K152521 · The Lifeguard Store, Inc. · CBP · Jun 3, 2016 · Anesthesiology
Device Facts
Record ID
K152521
Device Name
Seal Rite Non-Rebreathing Valve
Applicant
The Lifeguard Store, Inc.
Product Code
CBP · Anesthesiology
Decision Date
Jun 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5870
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Seal Rite Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask (without oxygen port) to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation ("CPR") rescue techniques. It is intended for over-the-counter use.
Device Story
Resuscitation valve for mouth-to-mask ventilation; used in hospitals and field emergency settings by rescuers. Device input: expired air from rescuer. Device output: rescue breaths delivered to patient. Operation: plastic body with silicone inspiratory one-way valve and rigid plastic expiratory one-way valve. During inspiration, rescuer air pressure closes expiratory valve and opens inspiratory valve to patient. During expiration, patient airway pressure closes inspiratory valve and opens expiratory valve, directing exhaled gases away from rescuer through exhaust ports. Protects rescuer from patient contamination. Connects to standard masks via 22 mm ID/15 mm ID fittings.
Clinical Evidence
Bench testing only. Device passed all performance requirements per AS 4259-1995, including ventilation performance, resistance testing (inspiratory/expiratory), dead space measurement, and environmental stress testing (temperature, immersion, contamination). Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-5, 10993-10, and 10993-12 passed.
Technological Characteristics
Materials: Polycarbonate, silicone. Principle: Mechanical one-way valve system (silicone inspiratory, rigid plastic expiratory). Dimensions: Standard 22 mm ID inlet, 22 mm OD/15 mm ID outlet. Connectivity: Standalone. Sterilization: Non-sterile. Conformance: AS 4259-1995.
Indications for Use
Indicated for adult and pediatric patients >18 months of age requiring mouth-to-mask CPR rescue ventilation.
Regulatory Classification
Identification
A nonrebreathing valve is a one-way valve that directs breathing gas flow to the patient and vents exhaled gases into the atmosphere.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, layered on top of each other to create a sense of depth and unity.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 3, 2016
The Lifeguard Store, Inc. % Jennifer J. Hennessy Attorney Quarles & Brady, Llp 33 E Main Street, Suite 900 Madison. Wisconsin 53703
Re: K152521
Trade/Device Name: Seal Rite Non-Rebreathing Valve Regulation Number: 21 CFR 868.5870 Regulation Name: Nonrebreathing Valve Regulatory Class: Class II Product Code: CBP Dated: May 4, 2016 Received: May 5, 2016
Dear Jennifer J. Hennessy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runna DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K152521
Device Name
Seal Rite Non-Rebreathing Valve
#### Indications for Use (Describe)
The Seal Rite Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask (without oxygen port) to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation ("CPR") rescue techniques. It is intended for over-the-counter use.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Seal Rite™ Non-Rebreathing Valve 510(k) Summary
| Submitter's Name | The Lifeguard Store, Inc.<br>2012 W. College Ave.<br>Normal, IL 61761<br>Telephone: (309) 451-5858<br>Fax: (309) 451-5959 |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration Number | 3003197958 |
| Contact Name | Amy Hilten, COO<br>The Lifeguard Store, Inc.<br>2012 W. College Ave.<br>Normal, IL 61761<br>Telephone: (309) 451-5858<br>Fax: (309) 451-5959<br>Email: amy@thelifeguardstore.com |
| Date Prepared | June 2, 2016 |
| Device Trade Name | Seal Rite™ Non-Rebreathing Valve |
| Device Common Name | Non-Rebreathing Valve |
| Classification Name | Valve, Non-Rebreathing |
| Product Code | CBP |
| Device Classification | Class II |
| Panel | Anesthesiology and Respiratory Devices |
| Regulatory Classification | 21 CFR 868.5870 |
| Type of 510(k) Submission | Traditional |
| Legally Marketed Predicate Device | Respironics Rescue Valve (K833748) |
Premarket Notification (510(k)) Number
K152521
{4}------------------------------------------------
### Description
The Seal Rite™ Non-Rebreathing Valve ("Seal Rite Valve") is a resuscitation valve designed for resuscitation using expired air for ventilation. It has fittings on the grooved patient end that will adapt to standard masks (22 mm ID) used for resuscitation. The Seal Rite Valve allows the rescuer to blow air into a mask through its silicone one-way valve and directs the patient's exhaled air away from the rescuer via a rigid one-way valve and multiple exhaust ports. The Seal Rite Valve helps protect the rescuer from potential patient contamination. The Seal Rite Valve does not have an oxygen port.
#### Indications for Use
The Seal Rite Non-Rebreathing Valve is a single-use, non-sterile device intended to be used with a ventilation mask (without oxygen port) to provide mouth-to-mask ventilation to health emergency victims requiring cardiopulmonary resuscitation ("CPR") rescue techniques. It is intended for over-the-counter use.
#### Predicate Device
The Seal Rite Valve is substantially equivalent to the Respironics Rescue Valve (K833748). The Seal Rite Valve and the Rescue Valve are both resuscitation valves utilized for mouth-to-mask emergency resuscitation using expired air from the rescuer through a valve for ventilation of the patient. Both valves direct the patient's exhaled gases away from the rescuer's face via a rigid one-way valve, and both valves have an ISO Standard 15/22 connector that allows use with a standard mask.
