K152983 · FUJIFILM Sonosite, Inc. · IYN · Nov 12, 2015 · Radiology
Device Facts
Record ID
K152983
Device Name
FUJIFILM SonoSite iViz Ultrasound System
Applicant
FUJIFILM Sonosite, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 12, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The FUJIFILM SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal - OB Abdominal Pediatric Cardiac Adult Cardiac Pediatric
Device Story
Handheld, battery-operated digital ultrasound system; consists of tablet and active transducer. Acquires real-time ultrasound data; displays high-resolution images. Used in clinical care by physicians/healthcare professionals. Supports B-mode, M-mode, Color Power/Velocity Doppler, and Tissue Harmonic Imaging. Includes SonoHD3 speckle reduction. Connectivity via Bluetooth/Wi-Fi for image transfer and OTA software updates. Provides visual output for diagnostic evaluation and fluid flow analysis; aids clinical decision-making by enabling portable, point-of-care imaging.
Clinical Evidence
No clinical data; bench testing only. Device evaluated for electrical, thermal, mechanical, and EMC safety, biocompatibility, and acoustic output per FDA-recognized standards (ISO 10993-1, IEC 60601-1, IEC 60601-2-37, NEMA UD 2-2004).
Indicated for qualified physicians and healthcare professionals for ultrasound imaging or fluid flow analysis of the human body, specifically for fetal-OB, abdominal, pediatric, adult cardiac, and pediatric cardiac applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing to the right, arranged in a stacked, overlapping manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 12, 2015
FUJIFILM Sonosite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street. NW BUFFALO MN 55313
Re: K152983
Trade/Device Name: FUJIFILM SonoSite iViz Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 8, 2015 Received: October 13, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ocks
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) BB K152983
Device Name FUJIFILM SonoSite iViz Ultrasound System
#### Indications for Use (Describe)
The FUJIFILM SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Fetal - OB Abdominal Pediatric Cardiac Adult Cardiac Pediatric
Type of Use (Select one or both, as applicable)
| <div> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <line stroke="black" stroke-width="2" x1="2" x2="14" y1="2" y2="14"></line> <line stroke="black" stroke-width="2" x1="14" x2="2" y1="2" y2="14"></line> </svg> Registration Use (Part 21 CFR 601 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> Over-The-Counter Use (21 CFR 601 Subpart C) </div> |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Table 1.3-1: Diagnostic Ultrasound Indications for Use Form – FUJIFILM SonoSite iViz Ultrasound System
| System: | FUJIFILM SonoSite iViz Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | N/A | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: | | | | | | |
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | N | N | | | N | B+M; B+CD | 1-3 |
| Abdominal | N | N | | | N | B+M; B+CD | 1-3 |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | | | N | B+M; B+CD | 1-3 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | | | | | | | |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | | | | | | | |
| Musculo-skel. (Superfic.) | | | | | | | |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | N | N | | | N | B+M; B+CD | 1-3 |
| Cardiac Pediatric | N | N | | | N | B+M; B+CD | 1-3 |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | | | | | | | |
| Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under this appendix
#### Additional Comments:
1: Color Doppler includes Power/Velocity/Variance
2: Tissue Harmonic Imaging (THI)
3: SonoHD3 Imaging (Speckle Reduction)
Prescription Use (Per 21 CFR 801.109)
{4}------------------------------------------------
| System: | FUJIFILM SonoSite iViz Ultrasound System | | | | | | |
|-------------------------------------------------------|-------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | P21v/5-1 MHz Transducer | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human | | | | | | |
| | body as follows: | | | | | | |
| Clinical Application | Mode of Operation | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | N | N | | | N | B+M; B+CD | 1-3 |
| Abdominal | N | N | | | N | B+M; B+CD | 1-3 |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | | | N | B+M; B+CD | 1-3 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | | | | | | | |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | | | | | | | |
| Musculo-skel. (Superfic.) | | | | | | | |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | N | N | | | N | B+M; B+CD | 1-3 |
| Cardiac Pediatric | N | N | | | N | B+M; B+CD | 1-3 |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | | | | | | | |
| Other (spec.) | | | | | | | |
Table 1.3-2: Diagnostic Ultrasound Indications for Use Form – P21v/5-1 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
#### Additional Comments:
- 1: Color Doppler includes Power/Velocity/Variance
2: Tissue Harmonic Imaging (THI)
3: SonoHD3 Imaging (Speckle Reduction)
Prescription Use (Per 21 CFR 801.109)
{5}------------------------------------------------
## 510(K) Summary
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
#### 1) Submitter's name, address, telephone number, contact person:
FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
| Corresponding Official: | Patricia Liau<br>Manager, Regulatory Affairs |
|-------------------------|----------------------------------------------|
| E-mail: | Patricia.Liau@sonosite.com |
| Telephone: | (425) 951-6870 |
| Facsimile: | (425) 951-1201 |
| Date prepared: | September 28, 2015 |
#### 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
#### Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Name
SonoSite iViz Ultrasound System (subject to change)
#### Classification Names
| Name | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### 3) Identification of the predicate or legally marketed device:
| SonoSite Edge Ultrasound System | K133454 |
|---------------------------------|---------|
| Philips Nuvis Ultrasound System | K133833 |
#### 4) Device Description:
The SonoSite iViz Ultrasound System is a highly featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. iViz is a custom fabricated digital electronic handheld tablet that is highly portable, battery-operated, and consists of an active transducer that connects to and is controlled by the tablet. iViz supports Bluetooth and wireless network connectivity for image transfer and over-the-air (OTA) software updates.
