← Product Code [LLZ](/productcode/LLZ) · K152822

# VidiStar HeartView (K152822)

_Vidistar, LLC · LLZ · Nov 25, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152822

## Device Facts

- **Applicant:** Vidistar, LLC
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 25, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The HeartView software application is intended to provide an automated processing, review, quantification, and multidimensional review of nuclear medicine cardiology medical images, and specifically, radionuclides distributed in the body using planar and tomographic short axis images. HeartView may be used in various clinical settings including a hospital, clinic, imaging center, physician office, or remote locations. HeartView implements algorithms for automatic quantification of myocardial perfusion single photon emission computerized tomography (SPECT) as well as quantification of ejection fraction, wall motion, and thickening from gated myocardial perfusion SPECT images.

## Device Story

HeartView is an internet-based application for processing and quantifying nuclear medicine cardiology images. Input consists of short axis slices reconstructed from raw gated and averaged SPECT datasets (rest and stress). The system processes datasets in 3D rather than individual slices, enforcing a constant mid-myocardium volume constraint across the heart cycle. Output includes 3D plots and quantitative metrics for cardiac function, such as ejection fraction, wall motion, wall thickening, transient ischemic dilation, and phase analysis. Used in hospitals, clinics, imaging centers, or remote locations by physicians to assist in clinical diagnostic assessments. The software provides automated analysis to support clinical decision-making regarding myocardial perfusion and cardiac performance.

## Clinical Evidence

No clinical trials; performance supported by bench testing. Comparative performance testing conducted using anonymized patient imaging studies. Quantitative precision compared via statistical software and qualitative output compared via blinded independent cardiologist review. Results demonstrated the subject device is comparable to the predicate device in both quantitative and qualitative outputs.

## Technological Characteristics

Internet-based software application. Implements 3D processing algorithms for SPECT data. Complies with NEMA PS 3.1-3.20 (DICOM) and ISO/IEC 10918-1 (JPEG) standards. Software level of concern: moderate.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Xeleris 3.1 Processing and Review Workstation ([K130884](/device/K130884.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure in profile, composed of three overlapping shapes.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 25, 2015

Vidistar, LLC % Ms. Kathryn Becker Principal Translational Science Solutions, LLC 92 Hasell Street #401 CHARLESTON SC 29401

Re: K152822

Trade/Device Name: Vidistar HeartView Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communication system Regulatory Class: II Product Code: LLZ Dated: September 29, 2015 Received: September 30, 2015

Dear Ms. Becker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ocks

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152822

Device Name

VidiStar HeartView

#### Indications for Use (Describe)

The HeartView system is an internet-based application intended for use by nuclear medicine or ractitioners and referring physicians for the automated processing, review, quantification, and multidimensional review of nuclear medicine cardiology medical images, and specifically, radionuclides distributed in the body using planar and tomographic short axis images.

HeartView may be used in various clinical settings including a hospital, clinic, imaging center, physician office, or remote locations.

The HeartView system implements algorithms for automatic quantification of myocardial perfusion single photon emission computerized tomography (SPECT) as well as quantification, wall motion, and thickening from gated myocardial perfusion SPECT images.

Gated results are presented as 3D plots that can be used to depict, localize, and/or quantify the distribution of radionuclide tracers and anatomical structures in scanned body tissue for clinical diagnostic purposes, including quantitative assessments of cardiac function (e.g., systolic and diastolic function, regional wall thickening, wall motion, transient ischemic dilation, and phase analysis).

| Type of Use (Select one or both, as applicable)                                      |                                                                                     |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY

| Submission Date:       | September 24, 2015                                                                                          |
|------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter Information: |                                                                                                             |
| Submitted By:          | VidiStar, LLC<br>PO Box 8539<br>Greenville, SC 29604                                                        |
| Contact Person:        | Craig Walker, MHA<br>Chief Executive Officer<br>VidiStar, LLC<br>Tel: (512) 797-1910<br>Fax: (864) 349-2169 |
| Email:                 | craig.walker@VidiStar.com                                                                                   |
| Device Information:    |                                                                                                             |

| Trade Name:          | HeartView                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Nuclear Medicine Workstation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification Name: | System, Image Processing, Radiological<br>21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulatory Class:    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code:        | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device:    | Xeleris 3.1 Processing and Review Workstation (K130884)<br>GE Healthcare<br>Class II (21 CFR 892.2050; Product Code LLZ)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Device Description:  | The HeartView platform is a comprehensive internet-based<br>application designed to process, review, and automatically<br>perform quantitative analysis of cardiac nuclear medicine<br>procedures. HeartView implements algorithms for<br>automatic quantification of myocardial perfusion SPECT,<br>as well as quantification of ejection fraction, wall motion,<br>and thickening from gated myocardial perfusion SPECT.<br>The algorithm takes short axis slices reconstructed from<br>raw datasets of gated and averaged acquisitions in rest and<br>stress, and operates in multi-dimension (3D), rather than<br>processing individual slices separately. For gated datasets, |

