← Product Code [IYN](/productcode/IYN) · K152962

# HS70A Diagnostic Ultrasound System (K152962)

_Samsung Medison Co., Ltd. · IYN · Nov 2, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152962

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Nov 2, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The HS70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

## Device Story

HS70A is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers (linear, curved, endocavity, phased array) and displays it in B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and 3D/4D modes. Operated by healthcare professionals in clinical settings, it provides real-time imaging, anatomical measurements, and analysis packages to assist in diagnosis. The system features advanced processing tools like S-Harmonic, Elastoscan, S-Detect, and various 3D volume rendering/calculation tools. Output is displayed on-screen for clinician review, facilitating clinical decision-making through visualization of structures and fluid flow. It includes real-time acoustic output display (mechanical/thermal indices) for patient safety.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including acoustic output measurements, biocompatibility (ISO 10993-1), and electrical/electromagnetic safety (IEC 60601 series).

## Technological Characteristics

Mobile ultrasound system; supports linear, curved, endocavity, and phased array transducers (1.0–20.0 MHz). Modes: B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, Tissue Harmonic Imaging, Elastoscan. Patient contact materials tested per ISO 10993-1. Compliance: IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, NEMA UD-2, NEMA UD-3. Software-controlled with analysis packages for measurements.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- WS80A Diagnostic Ultrasound System ([K143089](/device/K143089.md))
- RS80A Diagnostic Ultrasound System ([K151663](/device/K151663.md))
- A35 Diagnostic Ultrasound System ([K150478](/device/K150478.md))
- ACCUVIX XG Diagnostic Ultrasound System ([K103397](/device/K103397.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three abstract human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2015

SAMSUNG MEDISON CO., LTD. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313

Re: K152962

Trade/Device Name: HS70A Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: October 6, 2015 Received: October 7, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K152962

Device Name HS70A Diagnostic Ultrasound System

Indications for Use (Describe)

The HS70A Diagnostic Ultrasound System and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intra-operative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-raginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric and Peripheral vessel.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

### 510(k) No.:

Device Name: HS70A Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                                    |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)                   |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                                    |
|                           | Fetal/Obstetrics (See Note 3) | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 4, 5, 6, 7, 8, 9, 11      |
|                           | Abdominal (See Note 10)       | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 4, 6, 7, 8, 9, 11, 12, 13 |
|                           | Intra-operative (See Note 6)  | N | N                                                 | N   | N   | N                 | Note 1               | Notes 7, 9, 11                     |
|                           | Intra-operative (Neuro.)      | N | N                                                 | N   | N   | N                 | Note 1               | Notes 7, 9, 11                     |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                                    |
|                           | Pediatric                     | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 4, 6, 7, 8, 9, 11         |
|                           | Small Organ (See Note 5)      | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 12      |
|                           | Neonatal Cephalic             | N | N                                                 | N   | N   | N                 | Note 1               | Notes 7, 8, 9, 11                  |
|                           | Adult Cephalic                | N | N                                                 | N   | N   | N                 | Note 1               | Notes 7                            |
|                           | Trans-rectal                  | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 7, 8, 9, 12               |
|                           | Trans-vaginal                 | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 7, 8, 9, 12               |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                                    |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                                    |
|                           | Musculo-skel. (Convent.)      | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11         |
|                           | Musculo-skel. (Superfic.)     | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11         |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                                    |
|                           | Other (See Note 13)           | N | N                                                 | N   | N   | N                 | Note 1               | Notes 2, 7, 8, 9, 12               |
|                           | Cardiac Adult                 | N | N                                                 | N   | N   | N                 | Note 1               | Notes 4, 7                         |
| Cardiac                   | Cardiac Pediatric             | N | N                                                 | N   | N   | N                 | Note 1               | Notes 4, 7                         |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                                    |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                                    |
| Peripheral<br>Vessel      | Peripheral vessel             | N | N                                                 | N   | N   | N                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11          |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{4}------------------------------------------------

510(k) No.:

