K152737 · Focus Medical, LLC · GEX · Jun 29, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152737
Device Name
Lite Touch
Applicant
Focus Medical, LLC
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 29, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Lite Touch™ Er YAG Laser Treatment System is intended for the vaporization, excision, dissection, ablation and/or photocoagulation of soft tissue in the following surgical specialties: Dermatology; Cosmetic surgery; Plastic and general surgery; Oral surgery; Gynecology; Otorhinolaryngology; Podiatry.
Device Story
Lite Touch is an Er:YAG laser treatment system (2940nm wavelength) used for soft tissue surgery. System comprises a cabinet containing laser source, control system, power supply, and internal cooling; beam delivered via counterbalanced articulated arm and handpiece. Operated by physicians in clinical settings. Two-stage activation process ensures control. Device vaporizes, excises, dissects, ablates, or photocoagulates soft tissue to facilitate surgical procedures across multiple specialties. Benefits include precise tissue interaction via pulsed laser energy. Physician monitors procedure via touchscreen interface; output affects clinical decision-making by enabling controlled tissue removal.
Clinical Evidence
Bench testing only. Device complies with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-22 (surgical laser safety), IEC 60825-1 (laser safety), and IEC 62366/60601-1-6 (usability). No clinical data presented.
Technological Characteristics
Flashlamp-pumped solid-state Er:YAG rod laser; 2940nm wavelength; 24W power; 35-3000mJ energy; 300-1000us pulse duration; up to 16Hz frequency. Beam delivery via articulated arm. Water/air heat exchanger cooling. Touchscreen GUI. Complies with 21 CFR 1040.10/1040.11 and various IEC/EN standards for medical electrical equipment and laser safety.
Indications for Use
Indicated for soft tissue vaporization, incision, excision, dissection, ablation, and photocoagulation in dermatology, cosmetic, plastic, general, oral, and gynecological surgery, otorhinolaryngology, and podiatry. No specific patient population age or gender restrictions stated.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three wave-like lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 29, 2016
Focus Medical LLC Mr. John Lee Owner 23 Francis J. Clarke Circle Bethel, Connecticut 06801
Re: K152737
Trade/Device Name: Lite Touch Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 5, 2016 Received: May 23, 2016
Dear Mr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Jennifer R. Stevenson -A" in a large, bold font. The text is black and is set against a white background. In the background, there is a faint, gray watermark of the FDA logo.
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152737
#### Device Name
Lite Touch™ Er YAG Laser Treatment System
#### Indications for Use (Describe)
Lite Touch™ Er YAG Laser Treatment System is intended for the vaporization, excision, dissection, ablation and/or photocoagulation of soft tissue in the following surgical specialties:
- · Dermatology
- · Cosmetic surgery
- · Plastic and general surgery
- · Oral surgery
- · Gynecology
- · Otorhinolaryngology
- · Podiatry
Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510k Summary (21 CFR 807.92c)
| Submitter: | Focus Medical |
|----------------------------|---------------------------------------------------------|
| Address: | 23 Frances J. Clarke Circle<br>Bethel, CT, USA<br>06801 |
| Phone: | 203-730-8885 |
| Fax: | 203-730-8851 |
| Date Prepared: | 11 January 2016 |
| Name of Submission Device: | Lite Touch™, previously cleared under K974460 |
| Trade or Proprietary Name: | Lite Touch™ |
| Common/Usual Name: | Lite Touch™ Er:YAG Laser Treatment System |
| Classification: | Class II |
| Regulation: | 21 CFR 878.4810 |
| Product Code: | GEX |
| Classification Name: | Powered Laser Surgical Instrument |
| Review Panel: | General and Plastic Surgery |
| Predicate Device: | Asceplion MCL 31 Dermablate (K150140) |
{4}------------------------------------------------
### Device Description:
The Lite Touch™ is an Er:YAG Laser Treatment System intended for use in the vaporization, incision, excision, dissection, ablation and/or photocoagulation of soft tissue in the following surgical specialties: Dermatology; Cosmetic surgery; Plastic and general surgery; Gynecology; Otorhinolaryngology; and Podiatry.
