The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and facility information. The Sentinel Cardiology Information Management System provides download and storage of data from supported Holter recorders and feeds this data to separate and optional Holter analyzers. The completed reports are stored back in Sentinel. The Sentinel Cardiology Information Management System facilitates presenting, analyzing, reviewing cardiovascular information, including automatic interpretation of 12 lead ECG and stores the finished reports. The automatic interpretation of 12-lead ECG is not suitable for neonates. Sentinel supports importing reports from external systems and provides an HL7 interface to external systems. The Sentinel Cardiology Information Management System is intended to be used by a trained user and as an aid to a medical professional in hospital or primary care environment. The Sentinel Cardiology Information Management System is not intended for patient monitoring.
Device Story
Sentinel Cardiology Information Management System is a standalone software platform for capturing, storing, and managing cardiovascular data; inputs include ECG, ambulatory blood pressure (ABP), stress test, ergospirometry, spirometry, event, and plethysmography data from supported medical devices. System operates as a single installation or multi-user network; used by trained clinicians in hospitals or primary care. It integrates Glasgow ECG analysis for 12-lead ECG and HES analysis for stress tests to provide automated interpretations. Outputs include finished reports and reviewed waveforms; clinicians use these to aid in diagnosis and clinical decision-making. System supports HL7 interfaces for external system integration. Benefits include centralized data management, standardized reporting, and automated diagnostic support, improving clinical workflow efficiency.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation activities included software life cycle processes (IEC 62304), electrical safety (IEC 60601-1), usability (IEC 62366), and performance testing against specific standards for electrocardiographs (IEC 60601-2-25) and automated non-invasive sphygmomanometers (IEC 80601-2-30).
Technological Characteristics
Standalone software for cardiology data management. Connectivity via network; supports HL7 interface. Implements Glasgow ECG analysis and HES stress test analysis algorithms. Designed to comply with IEC 62304 (software life cycle), IEC 60601-1 (safety), and IEC 62366 (usability).
Indications for Use
Indicated for use by trained medical professionals in hospital or primary care environments to manage cardiovascular data (ECG, ABP, stress, ergospirometry, spirometry, event, plethysmography) for adult and pediatric patients. Automatic 12-lead ECG interpretation is contraindicated for neonates. Not for patient monitoring.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 1, 2016
Spacelabs Healthcare Ltd. % Thomas Kroenke Principal Consultant, Speed To Market, Inc. PO Box 3018 Nederland, Colorado 80466
Re: K152881
Trade/Device Name: Sentinel Cardiology Information Management System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK, DXH Dated: February 26, 2016 Received: February 29, 2016
Dear Thomas Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
FDA
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known) K152881 Page 1 of 1
Device Name
Spacelabs Healthcare Ltd. Sentinel Cardiology Information Management System
Indications for Use (Describe)
The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and facility information.
The Sentinel Cardiology Information Management System provides download and storage of data from supported Holter recorders and feeds this data to separate and optional Holter analyzers. The completed reports are stored back in Sentinel.
The Sentinel Cardiology Information Management System facilitates presenting, analyzing, reviewing cardiovascular information, including automatic interpretation of 12 lead ECG and stores the finished reports. The automatic interpretation of 12-lead ECG is not suitable for neonates.
Sentinel supports importing reports from external systems and provides an HL7 interface to external systems. The Sentinel Cardiology Information Management System is intended to be used by a trained user and as an aid to a medical professional in hospital or primary care environment.
The Sentinel Cardiology Information Management System is not intended for patient monitoring.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| <span> <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submission Date: | 25 February 2016 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare Ltd.<br>1 Harforde Court<br>John Tate Road<br>Hertford United Kingdom<br>SG13 7NW |
| Submitter Contact: | Mr. Al Van Houdt<br>Spacelabs Healthcare<br>Phone: +1 (425) 363-5970<br>Fax: +1 (425) 363-5762<br>Email: al.vanhoudt@spacelabs.com |
| Application<br>Correspondent: | Thomas Kroenke<br>Speed To Market, Inc.<br>PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>303 956 4232 |
| Manufacturing Site: | Spacelabs Healthcare<br>35301 SE Center Street<br>Snoqualmie, WA 98065 |
| Trade Name: | Sentinel Cardiology Information Management System |
| Common Name: | Programmable Diagnostic Computer |
| Primary Classification<br>Name: | Programmable Diagnostic Computer |
| Primary Classification<br>Regulation: | 21 CFR §870.1425 |
| Primary Product Code: | DQK |
| Secondary<br>Classification Name: | Transmitters and Receivers, Electrocardiograph, Telephone |
| Secondary<br>Classification<br>Regulation: | 21 CFR §870.2920 |
| Secondary Product<br>Code: | DXH |
{4}------------------------------------------------
| Substantially<br>Equivalent Devices: | New Spacelabs<br>Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------------------------------------------|
| | Sentinel Cardiology<br>Information Management<br>System | K062397 | DelMar Reynalds Medical Ltd.<br>(now part of Spacelabs<br>Healthcard Ltd.) / Sentinel |
| | | K130155 | GE Medical Systems<br>Information Technologies /<br>MUSE Cardiology Information<br>System |
| Device Description: | The Spacelabs Healthcare Ltd. (Spacelabs) Sentinel Cardiology<br>Information Management System (Spacelabs) is a stand-alone<br>software product which can capture, store, access and manage<br>electrocardiograph (ECG), ambulatory blood pressure (ABP), stress<br>test, ergospirometry, spirometry, event and plethysmography data<br>with Holter analyses and reports, together with the corresponding<br>patient information. Sentinel can be used as a single stand-alone<br>installation, or as part of a multi-user network. Sentinel may be used<br>for recording and reviewing of ECG of adults and pediatrics in a<br>clinical environment inside and outside of hospitals. | | |
| Intended Use: | The Sentinel Cardiology Information Management System is<br>intended to connect to supported medical devices and analyzers in<br>order to download, store, access and manage cardiovascular<br>information and to manage patient and facility information. | | |
| | The Sentinel Cardiology Information Management System provides<br>download and storage of data from supported Holter recorders and<br>feeds this data to separate and optional Holter analyzers. The<br>completed reports are stored back in Sentinel. | | |
| | The Sentinel Cardiology Information Management System<br>facilitates presenting, analyzing, reviewing cardiovascular<br>information, including automatic interpretation of 12 lead ECG and<br>stores the finished reports. The automatic interpretation of 12-lead<br>ECG is not suitable for neonates. | | |
| | Sentinel supports importing reports from external systems and<br>provides an HL7 interface to external systems. | | |
| | The Sentinel Cardiology Information Management System is<br>intended to be used by a trained user and as an aid to a medical<br>professional in hospital or primary care environment. | | |
| | The Sentinel Cardiology Information Management System is not<br>intended for patient monitoring. | | |
{5}------------------------------------------------
#### Technology Comparison:
The Spacelabs Sentinel Cardiology Information Management System employs the same technological characteristics as the predicate device.
