Weely Manual Wheelchair

K152827 · Rupiani · IOR · Feb 19, 2016 · Physical Medicine

Device Facts

Record IDK152827
Device NameWeely Manual Wheelchair
ApplicantRupiani
Product CodeIOR · Physical Medicine
Decision DateFeb 19, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.3850
Device ClassClass 1

Indications for Use

The indications for use for the Weelchair is to provide mobility to persons limited to a sitting position.

Device Story

The Weely Manual Wheelchair is a manually operated, user-propelled mobility device for individuals limited to a sitting position. It consists of an aluminum frame, seat/back upholstery, large rear wheels with hand rims, and front/rear swivel casters. The rear casters function as anti-tippers. A manual tilt mechanism, operated by rear release levers and gas springs, allows for seat/back adjustment to improve patient comfort or facilitate attendant access. The device is used in clinics, nursing homes, or home settings. It does not function autonomously; motion, speed, and direction are controlled by the user or caregiver via hand rims. The device provides mobility and positioning support, potentially improving patient comfort and accessibility.

Clinical Evidence

Bench testing only. The device was tested for static stability (ISO 7176-1), brake effectiveness (ISO 7176-3), dimensions/maneuvering (ISO 7176-5), seating/wheel dimensions (ISO 7176-7), static/impact/fatigue strength (ISO 7176-8), friction coefficients (ISO 7176-13), and labeling requirements (ISO 7176-15). Upholstery flammability was tested per ISO 8191-1 and ISO 8191-2.

Technological Characteristics

Aluminum frame/support members; steel fasteners; rubber/polyurethane tires; fabric-covered foam upholstery. Manual propulsion via hand rims. Manual tilt mechanism using gas springs. Rigid lower frame, folding backrest. No electronic components or software. Sterilization not applicable.

Indications for Use

Indicated for persons limited to a sitting position requiring mobility assistance. Suitable for user-propelled or attendant-propelled use in clinical or home environments.