### Device Name: CA1-7A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application     |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                          |  |  |
|--------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|--------------------------|--|--|
| General                  | Specific                      | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)         |  |  |
| (Track I only)           | (Tracks I & III)              |                                                   |   |     |     |                   |                      |                          |  |  |
| Ophthalmic               | Ophthalmic                    |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11  |  |  |
|                          | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11, 13 |  |  |
|                          | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                          |  |  |
| Fetal Imaging<br>& Other | Laparoscopic                  |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11     |  |  |
|                          | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Adult Cephalic                |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Trans-rectal                  |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Trans-urethral                |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Musculo-skel. (Convent.)      | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11     |  |  |
|                          | Musculo-skel. (Superfic.)     | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11     |  |  |
|                          | Intra-luminal                 |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                          |  |  |
| Cardiac                  | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Other (spec.)                 |                                                   |   |     |     |                   |                      |                          |  |  |
| Peripheral<br>Vessel     | Peripheral vessel             |                                                   |   |     |     |                   |                      |                          |  |  |
|                          | Other (spec.)                 |                                                   |   |     |     |                   |                      |                          |  |  |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

- Note 3: Includes infertility monitoring of follicle development
Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{5}------------------------------------------------

510(k) No.:

### Device Name: CA2-8A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11 |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11 |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                         |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                         |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                         |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                         |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                         |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                         |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                         |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                         |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                         |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                         |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |
| Peripheral<br>Vessel      | Peripheral vessel             |   |                                                   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate

{6}------------------------------------------------

### 510(k) No.:

### Device Name: CF4-9 for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application     |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |  |
|--------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|--|
| General                  | Specific                      | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |  |
| (Track I only)           | (Tracks I & III)              |   |                                                   |     |     |                   |                      |                  |  |
| Ophthalmic               | Ophthalmic                    |   |                                                   |     |     |                   |                      |                  |  |
|                          | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                  |  |
|                          | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 8, 9, 11   |  |
|                          | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                  |  |
|                          | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                  |  |
| Fetal Imaging<br>& Other | Laparoscopic                  |   |                                                   |     |     |                   |                      |                  |  |
|                          | Pediatric                     | P | P                                                 | P   | P   | P                 | Note 1               | Notes 8, 9, 11   |  |
|                          | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                  |  |
|                          | Neonatal Cephalic             | P | P                                                 | P   | P   | P                 | Note 1               | Notes 8, 9, 11   |  |
|                          | Adult Cephalic                |   |                                                   |     |     |                   |                      |                  |  |
|                          | Trans-rectal                  |   |                                                   |     |     |                   |                      |                  |  |
|                          | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                  |  |
|                          | Trans-urethral                |   |                                                   |     |     |                   |                      |                  |  |
|                          | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                  |  |
|                          | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                  |  |
|                          | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                  |  |
|                          | Intra-luminal                 |   |                                                   |     |     |                   |                      |                  |  |
|                          | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                  |  |
|                          | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                  |  |
| Cardiac                  | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                  |  |
|                          | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                  |  |
|                          | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |  |
| Peripheral<br>Vessel     | Peripheral vessel             | P | P                                                 | P   | P   | P                 | Note 1               | Notes 8, 9, 11   |  |
|                          | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |  |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 4: Color M-mode

Note 8: 3D imaging

{7}------------------------------------------------

510(k) No.:

Device Name: E3-12A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                   |  |  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------|--|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)  |  |  |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Notes 7, 8, 9     |  |  |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 7, 8, 9, 12 |  |  |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                   |  |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Trans-rectal                  | P | P                                                 | P   |     | P                 | Note 1               | Notes 7, 8, 9, 12 |  |  |
|                           | Trans-vaginal                 | P | P                                                 | P   |     | P                 | Note 1               | Notes 7, 8, 9, 12 |  |  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Other (See Note 13)           | P | P                                                 | P   |     | P                 | Note 1               | Notes 7, 8, 9, 12 |  |  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                   |  |  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                   |  |  |
| Peripheral<br>Vessel      | Peripheral vessel             |   |                                                   |     |     |                   |                      |                   |  |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                   |  |  |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{8}------------------------------------------------

510(k) No.:

### Device Name: VR5-9 for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                      |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)     |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                      |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9      |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                      |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                      |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                      |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                      |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                      |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                      |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-rectal                  | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Trans-vaginal                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                      |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                      |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                      |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                      |
|                           | Other (See Note 13)           | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                      |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                      |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                      |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                      |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                      |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                      |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate

{9}------------------------------------------------

510(k) No.:

Device Name: L3-12A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                            |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)           |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                            |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 4, 6, 7, 9, 11    |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                            |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                            |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                            |
|                           | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 4, 6, 7, 9, 11    |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11, 12 |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                            |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                            |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                            |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                            |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                            |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                            |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                            |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                            |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                            |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+CW, B+CW, B+C+PW, B+C+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{10}------------------------------------------------

