K152879 · Shina Med Corporation · FMF · Sep 9, 2016 · General Hospital
Device Facts
Record ID
K152879
Device Name
Sure-Fine Insulin Syringes
Applicant
Shina Med Corporation
Product Code
FMF · General Hospital
Decision Date
Sep 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
Device Story
Sure-Fine Insulin Syringes are single-use, disposable, sterile piston syringes designed for subcutaneous injection of U-100 insulin. The device consists of a graduated polypropylene barrel, plunger rod, isoprene rubber piston, and a stainless steel needle permanently affixed to the tip with epoxy. Available in various volumes (0.3cc, 0.5cc, 1cc) and needle gauges (28G-31G). The device is intended for use by patients or healthcare providers to deliver precise insulin doses. Operation follows standard manual piston syringe mechanics: the user draws insulin into the barrel and manually depresses the plunger to inject. The device provides a simple, reliable method for insulin administration, facilitating self-management of diabetes.
Clinical Evidence
No clinical data. Bench testing only. Performance verified per ISO 8537:2007 (freedom from extraneous matter, extraction, lubrication, graduated scale, barrel/plunger assembly, needle tubing). Biocompatibility verified per ISO 10993-1 (cytotoxicity, irritation, sensitization, systemic toxicity, blood interaction). Sterilization validated per ISO 11135.
Indicated for subcutaneous injection of U100 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
Shina Med Corporation c/o Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 2651 E. Chapman Ave., Ste. 110 Irvine, California 92831
Re: K152879
Trade/Device Name: Sure-Fine Insulin Syringes Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: August 4, 2016 Received: August 10, 2016
### Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows a signature and the name "Tina Kiang" in black font. The signature is illegible and consists of a series of loops and curves. There is a faint watermark in the background that is difficult to read.
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152879
Device Name
Sure-Fine Insulin Syringes
Indications for Use (Describe)
Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Registration (Part 21 CFR 201.1 Subpart D) | <input checked="checked" type="checkbox"/> Drug Use Compounding (21 CFR 201.1 Subpart G) |
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------|------------------------------------------------------------------------------------------|
__ Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
(K152879)
September 07, 2016
### 1. Submitted by:
Il-hwan, Jeong / Quality Management Representative SHINA MED CORPORATION 455-30, Bogaewonsam-ro, Bogae-myun, Anseong-si, Gyeonggido, 456-871, Republic of Korea Phone: +82 31 8057 2125 Fax : + 82 31 8057 2150
### 2. US Agent/Contact Person:
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton CA 92831
#### Device Name : 3.
- Trade Name : Sure-Fine Insulin Syringes
- Classification : Class II .
- Regulation Name : Piston syringe .
- Product Code . : FMF
- . Regulation Number : 21 CFR 880.5860
- · Review Panel : General Hospital
#### Predicate Device : 4.
Accu-Sure Insulin syringe (K091167) manufactured by SHINA CORPORATION
#### 5. Device Description:
Sure-Fine Insulin Syringes are designed for the subcutaneous injection of a dose of U-100 insulin. The syringe has a graduated barrel, a plunger rod, needle cap, protective end cap and needle permanently affixed to the tip of the syringe with epoxy. The syringes are available in the following sizes and cap color.
| Category | Insulin syringe | Needle<br>Gauge | Needle<br>Length | Cap color | |
|----------|-----------------------|-----------------|------------------|------------|-----------------|
| | | | | Needle cap | Plunger Cap |
| U-100 | 1/2cc and 1cc | 28Gauge | 1/2" | Orange | Orange or white |
| U-100 | 3/10cc, 1/2cc and 1cc | 29Gauge | 1/2" | Orange | Orange or white |
| U-100 | 3/10cc, 1/2cc and 1cc | 30Gauge | 1/2" | Orange | Orange or white |
| U-100 | 3/10cc, 1/2cc and 1cc | 30Gauge | 5/16" | Orange | Orange or white |
| U-100 | 3/10cc, 1/2cc and 1cc | 31Gauge | 5/16" | Orange | Orange or white |
This device operates in the principles of a piston syringe fluid path is sterile (EO gas sterilization), single use disposable.
{4}------------------------------------------------
#### 6. Indications for Use :
Sure-Fine Insulin Syringes are hypodermic insulin syringes for subcutaneous injection of U100 insulin.