Please refer to the Substantial Equivalence Summary Comparison Table below for a summary of the similarities and differences between the subject and predicate devices.
### Description of Operation
The Seal Rite Valve consists of a plastic body with two one-way valves: a silicone inspiratory one-way valve and a rigid, plastic expiratory one-way valve. Rescuers utilizing the Seal Rite Valve place their lips on the open end of the valve and blow expired air into the valve, which is attached to a standard mask, for the purpose of supplying rescue breaths to the patient. When the expired air enters the top of the Seal Rite Valve, the rigid one-way valve moves toward the patient, seals multiple exhaust ports, and the silicone one-way valve opens and provides air from the rescuer to the patient. The rigid, plastic expiratory valve remains closed when the rescuer is providing exhaled air to the patient because the pressure on the rescuer side of the plastic valve is
{5}------------------------------------------------
greater than the pressure inside the patient's airways. When the rescuer stops breathing exhaled air into the Seal Rite Valve, the silicone one-way valve closes and prevents the backflow of exhaled air from the patient to the rescuer. The rigid, plastic one-way valve lifts off of the exhaust ports due to the pressure inside the patient's airways being greater than the atmospheric pressure. The patient's exhaled air passes then through multiple exhaust ports.
#### Technological Characteristics Summary
The Seal Rite Valve is substantially equivalent to the Respironics Rescue Valve (K833748) with respect to intended use, design, principle of operation, technological characteristics and performance and the minor difference in device materials does not raise new questions of safety or effectiveness.
The Seal Rite Valve was tested to ensure conformance with the FDA Recognized AS 4259-1995 standard ("Ancillary devices for expired air resuscitation") and passed all testing requirements.
| Features | Seal Rite™ Non-Rebreathing<br>Valve (K152521) | Respironics Rescue Valve<br>(Predicate Device -- K833748) |
|---------------------------|-----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Intended Use | Mouth-to-mask ventilation | Mouth-to-mask ventilation |
| Product Code | CBP | CBP |
| Target Population | Adult/pediatric patients greater<br>than 18 months of age | Adult/pediatric patients greater<br>than 18 months of age |
| Environment of Use | Hospitals and field emergency | Hospitals and field emergency |
| Materials | Polycarbonate, silicone | Thermoplastic, silicone |
| Oxygen Port | Not available | Not available with standalone<br>valve; option available only with<br>accessory adapter cleared under<br>K842693 |
| Expiratory Resistance | 0.327 cm H2O (0.0321 kPa) at 50<br>LPM | 0.25 cm H2O at 50 LPM (per<br>K142402 510(k) Summary) |
| Inspiratory<br>Resistance | 1.15 cm H2O (0.113 kPa) at 50<br>LPM | 1.85 cm H2O at 50 LPM (per<br>K142402 510(k) Summary) |
| Inlet Connector | Standard 22 mm ID | Standard 22 mm ID |
| Outlet Connector | Standard 22 mm OD/15 mm ID | Standard 22 mm OD/15 mm ID |
| Sterile | No | No |
| Reusable | No - single patient use device | No - single patient use device |
| Duration of Use | Less than 24 hours | Less than 24 hours |
#### Substantial Equivalence Summary Comparison Table
{6}------------------------------------------------
| Principle of<br>Operation | Air flow through device used to<br>deliver inspiratory air and<br>exhaled expiratory air from<br>patient | Air flow through device used to<br>deliver inspiratory air and exhaled<br>expiratory air from patient |
|---------------------------|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| Compatibility | Designed for use with standard<br>resuscitation masks | Designed for use with standard<br>resuscitation masks |
### Performance Testing
As described below, a comprehensive battery of non-clinical tests was submitted to confirm product conformance with device requirements, including the FDA recognized AS 4259-1995 standard. These studies demonstrate that the Seal Rite™ Non-Rebreathing Valve is biocompatible, performs as intended and meets the requirements of the AS 4259-1995 standard and is therefore as safe and effective as the predicate device, the Rescue Valve.
## Biocompatibility Testing
The biocompatibility testing was conducted in accordance with FDA Blue Book Memo G95-1. Testing was conducted for cytotoxicity, sensitization and irritation per ISO 10993-5, 10993-10 and 10993-12.
| Test | Test Results |
|---------------|--------------|
| Cytotoxicity | Pass |
| Irritation | Pass |
| Sensitization | Pass |
### Bench Testing
| Test | |
|-----------------------------------------------------------------|------|
| Ventilation Performance | Pass |
| Expiratory Resistance for the Patient | Pass |
| Expiratory Resistance for the Rescuer | Pass |
| Inspiratory Resistance for a<br>Spontaneously Breathing Patient | Pass |
| Function After Contamination with<br>Stomach Contents | Pass |
{7}------------------------------------------------
| Function After Immersion in Water | Pass |
|--------------------------------------------------------------------------------------------------------------------------|------|
| Measurement of Dead Space | Pass |
| Resistance to Disengagement of Parts | Pass |
| High and Low Storage Conditions | Pass |
| High Temperature Operation | Pass |
| Low Temperature Operation | Pass |
| Attempted Displacement of Valve by<br>Finger, Function After Dropping,<br>Function After Potentially Deforming<br>Weight | Pass |
| Conformance of Connectors | Pass |
| ISTA Procedure 2A Shipping Testing | Pass |
In summary, the Seal Rite Valve is substantially equivalent to the predicate Respironics Rescue Valve with respect to intended use, technological characteristics and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.