The Powder Free Vinyl Patient Examination Gloves (Green) is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Disposable, non-sterile vinyl patient examination gloves; worn on examiner's hand or finger; acts as a flexible barrier film to prevent cross-contamination between patient and examiner; used in clinical settings; no electronic or software components.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including ASTM D5250-06(2011) for physical properties, ASTM D5151-06(2011) for pinholes, ASTM D6124-06(2011) for residual powder, and ISO 10993-10 for skin irritation and sensitization.
Technological Characteristics
Material: Polyvinyl Chloride (PVC). Form factor: Disposable, non-sterile, powder-free gloves. Sizes: S, M, L, XL. Standards: ASTM D5250-06(2011), ASTM D5151-06(2011), ASTM D6124-06(2011). Biocompatibility: ISO 10993-10. No electronic, software, or energy-based components.
Indications for Use
Indicated for use as a disposable, non-sterile medical glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 21, 2015
Supermax Plastic Products Co., Ltd % Ms. Kathy Liu Project Manager Hongray USA Medical Products Inc. 3973 Schaefer Ave. Chino, California 91710
Re: K152692
Trade/Device Name: Powder Free Vinyl Patient Examination Gloves (Green) Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: November 24, 2015 Received: November 27, 2015
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152692
Device Name
Powder Free Vinyl Patient Examination Gloves (Green)
Indications for Use (Describe)
The Powder Free Vinyl Patient Examination Gloves (Green) is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Supermax Plastic Products Co., Ltd.
314 Dragon River East Road, Luquan, Hebei Province, CHINA 050000
## 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR \$807.92.
The assigned 510(K) number is: K152692
#### 1. Owner's Identification :
Mr. Wu Zhigang Supermax Plastic Products Co., Ltd. 314 Dragon River East Road, Luquan, Hebei Province, CHINA 050000
Tel: 86-311-83601854 Fax: 86-311-83616934
Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611 Fax: 909-590-1511 Date Summary Prepared: December 15, 2015
## 2. Name of the Device:
Trade Name: Powder Free Vinyl Patient Examination Gloves (Green) Common Name: Patient Examination Gloves Classification Name: Patient Examination Glove Classification Regulation: 21 CFR 880.6250 Classification Panel: 80 General Hospitals Product Code: LYZ Device Class: Class I
## 3. Predicate Device Information:
Tangshan Hongyun Plastic Products Company Limited Powder-Free Yellow Vinyl Patient Examination Gloves (K141878)
#### 4. Device Description:
Powder Free Vinyl Patient Examination Gloves (Green) are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of vinyl materials and are powder free. The physical and performance characteristics of the devices meet all requirements of ASTM D5250-06 (Reapproved 2011)--Standard Specification For Poly (Vinyl Chloride) Gloves For Medical Application.
#### 5. Intended Use of the Device:
{4}------------------------------------------------
# Supermax Plastic Products Co., Ltd.
314 Dragon River East Road, Luquan, Hebei Province, CHINA 050000
The Powder Free Vinyl Patient Examination Gloves (Green) is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
#### 6. Technological Characteristics and Substantial Equivalence:
Supermax Plastic Products Co., Ltd.'s Powder Free Vinyl Patient Examination Gloves (Green) is substantially equivalent in safety and effectiveness to the Tangshan Hongyun Plastic Products Company Limited's Powder-Free Yellow Vinyl Patient Examination Gloves (K141878). The subject device and predicate device use a similar plastic flexible barrier film to achieve a device for the intended use.
