ENT BIOTECH SOLUTIONS TISSUE REMOVAL DEVICE (ELASSO)
K152702 · Ent Biotech Solutions · GEI · Feb 18, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152702
Device Name
ENT BIOTECH SOLUTIONS TISSUE REMOVAL DEVICE (ELASSO)
Applicant
Ent Biotech Solutions
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Elasso™ tissue removal device is intended to be used for the cutting and coagulation of soft tissue during otolaryngology (ENT) surgery including de-bulking of adenoid and tonsil tissue (Pharyngeal, Tubal, and Palatine).
Device Story
Elasso is a single-use, sterile surgical accessory for ENT procedures; used with external electrosurgical generator, monopolar cord, and patient return electrode. Device features forceps-like geometry with integrated loop electrode at distal tip; proximal end connects to generator via banana plug. Operated by surgeon via generator footswitch. Device delivers RF energy to soft tissue to perform cutting and coagulation. Forceps design facilitates tissue access and manipulation. Benefits include effective tissue de-bulking and coagulation during tonsillectomy and adenoidectomy. Used in clinical/surgical settings.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included mechanical, electrical safety/EMC, operational, in-vivo evaluation, ex-vivo thermal effects on tissue, thermal modeling, usability/access, biocompatibility, and sterility/shelf-life. Thermal spread of Elasso was compared to the predicate and found not to significantly exceed it.
Technological Characteristics
Electrosurgical instrument using RF energy. Materials: metal electrode, polymer insulator. Dimensions: 200.4mm length. Connectivity: wired to external generator. Sterilization: Gamma radiation (ISO 11137-1:2006, ISO 11137-2:2006). Electrical standards: IEC 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-2-2:2009. Biocompatibility: ISO 10993-1:2009.
Indications for Use
Indicated for patients requiring ENT surgery, specifically for cutting and coagulation of soft tissue including adenoid and tonsil tissue (Pharyngeal, Tubal, and Palatine).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
PEAK Plasmablade™ TNA Tonsil and Adenoid Tissue Dissection Device (K083415)
Reference Devices
St. Clair Forceps
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 18, 2016
Ent Biotech Solutions Laura Yamada Regulatory Affairs 75 Lewiston Road Gross Pointe Farms, Michigan 48236
Re: K152702
Trade/Device Name: ENT Biotech Solutions Tissue Removal Device (Elasso) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 15, 2016 Received: January 19, 2016
Dear Ms. Yamada:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K152702
Device Name
ENT Biotech Solutions Tissue Removal Device (Elasso™)
Indications for Use (Describe)
The ENT Biotech Solutions Tissue Removal Device (Elasso™) is intended to be used for the cutting and coagulation of soft tissue during otolaryngology (ENT) surgery including adenoid and tonsil tissue (Pharyngeal, Tubal, and Palatine).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Category | Comments |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared: | September 8, 2015 |
| Applicant: | ENT Biotech Solutions, INC<br>75 Lewiston Road<br>Grosse Pointe Farms, MI 48236<br>USA<br>Tel: 1.313.922.4615<br>Email: Adickson@entbiotechsolutions.com |
| Applicant's Contact Information: | Laura Yamada or Andrea Dickson<br>Email: Laura.Yamada@sbcglobal.net |
| Device Trade/ Proprietary Name: | ENT Biotech Solutions tissue removal device<br>(Elasso™) |
| Device Common Name: | Electrosurgical, Cutting & Coagulation Device<br>& Accessories |
| Device Product Code: | GEI |
| Device Classification / Name: | II / Electrosurgical, Cutting & Coagulation<br>Device & Accessories |
| CFR Citation: | 21 CFR 878.4400 |
#### Premarket Notification [510(k)] Summary
#### Substantial Equivalence Device Information
| Predicate Device(s): | PEAK Plasmablade™™ TNA Tonsil and Adenoid<br>Tissue Dissection Device |
|-----------------------------------------------------|---------------------------------------------------------------------------------|
| Predicate Device Manufacturers: | Medtronic (Formerly PEAK Surgical, Inc) |
| 510(k) Number: | K083415 |
| Predicate Device Common Name: | Electrosurgical, Cutting & Coagulation Device<br>& Accessories |
| Predicate Device Classification Name &<br>Citation: | Electrosurgical, Cutting & Coagulation Device &<br>Accessories, 21 CFR 878.4400 |
#### 510(k) Summary - Predicate Device Information
# Description of the Device
The ENT Biotech Solutions tissue removal device (also referred to as the "Elasso™") is a single-use, surgical accessory, used with a recommended electrosurgical generator (not supplied by ENT Biotech Solutions, INC) and associated equipment (monopolar cord and patient return electrode). The Elasso device configuration is similar to a standard forceps geometry (patterned after the commercially available reference device, St. Clair Forceps), but with an integrated loop electrode at the distal tip of one of the forceps. The proximal
{4}------------------------------------------------
end contains a single banana pluq. for connection to the generator. The hand piece is operated by the generator's footswitch controller.
