The ARC InstaTemp™ is an infrared thermometer that measures temperature from the forehead in both infants and adults without contacting the human body. The ARC InstaTemp MD™ is an infrared thermometer that measures temperature from the forehead in both infants and adults without contacting the human body. It is intended for use by a Health Care Professional in a clinical environment. This device is not intended for home use.
Device Story
Non-contact infrared thermometer; detects radiated infrared energy from temporal artery in forehead; distance 1-3cm. Internal algorithm converts forehead surface temperature to human body temperature; displays result on LCD. Used by adults at home (InstaTemp) or healthcare professionals in clinical settings (InstaTemp MD). Provides rapid, non-intrusive temperature readings; aids clinical decision-making by identifying febrile/non-febrile states. Benefits patient by enabling hygienic, non-invasive temperature monitoring.
Clinical Evidence
Clinical trial conducted on 136 subjects to verify clinical accuracy per ISO 80601-2-56, section 201.102. Results demonstrated reproducibility and accuracy for febrile and non-febrile patients consistent with existing clinical thermometers.
Technological Characteristics
Infrared sensor; polycarbonate housing; 2x AAAA battery power; LCD display; 1-3cm measurement distance; 32-42.8ºC range. Compliance with IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, ISO 80601-2-56, and ASTM E1965-98. Biocompatibility per ISO 10993-5 and ISO 10993-10.
Indications for Use
Indicated for measurement and display of human body temperature in infants and adults via non-contact forehead infrared sensing. ARC InstaTemp™ is for home or clinical use; ARC InstaTemp MD™ is for clinical use by healthcare professionals only.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Rycom Electron Technology Limited, RC002 Remote Infrared Thermometer (K090361)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 20, 2016
ARC Devices Limited Mr. Gary Brogan Director of Quality and Regulatory Affairs 2 Deerpark Industrial Estate Oranmore, Co. Galway Ireland
Re: K152905
Trade/Device Name: ARC InstaTemp™ Non-touch Thermometer, ARC InstaTemp™ MD Non-touch Thermometer Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: II Product Code: FLL Dated: June 10, 2016 Received: June 16, 2016
Dear Mr. Brogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang
-s
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K152905
Device Name ARC InstaTemp™ Non-Touch Thermometer
Indications for Use (Describe)
The ARC InstaTemp™ is an infrared thermometer that measures from the forehead in both infants and adults without contacting the human body.
Type of Use (Select one or both, as applicable)
J Prescription Use (Part 21 CFR 801 Subpart D)
2 Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K152905
Device Name
ARC InstaTemp MD™ Non-Touch Thermometer
#### Indications for Use (Describe)
The ARC InstaTemp MD™ is an infrared thermometer that measures from the forehead in both infants and adults without contacting the human body. It is intended for use by a Health Care Professional in a clinical environment. This device is not intended for home use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
view the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary (21 CFR 807.92) K152905
# I. SUBMITTER
ARC Devices Limited, 2 Deerpark Industrial Estate, Oranmore, Co Galway, Ireland
Phone: +353 91 653 155 Fax: +353 91 653 155
Contact Person: Gary Brogan Director or Quality and Regulatory Affairs,
Date Prepared: July 20, 2016
## II. DEVICE
Name of Devices: ARC InstaTemp™ Non-Touch Thermometer ARC InstaTemp MD™ Non-Touch Thermometer
Common or Usual Name: Infrared Thermometer
Classification Name: 21 CFR 880.2910 (Clinical Electronic Thermometer)
Regulatory Class: Class II
Product Code: FLL
## III. PREDICATE DEVICE
Rycom Electron Technology Limited, RC002 Remote Infrared Thermometer, K090361, cleared 16th June 2010.
## IV. DEVICE DESCRIPTION
The ARC Non-Touch Thermometer is a fast, accurate and non-intrusive device for measuring a person's temperature. It uses infrared sensor technology to detect the radiated infrared energy emitted from the forehead. Therefore as soon as the sensor is held near the forehead and activated, the temperature measurement is taken by detection of the emitted infrared heat generated by the arterial blood flow.
The ARC Non-Touch Thermometer uses sensor technology that detects the emitted heat from
{5}------------------------------------------------
the temporal artery in the forehead from a distance of approximately 1 inch. Once recorded an internal algorithm converts this forehead surface temperature into human body temperature and it is this temperature that is then presented to the user on the LCD screen. This temporal artery detection mechanism and the resulting conversion to body temperature characterizes the ARC Non-Touch Thermometers as an adjusted mode clinical thermometer as per ISO 80601-2-56. This is the same principle of operation as the listed predicate device, K090361.