{4}------------------------------------------------

|                      | it processes the dataset as a whole rather than processing             each frame separately, which adds additional knowledge to             the algorithm for its computations, and allows enforcement             of the constraint that the mid-myocardium volume is             constant during the whole heart beat cycle.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use:        | The HeartView software application is intended to provide             an automated processing, review, quantification, and             multidimensional review of nuclear medicine cardiology             medical images, and specifically, radionuclides distributed             in the body using planar and tomographic short axis             images. HeartView may be used in various clinical settings             including a hospital, clinic, imaging center, physician             office, or remote locations. HeartView implements             algorithms for automatic quantification of myocardial             perfusion single photon emission computerized             tomography (SPECT) as well as quantification of ejection             fraction, wall motion, and thickening from gated             myocardial perfusion SPECT images. |
| Indications for Use: | The HeartView system is an internet-based application             intended for use by nuclear medicine or radiology             practitioners and referring physicians for the automated             processing, review, quantification, and multidimensional             review of nuclear medicine cardiology medical images, and             specifically, radionuclides distributed in the body using             planar and tomographic short axis images.                                                                                                                                                                                                                                                                                                                                                                                             |
|                      | HeartView may be used in various clinical settings             including a hospital, clinic, imaging center, physician             office, or remote locations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
|                      | The HeartView system implements algorithms for             automatic quantification of myocardial perfusion single             photon emission computerized tomography (SPECT) as             well as quantification of ejection fraction, wall motion, and             thickening from gated myocardial perfusion SPECT             images.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|                      | Gated results are presented as 3D plots that can be used to             depict, localize, and/or quantify the distribution of             radionuclide tracers and anatomical structures in scanned             body tissue for clinical diagnostic purposes, including             quantitative assessments of cardiac function (e.g., systolic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{5}------------------------------------------------

### and diastolic function, regional wall thickening, wall motion, transient ischemic dilation, and phase analysis). Comparison to The device and the predicate device are both image Predicate Device post processing workstations devices that provide Technology: similar features of visualization and quantitative analysis and do not raise any new questions of safety or effectiveness. Performance Data: The following performance data were provided in support of the substantial equivalence determination: Software Verification and Software verification and validation testing were Validation Testing: conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern, as a malfunction of, or a latent design flaw in, the software device could lead to an erroneous diagnosis or a delay in delivery of appropriate medical care that would likely lead to minor injury. Comparative performance testing was conducted to Comparative Testing: demonstrate that the subject device is equivalent to the predicate device for the review of nuclear studies in terms of quantitative (software-calculated values) and qualitative (subjective clinical reading) output. To support this performance claim, anonymized patient imaging studies were compared quantitatively (precision utilizing statistical software) and qualitatively (clinical review by a blinded independent cardiologist) . The subject device was found to be comparable to the predicate device in terms of qualitative and quantitative output. Additional Information: Performance Standards: The subject device is in compliance with the following voluntary performance and safety standards: . NEMA PS 3.1 - 3.20 (2011) Digital Imaging and Communications in Medicine (DICOM) Set ISO/IEC 10918-1 First edition 1994-02-15 Information . technology- Digital compression and coding of continuous-tone still images: Requirements and

{6}------------------------------------------------

|                  | guidelines [including: Technical Corrigendum 1 (2005)]                                                                                                                                                                                                                                                                                                                                                     |
|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Other Standards: | The present 510(k) was prepared in consideration of the following guidance documents:                                                                                                                                                                                                                                                                                                                      |
|                  | Guidance for the Submission of Premarket Notifications for Medical Image Management Devices (2000) Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (2005) Content of Premarket Submissions for Management of Cybersecurity in Medical Devices (2014)                                                                                                           |
| Conclusion:      | Based on conformance to FDA recognized voluntary consensus standards, development under applicable FDA guidance, and the extensive product testing described within this 510(k) premarket notification, VidiStar provides evidence that its HeartView system is as safe and effective, and performs in a substantially equivalent manner to GE Healthcare's Xeleris 3.1 Processing and Review Workstation. |

{7}------------------------------------------------

---

**Source:** [https://fda-staging.innolitics.com/device/K152822](https://fda-staging.innolitics.com/device/K152822)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).