SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is intended to use as correction of refractive myopia in aphakic or non-aphakic persons with non-diseased eyes, who may have 1.00D of astigmatism or less. The lenses are available with a visibility-handling tint or with a decorative tint intended to enhance or alter the apparent color of the eye. Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lenses are not intended to be disinfected and should be discarded after a single use.
Device Story
Soft (hydrophilic) contact lens for refractive myopia correction; manufactured via cast molding. Material: Ocufilcon D (55% water content, HEMA/MAA copolymer cross-linked with EGDMA). Available in clear (visibility tint) or decorative (red/green) versions. Prescribed by eyecare practitioners for single-use daily wear; discarded after one use. Supplied sterile in isotonic phosphate buffered saline. Provides vision correction; decorative tints alter eye appearance.
Clinical Evidence
No clinical data. Bench testing only, including biocompatibility (cytotoxicity, ocular irritation, hypersensitization, systemic injection), physicochemical testing, and stability testing.
Technological Characteristics
Material: Ocufilcon D (HEMA/MAA copolymer, EGDMA cross-linked). Water content: 55%. Refractive index: 1.41. Oxygen permeability (Dk): 23 x 10^-11. Cast molded. Sterile, single-use. Visibility tint (Reactive Blue #19) or decorative tint (iron oxides/pigment green 7).
Indications for Use
Indicated for correction of refractive myopia in aphakic or non-aphakic persons with non-diseased eyes, with 1.00D of astigmatism or less. For single-use disposable daily wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
ACUVUE 2 COLOURS Brand (etafilcon A) Contact Lens with UV blocker (K010114)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2016
Smart Performer Corp. Ltd. Mr. Jeffersion Shen QA Dept. Manager & Management Representative No.13-1, Wuquan 1st Rd., Xinzhuang Dist. New Taipei City, 242 TW
Re: K152657
Trade/Device Name: Smart Performer 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL, MVN Dated: April 8, 2016 Received: April 8, 2016
Dear Mr. Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K152657
Device Name SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens
#### Indications for Use (Describe)
SMART PERFORMER 55 (Ocutilcon D) Soft (hydrophilic) Contact Lens is intended to use as correction of refractive myopia in aphakic or non-aphakic persons with non-diseased eyes, who may have 1.00D of astigmatism or less. The lenses are available with a visibility-handling tint or with a decorative tint intended to enhance or alter the apparent color of the eve.
Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lenses are not intended to be disanfected and should be discarded after a single use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995,
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This summary of 510(k) Safety and Effectiveness information is being submitted in accordance with the requirements of Safe Medical Device Act (SMDA) of 1990 and Title 21 CFR 807.92.
The assigned 510(k) number is
| | Submitter Information |
|-----------------|----------------------------------------------------------------|
| Company | SMART PERFORMER CORP. LTD. |
| | No.13-1, Wuquan 1st Rd., Xinzhuang Dist., New Taipei City 242, |
| | Taiwan (R.O.C.) |
| Contact Person | Jefferson Shen |
| | QA Dept. Manager & Management Representative |
| Telephone | 886-2-22999699 |
| Fax | 886-2-22996866 |
| E mail | jefferson@smartvision.tw |
| Date of Summary | July 8, 2015 |
### Device Information
| Name of Device | SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact<br>Lens |
|-------------------|---------------------------------------------------------------------|
| Common Name | Soft (hydrophilic) contact lens, daily wear |
| Classification | Class II |
| Regulation Number | 886.5925 |
| Product Code | LPL, MVN |
| Review Panel | Ophthalmic Devices |
### Predicate Device
- iLens@ (ocufilcon D) Daily Wear Soft (Hydrophilic) Contact Lens, K141280, submitted by SEINOH OPTICAL CO. LTD. .
- BIOMEDICS UV Colors (Ocufilcon D) Soft (hydrophilic) Contact Lens, K013377, submitted by Ocular Sciences Inc.
- ACUVUE 2 COLOURS Brand (etafilcon A) Contact Lens with UV blocker, K010114, submitted by VISTAKON, Division of Johnson & Johnson Vision Care, Inc.
{4}------------------------------------------------
#### Intended Use
SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is intended to use as correction of refractive myopia in aphakic or non-aphakic persons with non-diseased eyes, who may have 1.00D of astigmatism or less. The lenses are available with a visibility-handling tint or with a decorative tint intended to enhance or alter the apparent color of the eye. Eyecare practitioners may prescribe the contact lens for single-use disposable daily wear. The SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lenses are not intended to be disinfected and should be discarded after a single use.
#### Device Description
The SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is available as spherical lenses manufactured by cast molded method. The lens is made from a lens material that is approximately 55% water. The hydrogel lens' material is a copolymer composed of 2-hydroxyethyl methacrylate (HEMA) and methacrylic acid (MAA), which was cross-linked with ethyleneglycol dimethacrylate (EGDMA) via UV photo-polymerization. The SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is available visibility tinted and cosmetic tinted. The visibility tinted blue using Reactive Blue Dye #19 to make the lenses more visible for handling. The cosmetically tinted lens is tinted in an annular pattern, providing a clear optic zone, with iron oxides (red) or C.I. Pigment green 7. The cosmetically tinted lens is available in GREEN and RED. The lens is supplied sterile in sealed blister packers containing sterile isotonic phosphate buffered saline.
