K152627 · Metric Medical Devices, Inc. · HTY · Jan 13, 2016 · Orthopedic
Device Facts
Record ID
K152627
Device Name
IFS Fixation Scaffold System
Applicant
Metric Medical Devices, Inc.
Product Code
HTY · Orthopedic
Decision Date
Jan 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IFS Fixation Scaffold is indicated for small bone fixation, reconstruction and fusion such as inter-digital fusion of fingers and toes.
Device Story
IFS Fixation Scaffold System is a nitinol-based intramedullary implant for small bone fixation; used for inter-digital fusion of fingers and toes. Device consists of a hollow fenestrated cylinder with surface bulges and prongs; utilizes nitinol shape-memory properties to lock into medullary canal and provide compression. Implant is pre-loaded on an insertion instrument; released via knob rotation; instrument acts as depth stop. Upon release at room temperature, device transitions to austenitic state, shortening length and expanding diameter to engage bone. Provides mechanical fixation to facilitate bone healing. Used by surgeons in clinical settings.
Clinical Evidence
Bench testing only. Evaluated corrosion susceptibility (ASTM F2129), contraction/expansion forces, ultimate bending strength, and bone fixation strength. Results showed IFS performance equivalent or superior to predicate devices (K133520, K863734, K123363).
Technological Characteristics
Material: Nitinol. Form factor: Hollow fenestrated cylinder (2.5-4.5 mm diameter, 21-27 mm length). Principle: Shape-memory alloy transition to austenitic state at room temperature for bone fixation/compression. Features: Surface bulges and prongs for bone locking. Connectivity: None. Sterilization: Not specified. Standards: ASTM F2129 for corrosion testing.
Indications for Use
Indicated for small bone fixation, reconstruction, and fusion, specifically inter-digital fusion of fingers and toes.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other.
January 13, 2016
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Metric Medical Devices, Incorporated W.Casey Fox, Ph.D., P.E. President 846 Silver Springs Helotes, Texas 78023
Re: K152627
Trade/Device Name: IFS Fixation Scaffold System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: December 8, 2015 Received: December 14, 2015
Dear Dr. Fox:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications For Use
### Indications For Use
______________________________________________________________________________________________________________________________________________________________________________
K152627 510(k) Number (if Known):
Device Name: Intramedullary Fixation Scaffold (IFS) System
Indications for use: The IFS Fixation Scaffold is indicated for small bone fixation, reconstruction and fusion such as inter-digital fusion of fingers and toes.
Prescription Use X Over-The-Counter AND/OR Use (Per 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## Appendix I: 510(k) Summary of Safety and Effectiveness
510(k) Summary As required by 21 CFR 807.92: Name of Sponsor: Metric Medical Devices, Inc., 846 Silver Springs, Helotes, TX, 78023 Contact: W. Casey Fox, Ph.D., P.E., President 12/3/2015 Date Prepared: Proprietary Name: IFS Fixation Scaffold System Classification Name: Pin, Fixation, Bone Device classification: Class II per 21 CFR 888.3040 Product Device Code: HTY
Substantial Equivalence: The IFS is substantially equivalent to legally marketed intramedullary implants (K133520), k-wires (K863734) and bone staple (K123363) in indications for use, material, design and function. Any minor differences between these devices or their performance do not raise any questions of safety and effectiveness.
The IFS Fixation Scaffold is indicated for small bone fixation, Indications for use: reconstruction and fusion such as inter-digital fusion of fingers and toes.
Performance Data: Corrosion susceptibility, implant contraction and expansion forces, ultimate strength, amount of shape change and bone fixation strength was measured and compared between the IFS and three predicate devices.
Corrosion testing per ASTM F2129 "Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices" was performed. The corrosion behavior of the IFS was equivalent or better than the predicate device to which it was compared and presented high corrosion breakdown potentials for all IFS devices tested.
IFS contraction and expansion forces, ultimate strength and bone fixation strength were measured for the IFS and predicate devices. Contraction tests showed that the IFS contraction force causing bone segments to be put into compression was equivalent to or less than the 9x9 mm and 11x10 mm bone staple implants of K123363 and the expansion force causing retention in the medullary canal was equivalent or superior to the K133520 predicate device. Ultimate strength in bending was measured for the IFS. K133520 and K863734 and the IFS was found equivalent or superior in strength to the predicate devices tested. Bone fixation strength was measured for the IFS, K 133520 and K863734 and the IFS was found equivalent or superior to the predicate device.
Device Description: The IFS is a hollow fenestrated cylinder fabricated from nitinol with a section that increases the implants diameter to lock into the medullary compartment of bone while shortening its length. The nitinol's transition temperatures are set so that the material is in the austenitic state when released at room temperature. Features on the cylinder's surface, such as bulges and prongs, lock into bone.
{4}------------------------------------------------
The IFS is pre-loaded onto an insertion instrument and held elongated in position by a lock plate. During insertion, the IFS is released by turning a knob on its instrument while the lock plate both releases the IFS and limits the depth of implant insertion.
The IFS implants range in diameter from 2.5 to 4.5 mm and up to 9.3 mm when expanded and range in length from 21 to 27 mm depending on size and extent of contraction.
### Comparison of Technical Characteristics with the Predicate Device:
The IFS and predicates K123363 and K133520 are fabricated from nitinol. The IFS and K123363 change shape at room temperature and K133420 changes at body temperature. The IFS and predicate K123363 have essentially the same manufacturing thermal history and surface treatment. The K863734 predicate is fabricated from stainless steel.
The IFS and K123363 both change shape to pull bone together. The IFS shortens its cylindrical length and the K123363 brings its legs together and shortens its bridge to pull bone segments together and apply fixation forces to bone during healing. The fixation force of the IFS is in the same range as a small K123363 implant.
The IFS and K133520 both have sections that expand to lock into the medullary canal of bone. Features on the surface of these devices include barbs or prongs for locking into bone. The expansion force of the IFS is within the range of expansion forces of the predicate K133520.
The IFS and predicate K123363 and K133520 implants are held with an instrument during insertion. The IFS and K123363 instruments are fabricated from the same material. The instruments hold these implants from changing shape. When released from the instrument the shape changing features of the implants are enabled and the bone locking features engage bone. The IFS and K133520 instruments also act as a depth stop.
The K863734 predicate device is implanted with a drill motor, locks due to surface friction and interference, is round in cross section, solid, and has a trocar tip for cutting bone.
The IFS, K133520, K123363 and K863734 are small devices and similar in size: the IFS is 2.5 to 4.5 mm in diameter and up to 9.3 mm when expanded and its length ranges from 21 to 27 mm; the K 133520 device ranges in width from 4 mm to 7 mm when expanded and its length ranges12 to 25 mm; the K123363 device ranges in bridge and leg length from 9 mm to 20 mm; and the K863734 ranges in diameters of 1 mm to 2 mm and lengths exceeding several inches.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.