K152813 · Orich Medical Equipment (Tianjin) Co., Ltd. · KPR · Dec 2, 2016 · Radiology
Device Facts
Record ID
K152813
Device Name
FS-500DDR Medical Radiographic X-Ray System
Applicant
Orich Medical Equipment (Tianjin) Co., Ltd.
Product Code
KPR · Radiology
Decision Date
Dec 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Pediatric
Indications for Use
FS-500DDR Medical Radiographic X-Ray System is designed to be used by a qualified/trained doctor or technician on adult and pediatric for taking diagnostic images of the skull, spinal column, chest, abdomen, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
Device Story
FS-500DDR is a stationary digital radiographic (DR) X-ray system; utilizes high-voltage generator, X-ray tube, collimator, and flat-panel detector to capture diagnostic images. System configurations include UC arm or rail suspension assembly for tube mounting. Operated by trained clinicians/technicians in hospitals or research institutes; patient positioned sitting, standing, or lying. System produces digital images for diagnostic review; output assists clinicians in anatomical assessment and diagnosis. Benefits include high-quality diagnostic imaging for various body parts.
Clinical Evidence
Clinical concurrence study of image pairs performed to demonstrate equivalent diagnostic capability. Bench testing included electrical safety (EN 60601-1), EMC (EN 60601-1-2), and performance standards (IEC 60601-2-7, 2-28, 2-54, 1-3). Biocompatibility testing conducted per ISO 10993-1, 5, and 10.
Technological Characteristics
Stationary X-ray system; 50kW high-voltage generator (40-150kV, 10-630mA); Toshiba E7843 X-ray tube; FDX4343R flat-panel detector. Components include UC arm or rail suspension, radiographic table, and FU-800CCS digital acquisition/processing system. Power: 380Vac, 50/60Hz. Compliance with IEC/EN 60601 series standards.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic imaging of the skull, spinal column, chest, abdomen, and extremities. Used by qualified doctors or technicians with patients in sitting, standing, prone, or supine positions.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Revolution XR/d Digital Radiographic Imaging System (K012389)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three human profiles facing to the right, with flowing lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 2, 2016
ORICH MEDICAL EQUIPMENT (TIANJIN) CO., LTD. % Mr. Jun Peng Principal Consultant P&L Scientific, Inc. 6840 SW 45th Lane, Unit 5 MIAMI FL 33155
Re: K152813
Trade/Device Name: FS-500DR Medical Radiographic X-Ray System Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: September 26, 2016 Received: November 8, 2016
Dear Mr. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K152813
Device Name
FS-500DDR Medical Radiographic X-Ray System
#### Indications for Use (Describe)
FS-500DDR Medical Radiographic X-Ray System is designed to be used by a qualified/trained doctor or technician on adult and pediatric patients for taking diagnostic images of the skull, spinal column, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995
### *DO NOT SEND YOUR COMPEETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
510(k) summary of Safety and Effectiveness as required by the Safe Medical Devices Act of 1990 and codified in 21 CFR 807.92 upon which Substantial Equivalence is based: The Assigned 510(k) Number is: k152813
## 1. Submitter Information:
#### Sponsor/510(k) Owner: . ORICH MEDICAL EQUIPMENT (TIANJIN) CO., LTD. No. 16 Cuiming Road YAT-SEN Scientific Industrial Park, Tianjin Economic Technological Development Area, Tianjin 301726, China Phone: +86 22 8216 1240 Fax: +86 22 8216 0813
Date: September 26, 2016
#### 510(k) Contact Name: .
Mr. Jun Peng P&L SCIENTIFIC, INC. 6840 SW 45TH LN Unit 5 Miami, FL 33155 Phone: (305) 609 4701 Fax: (305) 397 0289 Email: jpeng@plscientificinc.com
## 2. Device Name and Classification
| Trade Name: | FS-500DDR Medical Radiographic X-Ray System |
|----------------------|----------------------------------------------|
| Common Name: | Direct Digital Radiographic X-Ray System |
| Classification Name: | Stationary X-Ray System, Product code 90 KPR |
| Classification: | 21 CFR 892.1680, Class II |
| Product code: | KPR |
## 4. Predicate Devices:
| Manufacturer Name | GE Medical Systems, Inc. | Orich Medical Equipment<br>(Tianjin) Co., Ltd. |
|-----------------------|--------------------------------------------------------|------------------------------------------------|
| Regulation Number | K012389 | K120034 |
| Trade Name | GE Revolution XR/d Digital Radiographic Imaging System | DF-211H Radiographic X-Ray System |
| Common Name | System, X-Ray, Stationary | System, X-Ray, Stationary |
| Classification Name | Stationary X-Ray System | Stationary X-Ray System |
| Classification Number | 21 CFR 892.1680 | 21 CFR 892.1680 |
| Product Code | KPR | KPR |
| Date of Clearance | Aug. 10, 2001 | Apr. 17, 2012 |
## 5. Description of Device
{4}------------------------------------------------
The FS-500DDR is a Medical Radiographic X-Ray System with digital radiography (DR) technology. It is consisting of a high voltage generator, control console, X-Ray tube assembly, collimator, UC bracket or Rail suspension assembly, solid-state X-Ray flat panel detector, work station and radiographic table. This product is applicable to clinical diagnostic radiography in all hospitals whether large or small. It is applicable to the radiography of various parts of human body. It can also be used in scientific research and education of medical scientific research institutes and medical colleges. Power supply requirements: three phase: 380Vac, 50Hz/60Hz; fluctuation range for voltage: 10%; the power source capacity: 50kVA, the range of tube voltage is 40-150kV, with current 10-630mA. The time range of exposure is 1ms-10s, the range of mAs is 0.1-630mAs.
