SpotSkan, a multi-parameter Spot Check Device, along with the appropriate accessories mentioned / supplied with the unit, is intended to measure a single adult or pediatric (but not neonatal) patient's vital signs at the physician's clinic. SpotSkan is not a Patient Monitor. Vital signs measured include SpO2, Non-Invasive Blood Pressure, Pulse Rate and Temperature. A thermal recorder is provided for printing measured data. The user is responsible to interpret the measured data made available, and shall be a professional health care provider. The monitor is not intended for home use.
Device Story
Portable spot-check device; measures SpO2, NIBP, pulse rate, temperature; intended for adult/pediatric patients in physician clinics. Operated by professional healthcare providers; not for home use; no continuous monitoring or alarms. Inputs: physiological sensors for vital signs. Device processes signals to display numeric values on 6.5" TFT LCD; stores last 12 patient readings. Output: numeric display and thermal printout of measurements/patient history. Battery backup (2 hours) allows operation during power loss. Healthcare providers interpret displayed data to inform clinical decisions. Benefits: provides quick, portable vital sign assessment for diagnostic support.
Clinical Evidence
Bench testing only. Compliance verified via IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, ISO 80601-2-61 (pulse oximetry), ISO 80601-2-56 (thermometry), IEC 80601-2-30 (NIBP), and ANSI/AAMI SP10. Software life cycle processes validated per IEC 62304 and risk management per ISO 14971.
Technological Characteristics
Portable spot-check monitor; 6.5" LED backlit TFT LCD (640x480). Measures SpO2, NIBP, pulse rate, temperature. Includes thermal recorder. Battery backup (2 hours). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, ISO 80601-2-61, ISO 80601-2-56, IEC 80601-2-30, ANSI/AAMI SP10, IEC 62304, ISO 14971.
Indications for Use
Indicated for adult and pediatric (excluding neonatal) patients in a physician's clinic for spot-check measurement of SpO2, non-invasive blood pressure, pulse rate, and temperature.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, with three human profiles forming the staff and a ribbon-like design representing the snakes.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 23, 2016
Skanray Technologies Private Limited % Yolanda Smith, Consultant Smith Associates 1468 Harwell Ave Crofton, Maryland 21114
Re: K152831
Trade/Device Name: SpotSkan Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, MWI Dated: January 19, 2016 Received: January 21, 2016
Dear Yolanda Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours
Mitchell Zeller
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K152831
Device Name
SpotSkan,
Indications for Use (Describe)
SpotSkan, a multi-parameter Spot Check Device, along with the appropriate accessories mentioned / supplied with the unit, is intended to measure a single adult or pediatric (but not neonatal) patient's vital signs at the physician's clinic. SpotSkan is not a Patient Monitor.
Vital signs measured include SpO2, Non-Invasive Blood Pressure, Pulse Rate and Temperature. A thermal recorder is provided for printing measured data. The user is responsible to interpret the measured data made available, and shall be a professional health care provider. The monitor is not intended for home use.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
□Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda. hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K152831
# 510(k) Summary (As Per 21 CFR 807.92)
| SPONSOR | |
|----------------------|----------------------------------------------------------------------------|
| Company Name: | Skanray Technologies PVt Ltd<br>Healthcare Division |
| Company Address: | #360, KIADB Industrial Area,<br>Hebbal, Mysore - 570018, Karnataka, India. |
| Telephone: | 91-821-2407202 |
| Fax: | 92-821-2407001 |
| Contact Person: | J. Mahadevan |
| Summary Prepared | May 08, 2015 |
| Trade Name: | SPOTSKAN |
| Common/Usual Name: | Spot Vital Signs Measurement Device |
| Classification Name: | Noninvasive Blood Pressure System |
| Product Code: | DXN |
| Device Class: | Class II |
| Regulation Number: | 870.1130 |
#### Traditional 510k – 510k Summary
{4}------------------------------------------------
#### K152831
#### Predicate Device
| Company | Product | 510(k) # |
|-------------------------|--------------------------------------------------|----------|
| Welch Allyn | Spot Vital Signs LXi | K101680 |
| Philips Medical Systems | SureSigns VS4 | K120132 |
| Suntech Medicals | Cycle BP Monitor & Pulse<br>Oximeter, Model 1060 | K060820 |
| L&T Medical & Systems | Stellar 300 | K103763 |
### Device Description
SpotSkan is SKANRAY's new portable Spot Check Device, which can be used to measure SpO2, Pulse rate, Blood Pressure non-invasively and Tympanic Temperature, for adult and pediatric (but not neonatal) patients. It is a compact device which has only numeric values displayed on a 6.5" LED Backlit TFT LCD display of resolution 640*480, containing SpO2, NIBP, Pulse Rate and Temperature values. It has last 12 patients' readings saved. It does not have continuous monitoring capability of the parameters or any alarms.
