The PXT701 Disposable Temporary Pacing Cable is designed to connect a pacing system analyzer (PSA) or temporary pacemaker to a cardiac lead.
Device Story
Disposable electrical extension cable; transmits signals/power between pacing system analyzer (PSA) or temporary pacemaker and patient-connected cardiac leads. Features bipolar design with color-coded alligator clips (red=positive, black=negative) for lead attachment and safety plug instrument connector. Length 2.5 meters. Used in clinical settings; operated by healthcare professionals. Acts as passive conduit for electrical signals. Facilitates temporary pacing therapy; enables connection of external pacing equipment to cardiac leads.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per ANSI/AAMI EC53:2013, including dielectric withstand voltage, defibrillation withstand, cable/leadwire noise, sink current, connector tensile strength, flex life, mate/unmate cycles, contact resistance, and retention force. Results met acceptance criteria; cable noise was equivalent to the predicate device.
Technological Characteristics
22-gauge metallic wire with plastic insulation; bipolar configuration; alligator clip patient-end connectors; safety plug instrument-end connector. 2.5-meter length. Sterile (Ethylene Oxide per ISO 10993-7). Performance testing per ANSI/AAMI EC53:2013. Disposable, single-patient use.
Indications for Use
Indicated for connecting a pacing system analyzer (PSA) or temporary pacemaker to a cardiac pacing lead in patients requiring temporary cardiac pacing.
Regulatory Classification
Identification
A patient transducer and electrode cable (including connector) is an electrical conductor used to transmit signals from, or power or excitation signals to, patient-connected electrodes or transducers.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a depiction of a caduceus, which is a traditional symbol of medicine, with a modern and abstract design.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 05, 2016
Tiger Medical Products % Pamela Papineau President Delphi Medical Device Consulting, Inc 5 Whitcomb Avenue Ayer, Massachusetts 01432
Re: K152947
Trade/Device Name: Disposable Temporary Pacing Cable Regulation Number: 21 CFR 870.2900 Regulation Name: Patient Transducer and Electrode Cable (Including Connector) Regulatory Class: Class II Product Code: DSA Dated: February 16, 2016 Received: February 19, 2016
Dear Pamela Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
#### K152947
Device Name
Disposable Temporary Pacing Cable
Indications for Use (Describe)
The PXT701 Disposable Temporary Pacing Cable is designed to connect a pacing system analyzer (PSA) or temporary pacemaker to a cardiac lead.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
PSC Publishing Services (301) 443-6740 FF
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
## 510(k) Summary (K152947) - Page 1 of 3
#### General Information
| Owner's Name: | Tiger Medical Products |
|-------------------------|---------------------------------------------------------------------------------------------------------------------|
| Address: | Jaje Plaza, Suite 703<br>1717 North Sichuan Road<br>Shanghai, China 200080 |
| Telephone: | 404-696-6999 |
| Contact Person: | Pamela Papineau, RAC (US, EU, CAN)<br>Delphi Medical Device Consulting, Inc.<br>5 Whitcomb Avenue<br>Ayer, MA 01432 |
| Telephone Number: | (978) 772-3552 |
| Fax Number: | (978) 796-5460 |
| Subject Device: | |
| Trade Name: | Disposable Temporary Pacing Cable |
| Common Name: | Component to diagnostic or physiological monitoring devices |
| Product Code: | DSA |
| FDA Regulation: | 21 CFR 870.2900 – Cable, Transducer and Electrode, Patient<br>(Including Connector) |
| Device Classification: | Class II |
| Predicate Device: | |
| Product Name: | Safe Connect Disposable Surgical Cable |
| Common Name: | Component to diagnostic or physiological monitoring devices |
| Product Code: | DSA |
| FDA Regulation: | 21 CFR 870.2900 – Cable, Transducer and Electrode, Patient<br>(Including Connector) |
| Device Classification: | Class II |
| Premarket Notification: | K971968 |
### Indications for Use
The PXT701 Disposable Pacing Cable is designed to connect a pacing system analyzer (PSA) or temporary pacemaker to a cardiac pacing lead.
