← Product Code [LZA](/productcode/LZA) · K152707

# Minghao Nitrile Powder Free Patient Examination Gloves, Blue Color (K152707)

_Minghao Medical Products Co., Ltd. · LZA · Mar 17, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152707

## Device Facts

- **Applicant:** Minghao Medical Products Co., Ltd.
- **Product Code:** [LZA](/productcode/LZA.md)
- **Decision Date:** Mar 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.6250
- **Device Class:** Class 1
- **Review Panel:** General Hospital

## Indications for Use

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

## Device Story

Nitrile examination gloves; disposable; non-sterile. Function as physical barrier against body fluids and bloodborne pathogens. Material properties allow for hand conformation and flexibility during medical procedures. Used in clinical settings by healthcare providers. Protects both patient and examiner from contamination. Performance verified via tensile strength, elongation, and pinhole leak testing.

## Clinical Evidence

No clinical data. Bench testing only. Compliance with ASTM D6319-10 (dimensions, physical properties), ASTM D5151-06 (pinhole freedom), ASTM D6124-06 (residual powder), and ISO 10993-10 (biocompatibility: skin irritation and dermal sensitization).

## Technological Characteristics

Material: Nitrile rubber. Form factor: Disposable, non-sterile, blue color, powder-free (polyurethane coating). Standards: ASTM D6319-10 (dimensions/physical properties), ASTM D5151-06 (pinhole), ASTM D6124-06 (powder residual), ISO 10993-10 (biocompatibility).

## Regulatory Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Nitrile Powder Free Patient Examination Gloves, Blue Color ([K120970](/device/K120970.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2016

Minghao Medical Products Co., Ltd. % Chu Xiaoan Official Correspondent Room 1606 Bldg. 1 Jianxiang Yuan No 209 Bei Si Huan Zhong Rd Haidian District Beijing, 100083 CHINA

Re: K152707

Trade/Device Name: Minghao Nitrile Powder Free Patient Examination Gloves, Blue Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Regulatory Class: I Product Code: LZA Dated: January 28, 2016 Received: February 1. 2016

Dear Chu Xiaoan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Teiashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152707

### Device Name

Minghao Nitrile Powder Free Patient Examination Gloves, Blue Color

Indications for Use (Describe)

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### Section 6 510(k) Summary

# 510(K) Summary

"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."

"The assigned 510(k) number is: ニ

### Premarket Notification [510(k)] Summary

#### 1.0 Submitter:

| Submitter's name :      | Minghao Medical Products Co.,Ltd.                                                   |
|-------------------------|-------------------------------------------------------------------------------------|
| Submitter's address :   | West Industrial Park, Luannan County,<br>Tangshan City, 063500 Hebei Province,China |
| Phone number :          | (86)315-8230716                                                                     |
| Fax number :            | (86)315-8010500                                                                     |
| Name of contact person: | Pang Cuiru                                                                          |

Date of preparation : 2016-03-01

#### 2.0 Name of the Device

| Device Name:            | Nitrile Powder Free Patient Examination<br>Gloves, Blue Color         |
|-------------------------|-----------------------------------------------------------------------|
| Proprietary/Trade name: | Minghao Nitrile Powder Free Patient<br>Examination Gloves, Blue Color |
| Common Name:            | Exam gloves                                                           |
| Classification Name:    | Patient examination glove                                             |
| Device Classification:  | I                                                                     |
| Regulation Number:      | 21 CFR 880.6250                                                       |
| Panel:                  | General Hospital (80)                                                 |
| Product Code:           | LZA                                                                   |

### 3.0 Predicate device

| Device Name:   | Nitrile Powder Free Patient Examination Gloves,<br>Blue Color |
|----------------|---------------------------------------------------------------|
| Company name:  | Tangshan Zhonghong Pulin Plastic Co.,Ltd.                     |
| 510(K) Number: | K120970                                                       |

## 4.0 Device Description:

#### How the device functions: 4.1

Nitrile films form a barrier to body fluids and bloodborne Pathogens

#### 4.2 Scientific concepts that form the basis for the device

The nitrile rubber is water tight under normal conditions of use. It's tensile properties cause it to conform to the hand, allowing movements necessary for a

{4}------------------------------------------------

medical procedure.

