← Product Code [DXN](/productcode/DXN) · K152659

# V10 (K152659)

_Mediana Co., Ltd. · DXN · Aug 25, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152659

## Device Facts

- **Applicant:** Mediana Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Aug 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation. Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers. Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.

## Device Story

V10 is a compact, lightweight vital signs monitor; inputs include NIBP cuffs, SpO2 sensors, and temperature probes; processes physiological signals to calculate systolic/diastolic/mean arterial pressure, SpO2, pulse rate, and body temperature; displays data via 7-segment and LED indicators; used in hospitals, clinics, and surgical centers by trained clinicians; provides real-time monitoring to assist clinical decision-making; enables continuous patient assessment; benefits patients through timely vital sign tracking and alarm notification for abnormal parameters.

## Clinical Evidence

Bench testing only. System verification and validation (V&V plan #MDR-EG140718-01) performed on production-equivalent units. Risk analysis and mitigation applied per design control procedures. No clinical trial data presented.

## Technological Characteristics

Compact monitor (298x218x172 mm, 3.4 kg); AC/battery powered; 7-segment and LED display; NIBP (oscillometric), SpO2 (pulse oximetry), and temperature sensing; complies with AAMI SP10 for NIBP; ISO 13485 certified manufacturing.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Mediana Co., Ltd. Vital Signs Monitors, Model ARGUS VCM ([K051375](/device/K051375.md))
- Mediana Co., Ltd. Pulse Oximeters, Model P10 ([K100225](/device/K100225.md))
- Covidien LLC. Pulse Oximeters, Model Nellcor Bedside SpO2 Patient Monitoring System ([K142825](/device/K142825.md))
- GE Healthcare. Vital Signs Monitors, Model CARESCAPE V100 ([K102426](/device/K102426.md))
- Tyco Healthcare Group, LP. Infrared Tympanic Thermometers, Model Kendall Genius 2 ([K060649](/device/K060649.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 25, 2016

Mediana Co., Ltd % Mr. Charlie Mack, President International Regulatory Consultants, LLC 77325 Joyce Way Echo. Oregon 97826

Re: K152659 Trade/Device Name: V10 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DQA, FLL Dated: July 11, 2016 Received: July 25, 2016

Dear Mr. Mack:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mude Jellman
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152659

Device Name

V10

Indications for Use (Describe)

The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation.

Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.

Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.

Type of Use (Select one or both, as applicable)

| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)                |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness

#### Submitter

Mediana Co., Ltd 132, Donghwagongdan-Ro, Munmak-Eup, Wonju-Si, Gangwon-Do, Korea Tel) +82-33-742-5400 Fax) +82-33-742-5483

Company Contact: Min-hye, Kim

Date Summary Prepared: July 31th, 2015

#### Device Name

| Trade Name:          | V10                                                                                                                                                                |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:         | Vital Signs Monitor                                                                                                                                                |
| Classification Name: | NIBP measurement system (21CFR870.1130), also contains<br>non-invasive pulse oximetry, SpO2 (21CFR870.2700) and<br>clinical electronic thermometer (21CFR880.2910) |
| Classification:      | Class II                                                                                                                                                           |
| Product Code:        | DXN, DQA, FLL                                                                                                                                                      |

#### Predicate Devices (Legally Marketed Devices)

The predicate device for the V10 are:

- . Mediana Co., Ltd. Vital Signs Monitors, Model ARGUS VCM Cleared by FDA through 510(k) No. K051375, and
- = Mediana Co., Ltd. Pulse Oximeters, Model P10 Cleared by FDA through 510(k) No. K100225, and
- " Covidien LLC. Pulse Oximeters, Model Nellcor Bedside SpO2 Patient Monitoring System, K142825, and
- = GE Healthcare. Vital Signs Monitors, Model CARESCAPE V100 Cleared by FDA through 510(k) No. K102426, and
- = Tyco Healthcare Group, LP. Infrared Tympanic Thermometers, Model Kendall Genius 2 Cleared by FDA through 510(k) No. K060649

#### Device Description

The V10 vital signs monitor is to monitor non-invasive blood pressure, pulse rate, non-invasive functional oxygen saturation of arterial hemoglobin (SpO2) for adult, pediatric and neonate patients and body temperature for adult and pediatric patients in general hospital and alternate care facilities by medically trained personnel. This monitor is available for sale only upon the order of a physician or licensed health care professional.

The Mediana V10 vital signs monitor is a lightweight and compact device (298 × 218 × 172 (mm) (W×H×D) and 3.4 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on 7-segment and LED displays.

{4}------------------------------------------------

#### Intended Use

The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation.

- Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.
- Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.

### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)

The Mediana vital signs monitor, Model V10 is substantially equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM, Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System, GE Healthcare vital signs monitors Model CARESCAPE V100 and Tyco Healthcare Group, LP. Infrared tympanic thermometers Model Kendall Genius2.

- The Non-Invasive Blood Pressure (NIBP) measurement specifications and = performance are equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM. Mediana Co., Ltd. vital signs monitors Model ARGUS VCM complies the same AAMI SP10 standard as Mediana vital signs monitor, Model V10.
- = The Pulse rate specifications and performance derived from either Non-Invasive Blood Pressure (NIBP) or Pulse Oximetry (SpO2) are equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM, Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System.
- = The Pulse Oximetry (SpO2) specifications and performance are equivalent to Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System.
- The Temperature specifications and performance are equivalent to GE Healthcare vital signs monitors Model CARESCAPE V100 and Tyco Healthcare Group, LP. Infrared tympanic thermometers Model Kendall Genius2.

{5}------------------------------------------------

### Summary of Performance Testing

The Mediana vital signs monitor, Model V10 substantially has been tested in accordance with the system V & V plan #MDR-EG140718-01 included with the submission using production equivalent units prior to release to market.

A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21CFR820, ISO13485 and CMDCAS ISO 13485 certified by DNV (Det Norske Veritas).

#### Conclusions

As stated above, the V10 is safe and effective, complies with the appropriate medical device guidance and standards and is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K152659](https://fda-staging.innolitics.com/device/K152659)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
