KYPHON HV-R Fenestrated Screw Cement, CD HORIZON Fenestrated Screw Set
K152604 · Medtronic Sofamor Danek USA, Incorporated · PML · Jan 6, 2016 · Orthopedic
Device Facts
Record ID
K152604
Device Name
KYPHON HV-R Fenestrated Screw Cement, CD HORIZON Fenestrated Screw Set
Applicant
Medtronic Sofamor Danek USA, Incorporated
Product Code
PML · Orthopedic
Decision Date
Jan 6, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
When used in conjunction with the CD HORIZON® Fenestrated Screws, KYPHON HV-R® Fenestrated Screw Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. KYPHON HV-R® Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
Device Story
System comprises high-viscosity, radiopaque, self-curing PMMA bone cement and cannulated multi-axial fenestrated screws; used in thoracic and lumbar spine. Physician mixes powder (methylmethacrylate-styrene co-polymer, barium sulfate, di-benzoyl peroxide) and liquid (methylmethacrylate monomer, N, N dimethyl-p-toluidine, hydroquinone) to create homogenous paste. Cement injected through screw fenestrations into targeted vertebral body to provide fixation and restore spinal integrity in tumor patients. Screws connect to standard spinal rods. Procedure performed by surgeons in clinical/OR setting. Output is stabilized spinal segment; aids clinical decision-making by providing temporary structural support where fusion is not feasible; benefits patients by addressing spinal instability caused by metastatic disease.
Clinical Evidence
Clinical and non-clinical performance data provided in support of substantial equivalence. No specific clinical study metrics (e.g., sensitivity, AUC) are detailed in the summary.
Technological Characteristics
PMMA-based bone cement (powder/liquid components); cannulated multi-axial screws with six distal fenestrations; diameters 4.5mm-10.5mm; lengths 30-100mm. Materials: PMMA, methylmethacrylate-styrene co-polymer, barium sulfate, methylmethacrylate monomer. Mechanical fixation via cement augmentation. Non-sterile screws; sterile cement components.
Indications for Use
Indicated for patients with advanced stage tumors of the thoracic and lumbar spine where life expectancy is insufficient for fusion and spinal structural integrity is not severely compromised.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2016
Medtronic Sofamor Danek USA, Incorporated Mr. Lee Grant Distinguished Regulatory Affairs Advisor 1800 Pyramid Place Memphis, Tennessee 38132
Re: K152604
Trade/Device Name: KYPHON HV-R® Fenestrated Screw Cement, CD HORIZON® Fenestrated Screw Set Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: PML, MNI Dated: December 8, 2015 Received: December 9, 2015
Dear Mr. Grant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K152604
Device Name KYPHON HV-R® Fenestrated Screw Cement
Indications for Use (Describe)
When used in conjunction with the CD HORIZON® Fenestrated Screws, KYPHON HV-R® Fenestrated Screw Coment is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. KYPHON HV-R® Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span> <span style="font-size:16px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152604
Device Name CD HORIZON® Fenestrated Screw Set
### Indications for Use (Describe)
When used in conjunction with KYPHON HV-R® Fenestrated Screw Cement, the CD HORIZON® Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in of insufficient duration to permit achievement of fusion. CD HORIZON® Fenestrated Screws augmented with KYPHON HV-R® Fenestrated Screw Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (*Select one or both, as applicable*)
