As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain. As a EMS device, FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for: - Prevention or retardation of muscle disuse atrophy. - Relaxation of muscle spasm. - Muscle re-education. - Maintaining or increasing range of motion. - Increasing local blood circulation. - Immediate post-surgical of calf muscle to prevent venous thrombosis. FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
Device Story
FD TENS 2090 and 2095 are transcutaneous electrical nerve stimulators (TENS) and powered muscle stimulators (EMS). Devices generate electrical pulses transmitted via skin-attached electrodes. TENS mode blocks pain signals to brain; EMS mode induces muscle contraction for therapeutic purposes (atrophy prevention, spasm relaxation, circulation, etc.). Devices are intended for prescription use. Healthcare providers or patients apply electrodes to affected body areas. Output parameters (pulse width, frequency, intensity) are adjustable or preset. Clinical benefit includes pain management and physical therapy support. No clinical data provided; substantial equivalence based on bench testing and comparison to predicate devices.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1) and electromagnetic compatibility (IEC 60601-1-2) testing.
Technological Characteristics
Symmetrical bi-phasic rectangular waveform; TENS/EMS stimulation; adjustable pulse width (50-250µs) and frequency (1-150Hz); powered by electrical current; electrodes for skin contact; safety compliance with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for patients requiring symptomatic relief/management of chronic intractable pain or adjunctive treatment for post-surgical/post-traumatic acute pain. EMS function indicated for muscle disuse atrophy prevention/retardation, muscle spasm relaxation, muscle re-education, range of motion maintenance/increase, local blood circulation increase, and post-surgical calf muscle stimulation to prevent venous thrombosis.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to create a sense of depth and connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 22, 2016
Fuji Dynamics Ltd. Man Man Chung Product Development Manager 1-3. 23/F.. Laws Commerical Plaza 788 Cheung Sha Wan Road Hong Kong, China
Re: K152676
Trade/Device Name: FD TENS 2090 and FD TENS 2095 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: GZJ, IPF Dated: December 16, 2015 Received: December 21, 2015
Dear Man Man Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152676
Device Name FD TENS 2090 and FD TENS 2095
#### Indications for Use (Describe)
As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
As a EMS device. FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for:
- Prevention or retardation of muscle disuse atrophy.
- · Relaxation of muscle spasm.
- · Muscle re-education.
- · Maintaining or increasing range of motion.
- · Increasing local blood circulation.
- · Immediate post-surgical of calf muscle to prevent venous thrombosis.
FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|--|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary [As required by 21 CFR 807.92]
#### Device 510(k) number: K152676
#### 1. Applicant Information
| Date prepared: | Aug 3, 2015 | |
|-----------------|----------------------------------------|--|
| Submitter: | Fuji Dynamics Ltd. | |
| | Unit 1-3, 23/F., Laws Commercial Plaza | |
| | 788 Cheung Sha Wan Road, Kowloon | |
| | Hong Kong | |
| Contact Person: | Man Man Chung | |
| Tel: | (852) 2786 4218 | |
| Fax: | (852) 2744 6775 | |
#### 2. General Device Information/Trade name
| Model No: | FD TENS 2090<br>FD TENS 2095 |
|--------------|------------------------------|
| Common Name: | Transcutaneous Elec |
trical Nerve Stimulator and Powered Muscle Stimulator GZJ, IPF Product code: Classification: Class II
#### 3. Predicate Device Information
| 510k number | Device Name | Manufacturer | Date cleared |
|-------------|--------------|---------------|--------------|
| K052813 | FD TENS 2030 | Fuji Dynamics | 2006 |
| K063642 | FD EMS | Fuji Dynamics | 2007 |
#### 4. Device Description
#### FD TENS 2090
FD TENS 2090 can provide both TENS treatments and EMS treatments. During TENS treatments, the FD TENS 2090 generates electrical pulses and transmit it to the electrodes which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain.
During EMS treatments, FD TENS 2090 generates electrical pulses and transmit it to the electrodes attached to the patient skin, causing the muscle to expand and contract. It is used to relax muscle spasms. prevent or retard atrophy, maintain or increase range of motion, increase local blood circulation, reeducate muscle and provide immediate post-surgical stimulation of calf muscle to prevent venous thrombosis.
#### FD TENS 2095
FD TENS 2095 can provide TENS treatment, During treatment, the FD TENS 2095 generates electrical pulses and transmit it to the electrodes, which are attached to the patient's skin. Consequently, the electrical pulses would then pass through the skin to the underlying peripheral nerves to aid in the blocking of pain signals traveling to the brain.
{4}------------------------------------------------
## 5. Intended Use:
#### FD TENS 2090
As a TENS device, FD TENS 2090 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
As a EMS device, FD TENS 2090 is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes on the affected body area. It is intended for:
- · Prevention or retardation of muscle disuse atrophy.
- Relaxation of muscle spasm.
- · Muscle re-education.
