SB Knife Jr type, SB Knife Short type, SB Knife Standard type, SB Knife Jr type - Long
Applicant
Sumitomo Bakelite Co., Ltd.
Product Code
KNS · Gastroenterology, Urology
Decision Date
Jun 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47703W, MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.
Device Story
SB Knife is a single-use, sterile, hand-held endoscopic electrosurgical knife. It connects to external Olympus or ERBE monopolar high-frequency (HF) generators, cables, and patient grounding plates. The device is operated by a physician during endoscopic procedures to perform tissue cutting, ablation, resection, incision, cauterization, hemostasis, coagulation, and vaporization within the digestive tract. The physician controls surgical outcomes by adjusting the generator's power, cycle, and duration. The device is intended for short-term, limited contact (less than 24 hours). It is not life-supporting or life-sustaining.
Clinical Evidence
No clinical data. Substantial equivalence supported by comparative bench and animal tissue testing evaluating tissue ablation, resection, incision, cauterization, hemostasis, coagulation, vaporization, and dissection performance against the predicate device.
Technological Characteristics
Single-use, EtO-sterilized endoscopic electrosurgical knife. Configurations include standard, short, and junior (crossing/grasping forceps) variants. Complies with IEC 60601-1, IEC 60601-2-2, and IEC 60601-2-18. Energy source: external monopolar HF generator. No software.
Indications for Use
Indicated for use in patients requiring tissue cutting within the digestive tract via high-frequency current. For prescription use only.
Regulatory Classification
Identification
An endoscopic electrosurgical unit and accessories is a device used to perform electrosurgical procedures through an endoscope. This generic type of device includes the electrosurgical generator, patient plate, electric biopsy forceps, electrode, flexible snare, electrosurgical alarm system, electrosurgical power supply unit, electrical clamp, self-opening rigid snare, flexible suction coagulator electrode, patient return wristlet, contact jelly, adaptor to the cord for transurethral surgical instruments, the electric cord for transurethral surgical instruments, and the transurethral desiccator.
Predicate Devices
Olympus Medical Systems, Corp. IT Knife 2/ Hook Knife/ Flex Knife/ Triangle Tip Knife Electrosurgical Knife (K092309)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing right, representing health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2016
Sumitomo Bakelite Co., Ltd. % Michael Beeltje Engineering Manager Vaupell Corp. 101 HP Almgren Dr. Agawam, MA 01001
Re: K152771
> Trade/Device Name: SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type, SB Knife Jr Type - Long Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KNS Dated: May 2, 2016 Received: May 6, 2016
Dear Michael Beeltje,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152771
#### Device Name
SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type, SB Knife Jr Type - Long
#### Indications for Use (Describe)
The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 005_ 510(k) Summary for SB Knife™
### 1. Submission Sponsor
Sumitomo Bakelite Co., Ltd. Medical Products Business Division 5-8, Higashi-Shinagawa 2-chome, Shinagawa-ku, Tokyo 140-0002, JAPAN
### 2. Date Prepared
May 2, 2016
### 3. Device Name
| Trade/Proprietary Name: | SB Knife Jr Type, SB Knife Short Type, SB Knife Standard Type,<br>SB Knife Jr Type - Long |
|-----------------------------------|-------------------------------------------------------------------------------------------|
| Common/Usual Name: | Endoscopic Electrosurgical Knife |
| Classification Name: | Endoscopic electrosurgical unit and accessories |
| Classification Regulation: | 876.4300 |
| Classification Panel: | Gastroenterology – Urology Devices |
| Product Code: | KNS |
| Device Class: | II |
| FDA Establishment Registration #: | Not Available at time of Submission |
### 4. Predicate Devices
Olympus Medical Systems, Corp. IT Knife 2/ Hook Knife/ Flex Knife/ Triangle Tip Knife Electrosurgical Knife (K092309)
### 5. Device Description
The SB Knife™ is a single-patient use EtO-sterilized hand-held device that connects to an external high frequency (HF) generator and is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current. Varying generator power, cycle and duration results in a control of these surgical capabilities. The SB Knives are designed to connect with Olympus and ERBE connectors and use the disposable or reusable Olympus or ERBE grounding plates and connecting cables intended for the generators.
The device is not life supporting or life sustaining. It is designed as a short-term limited contact device (less than 24 hours), has no software, and has been certified to comply with IEC 60601-1 and IEC 60601-2-2 and IEC 60601-2-18 by UL Japan
{4}------------------------------------------------
## 6. Indications for Use/Intended Use
The SB Knife™ is a single-use sterile Electrosurgical Knife (Model Numbers: MD-47703W, MD-47704, MD-47706 and MD-47703L). This device is designed to be used with Olympus and ERBE Electrosurgical monopolar HF generators, connectors and patient grounding plates to cut tissue within the digestive tract using high-frequency current.
## 7. Technological Characteristics and Substantial Equivalence
The SB Knife™ consists of four configurations:
- 7.1 SB Knife™ Standard (MD-47706), which has grasping forceps and standard length blade;
- 7.2 SB Knife™ Short (MD-47704), which has grasping forceps and shorter length blade;
- 7.3 SB Knife™ Jr (MD 47703W), which has crossing forceps and standard effective length;
- 7.4 SB Knife™ Jr Long (MD 47703L), which has crossing forceps and longer effective length.
Choice of SB Knives' configurations is depend on Surgeon's/Physician's strategy for procedure, and length of endoscope used with, etc.
The SB Knife™ is substantially equivalent to the Olympus Systems, Inc. FlexKnife (K092309), having similar design, materials and configurations, intended use, indications for use. The SB Knife™ has demonstrated equivalency of use and results in side-by-side bench testing completed to defined test protocols on animal tissue, and been in use in Japan since 2009.
### 8. Non-Clinical Testing
The following testing has been performed to support substantial equivalence:
Comparative bench and animal tissue testing was performed for tissue ablation, resection, incision, cauterization, hemostasis, coagulation, vaporization and dissection of tissues using Olympus and ERBE monopolar HF generators. Testing resulted in equivalent performance to the predicate Olympus surgical knife.
Accelerated Aging and Actual Aging of Sterile Package Seal and product efficacy was completed and resulted in determination of a three-year product shelf life.
### 9. Clinical Testing
Clinical testing was not performed for this medical device.
### 10. Conclusion
The SB Knife™, as designed and manufactured, is substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.