MHI-TM2000 Linear Accelerator System, VERO, Imavis
K152867 · Mitsubishi Heavy Industries, Ltd. · IYE · Oct 21, 2015 · Radiology
Device Facts
Record ID
K152867
Device Name
MHI-TM2000 Linear Accelerator System, VERO, Imavis
Applicant
Mitsubishi Heavy Industries, Ltd.
Product Code
IYE · Radiology
Decision Date
Oct 21, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Device Story
MHI-TM2000 is an image-guided linear accelerator for precision radiation therapy. It generates 6MV high-energy X-rays to treat tumors and lesions. System components include an O-ring gantry, multi-leaf collimator, electronic portal imaging device, couch, and X-ray-head-mounted gimbals. The device features a built-in imaging system and continuous beam alignment for moving targets. The subject version (Ver. 3.5) introduces 'Dynamic WaveArc' (DWA) therapy, which delivers radiation while simultaneously rotating the O-ring gantry and the rotational floor (or ring rotation only). Operated by clinical staff in a radiation therapy environment, the system provides precise dose delivery. Output is used by radiation oncologists and medical physicists to plan and execute treatment, potentially improving therapeutic outcomes for patients requiring localized radiation.
Clinical Evidence
Bench testing only. Verification and validation testing, including regression testing, were performed to ensure conformance to requirements and intended use. Testing confirmed that hazard safeguards functioned properly and the device met all applicable standards.
Indicated for radiation therapy of lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2015
Mitsubishi Heavy Industries, Ltd. % Mr. Katsuhisa Toyama Senior Administrator, Business Development Department 4-6-22 Kan-on-shin-machi, Nishi-ku Hiroshima-shi, Hiroshima-ken 733-8553 JAPAN
Re: K152867
Trade/Device Name: MHI-TM2000 Linear Accelerator System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 28, 2015 Received: September 30, 2015
Dear Mr. Toyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152867
Device Name
MHI-TM2000 Linear Accelerator System
#### Indications for Use (Describe)
MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image displays the logo and contact information for Mitsubishi Heavy Industries, LTD. The logo is a red emblem consisting of three diamond shapes arranged in a triangular formation. Below the logo, the text reads "MITSUBISHI HEAVY INDUSTRIES, LTD." followed by "MACHINERY, EQUIPMENT & INFRASTRUCTURE" and the address "6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU HIROSHIMA, 733-8553 JAPAN".
# 510(k) Summary MHI-TM2000 Linear Accelerator System
The following information is provided following the format of 21 CFR 807.92 for the MHI-TM2000 Linear Accelerator System.
Date prepared: September 17, 2015
- 1. Submitter: Mitsubishi Heavy Industries, Ltd. Machinery, Equipment & Infrastructure 4-6-22 Kan-on-shin-machi, Nishi-ku, Hiroshima 733-8553 Japan Establishment Registration Number: 3006942329 Contact person: Shuji Kaneko Manager, Medical System Engineering Section Phone: +81-82-291-2146 Fax: +81-82-294-9888
## 2. Name of the Device:
| Trade Name: | MHI-TM2000 Linear Accelerator System/ Vero/ Imavis |
|-----------------------|----------------------------------------------------|
| Common or Usual name: | Clinical Linear Accelerator |
| Classification Name: | Medical Charged Particle Radiation Therapy System |
| | 21CFR §892.5050, Class II |
| | Product Code: IYE |
## 3. Predicate devices to claim substantial equivalence:
MHI-TM2000 Linear Accelerator System: K122450
## 4. Description of the Device:
The MHI-TM2000 Linear Accelerator System (hereinafter called "MHI-TM2000 (Ver. 3.5)") is an Image-Guided Radiation Therapy device generating 6MV high energy X-ray for precision radiation therapy of tumors and conditions, where radiation treatment is indicated.
It includes a linear accelerator, multi-leaf collimator, electric portal imaging device and couch. It also includes O-ring-type gantry platform, X-ray-head-mounted gimbals, Built-in Imaging System, and function to continuously align treatment beam with moving target for precise radiotherapy of
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo and address of Mitsubishi Heavy Industries, LTD. The logo is a red emblem consisting of three diamond shapes arranged in a triangular formation. Below the logo is the company name in bold, followed by the description "MACHINERY, EQUIPMENT & INFRASTRUCTURE". The address is listed as 6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU HIROSHIMA, 733-8553 JAPAN.
moving targets, as notable characteristics.
The MHI-TM2000 (Ver. 3.5) has been developed based on MHI-TM2000 Linear Accelerator System previously cleared by FDA under K122450 (hereinafter called "MHI-TM2000 (K122450)'').
Modification was made for introducing a method of radiation therapy called "Dynamic WaveArc" ("DWA") that delivers radiation while rotating both O-ring-shaped gantry (Gantry rotation) and rotational floor of the gantry that moves entire gantry (Ring rotation), or rotating the Ring rotation only.
Radiation type, energy, and the method of generating radiation are identical between MHI-TM2000 (Ver. 3.5) and MHI-TM2000 (K122450)
# 5. Intended Use Statement:
MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
# Indications for Use Statement:
MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
## 6. Technological Characteristics:
Comparison to the predicate device is listed below:
| Feature | MHI-TM2000 (Ver. 3.5) | MHI-TM2000 (K122450) |
|-----------------------------------------------|-----------------------|----------------------|
| "Dynamic WaveArc" radiation<br>therapy method | Yes | No |
This is the only modification and all other technological characteristics are identical between MHI-TM2000 (Ver. 3.5) and MHI-TM2000 (K122450), such as Indications for use, Operation principles and Method of operation.
# 7. Summary of Performance Testing:
The design control procedures applied to the development of MHI-TM2000 (Ver. 3.5) includes requirements reviews, risk analysis, and verification and validation testing according to:.
-ISO14971: Risk Management Standard
-FDA Quality System regulation (21 CFR 820) and ISO13485: Quality Management System Standard
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo and address of Mitsubishi Heavy Industries, LTD. The logo is a red, three-diamond shape. Below the logo is the company name, followed by the description "MACHINERY, EQUIPMENT & INFRASTRUCTURE". The address is listed as 6-22, 4-CHOME, KAN-ON-SHIN-MACHI, NISHI-KU HIROSHIMA, 733-8553 JAPAN.
-Other international standards listed in "8. Standards conformance" below in 8.
Result of verification testing, including regression testing, showed conformance to applicable requirements specifications, assured hazard safeguards functioned properly.
Result of validation testing showed conformance to customer needs and intended use of the MHI-TM2000 (Ver. 3.5).
### 8. Standards conformance:
The MHI-TM2000 (Ver. 3.5) conforms with the FDA recognized consensus standards and other international standards listed below.
IEC 60601-1, IEC 60601-1-2, IEC 60601-2-28, IEC 60825-1, IEC 60976, IEC 62304 ISO13845, ISO13845, ISO14971
### 9. Conclusion:
The verification and validation testing results from bench testing, including regression testing, support the substantial equivalence of the MHI-TM2000 (Ver. 3.5) with MHI-TM2000 (K122450)
A special 510(k) is applicable to this application due to the following reasons:
-Change/ modification does not significantly affect safety and effectiveness of MHI-TM2000 (K122450);
-There is no change in basic design control activities specified in 21CFR 820.30;
-There is no change in the medical device information specified in 21CFR 807.87;
-There is no change in all Labeling including Indications for Use;
-There is no change in fundamental scientific technologies; and
-There is no change in material used.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.