Mammotome revolve Dual Vacuum Assist Biopsy System
K152989 · Devicor Medical Products, Inc. · KNW · Feb 25, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K152989
Device Name
Mammotome revolve Dual Vacuum Assist Biopsy System
Applicant
Devicor Medical Products, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Feb 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities. The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality. The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
Electromechanical breast biopsy system; comprises sterile single-use probe, reusable holster, and reusable control unit. Inputs: user activation via touchscreen, remote keypad, or footswitch. Operation: vacuum-assisted tissue acquisition combined with rotational/translational cutter movement. System uses closed-loop control for cutter position and speed. Output: tissue samples for histologic examination. Used in clinical settings for stereotactic and ultrasound-guided procedures. Healthcare providers operate the system to excise breast tissue; output informs clinical diagnosis of breast abnormalities. Benefits include reliable tissue sample collection for pathology.
Clinical Evidence
Bench testing and in vivo porcine model study. Evaluated sample weight, sampling reliability, and sample quality. Compared subject device against predicate systems. Results confirmed comparable performance in tissue retrieval and system functionality (holster-to-probe recognition, communication). No human clinical trials reported.
Technological Characteristics
Electromechanical biopsy system. Materials: stainless steel (needle, cutter), plastic (probe housing). Components: trocar-tipped needle, vacuum pump, power supply, valve actuators, touchscreen interface. Connectivity: wired (tethered vacuum, remote controls). Sterilization: Cobalt 60 irradiation. Software: microprocessor-based with closed-loop control for cutter position/speed.
Indications for Use
Indicated for patients with breast abnormalities requiring diagnostic tissue sampling for histologic examination; includes partial or complete removal of imaged abnormalities and partial removal of palpable abnormalities (e.g., fibroadenoma, fibrocystic lesions).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System (K123259)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, with flowing lines extending from the bottom profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2016
Devicor Medical Products, Inc. Ms. Shawna Rose Director, Regulatory Affairs 300 E-Business Way, Fifth Floor Cincinnati, Ohio 45241
Re: K152989
Trade/Device Name: Mammotome Revolve Dual Vacuum Assist Biopsy System Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: November 25, 2015 Received: November 27, 2015
Dear Ms. Rose:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152989
Device Name
Device Name: Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System
Indications for Use (Describe)
The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.
• The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
• The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures. In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 10px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following information is provided as required by 21 CFR § 807.92 for the Mammotome revolve® Vacuum Assisted Biopsy (VAB) System 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990 the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.
## Company:
Devicor® Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Establishment Registration Number: 3008492462
#### Contact:
Shawna Rose Director, Regulatory Affairs Devicor Medical Products, Inc. 300 E-Business Way, Fifth Floor Cincinnati, OH 45241 Ph: 513-864-9178 Fax: 513-864-9011 E-mail: srose@mammotome.com
Date of Submission: November 24, 2015
Proprietary Name: Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System
Common Name: Biopsy System
Regulation: 21 CFR 876.1075
Regulatory Class: II
Product Codes: KNW
Classification Name: Biopsy Instrument
- Predicate Devices: Mammotome® Biopsy System, K033700 Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System, K123259
{4}------------------------------------------------
Device Description: The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is an electromechanical breast biopsy device indicated to provide tissue samples for diagnostic sampling of beast abnormalities for histologic examination.
The Mammotome revolve® Dual VAB System is comprised of three primary subsystems:
- 1) a sterile, single-use probe
- 2) a reusable holster, and
- 3) a reusable control unit.
Intended Use: The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is indicated to provide tissue samples for diagnostic sampling of breast abnormalities.
- . The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.
- The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System is intended to provide breast tissue for histologic examination with partial removal of a palpable abnormality.
