The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation. Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers. Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.
Device Story
V10 is a compact, lightweight vital signs monitor; inputs include NIBP cuffs, SpO2 sensors, and temperature probes; processes physiological signals to calculate systolic/diastolic/mean arterial pressure, SpO2, pulse rate, and body temperature; displays data via 7-segment and LED indicators; used in hospitals, clinics, and surgical centers by trained clinicians; provides real-time monitoring to assist clinical decision-making; enables continuous patient assessment; benefits patients through timely vital sign tracking and alarm notification for abnormal parameters.
Clinical Evidence
Bench testing only. System verification and validation (V&V plan #MDR-EG140718-01) performed on production-equivalent units. Risk analysis and mitigation applied per design control procedures. No clinical trial data presented.
Technological Characteristics
Compact monitor (298x218x172 mm, 3.4 kg); AC/battery powered; 7-segment and LED display; NIBP (oscillometric), SpO2 (pulse oximetry), and temperature sensing; complies with AAMI SP10 for NIBP; ISO 13485 certified manufacturing.
Indications for Use
Indicated for monitoring NIBP, SpO2, pulse rate, and temperature in adult, pediatric, and neonatal patients in hospital and hospital-type clinical settings. Requires operation by trained clinicians (doctors, nurses, technicians).
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Mediana Co., Ltd. Vital Signs Monitors, Model ARGUS VCM (K051375)
Mediana Co., Ltd. Pulse Oximeters, Model P10 (K100225)
Covidien LLC. Pulse Oximeters, Model Nellcor Bedside SpO2 Patient Monitoring System (K142825)
GE Healthcare. Vital Signs Monitors, Model CARESCAPE V100 (K102426)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2016
Mediana Co., Ltd % Mr. Charlie Mack, President International Regulatory Consultants, LLC 77325 Joyce Way Echo. Oregon 97826
Re: K152659 Trade/Device Name: V10 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DQA, FLL Dated: July 11, 2016 Received: July 25, 2016
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mude Jellman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152659
Device Name
V10
Indications for Use (Describe)
The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation.
Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.
Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|------------------------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
#### Submitter
Mediana Co., Ltd 132, Donghwagongdan-Ro, Munmak-Eup, Wonju-Si, Gangwon-Do, Korea Tel) +82-33-742-5400 Fax) +82-33-742-5483
Company Contact: Min-hye, Kim
Date Summary Prepared: July 31th, 2015
#### Device Name
| Trade Name: | V10 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Vital Signs Monitor |
| Classification Name: | NIBP measurement system (21CFR870.1130), also contains<br>non-invasive pulse oximetry, SpO2 (21CFR870.2700) and<br>clinical electronic thermometer (21CFR880.2910) |
| Classification: | Class II |
| Product Code: | DXN, DQA, FLL |
#### Predicate Devices (Legally Marketed Devices)
The predicate device for the V10 are:
- . Mediana Co., Ltd. Vital Signs Monitors, Model ARGUS VCM Cleared by FDA through 510(k) No. K051375, and
- = Mediana Co., Ltd. Pulse Oximeters, Model P10 Cleared by FDA through 510(k) No. K100225, and
- " Covidien LLC. Pulse Oximeters, Model Nellcor Bedside SpO2 Patient Monitoring System, K142825, and
- = GE Healthcare. Vital Signs Monitors, Model CARESCAPE V100 Cleared by FDA through 510(k) No. K102426, and
- = Tyco Healthcare Group, LP. Infrared Tympanic Thermometers, Model Kendall Genius 2 Cleared by FDA through 510(k) No. K060649
#### Device Description
The V10 vital signs monitor is to monitor non-invasive blood pressure, pulse rate, non-invasive functional oxygen saturation of arterial hemoglobin (SpO2) for adult, pediatric and neonate patients and body temperature for adult and pediatric patients in general hospital and alternate care facilities by medically trained personnel. This monitor is available for sale only upon the order of a physician or licensed health care professional.
The Mediana V10 vital signs monitor is a lightweight and compact device (298 × 218 × 172 (mm) (W×H×D) and 3.4 kg) powered by AC mains (100-240VAC, 50-60Hz) and also powered by internal battery. The monitor provides patient data and monitoring status on 7-segment and LED displays.
{4}------------------------------------------------
#### Intended Use
The V10 is intended to be used to monitor noninvasive blood pressure (NIBP) - systolic, diastolic and mean arterial pressures, functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities. Monitor users should be skilled at the level of a technician, doctor, nurse or medical specialist. The V10 is suitable for continuous operation.
- Note: Hospital use typically includes such as general care floors, operating rooms, special procedure areas, intensive and critical care area, within the hospital. Hospital-type facilities include physician office-based facilities, sleep labs, skilled nursing facilities, surgical centers, and sub acute care centers.
- Note: The medically skilled and trained user can be clinicians like doctors and nurses who know how to take and interpret a patient's vital signs. These clinicians must take direct responsibility for the patient's life. This can include care-givers or medically trained interpreters who are authorized under the appropriate clinical facility procedures to support patient care. Any inappropriate setting, especially the alarm limit or alarm notification settings, can lead to a hazardous situation that injures the patient, harms the patient, or threatens the patient's life. This equipment should only be operated by trained users who can adjust the settings of the vital signs monitor.
### Summary of Technical Characteristics of the Device Compared to the Predicate Devices (Legally Marketed Devices)
The Mediana vital signs monitor, Model V10 is substantially equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM, Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System, GE Healthcare vital signs monitors Model CARESCAPE V100 and Tyco Healthcare Group, LP. Infrared tympanic thermometers Model Kendall Genius2.
- The Non-Invasive Blood Pressure (NIBP) measurement specifications and = performance are equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM. Mediana Co., Ltd. vital signs monitors Model ARGUS VCM complies the same AAMI SP10 standard as Mediana vital signs monitor, Model V10.
- = The Pulse rate specifications and performance derived from either Non-Invasive Blood Pressure (NIBP) or Pulse Oximetry (SpO2) are equivalent to Mediana Co., Ltd. vital signs monitors Model ARGUS VCM, Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System.
- = The Pulse Oximetry (SpO2) specifications and performance are equivalent to Mediana Co., Ltd. Pulse Oximeter Model P10, Covidien LLC. Pulse Oximeter Model Nellcor Bedside SpO2 Patient Monitoring System.
- The Temperature specifications and performance are equivalent to GE Healthcare vital signs monitors Model CARESCAPE V100 and Tyco Healthcare Group, LP. Infrared tympanic thermometers Model Kendall Genius2.
{5}------------------------------------------------
### Summary of Performance Testing
The Mediana vital signs monitor, Model V10 substantially has been tested in accordance with the system V & V plan #MDR-EG140718-01 included with the submission using production equivalent units prior to release to market.
A risk analysis identifying potential hazards and documenting mitigation of the hazards has been developed and applied as part of Mediana design control procedure. Mediana's quality system confirms to 21CFR820, ISO13485 and CMDCAS ISO 13485 certified by DNV (Det Norske Veritas).
#### Conclusions
As stated above, the V10 is safe and effective, complies with the appropriate medical device guidance and standards and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.