← Product Code [MQL](/productcode/MQL) · K152782

# SYDNEY IVF SPERM MEDIUM, SYDNEY IVF SPERM GRADIENT KIT, SYDNEY IVF SPERMIENT (K152782)

_William A. Cook Australia Pty. , Ltd. · MQL · Dec 21, 2015 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152782

## Device Facts

- **Applicant:** William A. Cook Australia Pty. , Ltd.
- **Product Code:** [MQL](/productcode/MQL.md)
- **Decision Date:** Dec 21, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.6180
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

Sydney IVF Sperm Medium is intended for use during in vitro fertilization procedures to process sperm. Sydney IVF Sperm Gradient Kit is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma. Sydney IVF Spermient is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.

## Device Story

Sydney IVF Sperm Medium, Sperm Gradient Kit, and Spermient are reproductive media used in IVF laboratories to process sperm. Sperm Medium provides a nutritious environment for sperm motility during 'swim-up' procedures. Sperm Gradient Kit and Spermient utilize density gradient centrifugation to separate motile sperm from seminal plasma and remove inhibitory compounds. These products contain Human Serum Albumin (HSA) and Gentamicin. They are supplied in sterile glass vials with Fluorotec-coated stoppers. The devices are used by embryologists or laboratory technicians in clinical IVF settings. By isolating high-quality, motile sperm, these media facilitate successful fertilization, potentially improving outcomes for patients undergoing fertility treatments.

## Clinical Evidence

Bench testing only. Stability studies validated a 20-week shelf-life at 2-8°C. Testing included endotoxin levels, Mouse Embryo Assay (MEA), Human Sperm Survival Assay (HSSA), pH, osmolality, sterility, and concentrations of pyruvate and ammonia.

## Technological Characteristics

Reproductive media containing 10 mg/mL HSA and 0.01 mg/mL Gentamicin. Sperm Gradient Kit/Spermient contain silane-coated silica particles in HEPES buffer. Osmolality: 285-295 mOsm/kg. Endotoxin: < 0.40 EU/mL. Packaging: Borosilicate type 1 glass vials with Fluorotec-coated rubber stoppers. Sterilization: Aseptic filtration.

## Regulatory Identification

Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).

## Special Controls

*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- Cook IVF Sperm Buffer, Cook IVF Sperm Medium, and Cook IVF Sperm Gradient Kit ([K002383](/device/K002383.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 21, 2015

William A. Cook Australia Pty Ltd Gordana Pozvek Senior Regulatory Affairs Specialist 95 Brandl Street Eight Mile Plains QLD 4113 Australia

Re: K152782

Trade/Device Name: Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit, Sydney IVF Spermient Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive media and supplements Regulatory Class: II Product Code: MOL Dated: November 24, 2015 Received: November 30, 2015

Dear Gordana Pozvek,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K152782

Device Name Sydney IVF Sperm Medium

Indications for Use (Describe)

Sydney IVF Sperm Medium is intended for use during in vitro fertilization procedures to process sperm.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K152782

Device Name Sydney IVF Sperm Gradient Kit

Indications for Use (Describe)

Sydney IVF Sperm Gradient Kit is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
| <input type="checkbox"/>                        |
| <input type="checkbox"/>                        |

X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K152782

Device Name Sydney IVF Spermient

Indications for Use (Describe)

Sydney IVF Spermient is intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

Attachment 5-2

Image /page/5/Picture/1 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, bold, sans-serif font, with the registered trademark symbol next to it. Below "COOK", the word "MEDICAL" is written in white, smaller, sans-serif font.

WILLIAM A. COOK AUSTRALIA PTY. LTD. 95 BRANDL STREET BRISBANE TECHNOLOGY PARK, EIGHT MILE PLAINS BRISBANE, QLD 4113, AUSTRALIA PHONE: 1800.777.222 FAX: +61.7.3841.1288 WWW.COOKMEDICAL.

