← Product Code [OWN](/productcode/OWN) · K152583

# KARL STORZ Endoscopic ICG Imaging System (K152583)

_KARL STORZ Endoscopy-America, Inc. · OWN · May 31, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152583

## Device Facts

- **Applicant:** KARL STORZ Endoscopy-America, Inc.
- **Product Code:** [OWN](/productcode/OWN.md)
- **Decision Date:** May 31, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The KARL STORZ Endoscopic ICG Imaging System is intended to provide real-time endoscopic visible and near-infrared fluorescence imaging. The KARL STORZ ICG Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging. Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

## Device Story

System provides real-time HD endoscopic video using visible (VIS) and near-infrared (NIR) light; utilizes indocyanine green (ICG) dye fluorescence. Components include rigid ICG endoscopes, filtered Xenon light source with footswitch, 3 CCD color camera head, and Image 1 SPIES Camera Control Unit. Used in minimally invasive surgery by surgeons; output displayed on medical-grade HD monitor. NIR mode enables visualization of vessels, blood flow, tissue perfusion, and biliary ducts. System projects 2D fluorescence image on dichromatic background. Provides surgeons with visual assessment to aid intraoperative decision-making; intended to improve tissue/vessel identification during surgery.

## Clinical Evidence

No clinical data. Evidence includes bench testing (spectral characteristics, light transmission, signal detection), a GLP animal study in a porcine minimally-invasive laparoscopy model, electrical safety (IEC 60601-1, 60601-2-18), electromagnetic compatibility (IEC 60601-1-2), biocompatibility (ISO 10993), and software/sterilization validation.

## Technological Characteristics

System uses rigid endoscopes, 3 CCD camera head, and filtered Xenon light source for VIS/NIR illumination. Connectivity via DVI-D or 3G-SDI to HD monitors. Biocompatibility per ISO 10993; electrical safety per IEC 60601-1/2-18; EMC per IEC 60601-1-2. Software-controlled image processing via Image 1 SPIES CCU.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Intuitive Surgical® da Vinci® Fluorescence Imaging Vision System ([K124031](/device/K124031.md))

## Reference Devices

- Image 1 SPIES Camera Control Unit ([K131953](/device/K131953.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 31, 2016

Karl Storz Endoscopy-America, Inc. Winkie Wong Senior Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

Re: K152583 Trade/Device Name: Karl Storz Endoscopic ICG Imaging System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: OWN Dated: April 13, 2016 Received: April 15, 2016

Dear Winkie Wong,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152583

Device Name

KARL STORZ Endoscopic ICG Imaging System

Indications for Use (Describe)

The KARL STORZ Endoscopic ICG Imaging System is intended to provide real-time endoscopic visible and nearinfrared fluorescence imaging. The KARL STORZ ICG Imaging System enables surgeons to perform minimally invasive surgery using standard endoscopic visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile duct, common bile duct and common hepatic duct), using near-infrared imaging.

Fluorescence imaging of biliary ducts with the KARL STORZ Endoscopic ICG Imaging System is intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The device is not intended for standalone use for biliary duct visualization.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|                                                 |  |

|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters. 

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

| Applicant:             | Karl Storz Endoscopy-America, Inc<br>2151 E. Grand Avenue<br>El Segundo, CA 90245                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:               | Winkie Wong<br>Senior Regulatory Affairs Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                        | Phone: (424) 218-8379<br>Fax: (424) 218-8519                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Date of Preparation:   | May 18th, 2016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Device Identification: | Trade Name: KARL STORZ Endoscopic ICG Imaging System<br>Common Name: Endoscopic Video Imaging System<br>Classification Name: Endoscope and accessories                                                                                                                                                                                                                                                                                                                                                                            |
| Product Code:          | OWN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation:            | 21 CFR part 876.1500 (Endoscope and Accessories)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Predicate Device(s):   | The KARL STORZ Endoscopic ICG Imaging System is<br>substantially equivalent to the Intuitive Surgical® da Vinci®<br>Fluorescence Imaging Vision System (K124031), marketed<br>by Intuitive Surgical, Inc.                                                                                                                                                                                                                                                                                                                         |
| Indications For Use:   | The KARL STORZ Endoscopic ICG Imaging System is<br>intended to provide real-time endoscopic visible and near-<br>infrared fluorescence imaging. The KARL STORZ ICG<br>Imaging System enables surgeons to perform minimally<br>invasive surgery using standard endoscopic visible light as<br>well as visual assessment of vessels, blood flow and related<br>tissue perfusion, and at least one of the major extra-hepatic<br>bile ducts (cystic duct, common bile duct and common<br>hepatic duct), using near-infrared imaging. |
|                        | Fluorescence imaging of biliary ducts with the KARL STORZ<br>Endoscopic ICG Imaging System is intended for use with<br>standard of care white light and, when indicated,<br>intraoperative cholangiography. The device is not intended<br>for standalone use for biliary duct visualization.                                                                                                                                                                                                                                      |