The ARC Non-Touch Thermometer is indicated for the measurement and display of human body temperature for patients of all ages. It can be used by responsible adults at home or by health care professionals in a clinical environment. In order to serve both these markets, ARC Devices have developed two separate models, the ARC InstaTemp™ Non-Touch Thermometer for home use and the ARC InstaTemp MD™ Non-Touch Thermometer for professional use. Both models have the same purpose, design, materials, energy source and function and also possess the same features related to safety and effectiveness. It is for this reason that the two models are included in the one 510k pre-market submission. This is in compliance with, Guidance for Industry and FDA Staff, Bundling Multiple Device or Multiple Indications in a Single Submission, June 22, 2007. The differences between both models while minor in nature and have been introduced to make the device more suitable for professional use as opposed to home use.
# V. INDICATIONS FOR USE
The ARC InstaTemp™ is an infrared thermometer that measures temperature from the forehead in both infants and adults without contacting the human body.
The ARC InstaTemp MD™ is an infrared thermometer that measures temperature from the forehead in both infants and adults without contacting the human body. It is intended for use by a Health Care Professional in a clinical environment. This device is not intended for home use.
The ARC InstaTemp™ and InstaTemp MD™ devices intentions for use are different from the predicate intentions for use statement because the user environments are specific to either home use or clinical use, whereas the predicate device can be used in either home use or clinical use. This difference does not raise new questions regarding safety and effectiveness because the technology is similar, the bodily measurement site is identical, and the user environments are the same but limited to specific devices.
{6}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Parameter | RC002 Remote Infrared<br>Thermometer | ARC InstaTemp™<br>Non-Touch<br>Thermometer | ARC InstaTemp MD™<br>Non-Touch<br>Thermometer | Comparison |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Measurement<br>Method | RC002 Remote Infrared<br>Thermometer uses an<br>infrared sensor to detect<br>and record the emitted<br>heat from the temporal<br>artery in the forehead. In<br>the 'body' function it<br>converts this forehead<br>surface temperature into<br>body temperature and it is<br>this temperature that is<br>then presented to the user<br>on the LCD screen. | ARC InstaTemp™<br>Non-Touch Thermometer<br>uses an infrared sensor to<br>detect and record the<br>emitted heat from the<br>temporal artery in the<br>forehead. It then converts<br>this forehead surface<br>temperature into human<br>body temperature and it is<br>this temperature that is then<br>presented to the user on<br>the LCD screen. | ARC InstaTemp MD™<br>Non-Touch Thermometer<br>uses an infrared sensor to<br>detect and record the<br>emitted heat from the<br>temporal artery in the<br>forehead. It then converts<br>this forehead surface<br>temperature into human<br>body temperature and it is<br>this temperature that is then<br>presented to the user on<br>the LCD screen. | Same |
| Range | In Body Mode: 32-42.9ºC | 32-42.8ºC | 32-42.8ºC | Same |
| Parameter | RC002 Remote Infrared<br>Thermometer | ARC InstaTemp™<br>Non-Touch<br>Thermometer | ARC InstaTemp MD™<br>Non-Touch<br>Thermometer | Comparison |
| Accuracy | $\pm$ 0.3ºC | $\pm$ 0.2ºC | $\pm$ 0.2ºC | Verified by<br>compliance<br>testing to IEC<br>60601-1. |
| Precision | $\pm$ 0.3ºC (34-35.9ºC)<br>$\pm$ 0.2ºC (36-39ºC)<br>$\pm$ 0.3ºC (39-42.5ºC) | $\pm$ 0.2ºC | $\pm$ 0.2ºC | Verified by<br>compliance<br>testing to IEC<br>60601-1. |
| Measurement<br>Distance | 5-8cm | 1-3cm | 1-3cm | Acceptable |
| Power supply | 2 x AA Battery | 2 x AAAA Battery | 2 x AAAA Battery | Acceptable |
| Operating | 10ºC to 40ºC | 10ºC to 40ºC | 10ºC to 40ºC | Same |
| Display | LCD | LCD | LCD | Same |
| Display | 0.1ºC | 0.1ºC | 0.1ºC | Same |
| Celsius /<br>Fahrenheit | Yes | Yes | Yes | Same |
| Led Backlight | Yes | Yes | Yes | Same |
| Automatic Stop | 5 seconds | 4 + 0.5 seconds | 4 + 0.5 seconds | Same |
| Dimensions | 196x150x50mm<br>(LxWxH) | 129x35x13mm<br>(LxWxH) | 129x35x13mm<br>(LxWxH) | Acceptable |
The following technological differences exist for the ARC Non-Touch Thermometers and its predicate device.
- The accuracy and precision is 0.2℃ across the full range as opposed to 0.3℃ for the ● predicate - This was verified during compliance testing to IEC 60601-1
- . The size of the battery used to power the device - The batteries were verified as being acceptable as part fulfillment of compliance testing to IEC 60601-1.
{7}------------------------------------------------
- . The recommended measurement distance - Reduction of the measurement distance reduces the noise due to ambient temperature. 1-3cm is a comfortable distance for the user to operate these thermometers.