- · Maintaining or increasing range of motion.
- · Increasing local blood circulation.
- · Immediate post-surgical of calf muscle to prevent venous thrombosis.
### FD TENS 2095
FD TENS 2095 is used for the symptomatic relief and management of chronic intractable pain and/or as an adjunctive treatment in the management of post-surgical and post-traumatic acute pain.
#### 6. Comparison to Predicate Device:
TENS Program comparison between FD TENS 2090, FD TENS 2095 and FD TENS 2030
| Characteristic | Subject device K152676 | Predicate device K052813 |
|---------------------------------------------------------|-------------------------------|---------------------------|
| Device | FD TENS 2090 and FD TENS 2095 | FD TENS 2030 |
| Waveform | Symmetrical Bi-Phasic | Symmetrical Bi-Phasic |
| Shape | Rectangular | Rectangular |
| Maximum Voltage<br>Load: 0.50 kohm | (0-peak voltage)<br>48.8V | (0-peak voltage)<br>31.3V |
| 02 kohm | 70.4V | 50.7V |
| 10 kohm | 88.8V | 85.0V |
| Maximum Current<br>Load: 0.50 kohm | 97.6 mA | 62.6 mA |
| 02 kohm | 35.2 mA | 25.35 mA |
| 10 kohm | 8.88 mA | 8.5 mA |
| Maximum pulse width | 250μs | 250μs |
| Maximum frequency | 150 Hz | 200 Hz |
| Maximum Output<br>Net charge per phase<br>Load: 500 ohm | 10.41 μC | 5.83μC |
| Maximum output<br>Charge per phase<br>Load: 500 ohm | 22.13μC | 15.77μC |
| 1kohm | 16.33μC | 13.60μC |
| Maximum output<br>RMS Current<br>Load: 500 ohm | 11.55 mA rms | 10.99 mA rms |
| 1kohm | 8.49 mA rms | 8.95 mA rms |
| Max Current Density | 0.15 | 0.1 |
{5}------------------------------------------------
| (mA/cm²) | Max Power Density (W/cm²) | |
|------------------------|----------------------------------|-----------------------------|
| | .00525 | .00182 |
| Burst Mode | Burst 1 | Burst II |
| a. Pulses per burst | 7 | 7 |
| b. Burst per second | 2 bursts per second | 2 |
| c. Burst duration | 250 ms | 250ms |
| d. Duty cycle | 50% | 50% |
| | Burst 2 | Burst I |
| a. Pulses per burst | 80 | 80 |
| b. Burst per second | 1 burst per 2 seconds | 1 burst per 2 seconds |
| c. Burst duration | 1 second | 1 second |
| d. Duty cycle | 50% | 50% |
| Continuous Mode | P1 CONST | CONSTANT |
| a. Pulse width | 59µs default | 50µs default |
| | Adjustable from 50µs to<br>250µs | Adjustable from 25 to 250µs |
| b. Pulse frequency | 120 Hz default | 120 Hz default |
| | Adjustable from 1 to 150 Hz | Adjustable from 1 to 200 Hz |
| Pulse width modulation | P4 MODUL 1 | W. MODUL |
| a. Pulse width | 50µs to 250µs in 5s | 25µs to 250µs in 8s |
| b. Pulse frequency | 120Hz default | 120 Hz default |
| | Adjustable from 1 to 150 Hz | Adjustable from 1 to 200 Hz |
| Frequency modulation | P5 MODUL 2 | F. MODUL |
| a. Pulse frequency | 20 to 100 Hz in 5s | 20 to 100Hz in 8s |
| b. Pulse width | 150µs default | 50µs default |
| | Adjustable from 50 to 250µs | Adjustable from 25 to 250µs |
## EMS Program comparison between FD TENS 2090 and FD EMS
| Characteristic | Subject device K152676 | Predicate device K063642 |
|--------------------|------------------------|--------------------------|
| Trade name | FD TENS 2090 | FD EMS |
| Waveform | Symmetrical Bi-Phasic | Symmetrical Bi-Phasic |
| Shape | Rectangular | Rectangular |
| Maximum Voltage | (0 to peak voltage) | (0 to peak voltage) |
| Load: 500 ohm | 48.4 V | 48.7 V |
| 2 kohm | 67.2 V | 60.0 V |
| 10 kohm | 87.2 V | 64.7 V |
| Max Output current | | |
| Load: 500 ohm | 96.8 mA | 97.40 mA |
{6}------------------------------------------------
| 2 kohm | 33.6 mA | 30.0 mA |
|---------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 10 kohm | 8.72 mA | 6.47 mA |
| Max pulse width | 300μs | 300μs |
| Max frequency | 40 Hz | 60 Hz |
| Biphasic waveform | Symmetrical Bi-Phasic<br>Rectangular Waveform | Symmetrical Bi-Phasic<br>Rectangular Waveform |
