Sydney IVF Cryopreservation Kit, Sydney IVF Thawing Kit
K152682 · Willian A. Cook Australia Pty, Ltd. · MQL · May 5, 2016 · Obstetrics/Gynecology
Device Facts
Record ID
K152682
Device Name
Sydney IVF Cryopreservation Kit, Sydney IVF Thawing Kit
Applicant
Willian A. Cook Australia Pty, Ltd.
Product Code
MQL · Obstetrics/Gynecology
Decision Date
May 5, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.6180
Device Class
Class 2
Indications for Use
Sydney IVF Cryopreservation Kit is intended for use during in vitro fertilization procedures for cryopreservation of 1-cell to 8-cell embryos. Sydney IVF Thawing Kit is intended for use during in vitro fertilization procedures for thawing of 1-cell to 8-cell embryos.
Device Story
Sydney IVF Cryopreservation and Thawing Kits are chemical solutions used in IVF laboratories to preserve supernumerary embryos. The Cryopreservation Kit uses sequential solutions containing propanediol and sucrose to dehydrate embryos, preventing intracellular ice crystal formation. The Thawing Kit uses sequential solutions with decreasing cryoprotectant concentrations to rehydrate embryos. Kits are used by embryologists in clinical IVF settings. Output is a cryopreserved or thawed embryo ready for future clinical use. Benefits include embryo viability maintenance for future implantation.
Clinical Evidence
Bench testing only. Performance validated via Mouse Embryo Assay (MEA) using 2-cell embryos (72hrs) with ≥80% blastocyst development rate compared to controls. Stability studies validated a 20-week shelf-life at 2-8°C, assessing MEA, endotoxin, osmolality, pH, sterility, proline concentration, and ammonia levels.
Technological Characteristics
Solutions contain propanediol, sucrose, 12 mg/mL Human Serum Albumin (HSA), and 0.01 mg/mL Gentamicin. Packaged in borosilicate type 1 glass vials with Fluorotec coated rubber stoppers and tamper-evident seals. Manufacturing via aseptic filtration. Specifications: pH 7.3-7.5, osmolality 285-295 mOsm/kg, endotoxin < 0.40 EU/mL.
Indications for Use
Indicated for cryopreservation and thawing of 1-cell to 8-cell human embryos during in vitro fertilization procedures.
Regulatory Classification
Identification
Reproductive media and supplement are products that are used for assisted reproduction procedures. Media include liquid and powder versions of various substances that come in direct physical contact with human gametes or embryos (including water, acid solutions used to treat gametes or embryos, rinsing solutions, sperm separation media, supplements, or oil used to cover the media) for the purposes of preparation, maintenance, transfer or storage. Supplements are specific reagents added to media to enhance specific properties of the media (e.g., proteins, sera, antibiotics, etc.).
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing). The device, when it is phosphate-buffered saline used for washing, and short-term handling and manipulation of gametes and embryos; culture oil used as an overlay for culture media containing gametes and embryos; and water for assisted reproduction applications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2016
Willian A. Cook Australia Pty Ltd Gordana Pozvek Senior Regulatory Affairs Specialist 95 Brandl Street Brisbane Technology Park, Eight Mile Plains Brisbane, QLD 4113, Australia
Re: K152682
> Trade/Device Name: Sydney IVF Cryopreservation Kit, Sydney IVF Thawing Kit Regulation Number: 21 CFR 884.6180 Regulation Name: Reproductive Media and Supplements Regulatory Class: Class II Product Code: MOL Dated: April 5, 2016 Received: April 7, 2016
Dear Gordana Pozvek,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a large, sans-serif font. The text is black and appears to be the main focus of the image. There is a faint, watermark-like design in the background, but it is not easily discernible.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K152682
Device Name Sydney IVF Cryopreservation Kit
Indications for Use (Describe)
Sydney IVF Cryopreservation Kit is intended for use during in vitro fertilization procedures for cryopreservation of 1-cell to 8-cell embryos.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K152682
Device Name Sydney IVF Thawing Kit
Indications for Use (Describe)
Sydney IVF Thawing Kit is intended for use during in vitro fertilization procedures for thawing of 1-cell to 8-cell embryos.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, bold, white letters against a red background. To the right of the word "COOK" is a small, superscripted "R" inside a circle, indicating a registered trademark. Below the word "COOK" is the word "MEDICAL" in smaller, white letters, positioned on a red angled shape.
WILLIAM A. COOK AUSTRALIA PTY. LTD. 95 BRANDL STREET BRISBANE TECHNOLOGY PARK, EIGHT MILE PL BRISBANE, QLD 4113, AUS HONE: 1800.777.222 FAX: +61.7.38 WWW.COOKMEDICAL
# Special 510(k) Summary - K152682
## SUBMITTED BY:
William A. Cook Australia Pty Ltd 95 Brandl Street Eight Mile Plains QLD 4113 Australia
| Contact Person: | Gordana Pozvek Ph.D. |
|-----------------|--------------------------------|
| Tel: | +61 (7) 3841 1188 |
| Fax: | +61 (7) 3841 3905 |
| E-mail: | Gordana.Pozvek@CookMedical.com |
Date Prepared: May 3, 2016
# DEVICE IDENTIFICATION:
| Trade Name: | Sydney IVF Cryopreservation Kit (K-SICS-5000) & |
|-------------------|---------------------------------------------------|
| | Sydney IVF Thawing Kit (K-SITS-5000) |
| Common Name: | Cryopreservation & Thawing Kits |
| Regulation No: | 21 CFR 884.6180, Reproductive Media & Supplements |
| Regulatory Class: | II |
| Product Code: | MQL - Media, Reproductive |
# PREDICATE DEVICE:
Cook IVF Cryopreservation Kit & Cook IVF Thaw Kit (K011157), cleared May 3, 2001.