- Dimensions of the devices The changes in dimensions have no effect on functionality . of the thermometer. The modified dimensions are deemed acceptable as the unit can be used by a single operator in the same way as the predicate.
Upon review of these technological differences, it can be concluded that they do not affect the core function of the device. In addition as both the subject and the predicate devices meet the same FDA recognized consensus standards, it can be concluded that the subject devices are substantially equivalent to the predicate device.
# VII. COMPARISON OF MATERIALS USED WITH THE PREDICATE DEVICE
| RC002 Remote Infrared<br>Thermometer | ARC InstaTemp™<br>Non-Touch Thermometer | ARC InstaTemp MD™<br>Non-Touch Thermometer |
|----------------------------------------------------------------|----------------------------------------------------------------|----------------------------------------------------------------|
| Materials of Components in contact<br>with the operators hand: | Materials of Components in contact<br>with the operators hand: | Materials of Components in contact<br>with the operators hand: |
| Acrylonitrile butadiene styrene<br>(ABS) | Upper Housing – Polycarbonate | Upper Housing - Polycarbonate |
| | Lower Housing - Polycarbonate | Lower Housing - Polycarbonate |
| | Button - Polycarbonate | Button - Polycarbonate |
| | LCD Screen - Polycarbonate | LCD Screen - Polycarbonate |
As can be reviewed in the table, the material in contact with the operators hand is ABS in the predicate device while it is Polycarbonate in the ARC Non-Touch Thermometer. The contact duration throughout the lifetime of the device would be <24hours and the device would always be in contact with intact skin.
Both ABS and Polycarbonate are used widespread in the medical device field and neither have any concerns from a user safety viewpoint. In addition the material does not affect the performance of the device. That aside biocompatibility testing, per ISO 10993-5:2009 and ISO 10993-10:2010 has been completed on both the ARC InstaTemp™ and the ARC InstaTemp MD™ to demonstrate that no biocompatibility concerns exist that affect the substantial equivalence claim.
# VIII.BIOCOMPATIBILITY
The ARC Non-Touch Thermometer is manufactured from an outer casing of polycarbonate that comes into direct contact with the operator's hand. For this reason, and in order to maintain compliance to ASTM E1965. GLP Biocompatibility testing in compliance to ISO 10993-5:2009 (cytotoxicity analysis) and ISO 10993-10:2010 (skin irritation analysis) was carried out on both the ARC InstaTemp™ and the ARC InstaTemp MD™. This testing concluded that no biocompatibility issues exist with the ARC Non-Touch Thermometer.
{8}------------------------------------------------
# IX. NON-CLINICAL PERFORMANCE DATA
The following non-clinical performance data was provided in support of the substantial equivalence determination:
# Software Verification and validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDAs Guidance for Industry and FDA Staff 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices'. The software for this device was considered as a 'moderate' level of concern since a malfunction of, or a latent design flaw in, the software device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to minor injury.
# Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing was conducted for the ARC Non-Touch Thermometer. This testing concluded that the device is in compliance with IEC 60601-1. IEC 60601-1-11 (2010) and ISO80601-2-56 standards for safety and IEC 60601-1-2 standard for EMC.
## Bench Performance Testing
The following bench testing was performed to support substantial equivalence
- ASTM E1965-98 (Reapproved 2009), Standard Specification for Infrared Thermometers . for Intermittent Determination of Patient Temperature.
- . Software / Algorithm report / Comparison Report with an identical device to the predicate.
# X. CLINICAL PERFORMANCE DATA
A clinical trial was completed on 136 subjects which verified the clinical performance by way of a clinical accuracy validation as defined in ISO 80601-2-56, section 201.102. The test report concluded that based on the required sample sizes as mandated by the standard that the ARC Devices Non-Touch Thermometer has levels of reproducibility and accuracy, for both febrile and non-febrile patients that are in keeping with the existing devices that are in widespread hospital practice. This clinical trial demonstrates that the ARC Devices Thermometer performs in the clinical setting and so substantial equivalence can be drawn to the predicate.
# XI. CONCLUSIONS
The ARC Non-Touch Thermometer, has the same intended use and similar characteristics as the predicate device. Moreover, the subject device demonstrates product safety through the successful testing to IEC 60601-1, IEC 60601-1-11 (2010) and ISO80601-2-56 along with successful testing to the electromagnetic standard IEC 60601-1-2. Furthermore, the subject device also meets the specific requirements of ASTM E1965-98.
Taking this information into account, it is concluded that the ARC Non-Touch Thermometer which includes both the ARC InstaTemp™ Non-Touch Thermometer and the ARC InstaTemp MD™ Non-Touch Thermometer, are substantially equivalent to the predicate device, the RC002 Remote Infrared Thermometer (K090361)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.