← Product Code [FRO](/productcode/FRO) · K152926

# AQUACEL Ag Surgical SP Dressing (K152926)

_Convatec Limited · FRO · Jan 21, 2016 · SU · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152926

## Device Facts

- **Applicant:** Convatec Limited
- **Product Code:** [FRO](/productcode/FRO.md)
- **Decision Date:** Jan 21, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Indications for Use

AQUACEL® Ag Surgical SP may be used for the management of wounds healing by primary intent (e.g., traumatic and elective post operative wounds/incisions) and as an effective barrier to bacterial penetration to help reduce infection.

## Device Story

Sterile, single-use post-operative wound dressing; comprises central Hydrofiber™ pad containing ionic silver, stitch-bonded with nylon/elastane yarns. Pad encapsulated between top hydrocolloid adhesive layer with outer polyurethane film and windowed skin-contact layer (polyurethane film, acrylic, and hydrocolloid). Dressing absorbs/retains serosanguinous exudate, transforming into soft gel to maintain moist healing environment. Outer polyurethane film acts as bacterial/viral barrier. Designed to conform to wound geometry during movement/edema. Applied by healthcare professionals in clinical settings. Benefits include infection reduction, secure adhesion, and non-traumatic removal.

## Clinical Evidence

Bench testing only. Functional and structural design parameters (fluid handling, adhesive properties) compared to predicate. Biocompatibility testing performed per ISO 10993.

## Technological Characteristics

Composite dressing; materials: Hydrofiber™ (sodium carboxymethylcellulose) with 1.2% ionic silver, nylon/elastane yarns, proprietary hydrocolloid adhesive, polyurethane film, and acrylic adhesive. Sterile (gamma irradiation). No energy source. Slimmer profile (27 mil border) compared to predicate.

## Predicate Devices

- AQUACEL™ Ag Surgical Dressing ([K091034](/device/K091034.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three intertwined snakes and a pair of wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 21, 2016

ConvaTec Limited Aaron Sibley Regulatory Affairs Manager First Avenue Deeside Industrial Park Flintshire, CH5 2NU United Kingdom

Re: K152926

Trade/Device Name: AQUACEL Ag Surgical SP Dressing Regulatory Class: Unclassified Product Code: FRO Dated: December 21, 2015 Received: December 23, 2015

Dear Mr. Sibley:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152926

#### Device Name AQUACEL® Ag Surgical SP Dressing

#### Indications for Use (Describe)

AQUACEL® Ag Surgical SP may be used for the management of wounds healing by primary intent (e.g., traumatic and elective post operative wounds/incisions) and as an effective barrier to bacterial penetration to help reduce infection.

Type of Use (Select one or both, as applicable)

| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for ConvaTec. The logo consists of the word "ConvaTec" in blue, followed by a blue circle containing three vertical lines topped with a horizontal line, resembling a simplified Greek temple. The logo is simple and clean, with a focus on the company name and a symbolic representation.

ConvaTec Limited First Avenue Deeside Industrial Park Flintshire CH5 2NU United Kingdom

| 510(k) Summary       | K152926                                                                                                                              |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device:      | AQUACEL™ Ag Surgical SP Dressing                                                                                                     |
| Date Prepared:       | January 21, 2016                                                                                                                     |
| Applicant:           | ConvaTec Limited<br>First Avenue<br>Deeside Industrial Park<br>Deeside<br>Flintshire CH5 2NU<br>UK                                   |
| Contact:             | Aaron Sibley<br>Regulatory Affairs Manager<br>Tel: +44 (0)1494 816054<br>Fax: +44 (0)1244 584311<br>Email: aaron.sibley@convatec.com |
| Device Trade Name:   | AQUACEL™ Ag Surgical SP Dressing                                                                                                     |
| Common Name:         | Surgical Cover Dressing                                                                                                              |
| Classification Name: | Dressing, Wound, Drug                                                                                                                |
| Product Code:        | FRO                                                                                                                                  |
| Device Class:        | Unclassified                                                                                                                         |
| Predicate Device     |                                                                                                                                      |
| Device Trade Name:   | AQUACEL™ Ag Surgical Dressing                                                                                                        |
| Common Name:         | Surgical Cover Dressing                                                                                                              |
| Classification Name: | Dressing, Wound, Drug                                                                                                                |
| Product Code:        | FRO                                                                                                                                  |
| Device Class:        | Unclassified                                                                                                                         |

510(k) Substantial Equivalence: K091034 - determined substantially equivalent on December 16, 2009

{4}------------------------------------------------

### Device Description:

AQUACEL™ Ag Surgical SP dressing is a sterile post-operative dressing comprising of an inner (wound contact) non-woven pad composed of Hydrofiber™ technology and ionic silver which is stitch bonded with nylon and elastane yarns. The pad is held in place between a top layer of hydrocolloid adhesive which is bound to an outer polyurethane film and a windowed skin contact layer consisting of polyurethane film sandwiched between one layer of acrylic and one layer of Hydrocolloid contacting the skin.

