← Product Code [IOR](/productcode/IOR) · K152536

# MyOn HC Manual Wheelchair (K152536)

_Invacare Corporation · IOR · Jan 5, 2016 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K152536

## Device Facts

- **Applicant:** Invacare Corporation
- **Product Code:** [IOR](/productcode/IOR.md)
- **Decision Date:** Jan 5, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3850
- **Device Class:** Class 1
- **Review Panel:** Physical Medicine

## Indications for Use

The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.

## Device Story

MyOn HC™ is a foldable, manually operated, adjustable lightweight wheelchair. It provides mobility and support to users limited to a sitting position, ranging from limited to full-day usage. The device features a horizontal folding mechanism, adjustable center of gravity (5"), and stepless backrest angle adjustments (-15° to +15°) operable while the user is seated. It accommodates various riggings (swing-in/out, detachable, fixed 80°). Accessories include posture belts, tipper aids, air pumps, passive illumination, cane holders, and luggage carriers. The device is intended for use by adolescents and adults (ages 12+) with weight capacities of 220 or 290 lbs depending on configuration. It is operated by the user or a caregiver to facilitate mobility in various environments.

## Clinical Evidence

No clinical data. Bench testing only. Testing included ANSI/RESNA WC/Volume 1 & 2 (2009) for static stability, dimensions, maneuvering space, seating dimensions, static/impact/fatigue strength, information disclosure, and ignition resistance. Flammability testing performed per CAL117:2013 and ISO 8191-1:1987/8191-2:1988.

## Technological Characteristics

Manual, foldable, lightweight wheelchair. Features horizontal folding frame, adjustable center of gravity, and stepless backrest adjustment. Materials include standard wheelchair components. Testing conducted per ANSI/RESNA WC/Volume 1 & 2 (2009) and ISO 8191 standards. No electronic or software components.

## Regulatory Identification

A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Action Patriot Manual Wheelchair ([K930803](/device/K930803.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 5, 2016

Invacare Corporation Elijah Wreh Regulatory Affairs Manager (Pre-Market) One Invacare Way Elyria, Ohio 44035

Re: K152536

Trade/Device Name: MyOn HCTM Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: December 4, 2015 Received: December 7, 2015

Dear Elijah Wreh:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Michael J. Hoffmann -A

for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152536

Device Name MyOn HCTM Manual Wheelchair

Indications for Use (Describe)

The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary

| SUBMITTER:            | Invacare Corporation<br>One Invacare Way<br>Elyria, OH 44035                                     |
|-----------------------|--------------------------------------------------------------------------------------------------|
|                       | Phone: (440) 329-6840<br>Email: <a href="mailto:ewreh@invacare.com">ewreh@invacare.com</a>       |
| CONTACT PERSON:       | Elijah N. Wreh<br>Regulatory Affairs Manager (Pre-Market)                                        |
| Date Prepared:        | January 5, 2016                                                                                  |
| DEVICE                |                                                                                                  |
| Name of Device:       | MyOn HCTM Manual Wheelchair                                                                      |
| Common or Usual Name: | Wheelchair, mechanical                                                                           |
| Classification Name:  | Mechanical wheelchair 21 CFR §890.3850                                                           |
| Regulatory Class:     | I                                                                                                |
| Product Code:         | IOR: Wheelchair, mechanical                                                                      |
| PREDICATE DEVICE:     | Action Patriot Manual Wheelchair (K930803)<br>No reference devices were used in this submission. |

### DEVICE DESCRIPTION

This Traditional [510(k)] submission is being supplied to the U.S. FDA to obtain authorization to market the MyOn HC™ Manual Wheelchair. The MyOn HC™ Manual Wheelchair is a foldable, manually operated fully adjustable lightweight wheelchair. It is indicated to provide mobility to persons which have limitations in mobility. It provides support and mobility to users that are seated in the wheelchair, for limited time up to permanent, full day usage. The design incorporates a horizontal folding mechanism connected to a left & right side frame on which many adjustments can be made to meet the individual user needs. The frame can be either equipped with swing in or swing out, detachable, riggings with an angle of 70°, 80° or the frame which has a fixed 80° front. It allows 5" of center of gravity (CG) adjustments and stepless backrest angle adjustments from -15° to + 15° which can be achieved whist the user is seated in the wheelchair. The subject device intended use is to provide mobility to persons ages 12 and up

{4}------------------------------------------------

(adolescents and adults) with a weight capacity of 220 & 2901bs depending on the seat width. There is no prior submission for the subject device.

