← Product Code [KPS](/productcode/KPS) · K152880

# Biograph Horizon PET/CT (K152880)

_Siemens Medical Solutions USA, Inc. · KPS · Jan 12, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152880

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [KPS](/productcode/KPS.md)
- **Decision Date:** Jan 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1200
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Siemens Biograph Horizon systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information. The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images. The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

## Device Story

Biograph Horizon PET/CT combines multi-slice X-ray CT and PET scanners for whole-body oncology, neurology, and cardiology imaging. System inputs include X-ray transmission data and PET radiopharmaceutical distribution signals. The device performs computer reconstruction of CT data and PET image processing to provide registered, fused metabolic and anatomic images. Operated by trained healthcare professionals in clinical settings, the system includes a patient handling unit and workstations for scan control, image reconstruction, and archival. Physicians use the fused output to detect, localize, and stage disease, and for radiotherapy planning. The system supports independent CT or PET modality operation. Benefits include precise anatomical reference for PET studies and fast attenuation correction via CT integration.

## Clinical Evidence

Bench testing only. Performance evaluated per NEMA NU2:2012 standards for PET (resolution, sensitivity, count rate, scatter fraction, mean bias) and FDA performance standards for ionizing radiation-emitting products for CT. All results met predetermined acceptance criteria.

## Technological Characteristics

Combined PET/CT system. PET: air-cooled, 70 cm bore. CT: 16-slice configuration. Connectivity: DICOM compliant. Software: command-based for patient/data management, scan control, and reconstruction. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-44, IEC 62304, NEMA NU 2-2012, NEMA XR 25.

## Regulatory Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

## Predicate Devices

- Biograph mCT PET/CT ([K151486](/device/K151486.md))

## Reference Devices

- Somatom Perspective ([K133590](/device/K133590.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 12, 2016

Siemens Medical Solutions USA, Inc. % Ms. Alaine Medio PET and PCS Regulatory Projects Manager 810 Innovation Drive KNOXVILLE TN 37932

Re: K152880

Trade/Device Name: Biograph Horizon PET/CT Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK Dated: December 1, 2015 Received: December 2, 2015

Dear Ms. Medio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152880

Device Name Biograph Horizon PET/CT

#### Indications for Use (Describe)

The Siemens Biograph Horizon PET/CT systems are combined Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and /or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

as required by 21 CFR Part 807.87(h)

| Identification of the Submitter |                                                                                                                                                                            |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:                      | M. Alaine Medio, RAC<br>PET and PCS Regulatory Projects Manager<br>Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932 |
| Manufacturers:                  | Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>810 Innovation Drive<br>Knoxville, TN 37932                                                                    |
|                                 | Siemens Medical Solutions USA, Inc.<br>Molecular Imaging<br>2501 North Barrington Road<br>Hoffman Estates, IL 60192                                                        |
| Telephone Number:               | (865)218-2703                                                                                                                                                              |
| Fax Number:                     | (865)218-3019                                                                                                                                                              |
| Date of Submission:             | September 29th, 2015                                                                                                                                                       |
| Identification of the product   |                                                                                                                                                                            |
| Device Proprietary Name:        | Biograph Horizon PET/CT                                                                                                                                                    |
| Common Name:                    | Positron Emission Tomography (PET) System<br>Computed Tomography (CT) System                                                                                               |
| Classification Name:            | Emission Computed Tomography System per 21 CFR<br>892.1200<br>Computed Tomography X-Ray System per 21 CFR<br>892.1750                                                      |
| Product Code:                   | 90 KPS and 90 JAK                                                                                                                                                          |
| Classification Panel:           | Radiology                                                                                                                                                                  |
| Device Class:                   | Class II                                                                                                                                                                   |

{4}------------------------------------------------

#### Marketed Devices to which Equivalence is claimed

Predicate:

| Device Proprietary Name: | Biograph mCT PET/CT                 |
|--------------------------|-------------------------------------|
| Manufacturer:            | Siemens Medical Solutions USA, Inc. |
| Product Code:            | 90 KPS and 90 JAK                   |
| Device Class:            | Class II                            |
| 510(k) Number:           | K151486                             |
| Reference Devices:       |                                     |
| Device Name:             | Somatom Perspective                 |
| 510(k) Numbers:          | K133590                             |

#### Device Description:

The Biograph Horizon PET/CT system is a combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanners. This system is designed for whole body oncology, neurology and cardiology examinations. The Biograph Horizon scanners provide registration and fusion of high-resolution metabolic and anatomic information from the two major components of each system (PET and CT). Additional components of the system include a patient handling system and acquisition and processing workstations with associated software.