The Lite Touch™ was previously cleared under K974460 in October 1997. This submission contains an expansion of the previously cleared indications for use. The Lite Touch™ consists of a streamlined, maneuverable cabinet that contains the laser source, control system, power supply and internal cooling system. The laser beam has a light wavelength of 2940nm and is delivered through a counterbalanced articulated arm and hand piece. There is a two stage process of dissimilar actions required to activate the laser beam. The operation of the Lite Touch™ are completely under the physician's control.
### Indications for Use:
Lite Touch™ Er YAG Laser Treatment System is intended for the vaporization, incision, dissection, ablation and/or photocoagulation of soft tissue in the following surgical specialties:
- . Dermatology
- Cosmetic surgery ●
- Plastic and general surgery
- . Oral surgerv
- . Gynecology
- Otorhinolaryngology
- Podiatry
{5}------------------------------------------------
Technological Characteristics:
| Performance Characteristic | Lite Touch™ | MCL 31 Dermablate |
|--------------------------------|-----------------------------------------|-----------------------------------------|
| 510K | Not assigned | K150140 |
| Manufacturer | Focus Medical LLC | Asclepion Laser Technologies |
| Power | 24w | 20w |
| Energy (pulse and super pulse) | 35-3000 mJ | 2500mj max |
| Length of Pulse | 300-1000us | 100-1000us |
| Frequency of Pulse | Up to 16 Hz | 20 Hz |
| Spot size at target | 4, 6, 8, 10 mm | 1 – 12mm |
| Wavelength | 2940 nm | 2940 nm |
| Output mode | Pulsed | Pulsed |
| Beam delivery | Articulated Arm | Articulated Arm |
| Aiming Beam | 635 nm | 650nm |
| Laser medium | Flashlamp pumped solid state Er:YAG rod | Flashlamp pumped solid state Er:YAG rod |
| Cooling methods | Water/air heat exchanger | Water/air heat exchanger |
| Display/User Interface | Graphic user interface touchscreen | Graphic user interface touchscreen |
| Power calibration | Internal | Internal |
The subject and predicate device K150140 use similar technology, and differences between the subject and predicate device do not raise any new types of questions regarding safety or effectiveness for the subject device's intended use.
### Testing Performance:
The Lite Touch TM device for this 510(k) submission is fundamentally unchanged from the device that was cleared under K974460. The Lite Touch TM is in compliance with the requirements of 21 CFR 1040.10 -Performance Standards for light-Emitting Products, and with 21 CFR 1040.11 - Performance Standards for Light-Emitting Products, Specific purpose laser products. The device has been tested for conformance to IEC 60601-1 Ed.3.0: 2005 + Corr 1. 2006 + Corr 2. 2007: EN 60601-1 :2006/AC: 2010 Medical electrical equ1pment. Part 1 General requirements for basic safety and essential performance, IEC 60601-1-2 2007, EN 60601-1-2: 2007; EN 60601-1-2 2007/AC 2010 Medical electrical equipment. Part 1-2 General requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility -Requirements and tests; IEC 60601-2-22 2007, EN 60601-2-22 2013 Medical electrical equipment - Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment; I EC 60825-1: 2007. EN 60825-1: 2007 Safety of laser products. Part 1: Equipment classification and requirements: IEC 62366: 2007/ A 1 :2014; EN 62366: 2008 Medical devices-Application of usability engineering to medical devices, and with IEC 60601-1-6: 2010: EN 60601-1-6: 2010 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance-Collateral standard: Usability.
{6}------------------------------------------------
### Summary of Substantial Equivalence:
The submission and predicate device have substantially equivalent technical and performance characteristics including: power level; energy delivered; aiming beam; laser medium; cooling methods; anatomical sites applicable; warnings and contraindications; and for use on the general population. The raw materials and electrical components of the submission and predicate device have a substantially equivalent design and operating controls.
Where there are differences between the submission and the predicate device K150140, a review of the data showed that the differences did not represent new unacceptable risk to the patient of the device. After review of the subject and predicate devices, including technical, performance, and design information as mentioned above, it is concluded that the device in this 510(k) and the predicate device are substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.