| - Symmetrical phases<br>- Phase Duration | Fixed Pulse width, max 300 μs | Fixed Pulse width, max 300 μs |
| Maximum Output Net<br>Charge Per Phase<br>Load: 500 ohm | 13.05μC | 3.57 μC |
| Maximum Output Charge<br>Per Phase<br>Load: 500 ohm<br>1 kohm | 24.11 μC<br>17.77 μC | 48.61 μC<br>28.77 μC |
| Maximum Output RMS<br>Current<br>Load: 500 ohm<br>1 kohm | 6.62 mA rms<br>4.91 mA rms | 12.5 mA rms<br>7.40 mA rms |
| Max Current Density<br>(mA/cm2) | 0.046 | 0.1966 |
| Max Power Density (W/cm2) | 0.00152 | 0.00614 |
| On Time | 5 or 10 sec fixed | 1 to 9 sec adjustable |
| Off Time | 10 or 30 sec fixed | 1 to 9 sec adjustable |
| Output pattern -<br>EMS Cycle Mode | Fixed pulse frequency, ramp<br>up/down and work/rest time.<br>Only pulse width is user<br>selectable.<br>(P6,P8,P9) | Fixed pulse widthandwork/rest time.<br>Selectable frequency and ramp<br>up/down time. |
| EMS Reciprocal Mode | Fixed pulse frequency, ramp<br>up/down and work/rest time.<br>Only pulse width is user<br>selectable<br>(P7). | Fixed pulse width andwork/rest time.<br>Selectable pulse frequency and ramp<br>up/down time. |
| Preset Program,P6 of subject<br>device | Program Name<br>P6<br>EMS Mode<br>Cycle mode<br>Pules Freq: 35Hz<br>Pulse width:300μs<br>Ramp up : 2sec<br>On time : 10sec<br>Ramp down : 2sec<br>Off time : 30sec | Program Name<br>CYCLE 2<br>EMS Mode<br>Cycle mode<br>Pules Freq: 50Hz*<br>Pulse width:300μs<br>Ramp up : 3sec<br>On time : 9sec<br>Ramp down : 3sec<br>Off time : 15sec |
| Preset Program,P7of subject<br>device | Program Name<br>P7<br>EMS Mode<br>Reciprocal mode<br>Pules Freq: 40Hz<br>Pulse width:300μs | Program Name<br>RECIPROCAL 3<br>EMS Mode<br>Reciprocalmode<br>Pules Freq: 50Hz*<br>Pulse width:300μs |
{7}------------------------------------------------
| | Ramp up : 4sec | Ramp up : 4sec |
|--------------------|----------------------------|----------------------|
| | On time : 10sec | On time : 8sec |
| | Ramp down : 4sec | Ramp down :4sec |
| | Off time : 10sec | Off time : 14sec |
| Preset Program, P8 | Program Name | Program Name |
| | P8 | CYCLE 2 |
| | EMS Mode | EMS Mode |
| | Cyclemode | Cycle mode |
| | Pules Freq: 40Hz | Pules Freq: 50Hz* |
| | Pulse width:300μs | Pulse width:300μs |
| | Ramp up : 4sec | Ramp up : 3sec |
| | On time : 10sec | On time : 9sec |
| | Ramp down : 4sec | Ramp down : 3sec |
| | Off time : 10sec | Off time : 15sec |
| Preset program P9 | Program Name | Program Name |
| | P9 | CYCLE 1 |
| | EMS Mode | EMS Mode |
| | Cyclemode | Cycle mode |
| | Pules Freq: 20Hz | Pules Freq: 50Hz* |
| | Pulse width:200μs | Pulse width:300μs |
| | Ramp up : 2sec | Ramp up : 2sec |
| | On time : 5sec | On time : 6sec |
| | Ramp down : 2sec | Ramp down : 2sec |
| | Off time : 10sec | Off time : 10sec |
| Preset Program, P1 | Program Name | Program Name |
| | P1 | CONSTANT 1 |
| | Pulse width: | Pulse width: |
| | 120µs default | 50 µs fixed |
| | Adjustable from | Pulse freq: |
| | 50µs to 250µs | 30Hz fixed |
| | Pulse freq: | -------------------- |
| | 120Hz default | Program Name |
| | adjustable from 1 to 150Hz | CONSTANT 2 |
| | | Pulse width |
| | | 200µs fixed |
| | | Pulse freq: |
| | | 50Hz |
#### 7. Non clinical Testing
IEC 60601-1:2005 + CORR. 1 (2006) + CORR. 2 (2007) IEC 60601-1-2:2007 (Edition 3)
Safety requirement EMC requirement
#### 8. Clinical testing
None
#### 9. Conclusion
FD TENS 2090 and FD TENS 2095 have the same intended use and the same technical characteristics as the predicate device(s), FD TENS 2030 (K052813) and FD EMS (K063642). FD TENS 2090, FD TENS 2095 are as safe and as effective as the predicate device. Therefore, FD TENS 2090 and FD TENS 2095 are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.