{4}------------------------------------------------

| Device Description:               | The KARL STORZ Endoscopic ICG Imaging System is used to<br>provide real-time high definition (HD) endoscopic video<br>images of visible (VIS) and near-infrared (NIR) indocyanine<br>green (ICG) dye fluorescence during minimally invasive<br>surgery. The system components are rigid ICG Endoscopes<br>for VIS and NIR illumination and imaging, a light source with<br>foot switch for emission of VIS and NIR illumination, a 3 CCD<br>(charge coupled device) color video camera head capable of<br>capturing both VIS and NIR imaging, and KARL STORZ<br>Indocyanine Green (ICG Kit). Additional accessories used<br>with the KARL STORZ Endoscopic ICG Imaging System<br>include a standard fiber-optic light cable for transmission of<br>VIS and NIR light and the Image 1 SPIES Camera Control<br>Unit (CCU) cleared in K131953. The KARL STORZ Endoscopic<br>ICG Imaging System can be used with any medical grade HD<br>monitor with a DVI-D or 3G-SDI input. The KARL STORZ<br>Endoscopic ICG Imaging System is a Class II device under<br>21 CFR 876.1500. |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological Comparison Summary: | The KARL STORZ Endoscopic ICG Imaging System is<br>substantially equivalent to the Intuitive Surgical® da Vinci®<br>Fluorescence Imaging Vision System (K124031) in terms of<br>its indications for use, design technology and performance<br>specification.<br><br>The differences between the subject and predicate device<br>are:<br>The subject device uses a Filtered Xenon Light source<br>whereas the Predicate (da Vinci) uses a Laser Diode<br>during NIR Illumination Mode. The subject device projects 2D fluorescence image on<br>a dichromatic background whereas the predicate<br>device (da Vinci) projects 3D fluorescence image on a<br>monochromatic background in NIR Illumination Mode.          Bench and comparative testing were used to demonstrate<br>substantial equivalence to the predicate Intuitive Surgical®<br>da Vinci® Fluorescence Imaging Vision System, K124031.<br>Therefore, the differences between the subject device and<br>the predicate device do not raise any new issues of safety<br>and effectiveness.                    |
| Non-Clinical<br>Performance Data: | The KARL STORZ Endoscopic ICG Imaging System has been<br>successfully tested for its functions and performance,<br>including verification that the spectral characteristics of the<br>ICG system illumination light source, light transmission<br>system and endoscope enable a selective visualization of the<br>ICG fluorescence signal as detected by the camera system.<br>Furthermore, a GLP animal study was successfully performed<br>by the NAMSA testing facility to evaluate the performance of<br>the KARL STORZ Endoscopic ICG Imaging System in the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{5}------------------------------------------------

|                            | porcine minimally-invasive laparoscopy model. <p>Safety testing was performed including electrical safety IEC 60601-1 and 60601-2-18, electromagnetic compatibility per IEC 60601-1-2, and biocompatibility of the patient contacting materials per ISO 10993. Additional validations were conducted for system software, the manual cleaning method, and the sterilization process.</p>                                   |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Performance Data: | Clinical testing was not required to demonstrate substantial equivalence to the predicate device.                                                                                                                                                                                                                                                                                                                          |
| Conclusion:                | The KARL STORZ Endoscopic ICG Imaging System is substantially equivalent to the predicate Intuitive Surgical® da Vinci® Fluorescence Imaging Vision System (K124031) with regards to device design, materials, operational principles, and performance characteristics. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as or better than the legally marketed device. |

---

**Source:** [https://fda-staging.innolitics.com/device/K152583](https://fda-staging.innolitics.com/device/K152583)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