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
### 510(k) Summary – K152604
# Medtronic Sofamor Danek USA, Inc.
### December 2015
| Submitter | Medtronic Sofamor Danek USA, Inc.<br>1800 Pyramid Place<br>Memphis, Tennessee 38132<br>Telephone: (901) 396-3133<br>Fax: (901) 346-9738 |
|------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact(s) | Lee Grant<br>Distinguished Regulatory Affairs Advisor<br>Direct Telephone – 901-344-0807<br><br>Pamela Edwards<br>Principal Regulatory Affairs Specialist<br>Direct Telephone -901-399-2125 |
| Date Prepared | December 30, 2015 |
| Trade Name | 1) KYPHON HV-R® Fenestrated Screw Cement<br>2) CD HORIZON® Fenestrated Screw Set |
| Regulatory Class | 1) KYPHON HV-R® Fenestrated Screw Cement<br>Class II |
| Regulation Number | 888.3027 Polymethylmethacrylate (PMMA) Bone Cement<br>PML |
| Regulation Name and Device<br>Product Classification Code | 2) CD HORIZON® Fenestrated Screw Set<br>Class II<br>21 CFR 888.3070 Pedicle Screw System<br>MNI |
| Predicate Devices | 1) KYPHON® HV-R® Bone Cement (Primary Predicate)<br>K150460 KYPHON® HV-R® Bone Cement (SE 04/28/15)<br><br>2) CD HORIZON® Spinal System<br>K042025 CD HORIZON® Spinal System (SE 08/25/04) and<br>K113174 CD HORIZON® Spinal System (SE 11/21/11) |
| | The predicate devices have not been subject to a design related recall |
| Description of Device | 1) KYPHON HV-R® Fenestrated Screw Cement is self-curing PMMA<br>based (high viscosity, radiopaque) bone cement. KYPHON HV-R®<br>Fenestrated Screw Cement will be provided sterile in two components:<br>20 grams of powder and nine grams of liquid. The powder contains<br>methylmethacrylate-styrene co-polymer, barium sulfate as a<br>radiopacifier, and di-benzoyl peroxide as an initiator. This liquid<br>contains methylmethacrylate monomer, N, N dimethyl-p-toluidine as a<br>promoter and hydroquinone as a stabilizer. The powder and liquid |
| | components are mixed, in the provided liquid-to-powder proportions, into a homogenous paste, to initiate the polymerization reaction of monomer into polymer. |
| | 2) The CD HORIZON® Fenestrated Screw Set consists of a variety of cannulated multi-axial screws (MAS) with fenestrations offered in diameters ranging from 4.5mm to 10.5mm, with lengths ranging from 30-100mm based on CD HORIZON® LEGACY™ and CD HORIZON® SOLERA™ implants contained in the CD HORIZON® Spinal System. The CD HORIZON® Fenestrated Screws are specifically designed to connect to 4.75mm, 5.5mm, and 6.0mm diameter rods and associated connecting components contained within the CD HORIZON® Spinal System. The screws contain six fenestrations near the distal tip of the screw which provides a controlled means to deliver a small amount of polymethylmethacrylate (PMMA) bone cement into a targeted vertebral body. These screws are provided non-sterile. |
| Indications for Use: | 1) KYPHON HV-R® Fenestrated Screw Cement: When used in conjunction with the CD HORIZON® Fenestrated Screws, KYPHON HV-R® Fenestrated Screw Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. KYPHON HV-R® Fenestrated Screw Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised. |
| | 2) CD HORIZON® Fenestrated Screw Set: When used in conjunction with KYPHON® HV-R Fenestrated Screw Cement, the CD HORIZON® Fenestrated Screws are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. CD HORIZON® Fenestrated Screws augmented with KYPHON HV-R® Fenestrated Screw Cement are for use at spinal levels where the structural integrity of the spine is not severely compromised. |
| Comparison of Technological<br>Characteristics with the<br>Predicate Devices | 1) The KYPHON® HV-R® Fenestrated Screw Cement is identical in composition, method of manufacture and sterilization to the primary predicate Kyphon® H-VR® Bone Cement cleared by the FDA in K150460 (SE 04/28/15). The predicate cement is intended to be used in patients diagnosed with pathological fractures in the vertebral body caused by conditions such as osteoporosis and cancer. The subject cement is also intended to be used in patients with metastatic cancer. Pathological fractures reflect one means in which spinal instability occurs and therefore in both instances the predicate and subject cement are intended to treat spinal instability. The only differences between the subject and predicate cement are the method in which the cement is delivered and that the subject cement is limited to patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of |
| | fusion. |
| | 2) The screws which comprise the CD HORIZON® Fenestrated Screw Set<br>have the same or similar indications, intended use, fundamental<br>scientific technology, and are manufactured from similar materials as the<br>pedicle screws found in the following FDA cleared CD HORIZON®<br>Spinal System applications: K042025 (SE 08/25/2004) and K113174<br>(SE 11/21/2011). The predicate screws in this application are cleared for<br>pedicle fixation in patients with spinal instability caused by tumors. The<br>subject pedicle screws are also intended to provide pedicle fixation in<br>patients diagnosed with spinal instability caused by tumors. Both the<br>predicate and subject screws are cannulated, are manufactured from the<br>same material, are similar in design, sizes and are intended to be used<br>with the same spinal rods and connecting components. The primary<br>difference is the subject screws contain fenestrations which allow<br>PMMA cement to flow in a controlled manner through the screw and<br>into the targeted pedicle. |
| Performance Data | Clinical data and Non-Clinical testing were provided in support of substantial<br>equivalence of the subject device. |
| Conclusion | Based on the provided performance data, the subject KYPHON® HV-R<br>Fenestrated Screw Cement is substantially equivalent to the KYPHON® HV-R®<br>Bone Cement (K150460, SE 04/28/15) and the CD HORIZON® Fenestrated<br>Screw Set is substantially equivalent to the CD HORIZON® Spinal System<br>(K042025 (SE 08/25/2004) and K113174 (SE 11/21/2011)). |
{5}------------------------------------------------
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.