And the properties between the subject device and the predicate device are compared in the following table:
| Characteristics | Standard | Device Performance | | Result of |
|-------------------------------------------------------------------------|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | | Predicate device | Subject Device | comparison |
| 510K number | | K141878 | K152692 | -- |
| Product name | 1 | Powder-Free Yellow Vinyl<br>Patient Examination<br>Gloves | Powder Free Vinyl<br>Patient Examination<br>Gloves (Green) | Similar |
| Product Code | | LYZ | LYZ | Same |
| Intended Use | | Powder-Free Yellow Vinyl<br>Patient Examination<br>Gloves is a non sterile<br>disposable device intended<br>for medical purposes that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner. | The Powder Free Vinyl<br>Patient Examination<br>Gloves (Green) is<br>disposable non-sterile<br>device intended for<br>medical purpose that is<br>worn on the examiner's<br>hand or finger to prevent<br>contamination between<br>patient and examiner. | Same |
| Size | | S/M/L/XL | S/M/L/XL | Same |
| Compare all<br>materials used<br>to fabricate the<br>devices | | PVC | PVC | Same |
| Dusting or<br>Donning<br>powder | | PU | PU | Same |
| Compare<br>performance<br>data supporting<br>substantial<br>equivalence | | Meets<br>ASTM D5151-06(2011)<br>ASTM D 5250-06 (2011)<br>ASTM D6124-06(2011) | Meets<br>ASTM D5151-06(2011)<br>ASTM D 5250-06 (2011)<br>ASTM D6124-06(2011) | Same |
| Color | | Yellow | Green | Similar |
| | Standard | Device Performance | | Result of |
| Characteristics | | Predicate device | Subject Device | comparison |
| Single patient<br>use | / | Single patient use | Single patient use | Similar |
| Dimensions:<br>Length | ASTM D 5250-<br>06 (2011) | Meets<br>230 mm min for all sizes | Meets<br>230 mm min for all sizes | Similar |
| Dimensions:<br>Width | ASTM D 5250-<br>06 (2011) | Meets<br>S: 83-86 mm<br>M: 94-97 mm<br>L: 104-107 mm<br>XL: 113-115 mm | Meets<br>S: 85±5 mm<br>M: 95±5 mm<br>L: 105±5 mm<br>XL: 115±5 mm | Similar |
| Dimensions:<br>Thickness | ASTM D 5250-<br>06 (2011) | Meets<br>Palm: 0.08 mm min<br>Finger: 0.05 mm min | Meets<br>Palm: 0.08 mm min<br>Finger: 0.05 mm min | Similar |
| Tensile<br>strength:<br>before<br>and<br>after aging | ASTM D 5250-<br>06 (2011) | Meets<br>Tensile Strength≥11MPa | Meets<br>Tensile Strength≥11MPa | Similar |
| Elongation<br>before and after<br>aging | ASTM D 5250-<br>06 (2011) | Meets<br>Elongation≥300% | Meets<br>Elongation≥300% | Similar |
| Freedom from<br>pinholes | ASTM D 5250-<br>06 (2011)<br>ASTM D5151-<br>06<br>(Reapproved<br>2011) | Meets<br>G-I, AQL2.5 | Meets<br>G-I, AQL2.5 | Similar |
| Residual<br>powder | ASTM D 5250-<br>06 (2011)<br>ASTM D6124-<br>06(Reaffirmati<br>on 2011) | Meets<br>Results below 2mg of<br>residual powder | Meets<br>Not more than 2mg per<br>glove | Similar |
| Biocompatibility | Skin irritation<br>ISO 10993-10 | Under conditions of<br>the<br>study, not an irritant | Under conditions of the<br>study, not an irritant | Same |
| | Sensitization<br>ISO 10993-10 | Under the conditions of the<br>study, not a sensitizer | Under the conditions of<br>the study, not a sensitizer | Same |
| Labeling | / | Powder Free<br>Devices color: Yellow<br>Patient examination Glove<br>Non Sterile<br>Single use only<br>Manufactured for:<br>Lot | Powder Free<br>Devices color: Green<br>Patient examination<br>Glove<br>Non Sterile Single<br>use only<br>Manufactured for:<br>Lot | Similar |
{5}------------------------------------------------
Supermax Plastic Products Co., Ltd.
314 Dragon River East Road, Luquan, Hebei Province, CHINA 050000
{6}------------------------------------------------
# Supermax Plastic Products Co., Ltd.
314 Dragon River East Road, Luquan, Hebei Province, CHINA 050000
Powder Free Vinyl Patient Examination Gloves (Green) shares the same or comparable technology characteristics compared to the predicate device. The subject device performs according to the glove performance standards ASTM D5250-06(2011), biocompatibility requirement and FDA requirements and the labeling claims for the product. It performs as well as the legally marketed predicate device.
#### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as Follows:
Powder Free Vinyl Patient Examination Gloves (Green) meet requirements per ASTM D 5250-06(2011), per ASTM D6124-06 (2011), ASTM D5151-06(2011), 21CFR 800.20 and ISO 10993-10 third edition 2010-08-01
#### 8. Discussion of Clinical Tests Performed:
Clinical data was not included in the submission. Substantial equivalence of the subject device was supported by non-clinical testing.
## 9. Conclusions:
Powder Free Vinyl Patient Examination Gloves (Green) conform fully to ASTM D 5250-06 (2011) standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data discussed above.
Drawn from the complete list of non-clinical tests, the subject device is as safe, as effective, and performs as well as the legally marketed predicate device. The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.