Biograph Horizon software is a command based program used for patient management, data management, scan control, image reconstruction and image archival and evaluation. All images conform to DICOM imaging format requirements.

The Biograph Horizon system which is the subject of this application is substantially equivalent to the commercially available Biograph mCT family of PET/CT systems (K151486). Differences compared to the Biograph mCT include:

- . The PET subsystem has been modified to provide for an air-cooled only PET system with a 70 cm bore size. The detectors, electronics, etc... are substantially equivalent to the current Biograph mCT system.
- . The commercially available Perspective 16 slice system (K133590) has been incorporated as the CT subsystem for the Biograph Horizon.
- . The software integrated into the Biograph Horizon systems is based upon the mCT software and includes a subset of features that are commercially available with the Biograph mCT systems. Major updates to the software system for use on the Biograph Horizon include:
	- o limiting of high end features available with the mCT

{5}------------------------------------------------

- changes in software to support the modified PET and CT subsystems O
- updates for workflow and anomaly corrections. O

#### Intended Use:

The Siemens Biograph Horizon systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

The CT component produces cross-sectional images of the body by computer reconstruction of X-Ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles. The PET subsystem images and measures the distribution of PET radiopharmaceuticals in humans for the purpose of determining various metabolic (molecular) and physiologic functions within the human body and utilizes the CT for fast attenuation correction maps for PET studies and precise anatomical reference for the fused PET and CT images.

The system maintains independent functionality of the CT and PET devices, allowing for single modality CT and / or PET diagnostic imaging.

These systems are intended to be utilized by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures.

#### Performance Testinq / Safety and Effectiveness:

PET Testing in accordance with NEMA NU2:2012 was conducted on two different configurations of the Biograph Horizon PET/CT systems, a 3 rinq version and a 4 ring version.

| Performance Criteria               | Results<br>(Average) | Acceptance<br>(Standard PET) | Acceptance<br>(TrueV PET) |
|------------------------------------|----------------------|------------------------------|---------------------------|
| Resolution – Full Size             |                      |                              |                           |
| Transverse Resolution FWHM @ 1 cm  | Pass                 | <= 4.7 mm                    | <= 4.7 mm                 |
| Transverse Resolution FWHM @ 10 cm | Pass                 | <= 5.5 mm                    | <= 5.5 mm                 |
| Transverse Resolution FWHM @ 20 cm | Pass                 | <= 7.6 mm                    | <= 7.6 mm                 |
| Axial Resolution FWHM @ 1 cm       | Pass                 | <= 5.0 mm                    | <= 5.0 mm                 |
| Axial Resolution FWHM @ 10 cm      | Pass                 | <= 7.0 mm                    | <= 7.0 mm                 |
| Axial Resolution FWHM @ 20 cm      | Pass                 | <= 11.3 mm                   | <= 11.3 mm                |
| Resolution - 256 x 256             |                      |                              |                           |
| Transverse Resolution FWHM @ 1 cm  | Pass                 | <= 7.3 mm                    | <= 7.3 mm                 |
| Transverse Resolution FWHM @ 10 cm | Pass                 | <= 7.6 mm                    | <= 7.6 mm                 |
| Transverse Resolution FWHM @ 20 cm | Pass                 | <= 8.9 mm                    | <= 8.9 mm                 |
| Axial Resolution FWHM @ 1 cm       | Pass                 | <= 6.1 mm                    | <= 6.1 mm                 |
| Axial Resolution FWHM @ 10 cm      | Pass                 | <= 7.3 mm                    | <= 7.3 mm                 |
| Axial Resolution FWHM @ 20 cm      | Pass                 | <= 11.9 mm                   | <= 11.9 mm                |