The associated accessories include:

- Posture belt .
- Tipper aid
- Air pump .
- Passive illumination ●
- . Cane holder
- . Luggage carrier

### INDICATIONS FOR USE

The MyOn HC™ Manual Wheelchair is intended to provide mobility to persons ages 12 and up (adolescents and adults) with a weight capacity of 220 & 290 lbs. depending on the seat width. The device is indicated to provide mobility to persons limited to a sitting position.

# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The device comparison showed that the subject device is substantially equivalent in intended use, design and operational principles to the previously cleared Action Patriot Manual Wheelchair (K930803). The subject device is substantially equivalent to the predicate device in regards to intended use, design, materials, and operational principles to provide mobility to persons limited to a sitting position.

{5}------------------------------------------------

### Design Characteristics Comparison

| Design                    | Action Patriot Manual Wheelchair<br>Predicate Device<br>(K930803) | MyOn HCTM Manual Wheelchair<br>Subject Device<br>Pending Submission                            |
|---------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Weight Limit              | 250Ibs                                                            | 220 & 290lbs*                                                                                  |
| Cross-brace Configuration | X-design                                                          | Folds lengthwise and backwards*                                                                |
| Seat Width                | 14"-20"                                                           | 12"-21" *                                                                                      |
| Seat Depth                | 16"& 18"                                                          | 14"-20" *<br>14"-16" adjustable*<br>16"-18" adjustable*                                        |
| Back Style                | Fixed or adjustable                                               | Fixed or adjustable                                                                            |
| Back Angle<br>Adjustment  | -5 to +20 degrees in 5 degree increments                          | -15 to +15 degrees*                                                                            |
| Back Height               | 15-19", 16", 20"                                                  | 12"-20"                                                                                        |
| Arm Types                 | Flip-back arm rests                                               | Flip-back arm rests, adjustable, desk length                                                   |
| Anti-Tippers              | Optional                                                          | Optional                                                                                       |
| Wheel Locks               | Push-to-Lock, Pull-to-Lock, Hill Holder                           | Push-to-Lock, Pull-to-Lock                                                                     |
| Rear Wheel<br>Sizes       | 20", 22", 24"                                                     | 22"-25" *                                                                                      |
| Caster Sizes              | 6", 8"                                                            | 4"-7" *                                                                                        |
| Front<br>riggings         | 60° straight<br>70° straight                                      | Pivot inwards and outwards<br>70° straight<br>80° straight*<br>90° straight*<br>Elevate 0-80°* |

### PERFORMANCE DATA

### Non-Clinical Test

Non-clinical laboratory testing was performed on the subject MyOn HCTM Manual Wheelchair to

determine substantial equivalence. The following testing was performed:

- ANSI / RESNA WC/Volume 2 2009, Section 1: Determination of Static Stability .
- ANSI / RESNA WC/Volume 1 2009, Section 5: Determination of Dimensions, Mass ● and Maneuvering Space
- ANSI / RESNA WC/Volume 1 2009, Section 7: Measurement of Seating and Wheel ● Dimensions
- ANSI / RESNA WC/Volume 1 2009. Section 8: Static. Impact And Fatigue Strengths ● Tests
- ANSI / RESNA WC/Volume 1 2009, Section15: Requirements for Information . Disclosure, Documentation and Labeling
- ANSI / RESNA WC/Volume 1 2009, Section16: Resistance to Ignition of Upholstered ● Parts
- CAL117:2013, Section 1: Flammability Testing ●
- ISO 8191-1:1987 & 8191-2:1988: Flammability Testing ●

{6}------------------------------------------------

Testing demonstrated that the subject MyOn HC™M Manual Wheelchair is substantially equivalent to the marketed predicate device.

## Animal Study

Animal testing was not required for this submission.

# Clinical Testing

Clinical testing was not required for this submission.

# CONCLUSIONS

The subject device has the same intended use and similar technological characteristics as the predicate device. The non-clinical laboratory data support the safety of the subject MyOn HC™ Manual Wheelchair and demonstrate that the subject device should perform as intended in the specified use conditions. Therefore, the subject MyOn HCTM Manual Wheelchair is substantially equivalent to the predicate device identified throughout this submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K152536](https://fda-staging.innolitics.com/device/K152536)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).