The Stryker Pi Drive Plus Motor is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, drilling, reaming, decorticating, shaping and smoothing of bone cement and teeth in a variety of surgical procedures, including but not limited to Dental, ENT (ear, nose and throat), Neuro, Spine and Endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
Device Story
The Stryker Pi Drive Plus Motor is a reusable, 40V DC electric surgical drill used with the Stryker CORE System. It functions as a rotary tool to transmit torque to various attachments and cutting accessories (burs, routers). The device is operated by a surgeon in an OR setting, with speed controlled via a foot switch or a hand-integrated sensor. An optional 'Extender' accessory provides additional length. The device facilitates bone and tissue manipulation, screw placement, and hardware cutting. It is non-patient contacting. Clinical benefit is derived from its ability to perform precise surgical tasks in diverse specialties. The device is cleaned and sterilized via moist heat (steam) between uses.
Clinical Evidence
No clinical testing was deemed necessary. Substantial equivalence is supported by non-clinical bench testing, including reliability, functional, graphic legibility, and temperature testing, demonstrating that the device meets performance requirements.
Technological Characteristics
Reusable 40V DC electric motor; stainless steel and aluminum housing; silicone and Slick Sil LSR coated cable. Rotary torque transmission; max speed 75,000 RPM. Connectivity: wired to CORE console. Sterilization: moist heat (steam) to SAL 10^-6. Dimensions: 109mm length, 17mm diameter. Weight: 403g. Features knurled grip and handswitch/footswitch control.
Indications for Use
Indicated for use in general patient populations requiring surgical cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement, and teeth, or placement/cutting of fixation hardware (screws, wires, pins) during dental, ENT, neuro, spine, and endoscopic procedures. No known contraindications.
Regulatory Classification
Identification
An ear, nose, and throat electric or pneumatic surgical drill is a rotating drilling device, including the handpiece, that is intended to drive various accessories, such as an ear, nose, and throat bur (§ 874.4140), for the controlled incision or removal of bone in the ear, nose, and throat area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2015
Stryker Corporation Ms. Julia Helgeson Staff Regulatory Affairs Specialist, Stryker Instruments 400 E. Milham Avenue Kalamazoo, MI 49001
Re: K152641
Trade/Device Name: Pi Drive Plus Motor, Extender Regulation Number: 21 CFR 874.4250 Regulation Name: Ear, Nose, and Throat Electric or Pneumatic Surgical Drill Regulatory Class: Class II Product Code: ERL, DZJ, HBE Dated: October 23, 2015 Received: October 26, 2015
Dear Ms. Helgeson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152641
Device Name Stryker Pi Drive Plus Motor
### Indications for Use (Describe)
The Stryker Pi Drive Plus Motor is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, drilling, reaming, decorticating, shaping and smoothing of bone cement and teeth in a variety of surgical procedures, including but not limited to Dental, ENT (ear, nose and throat), Neuro, Spine and Endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|--|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |
| <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Contact Details | | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---------------------|
| 510(k) Owner | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, Michigan 49001<br>US | | |
| FDA Establishment<br>Registration Number | 1811755 | | |
| Submitted By | Deval Patel, MS, RAC<br>Senior Regulatory Affairs Specialist<br>Ph: +1-269-389-5671<br>Fax: +1-269-389-5412<br>E-mail: Deval.Patel@Stryker.com | | |
| Date Submitted | September 14, 2015 | | |
| Device Name | | | |
| Trade Name | Stryker® Pi Drive Plus Motor | | |
| Common Name | Ear, Nose, and Throat Electric or Pneumatic Surgical Drill | | |
| Classification | Class II | | |
| Primary<br>Classification Name | Drill, Surgical, Ent (Electric or Pneumatic) including Handpiece<br>(21 CFR 874.4250, Product code ERL) | | |
| Secondary<br>Classification Name | Driver, Wire, and Bone Drill, Manual<br>(21 CFR 872.4120, Product code DZJ)<br>Drills, Burs, Trephines & Accessories (Simple, Powered)<br>(21 CFR 882.4310, Product code HBE) | | |
| Legally Marketed Predicate Device | | | |
| 510(k) Number | Product Code | Trade Name | Manufacturer |
| Predicate | | | |
| K141935 | ERL, HBE, DZJ | Stryker® S2 Drill | Stryker Instruments |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Stryker logo in black and white. The logo is a stylized wordmark with a unique font. The word "Stryker" is written in all lowercase letters, with the "r" having a distinctive shape. There is a registered trademark symbol to the right of the word.