### 7. Technological Characteristics :
The Sure-Fine Insulin Syringes are substantially equivalent to previously marketed device, Accu-Sure Insulin Syringe (K091167) manufactured by Shina Corporation. The device design, materials of construction, manufacturing method, release specifications and performance specifications are the same as the predicate device.
| Device Name | Subject Device | Predicate Device | |
|-----------------------------------|---------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| Manufacturer | SHINA MED CORPORATION | SHINA CORPORATION | |
| 510(k) Number | K152879 | K091167 | |
| Product Code | FMF | FMF | |
| Intended Use | Sure-Fine Insulin Syringes are<br>hypodermic insulin syringes for<br>subcutaneous injection of insulin. | Accu-Sure Insulin Syringe is a<br>hypodermic insulin syringe for<br>subcutaneous injection of insulin | |
| Design (Syringe) | Design | Needle / Barrel / Gasket /<br>Plunger / Needle Cap /<br>Protective end cap | Needle / Barrel / Gasket /<br>Plunger / Needle Cap /<br>Protective end cap |
| | Volume | 0.3cc, 0.5cc, 1.0cc | 0.3cc, 0.5cc, 1.0cc |
| | Gauge | 28, 29, 30,31Gauge | 28, 29, 30,31Gauge |
| | Length | 1/2", 5/16" | 1/2", 5/16" |
| Materials | Needle | STS304 | STS304 |
| | Barrel | Polypropylene | Polypropylene |
| | Plunger | Polypropylene | Polypropylene |
| | Piston | Isoprene Rubber | Isoprene Rubber |
| | Needle Cap | Polyethylene | Polyethylene |
| | Protective end cap | Polyethylene | Polyethylene |
| | Silicone | Polydimethylsiloxane | Polydimethylsiloxane |
| | Biocompatibility | Conform ISO10993-1 | Conform ISO10993-1 |
| Sterilization method<br>and S.A,L | Sterilized by ethylene<br>oxide gas SAL = 10-6 | Sterilized by ethylene<br>oxide gas SAL = 10-6 | |
{5}------------------------------------------------
#### 8. Non-clinical testing
The Sure-Fine Insulin Syringes have been designed and tested to meet the requirements of voluntary standards (listed below) and FDA Guidance documents (Guidance on the Content of Premarket Notification [510(K)] Submissions for Piston Syringes) applicable to the subject and predicate devices. The results of the non-clinical testing supported the conclusion of substantial equivalence.
### Performance Testing
The Sure-Fine Insulin Syringes have been designed and successfully tested to meet the applicable requirements outlined in ISO8537.
- Test performed per ISO8537:2007
- · ISO 8537-2007 Clause 5 Freedom from extraneous matter
- · ISO 8537-2007 Clause 6.Extraction testing
- · ISO 8537-2007 Clause 7 Lubrication of syringes and needles
- · ISO 8537-2007 Clause 8 Range of sizes
- ISO 8537-2007 Clause 9 Graduated scale
- ISO 8537-2007 Clause 10 Barrel
- · ISO 8537-2007 Clause 11 Piston/Plunger assembly
- · ISO 8537-2007 Clause 12 Nozzle
- · ISO 8537-2007 Clause 13 Needles tubing and needles
- · ISO 8537-2007 Clause 14 Performance of assembled syringe
### Biocompatibility Testing
The material of the Sure-Fine Insulin Syringes have successfully passed testing as outlined in ISO10993-1 for devices categorized as External communicating devices, Limited exposure.
- · ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for in vitro cytotoxicity
- · ISO 10993-10: 2010 Biological evaluation of medical devices Part 10: Test for Irritation and Sensitization 5.4 Intracutaneous reactivity test
- · ISO 10993-10 2010 Biological evaluation of medical devices Part 10: Teat for Irritation and Sensitization 6.2 Maximization sensitization test
- · ISO 10993-11 2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity 6.5 Acute systemic toxicity
- ISO 10993-4:2002 Biological evaluation of medical devices-Part 4: Selection of tests for interactions with blood.
- · The United States Pharmacopeia <151> (Pyrogen test)
### Sterilization and Shelf-life Testing
Sterilization of the Sure-Fine Insulin Syringes has been validated using the half-cycle method as outlined in ISO11135. Testing demonstrated maximum levels of ethylene oxide and ethylene chlorohydrins do not exceed the limits presented in IS010993-7. Shelf-life testing supports a shelf-life of 5-years after sterilization.
### Clinical Data
No prospective clinical trials were conducted in support of this Traditional 510(k)
### 9. Discussion
Sure-Fine Insulin Syringes have the same intended use, technological characteristics, materials of construction and performance specifications as the predicate device, Accu-Sure Insulin Syringe (K091167), and thus it performs in a substantial equivalent manner to the predicate device.
{6}------------------------------------------------
### 10. Conclusion
Based on the information provided in this premarket notification, SHINA MED CORPORATION concludes that the Sure-Fine Insulin Syringes are substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.