K152691 · Essential Dental Systems, Inc. · EMA · Feb 25, 2016 · Dental
Device Facts
Record ID
K152691
Device Name
EDS Universal Cement
Applicant
Essential Dental Systems, Inc.
Product Code
EMA · Dental
Decision Date
Feb 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
Final cementation of ceramic, composite, or metal inlays, onlays, crowns, bridges, posts, and screws. Final cementation of ceramic, composite, or metal restorations on implant abutments. Permament cementation of ceramic or composite veneers.
Device Story
EDS Universal Cement is an automix, dual-curing, self-adhesive resin cement used by dentists to affix ceramic, composite, or metal restorations to tooth surfaces (dentin/enamel) or implant abutments. The device eliminates the need for separate etchants or bonding agents. It consists of two pastes (base and catalyst) containing dimethacrylate resins, inorganic fillers (60% wt), and a photoinitiator system, packaged in a 5 mL dual-barreled syringe. The clinician dispenses the material via the automix system and applies it to the restoration; the cement cures via chemical reaction and/or light activation. The resulting bond provides radiopacity and resistance to mechanical stress and water damage. The device benefits patients by providing a stable, durable fixation for dental prosthetics.
Clinical Evidence
No clinical data. Bench testing only. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization), ISO 10993-6 (implantation), and ISO 10993-3 (genotoxicity). Performance testing included comparative shear bond strength to dentin, enamel, zirconia, and lithium disilicate surfaces. Fluoride release testing showed no fluoride release over 7 days.
Technological Characteristics
Automix, dual-curing, self-adhesive resin cement. Composition: dimethacrylate resins, 60% wt inorganic filler, photoinitiator. Form factor: 5 mL dual-barreled syringe. Non-sterile. Shelf-life: 2 years. Radiopaque.
Indications for Use
Indicated for final cementation of ceramic, composite, or metal inlays, onlays, crowns, bridges, posts, screws, and restorations on implant abutments, as well as permanent cementation of ceramic or composite veneers.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The graphic is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all capital letters.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 25, 2016
Essential Dental Systems, Inc. Mr. Jeffrey Wan Research and Development Manager 89 Leuning Street, Suite 8 South Hackensack, New Jersey 07606
Re: K152691
Trade/Device Name: EDS Universal Cement Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: January 25, 2016 Received: January 27, 2016
Dear Mr. Wan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 - Indications for Use
510(k) Number (if known): _ K152691
## Device Name: EDS Universal Cement
### Indications for Use:
- Final cementation of ceramic, composite, or metal inlays, onlays, crowns, bridges, posts, and screws.
- Final cementation of ceramic, composite, or metal restorations on implant abutments.
- Permament cementation of ceramic or composite veneers.
Prescription Use X (Per 21 CFR 801 Subpart D)
OR
Over the Counter Use (Per 21 CFR 807 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# Section 5 - 510(k) Summary
#### 1. Submitter's Identification:
Essential Dental Systems 89 Leuning Street South Hackensack, NJ 07606
Date Summary Prepared: February 18, 2016
Contact: Mr. Jeffrey Wan Contact Email: jwan@edsdental.com Contact Phone #: 201-487-9090 ext. 118 Contact Fax #: 201-487-5120
#### 2. Name of the Device:
Trade name: EDS Universal Cement Common name: Dental cement other than zinc oxide-eugenol Classification name: Dental Cement CFR Number: 872.3275 Device class: Class II Product Code: EMA
#### 3. Predicate Device Information:
Primary predicate: Unicem HM, 3M ESPE AG, K100908 Reference predicate: CALIBRATM CEMENT, Dentsply International, K040906
#### 4. Device Description:
EDS Universal Cement is an automix, dual-curing, self-adhesive resin cement. This multi-purpose device can be used to affix a variety of ceramic, composite, or metal restorations without the aid of an etchant or bonding agent. The device is comprised of two pastes (base and catalyst) that contain dimethacrylate resins, inorqanic filler particles (60% by wt), and a photoinitiator system. The mixing ratio is roughly 1 part base to 1 part catalyst. The device is packaged in a 5 mL dualbarreled syringe and comes in three shades: translucent (white), opaque (white), and A2 (tooth colored).
{4}------------------------------------------------
EDS Universal Cement provides a bond between tooth surfaces (dentin and enamel) and ceramic, composite, and metal restorations. The device provides radiopacity and is resistant to mechanical stress and water damage.
EDS Universal Cement is provided non-sterile and has a shelf-life of two years.