| Indication for Use | The Stryker Pi Drive Plus Motor is intended for use with the Stryker Consolidated Operating Room Equipment (CORE) System. When used with a variety of attachments and cutting accessories, the drill is intended for use in cutting, drilling, reaming, decorticating, shaping, and smoothing of bone, bone cement and teeth in a variety of surgical procedures, including but not limited to dental, ENT (ear, nose, and throat), neuro, spine, and endoscopic applications. It is also usable in the placement or cutting of screws, metal, wires, pins, and other fixation devices. |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description | The Stryker Pi Drive Plus Motor is a non-sterile, reusable, electric powered (40V DC) motor that rotates cutting accessories up to speeds of 75,000 RPM. It is powered by the CORE console. The cable of the Pi Drive Plus Motor is integrated into the proximal end and connects directly to the CORE console. The subject drill speed is controlled by a foot switch (connected to the CORE Console) or a removable handswitch which triggers a sensor within the handpiece. A variety of attachments and accessories can be used in conjunction with the drill. The motor (drill), cable, handswitch and footswitch are considered non-patient contacting device. |
| | The Extender is an optional direct drive accessory that attaches to the distal end of the drill and transmits torque from the drill to various attachments. The Extender provides an option to add length to the drill, based on the user preference. The Extender is delivered non-sterile, re-usable, constructed of stainless steel material and is considered a non-patient contacting device. |
| Performance Data (Non-Clinical Tests) | The results of the performance testing demonstrate that the functionality, integrity, and safety and effectiveness of the Stryker Pi Drive Plus Motor and Extender is sufficient for their intended use and support a determination of substantial equivalence to the predicate device. |
| Summary of Performance Testing | The following verification tests were performed on the subject device to demonstrate that device meets performance requirements under its indication for use conditions. |
- Reliability Testing- Motor (Drill), Handswitch Functional .
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a small circled "c" symbol to the upper right of the "r". The word is in black and the background is white.
### Range Testing and Extender
- Graphic Legibility Testing for Motor (Drill) and Extender ●
- . Temperature Testing
Results of these tests demonstrate that the functionality, integrity, and safety and effectiveness of the subject device are sufficient for their intended use, indications for use and support a determination of the substantial equivalence.
Clinical Tests No clinical testing was deemed necessary for this 510(k).