The extent of a histologic abnormality cannot always be readily determined from palpation or imaged appearance. Therefore, the extent of removal of the palpated or imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality, e.g., malignancy. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
In instances when a patient presents with a palpable abnormality that has been classified as benign through clinical and/or radiological criteria (e.g., fibroadenoma, fibrocystic lesion), the Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System may also be used to partially remove such palpable lesions. Whenever breast tissue is removed, histological evaluation of the tissue is the standard of care. When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
## Technological Characteristics:
The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System facilitates the diagnostic removal of tissue through a combination of vacuum and rotational/translational cutting functions. The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System utilizes the same primary subsystems as identified in the predicate devices to achieve its intended use:
{5}------------------------------------------------
- 1) a sterile single use probe containing a trocar tipped biopsy needle, rotating cutter specimen, collection chambers and vacuum tubing/valving;
- 2) a reusable holster, containing the drive motors, gear trains and user activation switches; and
- 3) a reusable control unit, containing the vacuum pump, power supply, valve actuators, user interface touchscreen, control electronics, and software.
In addition, several optional accessories are available including remote keypad and footswitch controls, a transport cart, and probe guides for various stereotactic (ST) tables.
The Mammotome revolve® Dual VAB System has been updated to integrate the Ultrasound modality as well as improvements on the performance reliability of the stereotactic functionality, while maintaining the basic technology, functionality and clinical outcomes of the predicate systems. Consistent with the configuration of the currently marketed systems, the Mammotome revolve® VAB System is configured for use in multiple imaging modalities, including Stereotactic and Ultrasound.
The system relies on software to operate many of its functions including utilization of closed loop control on cutter position. In both predicate software systems, the focus of the software is to aid in system set-up and facilitate biopsy functions. The Mammotome revolve® Dual VAB System software has been updated to include the Ultrasound functionality, as well as to provide improved GUI screen functionality and address market feedback.
A side-by-side comparison of the marketed and proposed devices is provided below.
{6}------------------------------------------------
## Side-by-Side Comparison to Legally Marketed Device
## Table 1a: Capital Equipment Side-by-Side Comparison of Mammotome revolve® Holster and Control Module to previous generation Mammotome® VAB Holster and Control Modules
| Device<br>Characteristics:<br>Holster / Control<br>Module | Marketed Device:<br>Mammotome® ST<br>System (K033700) | Marketed Device:<br>Mammotome® EX<br>System (K033700) | Marketed Device:<br>Mammotome revolve®<br>(ST) Biopsy System<br>(K123259) | Proposed Device:<br>Mammotome revolve®<br>(ST and U/S) Biopsy<br>System |
|-----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Breast Biopsy/complete,<br>partial removal of imaged<br>abnormality/ partial<br>removal of palpated<br>abnormality | Breast Biopsy/complete,<br>partial removal of imaged<br>abnormality/ partial<br>removal of palpated<br>abnormality | Breast Biopsy / complete,<br>partial removal of imaged<br>abnormality /partial<br>removal of palpated<br>abnormality | Breast Biopsy / complete,<br>partial removal of imaged<br>abnormality /partial<br>removal of palpated<br>abnormality |
| User interface<br>Mechanisms | Buttons on Remote Keypad,<br>Pedals on Footswitch,<br>Touchscreen | Buttons on Holster,<br>Buttons on Remote<br>Keypad, Pedals on<br>Footswitch, Touchscreen | Buttons on Holster, Buttons<br>on Remote Keypad, Pedals<br>on Footswitch, Touchscreen | Buttons on Holster, Buttons<br>on Remote Keypad, Pedals<br>on Footswitch, Touchscreen |
| Microprocessor and<br>upgradeable software | Yes | Yes | Yes | Yes |
| Display | Yes, LCD display | Yes, LCD display | Yes, LCD display | Yes, LCD display |
| Translational Cutter<br>Movement | Automatic or Semi<br>Automatic | Automatic or Semi<br>Automatic | Automatic | Automatic |
| Rotational Cutter<br>Movement | Automatic with cutter<br>advancement | Automatic with cutter<br>advancement | Automatic with cutter<br>advancement | Automatic with cutter<br>advancement |
| Rotational and<br>Translation Speed<br>control | Yes; Closed-loop control | Yes; Closed-loop control | Yes, Closed loop control | Yes, Closed loop control |