Interface of a thermal recorder to take print out of real time measurements and values from patient history make the data availability through a hard copy. It has got battery backup of 2 hours, which enables it to continue measurement even during mains power OFF condition. User can enter patient ID, name, age, height, weight; and select mode and sex; these details can be recorded in the thermal print out.
The scope of application is in the medical diagnostics, where a general physician will use the monitors in clinics. The user is responsible to interpret the measured data made available, and shall be a professional health care provider. The monitor is not intended for home use.
{5}------------------------------------------------
## Indications for Use
SpotSkan, a multi-parameter Spot Check Device, along with the appropriate accessories mentioned / supplied with the unit, is intended to measure a single adult or pediatric (but not neonatal) patient's vital signs at the physician's clinic. SpotSkan is not a Patient Monitor.
Vital signs measured include SpO2, Non-Invasive Blood Pressure, Pulse Rate and Temperature. A thermal recorder is provided for printing measured data. The user is responsible to interpret the measured data made available, and shall be a professional health care provider. The monitor is not intended for home use.
## Summary of Technological Characteristics
SpotSkan is designed and developed with reference to previous 510K cleared products as a Substantial Equivalent product.
The Technologies used and the processes followed are the same as described below and is comparable and in compliance to all the required Safety and performance standards.
### Similarities:
- Spotskan is compared with Suntech Medical's Cycle BP Monitor Model 1060 for ● the Pulse Oximeter. SpO2 technology and module used in these two products are same.
- Spotskan is compared with L&T Medical & Systems Stellar 300 for the NIBP technology and module used in these two products are same.
- Spotskan and Philips Medical Systems, SureSigns VS4 monitors are compared . for the Tympanic Temperature thermometer technology and module used in these two products are same.
- Spotskan and Welch Allyn, Spot Vital Signs LXi monitor is compared to show equivalence for Spot Check device.
{6}------------------------------------------------
### Differences:
Spotskan uses a different Power supply and Battery in comparison with predicate devices. These changes are verified and validated to confirm the performance, safety and efficacy of the device in meeting the standard requirements. Also the Power supply and Battery has got the CE and UL marking adding to the safety and performance compliance.
## Non Clinical Testing
Electrical and EMC Testing for the Star 90 included the following
| IEC 60601-1 | Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility -<br>Requirements and tests. |
| IEC 60601-1-6 | Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential performance -<br>Collateral standard: Usability |
| ISO 80601-2-61 | Medical electrical equipment - Part 2-61: Particular<br>requirements for basic safety and essential performance of<br>pulse Oximeter equipment |
| ISO 80601-2-56 | Medical electrical equipment - Part 2-56: Particular<br>requirements for basic safety and essential performance of<br>clinical thermometers for body temperature measurement. |
| IEC 80601-2-30 | Medical electrical equipment - Part 2-30: Particular<br>requirements for the basic safety and essential<br>performance of automatic cycling non-invasive blood<br>pressure monitoring equipment. |
{7}------------------------------------------------
### Additional Performance Testing:
| ANSI/SSMI SP10 | Manual, Electronic or Automated Sphygmomanometers |
|----------------|-----------------------------------------------------------------------------------|
| IEC 62304 | Medical device software – software life cycle processes<br>(Software/Informatics) |
| ISO 14971 | Application of risk management to medical devices |
## Substantial Equivalence
The SpotSkan is substantially equivalent to the predicate device in Indications for Use, Materials and Design. Safety and performance testing was performed and Skanray Technologies has concluded that the device does not introduce any significant questions of safety and efficacy and is substantially equivalent to the predicate devices.
Prakash U. P.
Head - Design & Development
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.