{6}------------------------------------------------
### Table 6-1: Summary of Predicate Comparison
| DESCRIPTION | STRYKER S2 DRILL [PREDICATE]<br>K141935 | STRYKER® Pi DRIVE PLUS MOTOR<br>[SUBJECT] |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for Use | The Stryker S 2 Drill is intended for<br>use with the Stryker Consolidated<br>Operating Room Equipment (CORE)<br>System. When used with a variety of<br>attachments and cutting accessories,<br>the drill is intended for use in cutting,<br>drilling, reaming, decorticating,<br>shaping, and smoothing of bone,<br>bone cement and teeth in a variety of<br>surgical procedures, including but not<br>limited to dental, ENT (ear, nose, and<br>throat), neuro, spine, and endoscopic<br>applications. It is also usable in the<br>placement or cutting of screws,<br>metal, wires, pins, and other fixation<br>devices. | The Stryker Pi Drive Plus Motor is<br>intended for use with the Stryker<br>Consolidated Operating Room<br>Equipment (CORE) System. When used<br>with a variety of attachments and<br>cutting accessories, the drill is intended<br>for use in cutting, drilling, reaming,<br>decorticating, shaping, and smoothing of<br>bone, bone cement and teeth in a<br>variety of surgical procedures, including<br>but not limited to dental, ENT (ear, nose,<br>and throat), neuro, spine, and<br>endoscopic applications. It is also usable<br>in the placement or cutting of screws,<br>metal, wires, pins, and other fixation<br>devices. |
| Classification of<br>Device | Class II | Class II |
| Primary Product Code | ERL<br>Drill, Surgical, ENT (Electric or<br>Pneumatic) including Handpiece | ERL<br>Drill, Surgical, ENT (Electric or<br>Pneumatic) including Handpiece |
| Primary Regulation | 21 CFR 878.4250<br>Ear, nose and throat electric or<br>pneumatic surgical drill | 21 CFR 878.4250<br>Ear, nose and throat electric or<br>pneumatic surgical drill |
| Condition of Use | Reusable | Reusable |
| Type of Use | Prescription Use Only | Prescription Use Only |
| Patient Population | General | General |
| Contraindications | None known | None known |
| Mode of Action | Rotary (transmits Torque) | Rotary (transmits Torque) |
| Power source | 40 V DC Electric Motor connected via<br>cable to CORE console | 40 V DC Electric Motor connected via<br>cable to CORE console |
| Diameter of Motor | 17mm | 17mm |
| Length of the Motor | 123.5mm | 109mm |
| DESCRIPTION | STRYKER S2 DRILL [PREDICATE]<br>K141935 | STRYKER® Pi DRIVE PLUS MOTOR<br>[SUBJECT] |
| Weight of the Motor | 313g | 403g |
| Maximum Speed | 75,000 rpm | 75,000 rpm |
| Accessories | CORE Console Attachments Cutting accessories Irrigation Clips | CORE Console Attachments Cutting accessories Irrigation Clips Extender |
| Means of Speed Control | Footswitch | Handswitch or Footswitch |
| Grip Design | Smooth | Knurled |
| Cutting Accessories<br>Retention Method by Drill | Burs: Mechanical Lock | Burs: Mechanical Lock |
| | Routers: Routers are not utilized<br>with the S2 Drill | Routers: Spring Collar Mechanism |
| Attachment Retention<br>Method by Drill | Attachments (Except Footed<br>Attachment):<br>Mechanical Lock<br>Footed Attachment:<br>Footed Attachments are not utilized<br>with the S2 Drill | Attachments (Except Footed<br>Attachment):<br>Friction<br>Footed Attachment:<br>Mechanical Lock |
| Housing Material of Drill | Stainless Steel and Aluminum | Stainless Steel and Aluminum |
| Cable Material | Silicone | Silicone and Slick Sil LSR coating |
| Method of Sterilization | Moist Heat (Steam) | Moist Heat (Steam) |
| Sterility Assurance Level<br>(SAL) | 10-6 | 10-6 |
| Cleaning Method | Manual and Mechanical<br>(automated) | Manual and Mechanical (automated) |
| Packaging | Packaged in a sealed retention insert | Packaged in a sealed retention insert |
Special 510(k) Submission
{7}------------------------------------------------
stryker®
### Table 6-1: Summary of Predicate Comparison (Continued)
STRYKER® Pi DRIVE PLUS MOTOR
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The letters are black and the background is white. There is a small circle with an R inside of it in the upper right corner of the image. The word "stryker" is a registered trademark of Stryker Corporation, a medical technology company.
### Conclusion/Substantial Equivalence Rationale
The Stryker Pi Drive Plus Motor is substantially equivalent in indications for use, intended use, technological characteristics, safety and effectiveness to the previously cleared Stryker S2 Drill. The subject device has the same operating principle, functional characteristics and applications as the predicate device. The modifications introduced raise no new issues of safety and effectiveness.
Therefore, the subject device is substantially equivalent to the existing predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.