#### 4.3 Physical and performance characteristics such as design, materials and physical properties:

Nitrile glove is known to create a barrier to bloodborne pathogens and body fluids. ASTM conforming tensile properties create a glove that is strong and flexible. The glove is manufactured in accordance with the requirements of ASTM D6319 and ASTM D5151 requirements.

# 5.0 Device Intended Use (Indication for use):

Nitrile Powder Free Patient Examination Gloves, Blue Color is a non-sterile disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

# 6.0 Summary of the Technological Characteristics of the Device:

The Nitrile Powder Free Patient Examination Gloves, Blue Color, non sterile are summarized with the following technological characteristics compared to ASTM or equivalent standard.

| Characteristics          | Standard                                                                             | Device performance                                                              |
|--------------------------|--------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Dimension                | ASTM standard D 6319-10.                                                             | Meets                                                                           |
| Physical Properties      | ASTM standard D 6319-10.                                                             | Meets                                                                           |
| Freedom from<br>pinholes | 21 CFR 800.20                                                                        | Meets                                                                           |
| Powder Residual          | ASTM standard D 6319-10 and<br>D6124-06(Reapproved 2011).                            | Meets<br><2mg/glove                                                             |
| Biocompatibility         | Primary Skin Irritation in rabbits<br>ISO 10993-10: Third Edition<br>2010-08-01.     | Under the conditions of the<br>study, the subject device is<br>non-irritating   |
|                          | Dermal sensitization in the<br>guinea pig ISO 10993-10: Third<br>Edition 2010-08-01. | Under the conditions of the<br>study, the subject device is<br>not a sensitizer |

## 7.0 Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:

Nitrile Powder Free Patient Examination Gloves, Blue Color, meet requirements per ASTM D6319-10.per ASTM D6124-06(Reapproved 2011), per 21 CFR 800.20 and ISO 10993-10: Third Edition 2010-08-01.

The performance test data of the non clinical tests that support a determination of substantial equivalent is the same as mentioned immediately above.

{5}------------------------------------------------

# 8.0 Substantial Equivalent Based on Assessment of Clinical Performance Data:

Clinical data was not needed for this submission.

