← Product Code [GWN](/productcode/GWN) · K152727

# ISEN-Goggles (K152727)

_Isen Tech & Trading Company, Ltd. · GWN · Apr 26, 2016 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152727

## Device Facts

- **Applicant:** Isen Tech & Trading Company, Ltd.
- **Product Code:** [GWN](/productcode/GWN.md)
- **Decision Date:** Apr 26, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1460
- **Device Class:** Class 2
- **Review Panel:** Neurology

## Indications for Use

The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.

## Device Story

ISEN-Goggles are opaque, head-mounted goggles used to record and display eye movements; includes two internal video cameras and infrared (IR) LEDs (940nm) for illumination. Device features green LEDs (573nm) to provide optional focal points for patients. Signals from cameras are processed by an internal image processor, allowing near-simultaneous viewing of both eyes on an integrated LCD screen. Data is stored on an internal SD-type memory card and can be transferred via a micro-USB port. Operated by clinicians in a clinical setting to observe nystagmus; output allows providers to visually assess eye movement patterns to aid in diagnostic decision-making. System is battery-powered and portable, representing a miniaturized, integrated version of traditional video eye movement recorders.

## Clinical Evidence

No clinical data. Bench testing included electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), lamp safety (IEC 62471), FCC compliance (47 CFR Part 15), battery safety (IEC 62133), biocompatibility (ISO 10993), and hardware/software verification and validation.

## Technological Characteristics

Opaque goggles with adjustable headband, two IR-sensitive video cameras, 940nm IR LEDs, 573nm green LEDs, integrated LCD display, SD-type memory, and micro-USB port. Powered by internal rechargeable Li-ion battery. Standards: IEC 60601-1, IEC 60601-1-2, IEC 62471, IEC 62133, ISO 10993, FCC Part 15. Software is embedded for image processing and display.

## Regulatory Identification

A nystagmograph is a device used to measure, record, or visually display the involuntary movements (nystagmus) of the eyeball.

## Predicate Devices

- ICS Medical VEMR Video Eye Movement Recorder ([K972631](/device/K972631.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 26, 2016

Isen Tech & Trading Company, Ltd. % Linda Bovard, RAC President Bovard Consulting LLC 378 Hardy Avenue Eugene, Oregon 97404

Re: K152727

Trade/Device Name: Isen-goggles Regulation Number: 21 CFR 882.1460 Regulation Name: Nystagmograph Regulatory Class: Class II Product Code: GWN Dated: March 22, 2016 Received: March 28, 2016

Dear Ms. Bovard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

William J. Heetderks -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K152727

Device Name ISEN-Goggles

Indications for Use (Describe)

The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

# Contact Details

| Applicant Name: | ISEN Tech & Trading Co., Ltd.                             |
|-----------------|-----------------------------------------------------------|
|                 | Rm509, Bldg A, Cyber Tower, No. 2 Zhongguancun South Ave. |
|                 | Beijing, 100086, China                                    |
|                 | +86-10-5162-7073                                          |
| Contact:        | Martin Liu, CEO                                           |
|                 | martin@isen.com.cn                                        |
| Date Prepared:  | March 20, 2016                                            |

## Device Name

Trade Name: ISEN-Goggles

Common Name: IR/Video Eye Movement Viewer/Recorder

Classification Name: 882.1460 Nystagmograph; GWN Nystagmograph

## Legally Marketed Predicate Device(s)

| 510(k) Number | Product Code | Trade Name                                      | Applicant         |
|---------------|--------------|-------------------------------------------------|-------------------|
| K972631       | GWN          | ICS Medical VEMR Video<br>Eye Movement Recorder | ICS Medical Corp. |

## Device Description

This device consists of patient eye goggles with an adjustable, elastic headband, an LCD display, two green LED lights to provide optional focal points, eight infrared LED lights to provide illumination for the internal cameras, an internal SD-type memory, and buttons to operate the device's function. There is also a micro-USB port to facilitate transfer of data captured and stored in the device's SD-type memory.

The patient eye goggles are an opaque, facially worn device fitted with two small video cameras, and two wavelengths of light emitting diodes (LEDs). One set of lights provides illumination for the cameras. The other allows the user to provide a focus point for the patient, if this is desired. These signals are processed by the image processor that allows for nearsimultaneous viewing of the video images from both eyes on the attached LCD screen.