#### 5. Indications for Use:
- Final cementation of ceramic, composite, or metal inlays, onlays, crowns, bridges, posts, and screws.
- Final cementation of ceramic, composite, or metal restorations on implant abutments.
- Permament cementation of ceramic or composite veneers.
#### 6. Comparison to Predicate Devices:
A comparison of EDS Universal Cement and the predicate devices indicates the following similarities and differences to the devices which received 510(k) clearance:
| | Proposed Device | Primary Predicate | Reference Predicate |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) | To be assigned | K100908 | K040906 |
| Device Name | EDS Universal Cement | Unicem HM | Calibra™ Cement |
| Manufacturer | Essential Dental<br>Systems | 3M ESPE | Dentsply International |
| Indications<br>for Use | - Final cementation of<br>ceramic, composite, or<br>metal inlays, onlays,<br>crowns, bridges, posts,<br>and screws.<br>- Final cementation of<br>ceramic, composite, or<br>metal restorations on<br>implant abutments.<br>- Permanent<br>cementation of ceramic<br>or composite veneers. | - Final cementation of<br>ceramic, composite or<br>metal inlays, onlays,<br>crowns, bridges, 2-3-unit<br>Maryland bridges and 3-unit<br>inlay/onlay bridges<br>(excluded for patients with<br>bruxism<br>or periodontitis).<br>- Final cementation of post<br>and screws.<br>- Final cementation of<br>ceramic, composite or<br>metal restorations on<br>implant abutments.<br>- Cementation of abutments<br>made of Lava™ zirconium<br>oxide ceramic. | - Adhesive cementation of<br>ceramic, porcelain, or<br>composite inlays/onlays,<br>veneers, or crowns.<br>- Adhesive cementation of all<br>metal crowns, bridges,<br>inlays/onlays including<br>precious, semi-precious, and<br>non-precious metals.<br>- Adhesive cementation of<br>PFM (porcelain fused to<br>metal) crowns and bridges.<br>- Adhesive cementation of<br>prefabricated and cast posts.<br>- Adhesive cementation of<br>resin-bonded retainer<br>bridges (Maryland bridges) |
| Composition | Methacrylate resins and<br>inorganic filler | Methacrylate resins and<br>inorganic filler | Methacrylate resins and<br>inorganic filler |
| Shades | Translucent, opaque<br>(white), A2 | Translucent, A2, A3 opaque | Translucent, light, medium,<br>opaque, bleach |
{5}------------------------------------------------
EDS Universal Cement is similar to the predicate devices Unicem HM and Calibra™ in that they are all automix, self-adhesive, dual-curing resin cements intended for ceramic, composite, or metal restorations. All devices are paste/paste systems with similar filler ratios. EDS Universal Cement is also similar to Calibra™ in that they are intended for the cementation of veneers.
EDS Universal Cement is different from the predicate devices in the number and type of shades available. EDS Universal Cement is also different from Unicem HM in that Unicem HM does not have an intended use for the cementation of veneers. The subject device and primary predicate have slightly different Indications for Use language. However, the difference in language does not change the intended use or substantial equivalence.
All of the components found in the predicate devices have been used in legally marketed devices. We believe that prior use of components in legally marketed devices, the performance and biocompatibility data provided support the substantial equivalence of EDS Universal Cement for the indicated uses.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
Testing of biocompatibility, physical properties, and adhesion to dentin and ceramic surfaces were conducted to determine equivalence of EDS Universal Cement to the predicate device Unicem HM.
Equivalence in safety to the predicate device was demonstrated by conducting four biocompatibility tests:
- Cytotoxicity testing according to ISO 10993-5
- Sensitization testing according to ISO 10993-10
- Implantation testing according to ISO 10993-6
- Genotoxicity testing according to ISO 10993-3
Equivalence in performance to the predicate device was demonstrated by conducting comparative shear bond strength testing to natural tooth structure and various crown materials.
Samples of the subject and predicate devices were cured onto dentin, enamel, zirconia, and lithium disilicate surfaces. Shear loading was applied at the cement-surface interface until debonding occurred.
Fluoride release testing was conducted on the subject device to determine its fluoride release profile.
{6}------------------------------------------------
Samples of the subject device were immersed in distilled water. The fluoride concentration was measured at 1, 3, and 7 days. No fluoride was released over a 7 day period.
#### 8. Discussion of Clinical Tests Performed:
Not Applicable
#### 9. Conclusions:
EDS Universal Cement is substantially equivalent to the cleared and marketed predicate device Unicem HM.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.