| Drive Train Type | Flexible Mechanical Drive<br>Cable | On board Motor and Gear<br>Train | On board Motor and Gear<br>Train | On board Motor and Gear<br>Train |
| Independent Lateral<br>and Axial Vacuum<br>System | Yes | Yes | Yes | Yes |
| Remote Footswitch /<br>Keypad Capability | Yes | Yes | Yes | Yes |
{7}------------------------------------------------
#### 510(k) Premarket Notification Mammotome revolve® Dual Vacuum Assisted Biopsy System
#### Disposables Side-by-Side Comparison of Mammotome revolve® Probes to Table 1b: previous generation Mammotome® VAB Probes
| Device Characteristics:<br>Probes | Marketed Device:<br>Mammotome® ST System<br>(K033700) | Marketed Device:<br>Mammotome® EX System<br>(K033700) | Marketed Device:<br>Mammotome revolve® (ST)<br>Biopsy System (K123259) | Proposed Device:<br>Mammotome revolve®<br>ST and U/S Biopsy System | | |
|-------------------------------------------|-------------------------------------------------------|-------------------------------------------------------|------------------------------------------------------------------------|--------------------------------------------------------------------|--|--|
| Needle Assembly | | | | | | |
| Needle Insertion<br>Method | Fired or Manual | Manual | Fired or Manual | Fired or Manual | | |
| Tip Type | Bladed trocar | Bladed trocar | Bladed trocar | Bladed trocar | | |
| Tip Material | Stainless steel | Stainless steel | Stainless steel | Stainless steel | | |
| Needle Configuration | Dual lumen | Dual lumen | Dual lumen | Dual lumen | | |
| Needle Material | Stainless steel | Stainless steel | Stainless steel | Stainless steel | | |
| Needle Diameter Sizes | 8G, 11G, and 14G | 8G, 11G, and 14G | 8G and 10G | 8G and 10G | | |
| Probe Housing | | | | | | |
| Vacuum Port<br>Attachment | Yes, tethered to control<br>module | Yes, tethered to control<br>module | Yes, tethered to Control<br>Module | Yes, tethered to Control<br>Module | | |
| Specimen Retrieval /<br>Collection Method | Manual | Manual | Automatic | Automatic | | |
| Housing Material | Plastic | Plastic | Plastic | Plastic | | |
| Probe Cutter Subassembly | | | | | | |
| Cutting Method | Rotation and translation of<br>inner cutter | Rotation and translation of<br>inner cutter | Rotation and translation of<br>inner cutter | Rotation and translation of<br>inner cutter | | |
| Tissue Transport<br>Method | Mechanical and Vacuum | Mechanical and Vacuum | Vacuum | Vacuum | | |
| Cutter Material | Stainless steel | Stainless steel | Stainless steel | Stainless steel | | |
| Packaging | | | | | | |
| Type | C-film with Tyvek Cover | C-film with Tyvek Cover | PETG with Tyvek Cover | PETG with Tyvek Cover | | |
| Sterilization | | | | | | |
| Method | Cobalt 60 Irradiation | Cobalt 60 Irradiation | Cobalt 60 Irradiation | Cobalt 60 Irradiation | | |
{8}------------------------------------------------
#### Performance testing:
To demonstrate substantial equivalence of the proposed device to the predicate devices, sideby-side comparison of tissue sample collection, using an in vivo porcine model, was performed. This model has historically been used to evaluate the ability of Mammotome® biopsy devices to collect tissue samples.
The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System and the predicate devices were each used to obtain tissue samples. Each sample was evaluated against the following criteria:
- Sample weight
- Sampling reliability
- Sample quality
Testing results confirmed that the Mammotome revolve® Vacuum Assisted Biopsy (VAB) System would retrieve a tissue sample comparable to that of the predicate device.
Additional testing was performed as well for integration of Ultrasound functionality and system enhancements. The Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System and the predicate device were each used to obtain test functionality and obtain tissue samples. Each sample was evaluated against the following criteria:
- . Sample weight
- . Sampling reliability
- Sample quality
- . Holster to Probe recognition
- . Holster/Control Module communication
Testing results confirmed that the Mammotome revolve® Vacuum Assisted Biopsy (VAB) System Ultrasound functionality and system enhancements was comparable to that of the predicate device.
## Conclusion:
Side-by-side comparison of the Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System to the predicate systems was performed on the technical characteristics of design, product components, materials of construction, system functionality, and clinical application (specifically, tissue sample characteristics). The test results of the Mammotome revolve® Dual Vacuum Assisted Biopsy (VAB) System demonstrated that it performed comparable to the safety and effectiveness of the predicate devices, and thus can be considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.