There are two types of configuration (1001-1002) for FS-500DDR; the difference is for the mounting of the X-ray tube. For X-ray tube mounting the configuration is either the UC arm or Rail suspension assembly.
| Component | Model | Manufacturer |
|-------------------------------------------------------|---------------|----------------------------------------------|
| Digital image<br>acquisition and<br>processing system | FU-800CCS | Orich Medical Equipment (Tianjin)Co., Ltd. |
| Medical monitor | RTV-5 | Orich Medical Equipment (Tianjin)Co., Ltd. |
| High-voltage generator | HK801-1(50kW) | Orich Medical Equipment (Tianjin)Co., Ltd. |
| X-ray tube assembly | DX7254/150 | Orich Medical Equipment (Tianjin)Co., Ltd. |
| X-ray tube | E7843 | Toshiba Electron Tubes and Devices Co., Ltd. |
| Flat panel detector | FDX4343R | Toshiba Electron Tubes and Devices Co., Ltd. |
| UC arm | FSU-I | Orich Medical Equipment (Tianjin)Co., Ltd. |
| Mobile radiographic<br>patient table | YDC-I | Orich Medical Equipment (Tianjin)Co., Ltd. |
| Beam limiting device | S/DS-2 | Orich Medical Equipment (Tianjin)Co., Ltd. |
Table 1-1 Components of FS-500DDR (1001)
Table 1-2 Components of FS-500DDR (1002)
| Component | Model | Manufacturer |
|-------------------------------------------------------|---------------|----------------------------------------------|
| Digital image<br>acquisition and<br>processing system | FU-800CCS | Orich Medical Equipment (Tianjin)Co., Ltd. |
| Medical monitor | RTV-5 | Orich Medical Equipment (Tianjin)Co., Ltd. |
| High-voltage generator | HK801-1(50KW) | Orich Medical Equipment (Tianjin)Co., Ltd. |
| X-ray tube assembly | DX7254/150 | Orich Medical Equipment (Tianjin)Co., Ltd. |
| X-ray tube | E7843 | Toshiba Electron Tubes and Devices Co., Ltd. |
| Flat panel detector | FDX4343R | Toshiba Electron Tubes and Devices Co., Ltd. |
| Rail suspension<br>assembly | FSU-III | Orich Medical Equipment (Tianjin)Co., Ltd. |
| Elevating photography<br>table | DFY-III | Orich Medical Equipment (Tianjin)Co., Ltd. |
{5}------------------------------------------------
| Beam limiting device | S/DS-2 | Orich Medical Equipment (Tianjin)Co., Ltd. |
|----------------------|--------|--------------------------------------------|
|----------------------|--------|--------------------------------------------|
## 6. Intended Use
FS-500DDR Medical Radiographic X-Ray System is designed to be used by a qualified/trained doctor or technician on adult and pediatric for taking diagnostic images of the skull, spinal column, chest, abdomen, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.
## 7. Comparison of Technological Characteristics
The FS-500DDR Medical Diagnostic X-Ray system is similar to the predicate devices:
- Has the same intended use and indications for use .
- . Utilizes the same operating principle
- Incorporates the same basic design ●
- Incorporates the same technological characteristics .
- Tested to the same electrical and electromagnetic safety standards for medical . electrical equipment
- . Manufactured under a quality system
The differences between the subject devices and predicate devices:
- The subject device utilizes a lower frequency X-ray generator which does not raise . any level of safety concern and affect any effectiveness; the generator frequency does affect the X-ray exposure parameters.
- . The subject device utilizes a radiographic table with smaller weight capacity which does not raise any level of safety concern and affect any effectiveness.
- . The inherent filtration for the collimator in the subject device is different from the predicate device; this difference does not affect the safety and effectiveness.
- . The subject device utilizes a different X-Ray flat panel detector, however, both the bench and clinical test demonstrate that does not raise the level of safety concern and affect any effectiveness.
### 8. Summary Performance Data
- EN 60601-1 Medical electrical equipment Part 1: General requirements for basic ● safety and essential performance 2006 + AC:2010
- EN 60601-1-2:2007 (Third Edition, 2007), Medical Electrical Equipment - Part 1-2: General Requirements for Safety - Collateral Standard: Electromagnetic
- IEC 60601-2-7:1998 Medical electrical equipment -Part 2-7: Particular requirements for the safety of high-voltage generators of diagnostic X-ray generators
- . IEC 60601-2-28:2010 Medical electrical equipment - Part 2-28: Particular requirements for the basic safety and essential performance of X-ray tube assemblies for medical diagnosis
- . IEC 60601-2-54:2009 - Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- . IEC 60601-1-3:2008 Medical electrical equipment - Part 1: General requirements for safety - 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment + AC2010
{6}------------------------------------------------
- . ISO 10993-5:1999: Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity
- . ISO 10993-10:2002, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity
- ISO 10993-1:2009 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing
- Clinical concurrence study of images pairs to demonstrate the equivalent diagnostic . capability
## 9. Conclusion
After analyzing both bench and external laboratory testing data, including electrical safety, mechanical safety and EMC compliance, the intended use and supporting data can conclude that the device in the submission is safe and effective as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.