{4}------------------------------------------------

# Intended Use/Indications for use

The ISEN-Goggles are used to record and visually display movements of the eyes including nystagmus.

# Substantial Equivalence Comparison

The predicate device for this submission is the ICS Medical VEMR Video Eye Movement Recorder (ICS VEMR) from ICS Medical that was cleared under K972631.

#### Table 5-1 Summary of Similarities and Differences

| Category                                          | ISEN-Goggles                                                                                                                               | ICS VEMR<br>K972631                                                                                                                   |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Goggles fitted with IR<br>sensitive video cameras | Yes                                                                                                                                        | Yes                                                                                                                                   |
| IR Source                                         | IR LEDs<br>Peak Wavelengths ( $λ$ ) = 940nm (red)                                                                                          | IR LEDs<br>Peak Wavelength ( $λ$ ) = 940nm (red)                                                                                      |
| Lights for visual focus point                     | Yes – 573nm (green)                                                                                                                        | No                                                                                                                                    |
| Light blocking opaque<br>goggles                  | Yes                                                                                                                                        | Yes                                                                                                                                   |
| Energy Source                                     | Internal Rechargeable Battery                                                                                                              | External Power Supply                                                                                                                 |
| Components                                        | Goggles<br>Mask 1 and 2<br>Battery Charger<br>Rechargeable Li-ion Battery (not<br>supplied but required)                                   | Goggles<br>VCR and Remote<br>Separate 17" Display Monitor                                                                             |
| Weight                                            | 17.6 oz. (mask plus goggles)                                                                                                               | 12-15oz – Goggles<br>9lb 4oz – VCR and Remote<br>37lb 5oz – 17" Display Monitor<br>TOTAL: 47lb 5oz – 47lb 8 oz                        |
| Interface to control device                       | Buttons located on goggle-mounted<br>LCD.                                                                                                  | Remote Control                                                                                                                        |
| Images Splitter/Power<br>supply interface         | Image Splitter not necessary with this<br>technology                                                                                       | Yes                                                                                                                                   |
| Video Recording Capability                        | Yes (built-in SD-type memory<br>drive). A micro-USB port is<br>provided to enable data transfer.                                           | Yes (built-in VCR)                                                                                                                    |
| Video-Goggle interface                            | An LCD display is attached to the<br>goggles, allowing for simultaneous<br>display of both eyes from two video<br>cameras within the mask. | An interface that allows for display<br>of both eyes on video monitor and<br>provides the power for the video<br>cameras and the LEDs |

Both devices are essentially goggles with internal cameras and LED lights. There are some differences. For example, the ICS VEMR product has only one type of LED that displays red light (940 nm) while the ISEN-Goggles has red (940nm) LEDs for illuminating the image

{5}------------------------------------------------

being recorded and green (573nm) LEDs to provide the option of a focal point for the patient during the test.

Many of the other differences relate to the miniaturization of components since the marketing of the ICS VEMR. The ISEN-Goggles combines many of the separate features of the ICS VEMR into one device instead of three separate devices that work together. This contributes to the reduction in overall weight and size of the system. The ISEN-Goggles weigh only 17.6 oz while the ICS VEMR system weighed 471b 5oz to 471b 8oz for the ICS VEMR system. The energy sources are also different.

However, these differences do not affect the substantial equivalence; they are still using the same technological basis to accomplish the same purpose.

# Non-clinical Testing

Non-clinical testing for the ISEN-Goggles included electrical safety testing of the Goggles and battery charger to IEC 60601-1, electromagnetic compatibility testing of the Goggles and battery charger to IEC 60601-1-2, lamp safety to IEC 62471, compatibility with FCC regulations per 47 CFR Part 15, battery testing to IEC 62133, and biocompatibility testing to ISO 10993. In addition, hardware and software verification and validation was completed.

# Clinical Testing

No clinical testing is included in this 510(k).

#### Summarv

The ISEN-Goggles are substantially equivalent to the ICS VEMR in intended use, device features, and use parameters.

---

**Source:** [https://fda-staging.innolitics.com/device/K152727](https://fda-staging.innolitics.com/device/K152727)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com