The Elasso™ is a single-use, sterile packaged device.
# Indications for Use / Intended Use
The Elasso™ tissue removal device is intended to be used for the cutting and coagulation of soft tissue during otolaryngology (ENT) surgery including de-bulking of adenoid and tonsil tissue (Pharyngeal, Tubal, and Palatine).
### Technological Characteristics
The Elasso™ device is similar to the predicate device in that they are both electrosurgical instruments used to cut and coagulate soft tissue utilizing RF energy. The Elasso device differs from the PEAK device in the hand piece design. However, both devices incorporate partially exposed electrode elements which allow for application of RF energy, such that the principles of operation involved in cutting and coagulation of tissue are the same on either device with respect to energy density. Also, the results of a comparison study indicate that the thermal spread produced by the Elasso device across its specified use settings does not significantly exceed the thermal spread produced by the predicate device. In addition, the Elasso's forceplike design is patterned after the geometry of the commercially available reference device. St. Clair Forceps device which is used to remove adult and pediatric tissue and has been over many decades as an acceptable tool for access to the indicated tissues. The differences identified do not present any new concerns of safety and effectiveness.
| Description | Predicate | ENT Biotech Solutions |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General | | |
| Intended Use | Intended to be used for the<br>cutting and coagulation of soft<br>tissue during otolaryngology<br>(ENT) surgery including<br>adenoidectomy and<br>tonsillectomy (Pharyngeal,<br>Tubal, Palatine). | Intended to be used for the<br>cutting and coagulation of soft<br>tissue during otolaryngology<br>(ENT) surgery including de-<br>bulking of adenoid and tonsil<br>tissue (Pharyngeal, Tubal, and<br>Palatine). |
| Principles of operation | Electrosurgical cutting and<br>coagulation device intended to<br>remove tissue and coagulate by<br>use of radio frequency (RF)<br>energy. | Same |
| Anatomical Site | Soft Tissue in ENT surgeries | Same |
| Target Population | Patients requiring ENT surgeries | Same |
| Use | Single-Use | Same |
| Materials | Metal electrode element/polymer<br>insulator | Same |
| | | |
| Specifications | | |
| Energy Delivered | Radio frequency | Same |
#### Comparison Table Table 6-1:
{5}------------------------------------------------
| Description | Predicate | ENT Biotech Solutions |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Rated Voltage | 5,000V peak to peak | 5,000V (or 10,000V peak to peak) |
| Max Power Output | 50W | 50W |
| Power Modes | Monopolar | Same |
| Design | Light weight hand piece (wand-like) connected via cable to generator | Light weight hand piece (forceps-like similar to the commercially available St. Claire Forceps) connected via cable to generator |
| Hand Piece Tip | Tonsil tip: curved electrode element housed in a plastic tip<br>Adenoid tip: straight electrode element housed in a plastic tip | Curved electrode element housed in a plastic tip |
| Overall Length | 210mm (Tonsil Tip)/245mm (Adenoid Tip)* | 200.4mm |
| Compliance | | |
| Sterilization | Provided Sterile (Ethylene Oxide) | Provided Sterile to SAL of 10-6 (Gamma Radiation, per ISO 11137-1:2006 and ISO 11137-2:2006) |
| Electrical Standards | Unknown | IEC 60601-1:2012, IEC 60601-1-2:2007, IEC 60601-2-2:2009 |
| Biocompatibility | Unknown | ISO 10993-1:2009, 510(k) Memorandum - #G95-1: Use of International Standard ISO-10993 |
*191.5mm (St. Claire Forceps)
# Summary of Studies
Performance testing was conducted on the Elasso device to demonstrate compliance with the product requirements and to demonstrate safety and substantial equivalence to the predicate. The following product performance tests are included: mechanical, electrical safety/electromagnetic compatibility, operational, in-vivo evaluation, ex-vivo thermal effects on tissue comparisons, thermal modeling, usability/device access, biocompatibility, sterility and packaging / shelf life evaluation. Like the predicate device, no clinical testing was deemed necessary to support substantial equivalence. The data submitted support the substantial equivalence claim for the proposed indications for use; the Elasso is as safe and effective as the predicate device.
# Conclusion
ENT Biotech Solutions, INC considers the Elasso device to be substantially equivalent to the legally marketed predicate device with respect to the device function, intended use, patient population and anatomical site. Any differences in technological characteristics between the Elasso device and the predicate device do not raise any new issues of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.