The Lens designs in the following parameter ranges:
| Chord Diameter | 13.8 mm to 14.8 mm |
|--------------------------------------------------------|---------------------------------------------------------------------------------|
| Center Thickness | 0.060 mm to 0.250 mm (varies with power) |
| Base Cure | 8.2 mm to 9.0 mm |
| Power | -0.25D to -12.00 D |
| The Lens designs in the following physical properties: | |
| Refractive Index | 1.41 |
| Water Content | 55 % |
| Oxygen Permeability (Dk)* | 23 x 10-11 (cm²/sec) (ml O2/ml x mm Hg ) at 35°<br>(revised Coulometric method) |
| Light Transmittance | >97% (clear)<br>>97% (tinted) |
| Specific Gravity | 1.10 (hydrated) |
{5}------------------------------------------------
## Non-Clinical Studies
The following tests were conducted as recommended by the Premarket Notification 510 (k) Guidance Document for Daily Wear Contact Lenses, revised May 1994.
- (a). Biocompatibility ( cytotoxicity, ocular irritation, delayed type hypersensitization and acute systemic injection)
- (b). Physicochemical Testing
- (c). Stability Testing
# Clinical Studies
Ocufilcon D lenses have been used widely. Their safety and effectiveness have been well documented. Their safety and effectiveness can be further exemplified by two lenses cleared by FDA
- iLens@ (ocufilcon D) Daily Wear Soft (Hydrophilic) Contact Lens, K141280
● BIOMEDICS UV Colors (Ocufilcon D) Soft (hydrophilic) Contact Lens, K013377 Clinical studies for SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens are not required for premarket notification as the USAN name and process are the same as the above mentioned predicate devices.
# Substantial Equivalence Summary
Information submitted in the 510(k) establishes that the SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens has comparable physicochemical properties to the predicate devices and do not raise questions of safety and effectiveness. Shelf life testing has shown the lenses remain sterile and that lens properties do not change before the expiration date. Results of cytotoxicity, ocular irritation, delayed type hypersensitization and acute systemic injection showed the SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is substantially equivalent to the predicate devices in safety and biocompatibility. Therefore, the SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is substantially equivalent to the predicate devices.
{6}------------------------------------------------
| Substantial Equivalence Comparison | | | | |
|------------------------------------|------------------------------------------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Device | SMART<br>PERFORMER 55<br>(Ocufilcon D) Soft<br>(hydrophilic)<br>Contact Lens | iLens@ (ocufilcon<br>D) Daily Wear<br>Soft (Hydrophilic)<br>Contact Lens | BIOMEDICS UV<br>Colors (Ocufilcon D)<br>Soft (hydrophilic)<br>Contact Lens | ACUVUE 2<br>COLOURS Brand<br>(etafilcon A) Contact<br>Lens with UV blocker |
| Material<br>(USAN name) | Ocufilcon D | Ocufilcon D | Ocufilcon D | Etafilcon A |
| Indication for use | correction of<br>refractive ametropia | correction of<br>refractive<br>ametropia | correction of<br>refractive ametropia | correction of<br>refractive ametropia |
| Manufacturing<br>Method | Cast Molded | Cast Molded | Cast Molded | Cast Molded |
| Water Content | 55% | 55% | 55% | 58% |
| Light<br>Transmittance | >97% | >95% | >97 % | > 70% |
| Dk (35 ℃) | $23 x 10^{-11}$ | $20 x 10^{-11}$ | $19.6 x 10^{-11}$ | $21.4 x 10^{-11}$ |
| Refractive Index | 1.41 | 1.405 | 1.41 | 1.40 |
| Toxicity | Non-Toxic | Non-Toxic | Non-Toxic | Non-Toxic |
| Color Additives | ● Iron oxides (red)<br>●<br>C.I. Pigment green 7 | Reactive blue dye #19 | ● Iron oxides<br>● Titanium Dioxide<br>● Carbazole Violet<br>● Copper<br>● Phthalocyanine Blue<br>● Copper<br>● Phthalocyanine Green | ● iron oxides<br>● titanium dioxide<br>● Reactive blue dye 4<br>● Vat orange 1<br>● Phthalocyanine Green<br>● Phthalocyaninato(2-) |
{7}------------------------------------------------
### Conclusion
A series of pre-clinical tests were performed to demonstrate the safety and effectiveness of the SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens and to establish substantial equivalence to the predicate devices. Information submitted in the 510(k) also establishes that the lens do not raise questions of safety and effectiveness. Therefore, the SMART PERFORMER 55 (Ocufilcon D) Soft (hydrophilic) Contact Lens is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.