{6}------------------------------------------------

| Performance Criteria          | Results<br>(Average) | Acceptance<br>(Standard PET) | Acceptance<br>(TrueV PET)  |
|-------------------------------|----------------------|------------------------------|----------------------------|
| Sensitivity @435 keV LLD      | Pass                 | >= 5.8 cps/MBq               | >= 10.9 cps/MBq            |
| Count Rate peak NECR          | Pass                 | >= 78 kcps @ <= 26 kBq/cc    | >= 135 kcps @ <= 26 kBq/cc |
| Count Rate peak trues         | Pass                 | >= 285 kcps @ <= 53 kBq/cc   | >= 465 kcps @ <= 42 kBq/cc |
| Scatter Fraction at peak NECR | Pass                 | <=40%                        | <=40%                      |
| Mean bias (%) at peak NEC     | Pass                 | <=6                          | <=6                        |

CT Testing in accordance with the requirements of the following FDA Performance Standards for lonizing Radiation Emitting Products and Light emitting products for CT equipment with a laser localizer was performed and documented in a Dosimetry and Imaging Performance Report. Further, each CT subsystem is tested and passes the Applicable Performance Standards prior to shipment:

- 21 CFR 1020.30 (a) ●
- 21 CFR 1020.30 (b)(36)(iii)-(v)
- 21 CFR 1020.30 (b)(58)-(62) ●
- 21 CFR 1020.30 (h)(3)(vi)-(viii) ●
- 21 CFR 1020.33
- 21 CFR 1040.10 ●
- 21 CFR 1040.11

Applicability Technique factors

CT, Scan, Scan Time, Tomogram, Dose

Information to be provided for users

Computed Tomography (CT) equipment

- Laser Products
- Specific purpose laser products

The Bioqraph Horizon PET/CT system was designed in accordance with Design Controls and in accordance with the following FDA recognized standards:

- Recognition Number 19-1: IEC 60601-1-2 Edition 3: 2007-03 ●
- Recognition Number 19-4: AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 .
- Recognition Number 5-85: IEC 60601-1-6 Edition 3.0 2010-01 ●
- Recognition Number 12-256: IEC 60601-2-44 Edition 3.1 2012-09 .
- . Recognition Number 12-269: IEC 60601-1-3 Edition 2.1 2013-04
- . Recognition Number 5-40: ISO 14971 Second Edition 2007-03
- . Recognition Number 13-8: IEC 62304 First Edition 2006-05
- Recognition Number 12-265: NEMA NU 2-2012 ●
- Recognition Number 12-238: NEMA PS 3.1 3.20 (2011) .
- . Recognition Number 12-225: NEMA XR 25
- . Recognition Number 12- 270: NEMA 61223-3-5 First Edition 2004
- . Recognition Number 12-226: NEMA 61223-2-6 Second Edition: 2006

{7}------------------------------------------------

All Performance testing performed met the predetermined acceptance values.

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.

Risk Management is ensured via a risk analysis in compliance with ISO 14971 to identify and provide mitiqation to potential hazards beginning early in the design cycle and continuing throughout the development of the product. Siemens Medical Solutions, USA Inc. adheres to recognized and established industry standards such as IEC 60601-1 series and 21 CFR 1020.30 and 21 CFR 1020.33 to minimize electrical, mechanical and radiation hazards.

Verification and validation of Siemens systems is performed in accordance with documented procedures, design and code reviews, test plans and specifications. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.

### Statement regarding Substantial Equivalence:

The Biograph Horizon PET/CT is based on the commercially available Biograph mCT Family of PET/CT scanners. There is no difference in Indications for use, nor is there any change in fundamental scientific technology as compared to the predicate device. The software and device features are a subset of the predicate device features, and pose no new questions of safety and / or efficacy.

Based on the information provided in the Premarket Notification Siemens Medical Solutions USA, Inc. considers the Biograph Horizon PET/CT system described in this Premarket Notification to be substantially equivalent to the currently commercially available predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K152880](https://fda-staging.innolitics.com/device/K152880)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).