← Product Code [KNS](/productcode/KNS) · K152771

# SB Knife Jr type, SB Knife Short type, SB Knife Standard type, SB Knife Jr type - Long (K152771)

_Sumitomo Bakelite Co., Ltd. · KNS · Jun 17, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152771

## Device Facts

- **Applicant:** Sumitomo Bakelite Co., Ltd.
- **Product Code:** [KNS](/productcode/KNS.md)
- **Decision Date:** Jun 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.4300
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47703W, MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.

## Device Story

SB Knife is a single-use, sterile, hand-held endoscopic electrosurgical knife. It connects to external Olympus or ERBE monopolar high-frequency (HF) generators, cables, and patient grounding plates. The device is operated by a physician during endoscopic procedures to perform tissue cutting, ablation, resection, incision, cauterization, hemostasis, coagulation, and vaporization within the digestive tract. The physician controls surgical outcomes by adjusting the generator's power, cycle, and duration. The device is intended for short-term, limited contact (less than 24 hours). It is not life-supporting or life-sustaining.

## Clinical Evidence

No clinical data. Substantial equivalence supported by comparative bench and animal tissue testing evaluating tissue ablation, resection, incision, cauterization, hemostasis, coagulation, vaporization, and dissection performance against the predicate device.

## Technological Characteristics

Single-use, EtO-sterilized endoscopic electrosurgical knife. Configurations include standard, short, and junior (crossing/grasping forceps) variants. Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18. Energy source: external monopolar HF generator. No software.

## Regulatory Identification

An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.

## Predicate Devices

- Olympus Medical Systems, Corp. IT Knife 2/ Hook Knife/ Flex Knife/ Triangle Tip Knife Electrosurgical Knife ([K092309](/device/K092309.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 17, 2016

Sumitomo Bakelite Co., Ltd. % Michael Beeltje Engineering Manager Vaupell Corp. 101 HP Almgren Dr. Agawam, MA 01001

Re: K152771

> Trade/Device Name: SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type, SB Knife Jr Type - Long Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KNS Dated: May 2, 2016 Received: May 6, 2016

Dear Michael Beeltje,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152771

#### Device Name

SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type, SB Knife Jr Type - Long

#### Indications for Use (Describe)

The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 005_ 510(k) Summary for SB Knife™

### 1. Submission Sponsor

Sumitomo Bakelite Co., Ltd. Medical Products Business Division 5-8, Higashi-Shinagawa 2-chome, Shinagawa-ku, Tokyo 140-0002, JAPAN

### 2. Date Prepared

May 2, 2016

### 3. Device Name

| Trade/Proprietary Name:           | SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type,<br>SB Knife Jr Type - Long |
|-----------------------------------|-------------------------------------------------------------------------------------------|
| Common/Usual Name:                | Endoscopic Electrosurgical Knife                                                          |
| Classification Name:              | Endoscopic electrosurgical unit and accessories                                           |
| Classification Regulation:        | 876.4300                                                                                  |
| Classification Panel:             | Gastroenterology – Urology Devices                                                        |
| Product Code:                     | KNS                                                                                       |
| Device Class:                     | II                                                                                        |
| FDA Establishment Registration #: | Not Available at time of Submission                                                       |

### 4. Predicate Devices

Olympus Medical Systems, Corp. IT Knife 2/ Hook Knife/ Flex Knife/ Triangle Tip Knife Electrosurgical Knife (K092309)

### 5. Device Description

The SB Knife™ is a single-patient use EtO-sterilized hand-held device that connects to an external high frequency (HF) generator and is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current. Varying generator power, cycle and duration results in a control of these surgical capabilities. The SB Knives are designed to connect with Olympus and ERBE connectors and use the disposable or reusable Olympus or ERBE grounding plates and connecting cables intended for the generators.

The device is not life supporting or life sustaining. It is designed as a short-term limited contact device (less than 24 hours), has no software, and has been certified to comply with IEC 60601-1 and IEC 60601-2-2 and IEC 60601-2-18 by UL Japan

{4}------------------------------------------------

## 6. Indications for Use/Intended Use

The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47703W, MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.

## 7. Technological Characteristics and Substantial Equivalence

The SB Knife™ consists of four configurations:

- 7.1 SB Knife™ Standard (MD-47706), which has grasping forceps and standard length blade;
- 7.2 SB Knife™ Short (MD-47704), which has grasping forceps and shorter length blade;
- 7.3 SB Knife™ Jr (MD 47703W), which has crossing forceps and standard effective length;
- 7.4 SB Knife™ Jr Long (MD 47703L), which has crossing forceps and longer effective length.

Choice of SB Knives' configurations is depend on Surgeon's/Physician's strategy for procedure, and length of endoscope used with, etc.

The SB Knife™ is substantially equivalent to the Olympus Systems, Inc. FlexKnife (K092309), having similar design, materials and configurations, intended use, indications for use. The SB Knife™ has demonstrated equivalency of use and results in side-by-side bench testing completed to defined test protocols on animal tissue, and been in use in Japan since 2009.

### 8. Non-Clinical Testing

The following testing has been performed to support substantial equivalence:

Comparative bench and animal tissue testing was performed for tissue ablation, resection, incision, cauterization, hemostasis, coagulation, vaporization and dissection of tissues using Olympus and ERBE monopolar HF generators. Testing resulted in equivalent performance to the predicate Olympus surgical knife.

Accelerated Aging and Actual Aging of Sterile Package Seal and product efficacy was completed and resulted in determination of a three-year product shelf life.

### 9. Clinical Testing

Clinical testing was not performed for this medical device.

### 10. Conclusion

The SB Knife™, as designed and manufactured, is substantially equivalent to the referenced predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K152771](https://fda-staging.innolitics.com/device/K152771)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).