#### 5) Intended Use:
The FUJIFILM SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-
{6}------------------------------------------------
continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
#### Fetal - OB Abdominal Pediatric Cardiac Adult Cardiac Pediatric
#### 6) Technological Characteristics:
SonoSite iViz and Edge, and Philips Nuvis Ultrasound Systems are Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.
| Feature | SonoSite iViz Ultrasound<br>System<br>(This submission) | SonoSite Edge<br>Ultrasound System<br>(K133454) | Philips Nuvis Ultrasound<br>System<br>(K133833) |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Intended Use | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body |
| Indications for<br>Use | Fetal - OB<br>Abdominal<br>Pediatric<br>Cardiac Adult<br>Cardiac Pediatric | Ophthalmic<br>Fetal - OB/GYN<br>Abdominal<br>Intraoperative (abdominal organs<br>and vascular)<br>Intra-operative (Neuro.)<br>Pediatric<br>Small Organ (breast, thyroid,<br>testicle, prostate)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-Rectal<br>Trans-Vaginal<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric<br>Trans-esophageal (cardiac)<br>Peripheral Vessel<br>Needle guidance | Fetal - OB<br>Abdominal<br>Other (Urology)<br>Other (Gynecology)<br>Other (Fetal Echo) |
| Transducer<br>Types | Phased Array | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array<br>Static Probes<br>Trans-esophageal | Curved Linear Array |
| Transducer<br>Frequency | 1.0 – 5.0 MHz | 1.0 - 15.0 MHz | 2.0 – 5.0 MHz |
| Modes of<br>Operation | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom<br>Combination Modes<br>SonoHD3 Noise Reduction<br>Velocity Color Doppler | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom<br>Combination Modes<br>Pulsed Wave (PW) Doppler<br>Continuous Wave (CW) Doppler<br>SonoHD2 Noise Reduction<br>SonoMB/MBe Image<br>Compounding<br>Steered CW Doppler<br>Velocity Color Doppler<br>Tissue Doppler Imaging (TDI) | B-mode Grayscale Imaging<br>Color Power Doppler<br>Combination Modes<br>Velocity Color Doppler |
{7}------------------------------------------------
| Feature | SonoSite iViz Ultrasound<br>System<br>(This submission) | SonoSite Edge<br>Ultrasound System<br>(K133454) | Philips Nuvis Ultrasound<br>System<br>(K133833) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| PW Doppler | Not available | Available | Not available |
| CW Doppler | Not available | Available | Not available |
| Patient<br>Contact<br>Materials | Transducers:<br>Polysulfone UDEL P1700<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber | Transducers:<br>Acrylonitrile-butadien-styrene<br>(ABS)<br>Cycoloy<br>Dow Medical Adhesive, Type A<br>Epoxy paste adhesive<br>Polyethylene (PE) Ionomer<br>Polyetheretherketone (PEEK)<br>Polysulfone UDEL P1700<br>Polyurethane<br>Poly-Vinyl-Chloride (PVC)<br>Silicone RTV Adhesive<br>Silicone Rubber<br>Urethane<br>Needle Guides:<br>Acetal copolymer<br>Acrylonitrile-butadien-styrene<br>(ABS) | Information not publicly available |
| System | iViz: | Edge: | Nuvis: |
| Characteristics | Handheld tablet<br>7", 1920 x 1200 pixels LCD<br>Operating system: Android<br>iViz ultrasound software running<br>as an "app" on tablet<br>System operates via battery<br>Wireless 802.11 support for image<br>transfer and over-the-air (OTA)<br>software updates | Handheld display and control<br>12.1", 800 x 600 pixels, LCD<br>Operating system: Windows CE<br>System operates via battery or AC<br>power<br>Wireless 802.11 support for image<br>transfer | Handheld tablet (COTS Nexus 7)<br>7", 1280 x 800 pixels LCD<br>Operating system: Android<br>Nuvis ultrasound software running<br>as an "app" on COTS tablet<br>System operates via battery<br>Wireless 802.11 support for image<br>transfer (non-DICOM) |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
## 7) Determination of Substantial Equivalence:
#### Summary of Non-Clinical Tests:
The iViz Ultrasound System has been evaluated for electrical, thermal, mechanical, and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development but were not limited to: Design Phase Reviews, Risk Assessment, Requirements Development, and Verification and Validation.
The iViz Ultrasound System is designed to comply with the following FDA recognized standards.
| Reference No. | Title |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1 | AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management process |
| IEC 60601-1 | AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 | AAMI / ANSI / IEC 60601-1-2:2007(R)2012, Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance - Collateral<br>standard: Electromagnetic compatibility - Requirements and tests (Edition 3) |
| IEC 60601-1-6 | IEC 60601-1-6 Edition 3.1 2013-10, Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential performance - Collateral standard:<br>Usability |
| IEC 60601-2-37 | IEC 60601-2-37:2007 Edition 2.0 2007-08, Medical electrical equipment - Part 2-37:<br>Particular requirements for the basic safety and essential performance of ultrasonic |
{8}------------------------------------------------
| Reference No. | Title |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | medical diagnostic and monitoring equipment |
| IEC 62304 | AAMI / ANSI / IEC 62304:2006, Medical device software - Software life cycle<br>processes |
| IEC 62359 | IEC 62359 Edition 2.0 2010-10-10, Ultrasonics – Field characterization – Test<br>methods for the determination of thermal and mechanical indices related to medical<br>diagnostic ultrasonic fields [Including: Technical corrigendum 1 (2011)] |
| ISO 14971 | ISO 14971:2007, Medical devices - Application of risk management to medical<br>devices |
| NEMA UD 2-2004 | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment |
## Summary of Clinical Tests:
The iViz Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.
## 8) Conclusion:
Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The iViz system and predicates meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence, and conform to applicable electromedical device safety standards. FUJIFILM SonoSite, Inc. believes that the iViz Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.