## Device Description:
The Disposable Temporary Pacing Cable is an electrical extension cable used to transmit signal from, or power or excitation signal to patient-connected electrodes. The cable is bipolar, with alligator clips at the end of the cable that will be attached to the patient leads, and a safety plug instrument connector at the other end that allows electrical connection to a pacing system analyzer or to an external pacing system (temporary pacemaker). The alligator clips are color coded (red = positive; black = negative). The cable is approximately 2.5 meters (8 feet) long, and is supplied sterile. The cable is a disposable, single-patient use device.
{4}------------------------------------------------
# 510(k) Summary (K152947) – Page 2 of 3
## Substantial Equivalence
The Tiger Medical Products Disposable Temporary Pacing Cable is substantially equivalent to the Remington Medical, Inc. Safe Connect Surgical Cable (K971968). Substantial equivalence, which is summarized below, is based on indications for use and device physical and technological characteristics.
| | Tiger Medical Products | Remington Medical, Inc. |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Disposable Temporary Pacing | Disposable Surgical Cable |
| | Cable (current submission) | (K971968) |
| Device Common/Usual | Component to diagnostic or<br>physiological monitoring devices | Component to diagnostic or<br>physiological monitoring devices |
| Name | | |
| Device Class | Class II | Class II |
| Product Code / Regulation | DSA / 21 CFR 870.2900 | DSA / 21 CFR 870.2900 |
| Regulation Name | Cable, Transducer and Electrode,<br>Patient (Including Connector) | Cable, Transducer and Electrode,<br>Patient (Including Connector) |
| Prescription Use | Rx Only | Rx Only |
| Intended Use | For temporary connection of a<br>pacing system analyser (PSA) or<br>temporary pace maker to a cardiac<br>pacing lead. | For temporary connection of a<br>pacing system analyser (PSA) or<br>temporary pace maker to a cardiac<br>pacing lead. |
| Indications for Use | To connect a pacing system<br>analyser (PSA) or temporary pace<br>maker to a cardiac pacing lead. | To act as a conduit for the electrical<br>signal generated from a Pacing<br>System Analyzer (PSA) or External<br>pacing Generator (EPG) to a<br>permanent pacemaker lead or<br>epicardial wire. |
| Statement | | |
| (ref. Proposed and | | |
| Predicate Device Labeling; | | |
| Appendix B) | | |
| Overall Cable Length | 2.5 meters / 8 feet | 2.5 meters / 8 feet |
| Lead Wire Cable | Alligator clips with color coded<br>boots (black = negative; red =<br>positive) | Alligator clips with color coded<br>boots (black = negative; red =<br>positive) |
| Connectors | | |
| Instrument Connector | "safe-connect" style | "safe-connect" style |
| Wire Type | 22-gauge metallic wire with<br>plastic insulation | 22-gauge metallic wire with plastic<br>insulation |
| Lead Compatibility | Single-Chamber Unipolar Leads;<br>Single-Chamber Bipolar Leads;<br>Dual Chamber Unipolar Leads;<br>Dual Chamber Bipolar Leads | Single-Chamber Unipolar Leads;<br>Single-Chamber Bipolar Leads:<br>Dual Chamber Unipolar Leads;<br>Dual Chamber Bipolar Leads |
| Performance Testing per | Yes | Yes |
| ANSI/AAMI EC53? | | |
| Sterile Device? | Yes | Yes |
| Sterilization Type | Ethylene Oxide | Ethylene Oxide |
| EO Sterilization Residuals | Per ISO 10993-7 | Per ISO 10993-7 |
| Disposable / Reusable | Disposable | Disposable |
| Shelf Life | 5 years | 3 years |
{5}------------------------------------------------
### 510(k) Summary (K152947) - Page 3 of 3
### Performance Testing:
Performance testing for the Tiger Medical Products Disposable Temporary Pacing Cable consists of physical integrity and electrical resistance in accordance with ANSI/AAMI EC53:2013 (dielectric withstand voltage, defibrillation withstand test, cable and leadwire noise, sink current, connector tensile strength, connector flex life testing, number of connector mate/unmate cycles, contact resistance, connector retention force and patient leadwire resistance). The Tiger Medical Products Disposable Temporary Pacing Cable has met all applicable acceptance criteria as defined in ANSVAAMI EC53:2013 with the exception of cable noise. The test results for cable noise were equivalent to the results of the same testing performed on the predicate device.
### Conclusion:
The Tiger Medical Products Disposable Temporary Pacing Cable has been demonstrated to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.