## DEVICE DESCRIPTION:
The Sydney IVF Cryopreservation and Thawing Kits are intended for cryopreservation and thawing of 1-cell to 8-cell human embryos. The Sydney IVF Cryopreservation and Thawing Kits provide users with the ability to cryopreserve supernumerary embryos created during the in vitro fertilization procedure and then to thaw them for use at a future point in time.
The Sydney IVF Cryopreservation Kit consists of three solutions containing increasing concentrations of cryoprotectant (both propanediol and sucrose are used). These buffers were designed to be used sequentially in order to remove water from embryos prior to cryopreservation. The removal of water prevents ice crystal formation inside the embryo thereby limiting damage and improving viability. It contains 12 mg/mL Human Serum
AORTIC CRITICAL WOMEN'S INTERVENTIONAL LEAD SURGERY UROLOGY ENDOSCOPY INTERVENTION MANAGEMENT INTERVENTION HEALTH CARE RADIOLOGY William A. Cook Australia Pty Ltd K152682 Page 1 of 3
{5}------------------------------------------------
Albumin (HSA) and 0.01mg/mL Gentamicin. Sydney IVF Cryopreservation Kit is designed for use with Sydney IVF Thawing Kit.
The Sydney IVF Thawing Kit consists of four solutions with decreasing concentrations of cryoprotectants (propanediol and sucrose) which are used sequentially throughout the thawing process. It contains 12 mg/mL Human Serum Albumin (HSA) and 0.01mg/mL Gentamicin. It is designed for use with Sydney IVF Cryopreservation Kit.
Sydney IVF Cryopreservation and Thawing Kits are provided in glass vials with Fluorotec coated rubber stoppers held in place with a tamper evident seal. Sydney IVF Cryopreservation Kit is packaged in a carton box containing 2 x 10mL and 1 x 20mL solutions per kit. The Sydney IVF Thawing Kit is packaged in a carton box containing 4 x 10mL solutions per kit.
# INDICATIONS FOR USE:
Sydney IVF Cryopreservation Kit is intended for use during in vitro fertilization procedures for cryopreservation of 1-cell to 8-cell embryos.
Sydney IVF Thawing Kit is intended for use during in vitro fertilization procedures for thawing of 1-cell to 8-cell embryos.
The only differences in the indications for use listed above and that of the predicate device are:
- An update to clarify the intended use of the device as the original statements refers to "zygotes or embryos", and this has been updated to "1-cell to 8-cell embryos". The update serves to describe the intended use more accurately than the previous statement. There is no change to the clinical use of the device.
This clarification does not represent a new intended use nor does it pose any safety or effectiveness issues.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE:
The Sydney IVF Cryopreservation Kit & Sydney IVF Thawing Kit and the predicate device (K011157) have the same fundamental technology and similar technological characteristics including the following:
{6}------------------------------------------------
- Similar chemical formulation
- Similar performance specifications: ●
- pH 7.3 7.5 ー
- Osmolality 285 - 295 mOsm/kg
- ー Endotoxin < 0.40 EU/mL
- A Mouse Embryo Assay (MEA) is used to screen the product for embryo । toxicity.
- Same method of manufacturing process aseptic filtration ●
- . Same packaging - borosilicate type 1 vials with FluroTec coated stopper and tamper evident seals.
The modification that was made to the predicate device was a change in shelf-life from 6 weeks at 2-8°C (for predicate device) to 20 weeks at 2-8°C. In addition, minor changes were made to the formulation and specifications.
The technological characteristics of Sydney IVF Cryopreservation Kit & Sydney IVF Thawing Kit are comparable to the predicate device.
# PERFORMANCE DATA:
The product specifications for the Sydney IVF Cryopreservation Kit & Sydney IVF Thawing Kit and the predicate device are the same regarding sterility, pH, osmolality and endotoxin.
The MEA is used for both the predicate and the proposed device, but the test assay and specification has changed from 1-cell MEA (96hrs) with ≥75% of control that develop to blastocyst (predicate) to 2-cell MEA (72hrs) with ≥80% of control that develop to blastocyst.
## Stability & Shelf Life
The shelf-life of Sydney IVF Cryopreservation Kit & Sydney IVF Thawing Kit has been validated in stability studies to 20 weeks at 2 - 8°C. Stability tests included MEA, endotoxin, osmolality, pH, sterility and the concentrations of the amino acid proline and the HSA degradation by-product ammonia.
## CONCLUSION:
The results of the testing provide reasonable assurance that the Sydney IVF Cryopreservation Kit & Sydney IVF Thawing Kit is as safe and effective as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.