| Characteristic | GE Medical Systems<br>MUSE | DelMar Reynalds<br>Medical Ltd.<br>Sentinel | Proposed Device |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Acquisition<br>and Analysis | All compatible<br>devices ECG<br>measurements View ECG<br>waveforms Print ECG<br>waveforms Edit interpretation Serial presentation<br>GE devices ECG interpretation<br>- Marquette 12SL Serial comparison 15 lead<br>adult/pediatric Right ventricular<br>involvement ECG<br>interpretation Third party devices<br>different than above ECG interpretation<br>- Vendor specific | ABP: TrackerNIBP2,<br>P6, 90207, 90217 Acquire test Edit test Print test View test Holter: Lifecard CF,<br>Aria View ECG<br>waveforms Print ECG<br>waveforms Edit interpretation 12 Lead: CD12-<br>Serial, CardioCollect ECG interpretation View ECG<br>waveforms Print ECG<br>waveforms Edit interpretation Carts: Voyager, GE-<br>Mac 5000, GE-Mac<br>5500, Philips<br>Pagewriter,<br>CarioCollect View ECG<br>waveforms Print ECG<br>waveforms Edit interpretation | ABP: TrackerNIBP2,<br>90207, 90217,<br>90217A, 90227 Acquire test Edit test Print test Review test Serial presentation Holter, Lifecard CF,<br>Aria. EVO Review ECG<br>waveforms Print ECG<br>waveforms Edit interpretation 12-Lead: CD12-<br>Serial, CD12-USB ECG interpretation View ECG<br>waveforms Print ECG<br>waveforms Edit interpretation Serial presentation Carts: Voyager, GE-<br>Mac 5000, Philips<br>Pagewriter Review ECG<br>waveforms Print ECG<br>waveforms Edit interpretation Third party devices<br>different than above ECG interpretation<br>- Vendor specific |
| Modalities | 12 lead resting<br>ECG Bedside 12 lead<br>ECG Defibrillators | Ambulatory BP 12 lead ECG:<br>Resting Holter Stress | Ambulatory BP 12 lead ECG:<br>Resting and<br>Rhythm Holter Stress Event |
| Characteristic | GE Medical Systems<br>MUSE | DelMar Reynalds<br>Medical Ltd.<br>Sentinel | Proposed Device |
| Arrhythmia<br>Analysis | • Marquette 12SL<br>ECG analysis | • Glasgow ECG<br>analysis for 12 lead<br>ECG<br><br>• HES analysis for<br>stress tests | • Glasgow ECG<br>analysis for 12 lead<br>ECG<br><br>• HES analysis for<br>stress tests |
Page 3 of 5
{6}------------------------------------------------
### Summary of Performance Testing:
| Software | Software was designed and developed according to a robust software development process, and was rigorously verified and validated. Software information is provided in accordance with |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | • FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05; |
| | • FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99; |
| | • FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; |
| | • FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, 02 Oct 14; and |
| | • IEC 62304: 2006, Medical device software – Software life cycle processes. |
| | Test results indicated that Spacelabs Sentinel complies with predetermined specifications and the applicable standard. |
| Electrical Safety | Spacelabs Sentinel is a software product only, and designed to comply with the applicable requirements of: |
| | • IEC 60601-1: 2005, Am1: 2012, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance |
| | Verification results indicated that Spacelabs Sentinel complies with the applicable requirements of the standard. |
{7}------------------------------------------------
Performance Testing -Spacelabs Sentinel was tested in accordance with: Bench • IEC 60601-1-6: 2010, Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance --Collateral standard: Usability; · IEC 60601-2-25: 2011, Medical electrical equipment – Part 2-25: Particular requirements for the basic safety and essential performance of electrocardiographs; • IEC 80601-2-30: 2013, Medical electrical equipment – Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers; and · IEC 62366: 2006, Am1: 2014, Medical devices - Application of usability engineering to medical devices. · Test results indicated that Spacelabs Sentinel complies with predetermined specifications and the applicable standards. Conclusion Verification and validation activities were conducted to establish the performance and safety characteristics of Spacelabs Sentinel. The results of these activities demonstrate that Spacelabs Sentinel is as safe, as effective, and performs as well as or better than the predicate device. Therefore, Spacelabs Sentinel is considered substantially equivalent
to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.