Regulatory Classification

Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a design that resembles an abstract bird or wing above them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 19, 2016 Rupiani % Edward A. Kroll President Spectre Solutions, Inc. 5905 Fawn Lane Cleveland, Ohio 44141 Re: K152827 Trade/Device Name: Weely Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: January 13, 2016 Received: January 19, 2016 Dear Edward A. Kroll: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements {1}------------------------------------------------ as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Michael J. Hoffmann -A for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K152827 Device Name Weely Manual Wheelchair Indications for Use (Describe) The indications for use for the Weelchair is to provide mobility to persons limited to a sitting position. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) |X | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for RUPIANI. The logo consists of a red circle above a gray circle with a white stripe, followed by the word "RUPIANI" in red, bold letters. Below the logo is the text "510(k) Premarket Notification K152827". RUPIANI Weely Manual Wheelchair ### Rupiani Weely Manual Wheelchair 510(k) Summary (Modified February 12, 2016) #### l. SUBMITTER Rupiani 74, avenue du 8 mai 1945 - 69120 Vaulx en Velin - FRANCE Tel: +33(0)4 37 45 02 54 Fax +33(0)4 37 45 02 55 Web: www.rupiani.fr Email: info@rupiani.fr #### II. DEVICE Name of Device: Weely Manual Wheelchair Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical (21 CFR 890.3850) Regulatory Class: I Product Code: IOR #### III. PREDICATE DEVICE PDG Product Design Group Fuze T50 Manual Wheelchair (K063736) #### IV. DEVICE DESCRIPTION Device Description: The Rupiani Weely Manual Wheelchair (Weely) is a manually operated, user propelled, manual, mechanical wheelchair. It's intended function and use is to provide mobility to persons limited to a sitting position. It may also be used in as an attendant propelled transport device in a health care environment such as a hospital, nursing home or extended care facility. The Weely consists primarily of an upper and lower frame assembly, upholstery for seating, large wheels for propelling the chair and rear swivel type pivoting casters for turning. The wheels and casters are mounted to the lower frame assembly. The wheelchair seat and back mount to the upper frame assemblies are constructed from aluminum tubing. The rear casters act as "anti-tippers and provide added stability The lower frame is of rigid (non-folding) design. The back rest can be folded to allow for storage and transport of the chair when it is not in use. The Weely is equipped with a Varilite rigid backrest and a Varilite EVOLUTION cushion covered by SooFleece III mesh fabric {4}------------------------------------------------ Image /page/4/Picture/0 description: The image contains the logo for RUPANI. The logo consists of a stylized letter "i" on the left and the word "RUPIANI" on the right. The "i" is a sphere with a red dot on top and a gray stripe running through it. The word "RUPIANI" is in red and is in a bold sans-serif font. provided by Ventex. The upholstery materials meet the requirements of ISO 8191-1 Part 1:Ignition Source: Smoldering Cigarette and - ISO 8191-2 Part 2: Ignition Source: Flame Match-Flame Equivalent Standard for Resistance to Ignition of Upholstered Parts. The Weely includes a tilt mechanism which allows the seat and back of the wheelchair to be tilted. This feature is used to provide comfort to those users who may be confined to the wheelchair for extended periods of time. The tilt mechanism can also serve as an attendant aid in those situations where a patient needs to be tilted for attendant access. ## Device Function Device function is dependent solely upon the wheelchair user or caregiver. It does not function on its' own in any manner. The wheelchair user or caregiver controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels. ### Scientific Concepts There are no complex scientific concepts related to the Weely manual wheelchair. It is a simple, basic, manually operated mobility device. ### Significant Physical and Performance Characteristics: ### Design: The Weely manual wheelchair consists primarily of an upper and lower frame assembly, upholstery for seating, large wheels for propelling the chair and rear swivel type pivoting casters for turning. The rear casters also serve as an "anti-tipper" system which minimizes the potential for the chair to tip over backwards. The wheels and casters are mounted to the lower frame assembly. The wheelchair seat and back mount to the upper frame assembly. Both frame assemblies are constructed from aluminum tubing. The Weely features a tilt mechanism which allows the seat and back of the wheelchair to be tilted. This feature is used to provide comfort to those users who may be confined to the wheelchair for extended periods of time. The tilt mechanism can also serve as an attendant aid in those situations where a patient needs to be tilted for attendant access. ### Materials: Materials used are: - . Aluminum frame, support members, wheels and components - Steel fasteners and components - Rubber and Polyurethane tires - . Fabric covered foam upholstery {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for RUPANI. The logo consists of a stylized letter "i" with a red dot on top, followed by the word "RUPIANI" in red, bold letters. The logo is set against a white background. ### Physical Properties: The Weely manual wheelchair consists primarily of an aluminum frame assembly, a back rest frame, seat and back rest upholstery, large rear wheels with hand rims for self-propelling the chairs and front and rear swivel pivoting casters for turning. #### V. INDICATIONS FOR USE The indications for Use for the Weely Manual Wheelchair are to provide mobility to persons limited to a sitting position. This identical to the predicate device. ### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The technology and principle of operation for the Weely and the Fuze T50 are identical. They both consist of aluminum frames with a seat and large wheels with hand rims for propelling the device. Smaller, pivotinq type casters are mounted on the front of the chairs for steering and turning. Device function is dependent solely upon the wheelchair user or the caregiver. They do not function on their own in any manner. The wheelchair user or caregiver controls motion, speed and direction by propelling the wheelchair. If both rear wheels are propelled at the same time, the chair will move forward in a straight direction. If only the left rear wheel is propelled the chair will turn to the right. If only the right rear wheel is propelled the chair will turn to the left. The tilt feature is manually operated and is activated by two release levers located at the rear of the wheelchair. It utilizes standard gas springs for maintaining the static position of the chair. The tilt angle is adjusted by squeezing the levers inward simultaneously until the desired tilt angle has been reached. Once any of the two levers are released, the chair will remain at the angle chosen. ### Discussion of Similarities and Differences: The Weely manual wheelchair is substantially equivalent to the predicate PDG Fuze T50 wheelchair (K063736) in technology, function, performance and materials. It has the same indications for use which is to provide mobility for persons restricted in a sitting position. The frame width, depth and the back cane heights vary slightly between the devices, but since these are only used to better fit the device to the user's needs, these slight differences do not cause any concerns for the safety and effectiveness of the weight limit differs slightly as the maximum patient weight for the Weely is 280 lbs. where the predicate is 250 lbs. This difference has no effect on safety or effectiveness and both devices are clearly labeled with maximum patient weight capacity. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for Rupiani. The logo consists of a red sphere above a gray sphere with a white line running through it on the left side. To the right of the sphere is the word "RUPIANI" in red, bold letters. With regard to accessories and add-ons, both devices offer the same types of armrests, backrests, hangers and footplates. These accessories are made of the exact same materials for both devices and thus do not raise any questions for the safety and effectiveness of the Weely wheelchairs. #### VII. PERFORMANCE DATA The Weely Manual Wheelchairs have been tested to the following standards; - ISO 7176-1:1999 Determination of Static Stability - ISO 7176-3:2012 Determination of Effectiveness of Brakes - ISO 7176-5:2008 Determination of Overall Dimensions, Mass and Maneuvering - Space - ISO 7176-7:1998 Determination of Seating and Wheel Dimensions - ISO-7176-8:1998 Requirements and Test Method for Static Impact and Fatique Strength - ISO 7176-13:1989 Determination of Coefficients of Friction of Test Surfaces - ISO 7176-15 Requirements for Information Disclosure, Documentation and Labeling - ISO 8191-1: 1987 Furniture Assessment of the Iqnitability of Upholstered Furniture (Part 1:Ignition Source: Smoldering Cigarette) - ISO 8191-2: 1989 Furniture Assessment of the Ignitability of Upholstered Furniture (Part 2: Ignition Source: Flame Match-Flame Equivalent) The performance data published by PDG Product Design Group states that PDG Wheelchairs are tested to ISO Standards. Test reports are published on the PDG web site but cannot be reproduced without permission. The Fuze T50 510(k) Summary states that the Fuze T50 has been tested to and meets the requirements of ISO 7176 Determination of Static Stability. Further, manual wheelchairs are subject to the 1995 FDA quidance document entitled: "Guidance Document for the Preparation of Premarket Notification [(510(k)] Applications Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles." This quidance includes various safety and performance requirements that the Fuze T50 would have to have met or addressed prior to being cleared by FDA. Meeting these requirements also serves as a basis for substantial equivalence of the Weely to its' predicate. #### VIII. CONCLUSIONS: Based on the above discussion, Rupiani believes that the Weely wheelchair is substantially equivalent to the PDG Fuze T50 (K063736). While there are some differences between the Weely and its predicate, these differences are minor and do not alter the intended function and use of the device, nor do they raise any new questions pertaining to safety or effectiveness. Performance testing and compliance with FDA guidance for manual wheelchairs supports the safety of the Weely manual wheelchairs. Since the predicate device was cleared based in part on {7}------------------------------------------------ Image /page/7/Picture/0 description: The image features the logo for RUPIANI, with a stylized icon to the left and the company name in bold, red letters to the right. The icon consists of a gray sphere with a curved, dark gray line running through it, topped with a red dot. A thin, red line runs along the bottom edge of the logo, providing a subtle border. 510(k) Premarket Notification K152827 these data, conformance with the same requirements demonstrate that the Weely will perform as intended in the specified use conditions. Further, technology, principle of operation and indications for use between Weely and the Fuze T50. This demonstrates that they perform comparably to the predicate device that is currently marketed for the same intended use. Therefore, the Weely manual wheelchairs are substantially equivalent to the Fuze T50 device. {8}------------------------------------------------ Image /page/8/Picture/0 description: The image contains the word "RUPIANI" in red, bold letters. To the left of the word is a logo that features a red dot above a gray circle with a white design inside. The logo and word are on a white background. There is a red line at the bottom of the image. 510(k) Premarket Notification K152827 RUPIANI Weely Manual Wheelchair | Feature/Specification | Fuze T50 | Rupiani<br>Weely | Discussion of Differences | |-------------------------|-------------------------------------------------------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Accession Number | K063736 | TBD | NA | | Clearance Date | January 17, 2007 | TBD | NA | | Intended Use | To provide mobility to<br>persons limited to a<br>seated position | Same | No difference | | Seat to Floor Height | 13"-20" | 16.5" -<br>18.5" | No difference. The Weely is within the same range as<br>the Fuze. | | Seat Width | 15" - 20" | 15", 17"<br>and 19" | No difference. The Weely is within the same range as<br>the Fuze | | Seat Depth | 15"-20" | 17" - 22" | The Weely is slightly wider due to its' higher weight<br>limitation for which it is labeled. Chair width does not<br>affect safety or effectiveness. It indicates that the Weely<br>will accommodate slightly larger patients. | | Overall Length | 45" with footrest | 46" with<br>footrest | A one inch difference in overall length does not affect<br>safety or effectiveness. | | User Weight Limit | 250 lbs. | 280 lbs. | This has no effect on safety or effectiveness. Both<br>products are labeled with their maximum weight limit. | | Tilt Feature | Yes | Yes | No difference. Both have the tilt feature. | | Tilt Range | Zero to 50 degrees. | Zero to 35<br>degrees | No difference. The Weely is within the same range as<br>the Fuze | | Frame Material | Aluminum | Same | No difference | | Frame Type | Rigid lower frame with<br>folding backrest | Same | No difference | | Front Caster | 5",6" or 8" | 6" | No difference. Both offer 6" diameter casters | | Large Wheel | 12", 16", 20", 22", or<br>24" | 12", 22" or<br>24" | No difference. The Fuze offers more sizes however the<br>Weely sizes are with the range of the Fuze. | ### Weely Manual Wheelchair Technological Comparison to Predicate Device
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...