510(k) No.:

### Device Name: LA3-16A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application     |                              | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                            |  |
|--------------------------|------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------------|--|
| General                  | Specific                     | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)           |  |
| (Track I only)           | (Tracks I & III)             |                                                   |   |     |     |                   |                      |                            |  |
| Ophthalmic               | Ophthalmic                   |                                                   |   |     |     |                   |                      |                            |  |
|                          | Fetal (See Note 3)           |                                                   |   |     |     |                   |                      |                            |  |
|                          | Abdominal (See Note 10)      |                                                   |   |     |     |                   |                      |                            |  |
|                          | Intra-operative (See Note 6) |                                                   |   |     |     |                   |                      |                            |  |
|                          | Intra-operative (Neuro.)     |                                                   |   |     |     |                   |                      |                            |  |
| Fetal Imaging<br>& Other | Laparoscopic                 |                                                   |   |     |     |                   |                      |                            |  |
|                          | Pediatric                    |                                                   |   |     |     |                   |                      |                            |  |
|                          | Small Organ (See Note 5)     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11, 12 |  |
|                          | Neonatal Cephalic            |                                                   |   |     |     |                   |                      |                            |  |
|                          | Adult Cephalic               |                                                   |   |     |     |                   |                      |                            |  |
|                          | Trans-rectal                 |                                                   |   |     |     |                   |                      |                            |  |
|                          | Trans-vaginal                |                                                   |   |     |     |                   |                      |                            |  |
|                          | Trans-urethral               |                                                   |   |     |     |                   |                      |                            |  |
|                          | Trans-esoph. (non-Cardiac)   |                                                   |   |     |     |                   |                      |                            |  |
|                          | Musculo-skel. (Convent.)     | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                          | Musculo-skel. (Superfic.)    | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                          | Intra-luminal                |                                                   |   |     |     |                   |                      |                            |  |
|                          | Other (See Note 13)          |                                                   |   |     |     |                   |                      |                            |  |
|                          | Cardiac Adult                |                                                   |   |     |     |                   |                      |                            |  |
| Cardiac                  | Cardiac Pediatric            |                                                   |   |     |     |                   |                      |                            |  |
|                          | Trans-esophageal (Cardiac)   |                                                   |   |     |     |                   |                      |                            |  |
|                          | Other (spec.)                |                                                   |   |     |     |                   |                      |                            |  |
| Peripheral<br>Vessel     | Peripheral vessel            | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |  |
|                          | Other (spec.)                |                                                   |   |     |     |                   |                      |                            |  |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development
- Note 4: Color M-mode
- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

- Note 11: Panoramic imaging
Note 12: ElastoScan

{11}------------------------------------------------

510(k) No.:

### Device Name: LA3-16AI for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                              |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |
|---------------------------|------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III) | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                   |   |                                                   |     |     |                   |                      |                  |
|                           | Fetal (See Note 3)           |   |                                                   |     |     |                   |                      |                  |
|                           | Abdominal                    |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6) | P | P                                                 | P   |     | P                 | Note 1               | Note 7, 9, 11 b) |
|                           | Intra-operative (Neuro.)     | P | P                                                 | P   |     | P                 | Note 1               | Note 7, 9, 11 b) |
| Fetal Imaging<br>& Other  | Laparoscopic                 |   |                                                   |     |     |                   |                      |                  |
|                           | Pediatric                    |   |                                                   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)     |   |                                                   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic            |   |                                                   |     |     |                   |                      |                  |
|                           | Adult Cephalic               |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-rectal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-vaginal                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-urethral               |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)   |   |                                                   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)     | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11 a) |
|                           | Musculo-skel. (Superfic.)    | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11 a) |
|                           | Intra-luminal                |   |                                                   |     |     |                   |                      |                  |
|                           | Other (See Note 13)          |   |                                                   |     |     |                   |                      |                  |
|                           | Cardiac Adult                |   |                                                   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric            |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)   |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel            | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11 a) |
|                           | Other (spec.)                |   |                                                   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K1516639 and K103397 9) (LS5-13); E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

Note 2: Includes imaging for guidance of biopsy

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: MultiVision (Spatial Compound Imaging)

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

{12}------------------------------------------------

510(k) No.:

### Device Name: LA2-9A for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                           |  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|---------------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)          |  |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                           |  |
|                           | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                           |  |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11 |  |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                           |  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                           |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                           |  |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                           |  |
|                           | Small Organ (See Note 5)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11 |  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                           |  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                           |  |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                           |  |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                           |  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                           |  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                           |  |
|                           | Musculo-skel. (Convent.)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11 |  |
|                           | Musculo-skel. (Superfic.)     | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11 |  |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                           |  |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                           |  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                           |  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                           |  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                           |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                           |  |
| Peripheral<br>Vessel      | Peripheral vessel             | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11 |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                           |  |

N= new indication; P= previously cleared by FDA K151663; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development

- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate

Note 4: Color M-mode

{13}------------------------------------------------

### 510(k) No.:

### Device Name: PE2-4 for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |          |           |           |  |  |
|----------------------|-------------------------------|---|---------------------------------------------------|-----|-----|----------|-----------|-----------|--|--|
| General              | Specific                      | B | M                                                 | PWD | CWD | Color    | Combined* | Other     |  |  |
| (Track I only)       | (Tracks I & III)              |   |                                                   |     |     | Doppler* | (Spec.)   | (Spec.)   |  |  |
| Ophthalmic           | Ophthalmic                    |   |                                                   |     |     |          |           |           |  |  |
|                      | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |          |           |           |  |  |
|                      | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P        | Note 1    | Note 7    |  |  |
|                      | Intra-operative (See Note 6)  |   |                                                   |     |     |          |           |           |  |  |
|                      | Intra-operative (Neuro.)      |   |                                                   |     |     |          |           |           |  |  |
| Fetal Imaging        | Laparoscopic                  |   |                                                   |     |     |          |           |           |  |  |
| &<br>Other           | Pediatric                     |   |                                                   |     |     |          |           |           |  |  |
|                      | Small Organ (See Note 5)      |   |                                                   |     |     |          |           |           |  |  |
|                      | Neonatal Cephalic             | P | P                                                 | P   | P   | P        | Note 1    | Note 7    |  |  |
|                      | Adult Cephalic                | P | P                                                 | P   | P   | P        | Note 1    | Note 7    |  |  |
|                      | Trans-rectal                  |   |                                                   |     |     |          |           |           |  |  |
|                      | Trans-vaginal                 |   |                                                   |     |     |          |           |           |  |  |
|                      | Trans-urethral                |   |                                                   |     |     |          |           |           |  |  |
|                      | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |          |           |           |  |  |
|                      | Musculo-skel. (Convent.)      |   |                                                   |     |     |          |           |           |  |  |
|                      | Musculo-skel. (Superfic.)     |   |                                                   |     |     |          |           |           |  |  |
|                      | Intra-luminal                 |   |                                                   |     |     |          |           |           |  |  |
|                      | Other (See Note 13)           |   |                                                   |     |     |          |           |           |  |  |
|                      | Cardiac Adult                 | P | P                                                 | P   | P   | P        | Note 1    | Note 4, 7 |  |  |
| Cardiac              | Cardiac Pediatric             | P | P                                                 | P   | P   | P        | Note 1    | Note 4, 7 |  |  |
|                      | Trans-esophageal (Cardiac)    |   |                                                   |     |     |          |           |           |  |  |
|                      | Other (spec.)                 |   |                                                   |     |     |          |           |           |  |  |
| Peripheral           | Peripheral vessel             |   |                                                   |     |     |          |           |           |  |  |
| Vessel               | Other (spec.)                 |   |                                                   |     |     |          |           |           |  |  |

N= new indication; P= previously cleared by FDA K143089; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+CW, B+C+PW, B+PD+PW, B+DPD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E

- Note 2: Includes imaging for guidance of biopsy
- Note 3: Includes infertility monitoring of follicle development

- Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients
- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate

Note 4: Color M-mode

{14}------------------------------------------------

510(k) No.:

### Device Name: PA3-8B for use with HS70A

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                  |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Note 7           |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                  |
|                           | Laparoscopic                  |   |                                                   |     |     |                   |                      |                  |
|                           | Pediatric                     | P | P                                                 | P   | P   | P                 | Note 1               | Note 7           |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             | N | N                                                 | N   | N   | N                 | Note 1               | Note 7           |
|                           | Adult Cephalic                | P | P                                                 | P   | P   | P                 | Note 1               | Note 7           |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |   |                                                   |     |     |…

---

**Source:** [https://fda-staging.innolitics.com/device/K152962](https://fda-staging.innolitics.com/device/K152962)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).