple electrode electrophysiological mapping of cardiac structures, specifically stimulation and recording of electrograms in the atrial regions of the heart.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 23, 2016 Sterilmed. Inc. Patricia Kaufman Regulatory Affairs Specialist 5010 Cheshire Parkway, Suite 2 Plymouth, Minnesota 55446 Re: K153006 Trade/Device Name: Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Catheter Electrophysiology (EP) Catheters (See attached list of models) Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: NLH Dated: May 23, 2016 Received: May 24, 2016 Dear Patricia Kaufman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Mitchell Stein for Bram D. Zuckerman, M.D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | Catheter | Model | Poles | Loop | | Curve | Electrode<br>Spacing (mm) | Shaft | | |--------------------------------------|---------|-------|---------------|-------------|-------|---------------------------|-------------|-------------| | | | | Diameter (mm) | Profile (F) | | | Profile (F) | Length (cm) | | LASSO 2515 NAV eco Variable Catheter | | | | | | | | | | 1 | D134301 | 20 | 25 - 15 | 4 | D | 2-6-2 | 7 | 115 | | 2 | D134302 | 10 | 25 - 15 | 4 | D | 8 | 7 | 115 | | LASSO NAV eco Catheter | | | | | | | | | | 3 | D134901 | 10 | 15 | 4.5 | D | 4.5 | 7 | 115 | | 4 | D134902 | 20 | 15 | 4.5 | D | 4.5 pairs | 7 | 115 | | 5 | D134903 | 10 | 20 | 4.5 | D | 6.0 | 7 | 115 | | 6 | D134904 | 20 | 20 | 4.5 | D | 6.0 pairs | 7 | 115 | | 7 | D134905 | 10 | 25 | 4.5 | D | 8.0 | 7 | 115 | | 8 | D134906 | 20 | 25 | 4.5 | D | 8.0 pairs | 7 | 115 | | 9 | D134909 | 10 | 20 | 4.5 | D | 6.0 | 7 | 115 | # K153006 MODELS INTENDED FOR REPROCESSING {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153006 #### Device Name Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters Indications for Use (Describe) The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. stimulation and recording only. The catheters are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters with navigation (Nav) capabilities provide location information when used with compatible Carto® 3 EP Navigation Systems. (These catheters are not compatible with Carto® 3 EP Navigation Systems prior to version 2.3.) Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ ### 2. 510(K) SUMMARY This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT | Sterilmed, Inc.<br>5010 Cheshire Pkwy N, Suite 2<br>Plymouth, MN 55446<br>www.sterilmed.com<br>Tel: (888) 541-0078 | |---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------| | OFFICIAL<br>CORRESPONDENT | Patricia F. Kaufman<br>Sterilmed, Inc.<br>5010 Cheshire Pkwy N, Suite 2<br>Plymouth, MN 55446<br>pkaufman@sterilmed.com<br>Tel: (763) 488-3211<br>Fax: (763) 488-3350 | | DATE PREPARED | 12 October 2015 | | TRADE NAME | Reprocessed LASSO NAV eco and Reprocessed LASSO<br>2515 NAV eco Variable Electrophysiology (EP) Catheters | | COMMON NAME | Electrophysiology Diagnostic Catheters | | 510(K) NUMBER | K153006 | | DEVICE CLASSIFICATION | Name: Catheter, Recording, Electrode, Reprocessed<br>Regulation No: §870.1220<br>Product Code: NLH<br>Class: II<br>Panel: Cardiovascular | | PREDICATE DEVICE | Lasso 2515 NAV eco Variable Catheter, Lasso NAV eco<br>Catheter - K113213 | # SUBSTANTIALLY EQUIVALENT TO: The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are substantially equivalent to the single use Biosense Webster Lasso 2515 NAV eco Variable Catheter and Lasso NAV eco Catheter in intended use and technological features. ### DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: {5}------------------------------------------------ The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are designed to facilitate electrophysiological mapping of the atria of the heart. They are deployed in the right or left atrium through an 8F guiding sheath. These deflectable catheters consist of a circular spine on the distal tip with platinum/iridium electrodes that can be used for stimulation and recording. Models designated as NAV are compatible with the Carto® 3 EP Navigation System. | | Catheter<br>Model | Poles | Loop | | Curve | Electrode<br>Spacing (mm) | Shaft | | |---------------------------------------|-------------------|-------|---------------|-------------|-------|---------------------------|-------------|-------------| | | | | Diameter (mm) | Profile (F) | | | Profile (F) | Length (cm) | | LASSO® 2515 NAV eco Variable Catheter | | | | | | | | | | 1 | D134301 | 20 | 25 - 15 | 4 | D | 2 - 6 - 2 | 7 | 115 | | 2 | D134302 | 10 | 25 - 15 | 4 | D | 8 | 7 | 115 | | LASSO® NAV eco Catheter | | | | | | | | | | 3 | D134901 | 10 | 15 | 4.5 | D | 4.5 | 7 | 115 | | 4 | D134902 | 20 | 15 | 4.5 | D | 4.5 pairs | 7 | 115 | | 5 | D134903 | 10 | 20 | 4.5 | D | 6.0 | 7 | 115 | | 6 | D134904 | 20 | 20 | 4.5 | D | 6.0 pairs | 7 | 115 | | 7 | D134905 | 10 | 25 | 4.5 | D | 8.0 | 7 | 115 | | 8 | D134906 | 20 | 25 | 4.5 | D | 8.0 pairs | 7 | 115 | | 9 | D134909 | 10 | 20 | 4.5 | D | 6.0 | 7 | 115 | The following models are the subject of this submission: # INDICATIONS FOR USE: The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. stimulation and recording only. The catheter is designed to obtain electrograms in the atrial regions of the heart. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters with navigation (NAV) capabilities provide location information when used with compatible Carto® 3 EP Navigation Systems. (These catheters are not compatible with Carto® 3 EP Navigation Systems prior to version 2.3.) # TECHNICAL CHARACTERISTICS: The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are identical to the predicate devices in design, materials of construction, and intended use. There are no changes to the clinical applications, patient population, performance specifications, or method of operation. {6}------------------------------------------------ # PERFORMANCE STANDARDS: No applicable performance standards have been issued under 514 of the Food, Drug and Cosmetic Act for a Catheter, Recording, Electrode, Reprocessed §870.1220. # FUNCTIONAL AND SAFETY TESTING: Representative samples of Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters were tested to demonstrate appropriate functional characteristics. Process validation testing was performed to validate the cleaning and sterilization procedures as well as device packaging. In addition, the manufacturing process includes visual and validated functional testing of 100% of products reprocessed. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are reprocessed no more than one (1) time. Each device is marked and tracked through each reprocessing cycle. After the device has reached the maximum number of reprocessing cycles (1), the device is rejected from further reprocessing. Reprocessing is performed only by the manufacturer Sterilmed. # SUMMARY OF NONCLINICAL TESTING: Specific non-clinical tests performed included: cleaning validation, sterilization validation (ISO 11135, USP <71>), biocompatibility testing (ISO 10993-1), ethylene oxide residual testing (ISO 10993-7), packaging validation (ASTM D 4169, ASTM F 88, ASTM F 2096), and shelf life validation (ASTM F 1980). In addition, validation of functional performance (bench testing) was performed through simulated use, visual inspection, fatigue testing, and function testing. Performance testing shows the Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters to perform as originally intended. # BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE: Sterilmed concludes that the Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are safe, effective, and substantially equivalent to the predicate devices, Biosense Webster Lasso Catheters (K113213), as described in this premarket notification submission. The Reprocessed LASSO NAV eco and Reprocessed LASSO 2515 NAV eco Variable Electrophysiology (EP) Catheters are substantially equivalent to the listed predicate devices with respect to their indications for use (intended use) and technical characteristics. The information and data provided in this 510(k) submission identifies no new safety or effectiveness issues.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...