CAMLOG and CONELOG Abutments for Screw-retained Restorations (ASR)
Applicant
Altatec GmbH
Product Code
NHA · Dental
Decision Date
Jan 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw-retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
Device Story
ASR abutments are dental components used to support screw-retained prosthetic restorations (crowns, bridges, bars) on CAMLOG or CONELOG implants. Available in straight and angled (17°/30°) configurations with various platform diameters and gingival heights; includes Type A/B anti-rotation cam options. Components are fabricated from titanium alloy (ASTM F136). Used in dental clinics by clinicians to restore chewing function. The device acts as a mechanical interface between the endosseous implant and the final prosthesis. Benefits include restoration of oral function and aesthetics for patients with missing teeth.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing, including static and dynamic compression-bending fatigue testing per ISO 14801, and validation of sterilization processes per ISO 11137 and ISO 17665. Biocompatibility was evaluated per ISO 10993-1.
Technological Characteristics
Materials: Titanium-6 Aluminum-4 Vanadium ELI alloy (ASTM F136). Design: Straight and angled (17°, 30°) abutments; multiple platform diameters (3.3-6.0 mm) and gingival heights. Interface: CAMLOG or CONELOG implant/abutment connections. Sterilization: Radiation and moist heat. No software or electronic components.
Indications for Use
Indicated for patients requiring support for occlusal screw-retained prosthetic restorations in the upper or lower jaw in conjunction with CAMLOG® or CONELOG® dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of an eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 12, 2016
ALTATEC GmbH c/o Ms. Linda K. Schulz Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K152509
Trade/Device Name: CAMLOG® and CONELOG® Abutments For Screw-retained Restorations (ASR) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 10, 2015 Received: December 11, 2015
Dear Ms. Schulz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory. Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### Device Name
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)
#### Indications for Use (Describe)
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw- retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
# Altatec GmbH CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)
January 11, 2016
### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Altatec GmbH<br>Maybachstrasse 5<br>D-71299 Wimsheim, Germany | | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|--|--|
| | Telephone: | +49 7044 9445 0 | | |
| | Fax: | +49 7044 9445 723 | | |
| Official Contact | Colleen Boswell<br>Director of RA/QA Ortho Organizers, Inc. | | | |
| Representative/Consultant | Linda K. Schulz, BSDH, RDH<br>Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | | | |
| | Telephone: | +1 (858) 792-1235 | | |
| | Fax: | +1 (858) 792-1236 | | |
| | Email: | lschulz@paxmed.com<br>flarson@paxmed.com | | |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | CAMLOG® and CONELOG® Abutments for Screw-retained<br>Restorations (ASR) |
|----------------------------|-------------------------------------------------------------------------|
| Common Name | Endosseous dental implant abutment |
| Classification Name | Endosseous dental implant abutment |
| Classification Regulations | 21 CFR 872.3630, Class II |
| Product Code | NHA |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Devices Branch |
### INTENDED USE
CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR) are intended to support occlusal screw-retained prosthetic restorations in the upper jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants.
{4}------------------------------------------------
#### PREDICATE INFORMATION
The primary predicate device is K133421.
Reference Predicates
K141871 - Straumann® Screw Retained Abutments K073553 - CAMLOG Implant System Abutments K083496 - CAMLOG Implant System Modified Implants and Abutments K113779 - CONELOG® Implant System
### DEVICE DESCRIPTION
Abutment for Screw-retained Restorations (ASR) is available with the Camlog or the Conelog implant/abutment interface. ASR Straight is provided in five platform diameters (3.3. 3.8. 4.3. 5.0 and 6.0 mm) and three gingival heights (0.5, 2.0 and 4.0 mm), for a total of 15 abutment options. ASR Angled is provided in four platform diameters (3.3, 3.8, 4.3, and 5.0 mm), four gingival heights (2.5, 3.5, 4.0 and 5.0 mm), and two angles (17° and 30°). Each of the angled designs is provided in a Type A option (anti-rotation cam alignment away from the abutment cone angle) or Type B option (anti-rotation cam alignment in the direction of the abutment cone angle), for a total of 32 abutment options. A healing cap, and a titanium cap for temporary or permanent restoration fabrication are available for all sizes. All ASR components are made from titanium alloy conforming to ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401).
### PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: sterilization validation according to : ISO 11137-1 Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices and ISO 11137-2 Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose, ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices; ISO 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1; ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process, and static and dynamic compression-bending testing according to ISO 14801 Dentistry - Implants - Dynamic fatigue test for endosseous dental implants.
Clinical data were not submitted in this premarket notification.
{5}------------------------------------------------
# EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles predicate devices shown above. Below are summary tables showing technical comparison between the subject device and the primary predicate devices.
| | Indications for Use |
|---------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | |
| Altatec Gmbh<br><br>CAMLOG® and<br>CONELOG® Abutments<br>for Screw-retained<br>Restorations (ASR) | CAMLOG® and CONELOG® Abutments for Screw-retained Restorations (ASR)<br>are intended to support occlusal screw-retained prosthetic restorations in the upper<br>jaw and lower jaw in conjunction with CAMLOG® and CONELOG® Implants. |
| Primary Predicate Device | |
| Straumann USA, LLC<br><br>Straumann® Magellan™<br>Abutment System<br><br>K133421 | The Straumann® Magellan™ abutments are indicated to be placed into Straumann®<br>dental implants to provide support for prosthetic reconstructions such as crowns,<br>bridges and bars.<br>The final processed devices have the purpose of restoring chewing function.<br>Magellan™ abutments are indicated for screw-retained restorations. |
| | Subject<br>Device | Primary<br>Predicate | Reference Predicate Devices | | |
|-----------------------------------------|----------------------------------------------------------------------------------------|-----------------------------------------------|----------------------------------------------|------------------------------------------|-------------------------------|
| | Altatec Gmbh | Straumann<br>USA, LLC | Straumann<br>USA, LLC | Altatec<br>GmbH | Altatec Gmbh |
| | CAMLOG®<br>and<br>CONELOG®<br>Abutments for<br>Screw-retained<br>Restorations<br>(ASR) | Straumann®<br>Magellan™<br>Abutment<br>System | Straumann®<br>Screw<br>Retained<br>Abutments | CAMLOG<br>Implant<br>System<br>Abutments | CONELOG®<br>Implant<br>System |
| | | K133421 | K141871 | K073553 | K113779 |
| Design | | | | | |
| Prosthesis<br>Attachment | Screw-retained | Screw-retained | Screw-retained | Screw-retained | Screw-retained |
| Abutment<br>Platform<br>Diameter,<br>mm | 3.3 - 6.0 | 3.5 - 4.6 | 4.6 | 3.3 - 6.0 | 3.3 - 5.0 |
| Abutment<br>Angle | Straight to 30° | Straight to 30° | Straight to<br>30° | Straight to<br>20° | Straight to<br>20° |
| Material | | | | | |
| Abutment | Ti-6AL-4V<br>ELI | Ti-6AL-7Nb | Ti-6AL-7Nb | Ti-6AL-4V<br>ELI | Ti-6AL-4V<br>ELI |
{6}------------------------------------------------
Abutment for Screw-retained Restorations (ASR) is substantially equivalent to the predicate devices in K133421 and K141871. K133421 is the primary predicate having straight and angled abutments (17° and 30°), multiple gingival heights (1.0 - 4.0 mm) and Type A and B indexing orientations. Straumann® Screw Retained Abutments K141871 adds angled abutments for the NC diameter implant. Both ASR and Straumann Screw-retained Abutment have a titanium cap designed for temporary or permanent restorations. The subject device and primary predicate have slightly different Indications for Use language. However, the difference in language does not change the intended use of abutments. Both abutments are being used in conjunction with dental implants for the purpose of supporting screw-retained dental prostheses.
## CONCLUSIONS
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject device and predicate devices encompass the same range of physical dimensions, including length, diameter, height, and angle of the abutment. The subject and predicate devices are packaged in similar materials and sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.