# 510(k) Summary

### SUBMITTED BY:

William A. Cook Australia Pty Ltd 95 Brandl Street Eight Mile Plains QLD 4113 Australia

| Contact Person: | Gordana Pozvek Ph.D.           |
|-----------------|--------------------------------|
| Tel:            | +61 (7) 3841 1188              |
| Fax:            | +61 (7) 3841 3905              |
| E-mail:         | Gordana.Pozvek@CookMedical.com |

#### Date Prepared: September 22, 2015

### DEVICE IDENTIFICATION:

| Trade Name:       | Sydney IVF Sperm Medium (Product code: K-SISM)       |
|-------------------|------------------------------------------------------|
|                   | Sydney IVF Sperm Gradient Kit (Product code: K-SISG) |
|                   | Sydney IVF Spermient (Product code: K-SISP)          |
| Common Name:      | Sperm Processing Solutions                           |
| Regulation No:    | 21 CFR 884.6180, Reproductive Media & Supplements    |
| Regulatory Class: | II                                                   |
| Product Code:     | MQL - Media, Reproductive                            |

# PREDICATE DEVICE:

Cook IVF Sperm Buffer, Cook IVF Sperm Medium, and Cook IVF Sperm Gradient Kit (K002383), cleared September 13, 2000.

# DEVICE DESCRIPTION:

The Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient are used for processing sperm during in vitro fertilization procedures.

Sperm Medium is used to provide a "liquid" and nutritious environment for the sperm to maintain its motility for the "swim up" procedures and the following fertilization process. The medium contains 10 mg/mL Human Serum Albumin (HSA) and 0.01 mg/mL Gentamicin. It is available in 20, 50 or 100 mL vials.

{6}------------------------------------------------

Sydney IVF Sperm Gradient Kit is used to separate sperm based on density, using density gradient solutions. Sperm Gradient Kit contains both 40% density solutions and is comprised of silane coated silica particles which are diluted in a HEPES buffer. The sperm gradient has the dual purpose of enriching motile sperm populations and removing sperm inhibitory compounds normally present in the eiaculate. It contains 10 mg/mL HSA and 0.01 mg/mL Gentamicin. Sydney IVF Sperm Gradient Kit is available in 20 or 50 mL vials, packaged in a carton box, consisting of one vial filled with the 40% sperm gradient solution, and a second vial filled with the 80% sperm gradient solution.

Sydney IVF Spermient is intended to separate sperm based on density, using density gradient solutions. This is achieved by diluting the Spermient into various concentrations with a HEPES buffered solution. It contains colloidal silica coated with an inert polymer (silane) and has been formulated to be isotonic. Spermient contains 10 mg/mL HSA and 0.01 mg/mL Gentamicin. It is available in 20 or 100 mL vials.

The Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient are provided in glass vials with Fluorotec® coated rubber stoppers held in place with a tamper evident seal.. These products are single use, sterile (aseptic filtration) devices.

# INDICATIONS FOR USE:

Sydney IVF Sperm Medium is intended for use during in vitro fertilization procedures to process sperm.

Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient are intended for use during in vitro fertilization procedures to separate motile sperm from seminal plasma.

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:

The Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit & Sydney IVF Spermient, and the predicate device (K002383) have the same fundamental technology and similar technological characteristics including the following:

- . Similar chemical formulation
- . Similar performance specifications:
	- Osmolality 285 295 mOsm/kg —
	- -Endotoxin < 0.40 EU/mL
	- -A Mouse Embryo Assay (MEA) is used to screen the product for embryo toxicity.
- Same method of manufacturing process - aseptic filtration.
- Same packaging - borosilicate type 1 vials with FluroTec coated stopper and tamper evident seals.

{7}------------------------------------------------

The modification that was made to the predicate device was a change in shelf-life from 8 weeks at 2-8°C (for predicate device) to 20 weeks at 2-8°C. In addition, minor changes vere made to the formulation and specifications.

The technological characteristics of Sydney IVF Sperm Medium, Sperm Gradient Kit & Spermient are comparable to the predicate device.

# PERFORMANCE DATA:

The shelf-life of Sydney IVF Sperm Medium, Sperm Gradient Kit and Spermient has been validated in stability studies to 20 weeks at 2 - 8°C. Stability tests included endotoxin, MEA, Human Sperm Survival Assay (HSSA), pH, osmolality, sterility and the concentrations of pyruvate and the HSA by-product ammonia.

### CONCLUSION:

The results of the testing provide reasonable assurance that the Sydney IVF Sperm Medium, Sydney IVF Sperm Gradient Kit and Sydney IVF Spermient Kit is as safe and effective as the predicate device and supports a determination of substantial equivalence.

---

**Source:** [https://fda-staging.innolitics.com/device/K152782](https://fda-staging.innolitics.com/device/K152782)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