#### 9.0 Substantial Equivalence Comparison:

| Features & Description                                                                               | Predicate Device                                                                                                                                                                                                                             | Subject Devic                                                                                                                                                                                                                     | Result of Comparison        |
|------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Company                                                                                              | Tangshan Zhonghong Pulin<br>Plastic Co.,Ltd.                                                                                                                                                                                                 | Minghao Medical Products<br>Co.,Ltd.                                                                                                                                                                                              | --                          |
| 510(K) Number                                                                                        | K120970                                                                                                                                                                                                                                      | K152707                                                                                                                                                                                                                           |                             |
| Device name                                                                                          | Nitrile Powder Free Patient<br>Examination Gloves, Blue<br>Color                                                                                                                                                                             | Nitrile Powder Free Patient<br>Examination Gloves, Blue Color                                                                                                                                                                     | same                        |
| Product Code                                                                                         | LZA                                                                                                                                                                                                                                          | LZA                                                                                                                                                                                                                               | same                        |
| Size                                                                                                 | Small/ Medium/<br>Large/X large                                                                                                                                                                                                              | Small/ Medium/<br>Large/X large                                                                                                                                                                                                   | same                        |
| Intend for use                                                                                       | Nitrile Powder Free Patient<br>Examination Gloves, Blue<br>Color is a disposable device<br>intended for medical<br>purposes that is worn on the<br>examiner's hand or finger to<br>prevent contamination<br>between patient and<br>examiner. | Nitrile Powder Free Patient<br>Examination Gloves, Blue Color<br>a disposable device intended fo<br>medical purposes that is worn on<br>the examiner's hand or finger t<br>prevent contamination between<br>patient and examiner. | Substantially<br>equivalent |
| Device<br>Description and<br>Specifications                                                          | Meets ASTM D 6319-10                                                                                                                                                                                                                         | Meets ASTM D 6319-10                                                                                                                                                                                                              | Substantially<br>equivalent |
| Dimensions<br>--Length                                                                               | Meets ASTM<br>D6319-10<br>$≥230$ mm min                                                                                                                                                                                                      | Meets<br>ASTM D6319-10<br>$≥230$ mm min for all sizes                                                                                                                                                                             | Substantially<br>equivalent |
| Dimensions<br>-- Width                                                                               | Small 70-90 mm<br>Medium 85-105mm<br>Large 100-120mm<br>Xlarge 110-130 mm<br>Meets ASTM D6319-10                                                                                                                                             | Small 70-90 mm<br>Medium 85-105 mm<br>Large 100-120 mm<br>X large 110-130 mm<br>Meets ASTM D6319-10                                                                                                                               | Substantially<br>equivalent |
| Dimensions<br>--Thickness                                                                            | Finger 0.05mm min.<br>Palm 0.05mm min.<br>Meets ASTM D6319-10                                                                                                                                                                                | Thickness (mm) min.<br>Finger 0.05 mm min.<br>Palm 0.05 mm min.<br>Meets ASTM D6319-10                                                                                                                                            | Substantially<br>equivalent |
| Physical<br>Properties                                                                               | Meets ASTM D6319-10<br>Before aging/after aging<br>$Elongation≥500%$<br>$Tensile Strength≥14$ MPa                                                                                                                                            | Meets ASTMD D6319-10<br>Before aging/after aging<br>$Elongation≥500%$<br>$Tensile Strength≥ 14$ MPa                                                                                                                               | Substantially<br>equivalent |
| Freedom<br>from<br>Pinholes                                                                          | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10<br>• ASTM D 5151-06<br>(Reapproved 2011)                                                                                                                                                         | Meets<br>• 21 CFR 800.20<br>• ASTM D6319-10<br>• ASTM D5151-06<br>(Reapproved 2011)                                                                                                                                               | Substantially<br>equivalent |
|                                                                                                      |                                                                                                                                                                                                                                              | Inspection Level I<br>AQL2.5                                                                                                                                                                                                      |                             |
| Residual<br>Powder                                                                                   | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>below 2mg of residual<br>powder                                                                                                                                                              | Meets ASTM<br>D 6124-06<br>(Reapproved 2011)<br>Results generated values below<br>2mg of residual powder                                                                                                                          | Substantially<br>equivalent |
| Materials used<br>to fabricate the<br>devices                                                        | Nitrile                                                                                                                                                                                                                                      | Nitrile                                                                                                                                                                                                                           | Substantially<br>equivalent |
| Dusting or<br>Donning<br>Powder: name                                                                | PU                                                                                                                                                                                                                                           | Polyurethane                                                                                                                                                                                                                      | Substantially<br>equivalent |
| Compare<br>performance<br>data supporting<br>substantial<br>equivalence                              | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reapproved 2011)                                                                                                                                           | Meets<br>ASTM D5151-06<br>(Reapproved 2011)<br>ASTM D6319-10<br>ASTM D6124-06<br>(Reapproved 2011)                                                                                                                                | Substantially<br>equivalent |
| Single Patient<br>Use                                                                                | Single Patient Use                                                                                                                                                                                                                           | Single Patient Use                                                                                                                                                                                                                | Substantially<br>equivalent |
| Biocompatibility                                                                                     | SKIN IRRITATION<br>DERMAL and<br>SENSITIZATION<br>STUDIES Meets ISO<br>10993-10:2002/Amd.1:2006                                                                                                                                              | SKIN IRRITATION DERMAL<br>and SENSITIZATION STUDIES<br>Meets ISO 10993-10: Third<br>Edition 2010-08-01.<br><br>Under the conditions of the study,<br>not an irritant and under<br>conditions of the study, not a<br>sensitizer.   | Substantially<br>equivalent |
| Labeling for the<br>legally marketed<br>device to which<br>substantial<br>equivalence is<br>claimed. | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-Blue color<br>Non sterile                                                                                                                 | -Powder Free<br>-Patient Examination Glove<br>-Single Use Only<br>- Manufactured For:<br>- Lot<br>-Blue color<br>Non sterile                                                                                                      | Substantially<br>equivalent |

{6}------------------------------------------------

#### 10.0 Substantial Equivalence Comparison:

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue Color meet the ASTM standard or equivalent standard and FDA requirements for waterleak test on pinhole AQL., meet labeling claims.

It can be concluded that the Nitrile Powder Free Patient Examination Gloves, Blue is as safe, as effective, and performs as well as the predicate device, Nitrile Color Powder Free Patient Examination Gloves, Blue Color, Tangshan Zhonghong Pulin Plastic Co., Ltd.. K120970.

---

**Source:** [https://fda-staging.innolitics.com/device/K152707](https://fda-staging.innolitics.com/device/K152707)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