The outer layer of polyurethane film of AQUACEL™ Ag Surgical SP dressing provides a bacterial and viral barrier to reduce the risk of infection provided the dressing remains intact and there is no leakage.

The highly absorbent central Hydrofiber™ pad absorbs and retains serosanguinous leakage and transforms into a soft gel which maintains a moist environment to support the body's healing process.

The hydrocolloid component of the dressing provides secure, skin friendly adhesion and supports non traumatic removal of the dressing while minimizing damage to the intact surrounding skin.

AQUACEL™ Ag Surgical SP dressing is a waterproof, bacterial and viral barrier dressing which is designed to conform to changes in the wound/incision geometry during body movement and post-operative edema.

The dressing is intended for single use and is provided sterile (via a gamma irradiation process).

This Special 510(k) details the modifications of the predicate device, AQUACEL™ Ag Surgical dressing, which form the AQUACEL™ Ag Surgical SP dressing. These modifications do not alter the intended use or indications for use.

#### Indications for Use:

Under the supervision of a healthcare professional, AQUACEL™ Ag Surgical SP cover dressing may be used for the management of:

Wounds healing by primary intent (e.g. traumatic and elective post operative wounds/incisions) and as an effective barrier to bacterial penetration to help reduce infection.

The indications for use are identical to the predicate device.

# Summary of Technological Characteristics:

AQUACEL™ Ag Surgical SP dressing has the same intended use and indications for use as the predicate device but has a slimmer profile than the predicate device. This is achieved by replacing an internal hydrocolloid layer of the composite dressing with a thinner acrylic adhesive, and reducing the profile of the two remaining Hydrocolloid layers. The differences in the composition, as detailed below, is explained by the reduction of Hydrocolloid adhesive in the modified device as compared to the predicate device. This necessarily increases the %w/w of the other components overall.

{5}------------------------------------------------

There are no differences in operation and directions for use between AQUACEL™ Ag Surgical SP dressing and the predicate device.

|                         | Predicate Device                                                                                                                                                                                                              | Modified Device                                                                                                                                                                                                                                                                                                                                      |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                         | AQUACEL™ Ag Surgical                                                                                                                                                                                                          | AQUACEL™ Ag Surgical                                                                                                                                                                                                                                                                                                                                 |
|                         | Dressing K091034                                                                                                                                                                                                              | SP Dressing                                                                                                                                                                                                                                                                                                                                          |
| Design                  | A composite dressing incorporating a<br>central silver containing absorbent<br>pad within a Hydrocolloid adhesive<br>construct with an approximate<br>border thickness of 48 mil.                                             | A composite dressing incorporating a<br>central silver containing absorbent<br>pad within a Hydrocolloid adhesive<br>construct with an approximate<br>border thickness of 27 mil. This<br>reduction is achieved by reducing<br>thickness of two Hydrocolloid layers<br>and replacing an internal<br>Hydrocolloid layer with an acrylic tie<br>layer. |
| Materials               | Layer 1: Polyurethane and<br>Hydrocolloid adhesive<br>Layer 2: Hydrofiber™ pad with 1.2%<br>ionic silver<br>Layer 3: Hydrocolloid adhesive<br>Layer 4: Polyurethane<br>Layer 5: Hydrocolloid adhesive                         | Layer 1: Polyurethane and<br>Hydrocolloid adhesive<br>Layer 2: Hydrofiber™ pad with 1.2%<br>ionic silver<br>Layer 3: 100% coat acrylic adhesive<br>Layer 4: Polyurethane<br>Layer 5: Hydrocolloid adhesive                                                                                                                                           |
| Chemical<br>Composition | Approximates:<br>13.32% Silver Hydrofiber™ (sodium<br>carboxymethylcellulose with silver)<br>0.71% Nylon yarn<br>0.89% Elastane (polyurethane) yarn<br>83.83% proprietary Hydrocolloid<br>adhesive<br>1.26% Polyurethane Film | Approximates:<br>19.65% Silver Hydrofiber™ (sodium<br>carboxymethylcellulose with silver)<br>0.93% Nylon yarn<br>1.12% Elastane (polyurethane) yarn<br>71.47% proprietary Hydrocolloid<br>adhesive<br>3.43% Polyurethane Film<br>3.4% Acrylic adhesive                                                                                               |
| Energy                  | No energy sources are                                                                                                                                                                                                         | No energy sources are                                                                                                                                                                                                                                                                                                                                |
| Source                  | utilized                                                                                                                                                                                                                      | utilized                                                                                                                                                                                                                                                                                                                                             |

# Summary of Performance (Non-Clinical Testing) Data:

Non-clinical testing of the subject device for functional and structural design parameters (such as fluid handling and adhesive properties) has been performed. In this testing, the device's performance has been found to be comparable to the aforementioned predicate device.

The device has also been biologically evaluated in accordance with the US Food and Drug Administration's guidance entitled Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', issued May 1, 1995, and has been found safe in such respect for its intended use.

In conclusion, the subject device has been demonstrated as safe and effective and substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K152926](https://fda-staging.innolitics.com/device/K152926)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).